Atch 2 - LINAC Requirements and Specifications.pdf

PDF 190 KB Posted

Attached to
Linear Accelerator Federal contract opportunity
Solicitation number
HU000125Q0028
Issued by
DOD Uniformed Services University of the Health Sciences

About this file

Attachment 2 provides detailed technical specifications for a Linear Accelerator (LINAC) procurement for the Armed Forces Radiobiology Research Institute (AFRRI). The document specifies precise installation requirements, including a room space of 20 ft 8-1/8 inches wide, 22 ft deep, with a 13 ft 1-1/2 inch machine opening and 8 ft 6-3/8 inches height, located in Building 42, Room 1127.

The LINAC must include specific technical capabilities: photon energies of 4 MV (mandatory), 10 MV, and 15 MV with a 40 cm x 40 cm field size, high-resolution beam shaping, multi-leaf collimator, and MV high-dose rate mode. Electron specifications require energies of 6, 9, 12, and 15 MeV with multiple aperture plate applicators. Additional required systems include MV portal imaging, kV cone-beam CT imaging, laser alignment, CCTV with pan/tilt/zoom capabilities, and an intercom system. The procurement also involves removing the current Elekta Infinity LINAC and providing onsite training.

View the file

Other files for this federal contract opportunity

Other files attached to Linear Accelerator, newest first.
File Type Posted
HU000125Q0028 0002.pdf PDF
Linear Accelerator Solicitation and Site Visit Questions.pdf PDF
Atch 5 Elevator Dimensions.pdf PDF
Atch 6 LINAC - Alternate Route.pdf PDF
Atch 4 AFRRI Floor Plan.pdf PDF
RFQ AFRRI LINAC Amendment 0001.pdf PDF
HU000125Q0028 0001.pdf PDF
Attch 1 - Provisions and Clauses.pdf PDF
Atch 3 - NSAB Map.pdf PDF
RFQ AFRRI LINAC.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment 2 – LINAC Requirements and Specifications

AFRRI LINAC Requirements and Specifications

• FOB Delivery

Site visit will identify the path from the loading dock (via 10,000 lb freight elevator) to LINAC exposure room. Minimum door opening is 48 inches.

• Installation

Location: Bldg 42, Room 1127

The space requirement is:

Width: 20 ft 8-1/8 inches. Opening for the machine is 13 ft 1-1/2 inches.

Depth: 22 ft

Height: 8 ft 6-3/8 inches

20’ 8-1/8”

13’ 1-1/2”

22’

Attachment 2 – LINAC Requirements and Specifications

• Commissioning

• Removal and disposal of current LINAC: Elekta Infinity with Agility Head and XVI kV

Cone Beam CT and iView MV Portal Imaging

• Onsite training

LINAC Minimum Specifications

• Photon specifications o Minimum of 3 flattened energies o Photon energies: 4 MV, 10 MV and 15MV

NOTE the 4 MV is an absolute must o Minimum 40 cm x 40 cm field size o High resolution beam shaping (5 mm at isocenter) across full field size o Multi-leaf collimator (MLC) with fine resolution leaves (0.5 cm wide) across the full field size o MV high-dose rate mode (flattering filter free)

• Electron specifications o Electron flatness according to U.S.A. standards, optimized at 100 cm.

o Electron energies: 6 MeV, 9 MeV, 12 MeV, 15 MeV o Aperture plate electron beam applicators

6 x 6 cm, surface-to-surface distance (SSD) 95 cm 10 x 10 cm, SSD 95 cm 14 x 14 cm, SSD 95 cm 20 x 20 cm, SSD 95 cm

• MV portal imaging system

• kV cone-beam CT imaging system

• Laser alignment system

• CCTV system o Pan, tilt, zoom, high definition, full color, megapixel class o Minimum 2 cameras located in treatment area o Video displayed and control from control area

• Intercom system – control area to treatment area

File details come from the government source that posted it. Updated .