Attach 001 P09 Statement of Need_WN CBOC Sound Booth.docx
DOCX document 19 KB Posted
- Solicitation number
- 36C24423Q1147
About this file
This statement of need document outlines the requirements for the delivery and installation of an audiometric sound booth at the West Norriton Community Based Outpatient Clinic in Norristown, Pennsylvania. The project requires furnishing and installing a single-walled Eckel ADA C-14A LP audiometric testing booth with specified interior and exterior dimensions, a low profile ADA compliant entry, lighting, electrical outlets, data ports, windows, and noise reduction characteristics. The product must be compatible for use with specified audiology equipment. Delivery and installation must take place at the West Norriton CBOC during normal business hours, and the contractor shall coordinate access for inspection and acceptance. The related federal contract opportunity is a Veterans Affairs VOSB set-aside solicitation for the sound booth project located at the West Norriton CBOC.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attach 06 20939 VA PrivTraining for Contractor Personnel-No Access to PHI or VA Sens Info 09212018.pdf | ||
| Attach 004 52.225-18-Place of Manufacture.docx | DOCX document | |
| Attach 002A Price Cost Scehdule--Mandatory.xlsx | XLSX spreadsheet | |
| Attach 005 852.219-76 VA NOTICE OF LIMITATIONS SubKTR.docx | DOCX document | |
| Attach 003 52.225-2 BUY AMERICAN CERTIFICATE.docx | DOCX document | |
| Attach 002B Price Cost--Informational.docx | DOCX document | |
| 36C24423Q1147_1.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF NEED FOR
DELIVERY AND INSTALLATION OF AUDIOBOOTH
WEST NORRITON COMMUNITY BASED OUTPATIENT CLINIC
2495 GENERAL ARMISTEAD AVENUE
NORRISTOWN, PA 19403-3647
1. Background.
1.1 The Audiology Section of the West Norriton Community Based Outpatient Clinic (CBOC).
1.2 All products shall conform to commercial industry standards. Upon award of the purchase order, fabrication of the booth shall be on a standard schedule lead time of no longer than 6-8 weeks. The contractor shall provide a solid schedule and coordinate with VA point of contact. The Contractor shall complete a walk through with VA Government Technical Point of Contact (GTPOC) for a physical inspection of the installation of the booth and accessories prior to acceptance by the VA.
2. Objective Install a single-walled Audiometric Testing Booth in West Norriton, room 131. The West Norriton CBOC is located at 2495 General Armistead Avenue, Norristown, PA 19403.
3. Description of Supplies/Services
3.1 Supplies/Services will include fabrication of the booth, shipping of the booth to site, and installation of the booth.
3.1.1 Deliver, unpack, and remove the packing and shipping debris.
3.1.2 Provide serial #’s for booth at least 1 workday in advance to the following to allow equipment records to be created and Biomed to be notified: To Be Published.
3.1.3 Install, configure, test, and calibrate booth to ensure the system functions as designed and the desired outcomes are achieved.
4. Specific Task Requirements/Salient Characteristics: Brand Name or Like Eckel ADA C-14A LP meeting the following salient characteristics.
4.1 Booth shall have the following outside dimensions width (W) x depth (D) x height (H): 2159mm x 1880mm x 2372mm
4.2 Booth shall have the following inside dimensions width (W) x depth (D) x height (H): 1956mm x 1676mm x 2134mm
4.3 Booth shall have 14mm rise- low profile entry, Americans with Disabilities Act (ADA) compliant entry
4.4 Booth shall have an ADA compliant 36” clear opening
4.5 Booth shall have a minimum of 4 duplex outlets on the inside
4.6 Booth shall have a minimum of 1 duplex ethernet data port on the inside
4.7 Booth shall have a 30” window
4.8 Booth shall have a 20” x 60” window in the door
4.9 Booth shall have dimmable LED lighting
4.10 Booth shall be compatible for use with Titan/AOE with touch screen, Audiostar PRO 1 without AMTAS, VeriFit2
4.11 Sprinkler Port and empty rough-ins will allow sprinkler to be installed by the Government
4.12 Booth shall have a minimum noise reduction as follows:
4.10.1 OCTAVE BAND FREQUENCY (HZ)
| 125 |
| 250 |
| 500 |
| 1K |
| 2K |
| 4K |
| NIC |
| 24 |
| 32 |
| 44 |
| 53 |
| 58 |
| 61 |
| 45 |
5. Place of Performance
5.1 Audiometric Sound Booth to be shipped directly to West Norriton CBOC:
West Norriton 2495 General Armistead Avenue Norristown, PA 19403-3647
5.2 Equipment removal, delivery, and installation must be during the West Norriton CBOC business hours: Monday – Friday, 8:00 AM – 4:30 PM, excluding government holidays.