Att 1 Salient Characteristics.pdf

PDF 482 KB Posted

Attached to
Millimeter Electrical Analyzer System Federal contract opportunity
Solicitation number
N00173-21-P-6000
Issued by
Department of the Navy Secretary of the Navy Office of Naval Research

About this file

This document outlines salient characteristics for a millimeter electrical network analyzer system procurement. The Naval Research Laboratory requires a 4-port network analyzer unit with frequency extender modules to operate from 10 MHz to 220 GHz for electrical characterization of components in receiver systems. The system must meet specified dynamic range, noise floor, power output, trace noise, reflection and transmission tracking parameters, and temperature stability. Additional requirements include noise figure measurement features, spectrum analysis software to 120 GHz, and the ability to perform scattering parameter and active device measurements. Contract N00173-21-P-6000 was awarded to Technical Communities, Inc. on February 25, 2021 for $543,356.70 to fulfill these requirements.

View the file

Other files for this federal contract opportunity

Other files attached to Millimeter Electrical Analyzer System, newest first.
File Type Posted
Att 2 Quote 472382-1.pdf PDF
Redacted Approved J and A.pdf PDF
2-10 N00173-21-P-6000.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PR# 72-7203-20

SALIENT CHARACTERISTICS

The Naval Research Laboratory (NRL) has a requirement for the following pieces of equipment. An electrical network analyzer systems is required for the electrical characterization of components used in receiver systems in the frequency bands from 10

MHz to 220 GHz (specific bands denoted in performance requirements). The electrical network analyzer system shall consist of a base 4-port network analyzer unit with frequency extender modules as required to meet frequency spectrum, WR-05 calibration kit, 2.4 mm test cables, and integrated analysis software. The specifications below shall be met in order to be a compatible electrical network analyzer systems. For responses to the RFI, please provide technical backup that the product offered meets all required specifications and is fully compatible electrical network analyzer systems bits.

Millimeter-wave Electrical Network Analyzer System (MENAS)

-Frequency requirements, base unit: 10MHz – 43 GHz; extension bands to 75-110 GHz, and 140-220 GHz

-System dynamic range at 20 GHz, 10 Hz IFBW: 125 dB

-Noise floor at 20 GHz, 10 Hz IFBW: -114 dBm

-Power output at 20 GHz, adjustable; up to or greater than: -14 dBm

-Trace noise, at 10 KHz IFBW Mag (dBrms)/ Phase (degrms): 0.01/0.07

-Test set: 4 ports, with reference mixer switch

(The 4 port capability is only up to 43 GHz, not extension modules)

-Reflection tracking magnitude (10-20 GHz): +/- 0.01

-Reflection tracking phase (10-20 GHz): +/- 0.05

-Transmission tracking magnitude (10-20 GHz): +/- 0.05

-Transmission tracking phase (10-20 GHz): +/- 0.4

-Temperature stability (20 GHz):

Magnitude: 0.017 dB/deg C

Phase: 0.45 deg/deg C

-Standard wall power, 110-120 VAC, 60 Hz operation

Supplied noise figure measurement feature

Spectrum analysis software for multichannel analysis at millimeter-wave frequencies to support frequencies to 120 GHz, minimum.

System shall be capable of making standard scattering parameter measurements, PR# 72-7203-20

SALIENT CHARACTERISTICS

as well as active measurement applications that include swept-frequency gain compression, two-tone IMD, pulse, noise figure measurements for amplifiers and frequency converters

(The Noise Figure, Pulse, IMD, Gain Compression measurements only pertain to the mainframe unit, up to 43 GHz.)

Expansion capability to employ automatic calibration modules

System shall be contained as a desktop or tabletop configuration.