APPENDIX I.pdf
PDF 130 KB Posted
- Attached to
- Laboratory Specialty Gases
- Solicitation number
- ITB-05809
- Issued by
- Leon County, Florida
About this file
This document is an Appendix I related to a state and local contract opportunity from the Florida Department of Agriculture and Consumer Services (FDACS) for the outright purchase and delivery of pure, specialty/liquid, and industrial gases to be used in various laboratories within the Department. The contract requires ultrapure grades of helium, hydrogen, nitrogen, and air, with detailed specifications for purity, cylinder contents, and allowable contaminants. The contract covers multiple jurisdictions in Florida, including Leon County and Tampa City.
The contract seeks competitive bids for these laboratory specialty gases, with no mention of set-asides, incumbents, or funding details. Pricing terms and any other salient commercial information are not provided in the appendix.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| #25 LABORATORY SPECIALTY GASES Draft V3.pdf | ||
| Addendum 1.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
APPENDIX I
UPC Gas
1. Helium (UPC)
Purity: Ultrapure carrier - 99.9995% Cylinder contents: 291 cubic feet Cylinder valve outlet: CGA 580
Analytical applications: Must be free of all contaminants, which can be detected on gas chromatography equipped with these detectors: flame ionization; electron capture; Hall;
nitrogen-phosphorus; photoionization; mass selective. General categories of contaminants include: petroleum hydrocarbons; halogenated compounds; nitrogen or phosphorous containing compounds; industrial solvents; unsaturated hydrocarbons; components having a high electron affinity; or any other component that the laboratories consider as a contaminant in the analytical applications.
Minimum requirements:
N2 < 3.0 ppm
02 < 1.0 ppm THC < 0.5 ppm
H20 < 1.0 ppm Dew Point -105 F
2. Hydrogen (UPC)
Purity: Ultrapure carrier - 99.9995% Cylinder contents: 260 cubic feet Cylinder valve outlet: CGA 350
Analytical requirements: Must be free of the same components listed for ultrapure helium in number 1 above.
N2 < 5.0 ppm O2 < 1.0 ppm
THC < 0.5 ppm H2O < 1.0 ppm
Dew Point -105 F
3. Nitrogen (UPC)
Purity: Ultrapure carrier - 99.9993% Cylinder contents: 304 cubic feet Cylinder valve outlet: CGA 580
02 < 1.0 ppm THC < 0.5 ppm 20 < 1.0 ppm CO < 0.5 ppm
CO2 < 0.5 ppm Dew Point - 105 F
4. Air (Ultra Zero)
Purity: Ultra Zero Grade Cylinder contents: 311 cubic feet.
Cylinder valve outlet: CGA 590
O2 20 - 22% THC < 0.1 ppm H2O < 2.0 ppm *CO < 1.0 ppm *CO2 < 1.0 ppm Dew point -70 F *Combined total
5. Oxygen - (UPC)
Purity: 99.996% Cylinder Size 282 cubic ft.
H2O <1 ppm THC <0.5 ppm CO <1 ppm CO2 <1 ppm Ar <20 ppm N2 <10 ppm
Minimum requirements:
THC <0.5
H20 <2.0
6. Argon (UHP)
Purity: Ultrahigh purity / Zero grade - 99.999% Cylinder contents: 337 cubic feet Cylinder valve outlet: CGA580
Analytical requirements: Analytical requirements: Must contain no trace metal contaminants that might interfere with trace metals analysis. Shall contain no detectable traces of the following: Hg, Al, Ba, Be, Sb, Tl, Fe, Zn, Cd, Cr, Pb, Cu, Co, Mn, Ni, Ca, Mg, K, Na, Sr, As, Se, V, Ti, Mo.
THC < 0.5 ppm O2 < 4.0 ppm H2O <3.5 ppm
7. Atomic Absorption Acetylene
Purity: 99.6% Total Sulfurs <15 ppm Phosphine <10 ppm
8. Research Carbon Dioxide
Purity: 99.999% Ar+O2+CO <1 ppm H2O <3 ppm THC <1 ppm N2 < 5 ppm
Specialty/Liquid Gas
9. Liquid Argon (Dewar=s)
Purity - 99.997% Cylinder contents: 180 L high-pressure Dewar Cylinder valve outlet: CGA 580-valve outlet for high-pressure service, pressure regulator set at 135 psi with a 235 psi pressure relieve valve.
CO2 < 1 ppm N2 <20 ppm O2 <5 ppm THC <1 ppm H2O <10.5 ppm Dew Point -76 degrees F
10. Liquid Nitrogen
Purity - 99.998% Cylinder size: 180 L low pressure (below 30 psi) Dewar on casters.
Cylinder valve outlet: 1 / 2" ODTx45 flair fitting with 22 psi liquid control valve for liquid service.
11. 5% UHP Methane/ 95% UHP Argon (P5)
Cylinder contents: 223 cubic feet Cylinder valve outlet: CGA 350
Analytical requirements: Must be suitable for ECD analysis.
12. Liquid Nitrogen 10L Transfill.
Industrial Gas
13. CO2 Industrial Grade - Dip Tube
14. CO2 Industrial Grade- No Dip Tube
15. Industrial Nitrogen
Purity: 99.998%
O2 <1 ppm H2O <4 ppm Dew Point -90 degrees F
16. Industrial Argon
Purity: 99.997% H2O < 10.5 ppm CO2 < 1 ppm N2 < 20 ppm O2 < 5 ppm THC < 1 ppm Dew Point -76 degrees F
File details come from the government source that posted it. Updated .