Sensor Calibration Requirements Document
For 704th Test Group/846th Test Squadron
1 Purpose
1.1 The 704th Test Group (704 TG) mission is to be the premier provider of unique test and evaluation capabilities relevant to tomorrow’s warfighter through world-class people improving technical excellence, cost-effectiveness, and agility. The 846th Test Squadron
(846 TS) is one of the units assigned to the 704 TG. The 846 TS’s primary test capability is the Holloman High Speed Test Track (HHSTT) where rocket powered sled trains are utilized to simulate flight-like environments and scenarios with the control and recovery of a ground-based system. In order to collect data from these systems, various sensors are essential. These sensors are required for test mission validation of various avionics platforms, payloads, and equipment for the Department of Defense in conjunction with its partners over a number of differing rocket sled configurations. The 846 TS must have sensors for each mission and have them calibrated at scheduled times. The following requirements include calibration requirements for each sensor.
2 Scope
2.1 The Contractor shall deliver calibrated sensors as needed.
3 Estimated Annual Sensor Calibration Requirements:
Description MFR Part Number
Pico 64 DAS EME Pico64DAS
Pico 16 DAS EME Pico16DAS
3.1 Sensor Calibration Specifications:
3.1.1 Pico 64 DAS:
Perform the following tests:
3.1.1.1 Visual Inspection
3.1.1.2 Data Collection All Channels
3.1.1.3 Static Gain
3.1.1.4 Static Non-Linearity
3.1.1.5 Class 1000 Frequency Response Gain 1.5
3.1.1.6 Class 1000 Frequency Response Gain 300
3.1.1.7 Channel Step Phase Response
3.1.1.8 Channel Phase Delay
3.1.1.9 Static Excitation
3.1.1.10 Excitation Short 2.5 VDC
3.1.1.11 Excitation Short 5 VDC
3.1.1.12 Excitation Short 10 VDC
3.1.1.13 Input Offset
3.1.1.14 Output Offset
3.1.1.15 Shunt and Rcal Resistors
3.1.1.16 Trigger, Digital Output/Input Timing
3.1.1.17 Sample Clock Stability
Calibration performed in compliance with ISO/IEC 17025:2005
All tests and calibrations are done in accordance with traceability to National
Metrology Institue such as National Institue of Standards and Technology (NIST)
Provide calibration certificate with test results
3.1.2 Pico 16 DAS:
3.1.2.1 Perform the following tests:
3.1.2.1.1 Visual Inspection
3.1.2.1.2 Data Collection All Channels
3.1.2.1.3 Static Gain
3.1.2.1.4 Static Non-Linearity
3.1.2.1.5 Class 1000 Frequency Response Gain 1.5
3.1.2.1.6 Class 1000 Frequency Response Gain 300
3.1.2.1.7 Channel Step Phase Response
3.1.2.1.8 Channel Phase Delay
3.1.2.1.9 Static Excitation
3.1.2.1.10 Excitation Short 2.5 VDC
3.1.2.1.11 Excitation Short 5 VDC
3.1.2.1.12 Excitation Short 10 VDC
3.1.2.1.13 Input Offset
3.1.2.1.14 Output Offset
3.1.2.1.15 Shunt and Rcal Resistors
3.1.2.1.16 Trigger, Digital Output/Input Timing
3.1.2.1.17 Sample Clock Stability
Calibration perform in compliance with ISO/IEC 17025:2005
All tests and calibrations are done in accordance with traceability to National
Metrology Institue such as National Institue of Standards and Technology (NIST)
Provide calibration certificate with test results
4 Milestones
4.1 The contractor shall deliver the required sensors calibrated to the 846 TS following purchase order NLT 60 days ARO.
5 Deliverables
5.1 The contractor shall deliver all hardware and documentation to 704 TSS/XPL, Attn:
Jim Bigas at 1644 Vandergrift Road, Building 1260, Holloman Air Force Base, New
Mexico 88330-7850.