ATCH 1 SCHEDULE SALIENTS 36C24822Q1352.docx

DOCX document 27 KB Posted

Attached to
Hospital Bed Mover **Brand Name or Equal**
Solicitation number
36C24822Q1352
Issued by

View the file

Other files

File Type Posted
COMBINED 36C24822Q1352.docx DOCX document
ATCH 3 LIMITATIONS OF SUBCONTRACTING 36C24822Q1352.docx DOCX document
ATCH 2 CLAUSES AND PROVISIONS 36C24822Q1352.docx DOCX document
Cover Page - 36C24822Q1352_1.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ACKNOWLEDGMENT OF AMENDMENTS: The offeror acknowledges receipt of amendments to the Solicitation numbered and dated as follows:

AMENDMENT NO
DATE

B.1 PRICE/COST SCHEDULE

ITEM INFORMATION

ITEM NUMBER
DESCRIPTION OF SUPPLIES/SERVICES
QUANTITY
UNIT
UNIT PRICE
AMOUNT
1.00
EA
__________________
__________________

StaminaLift Bariatric Bed Mover, Model # TS6000 (brand name or equal) Contract Period: Base POP Begin: 07-25-2022 POP End: 09-26-2022 PRINCIPAL NAICS CODE: 339113 - Surgical Appliance and Supplies Manufacturing PRODUCT/SERVICE CODE: 6515 - Medical and Surgical Instruments, Equipment, and Supplies

MANUFACTURER PART NUMBER (MPN): TS6000

LOCAL STOCK NUMBER: TS6000

GRAND TOTAL
__________________

B.2 DELIVERY SCHEDULE

ITEM NUMBER
QUANTITY
DELIVERY DATE
0001
SHIP TO:
James A. Haley VA Medical Center (JHVAMC)

13000 Bruce B. Downs Boulevard Tampa, FL 33612

USA

1.00
09/26/2022
MARK FOR:
ATTN: Logistics

813-972-2000

ATTACHMENT 2 – SCHEDULE/SALIENTS 36C24822Q1352

Page 1 of Page 1 of

Technical Specifications Salient Characteristics Hospital Bed Mover (Brand Name or Equal)

StaminaLift, Product #TS 6000, Quantity = 1. The specifications established in this document are the minimum requirements to be accepted. Products furnished to these specifications must meet or exceed all requirements herein. Modifications or additions to basic standard products or less size or capability to meet these requirements will not be accepted. Must be able to provide single person operator.

1) Essential/significant physical, functional, or performance characteristics.

a) Compatibility – Must be compatible with bariatric beds, sand beds, and acute care beds, and stretchers.

b) Networking – N/A.

c) Power Needs- GEL batteries, DC electric motors.

d) Measurements (height, weight)- 33in. x 36in. x 48in. Lx Wx H - These dimensions exactly accommodate hallways, doorways, and elevators in the facility.

Must Comply with the relevant requirements of the Applicable Standard EN60601-1 and the following:

· Minimum lift 993 lbs. not to exceed 1000 lbs., minimum push or pull 1,985 lbs. not to exceed 1,987 lbs.

· Minimum weight of machine 440 lbs., not exceed 445 lbs.

· In Use: minimum 30in x 30 in x 45 in, not to exceed maximum 33 in. x 36 in. x 48 in.

· Sealed maintenance free GEL batteries.

· Horn (warning device) on control unit.

· Must be suitable for indoor use on a variety of surfaces.

· Able to control speed, jaw movement, and sensitivity of joystick.

· Shall provide electric locking function.

· Shall provide automatic brakes / emergency shut-down button.

· Emergency jaw release (allows override of lowering mechanism emergency brake release – and override of brake mechanism safety features to prevent unauthorized use).

· Two variable drive DC electric motors. Calcium oxide concentration greater than 16.7-17.4% (Calcium oxide batteries do not heat up).