APPENDIX I.pdf

PDF 130 KB Posted

Attached to
Laboratory Specialty Gases
Solicitation number
ITB-05809
Issued by
Leon County, Florida

About this file

This document is an Appendix I related to a state and local contract opportunity from the Florida Department of Agriculture and Consumer Services (FDACS) for the outright purchase and delivery of pure, specialty/liquid, and industrial gases to be used in various laboratories within the Department. The contract requires ultrapure grades of helium, hydrogen, nitrogen, and air, with detailed specifications for purity, cylinder contents, and allowable contaminants. The contract covers multiple jurisdictions in Florida, including Leon County and Tampa City.

The contract seeks competitive bids for these laboratory specialty gases, with no mention of set-asides, incumbents, or funding details. Pricing terms and any other salient commercial information are not provided in the appendix.

View the file

Other files

File Type Posted
#25 LABORATORY SPECIALTY GASES Draft V3.pdf PDF
Addendum 1.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

APPENDIX I

UPC Gas

1. Helium (UPC)

Purity: Ultrapure carrier - 99.9995% Cylinder contents: 291 cubic feet Cylinder valve outlet: CGA 580

Analytical applications: Must be free of all contaminants, which can be detected on gas chromatography equipped with these detectors: flame ionization; electron capture; Hall;

nitrogen-phosphorus; photoionization; mass selective. General categories of contaminants include: petroleum hydrocarbons; halogenated compounds; nitrogen or phosphorous containing compounds; industrial solvents; unsaturated hydrocarbons; components having a high electron affinity; or any other component that the laboratories consider as a contaminant in the analytical applications.

Minimum requirements:

N2 < 3.0 ppm

02 < 1.0 ppm THC < 0.5 ppm

H20 < 1.0 ppm Dew Point -105 F

2. Hydrogen (UPC)

Purity: Ultrapure carrier - 99.9995% Cylinder contents: 260 cubic feet Cylinder valve outlet: CGA 350

Analytical requirements: Must be free of the same components listed for ultrapure helium in number 1 above.

N2 < 5.0 ppm O2 < 1.0 ppm

THC < 0.5 ppm H2O < 1.0 ppm

Dew Point -105 F

3. Nitrogen (UPC)

Purity: Ultrapure carrier - 99.9993% Cylinder contents: 304 cubic feet Cylinder valve outlet: CGA 580

02 < 1.0 ppm THC < 0.5 ppm 20 < 1.0 ppm CO < 0.5 ppm

CO2 < 0.5 ppm Dew Point - 105 F

4. Air (Ultra Zero)

Purity: Ultra Zero Grade Cylinder contents: 311 cubic feet.

Cylinder valve outlet: CGA 590

O2 20 - 22% THC < 0.1 ppm H2O < 2.0 ppm *CO < 1.0 ppm *CO2 < 1.0 ppm Dew point -70 F *Combined total

5. Oxygen - (UPC)

Purity: 99.996% Cylinder Size 282 cubic ft.

H2O <1 ppm THC <0.5 ppm CO <1 ppm CO2 <1 ppm Ar <20 ppm N2 <10 ppm

Minimum requirements:

THC <0.5

H20 <2.0

6. Argon (UHP)

Purity: Ultrahigh purity / Zero grade - 99.999% Cylinder contents: 337 cubic feet Cylinder valve outlet: CGA580

Analytical requirements: Analytical requirements: Must contain no trace metal contaminants that might interfere with trace metals analysis. Shall contain no detectable traces of the following: Hg, Al, Ba, Be, Sb, Tl, Fe, Zn, Cd, Cr, Pb, Cu, Co, Mn, Ni, Ca, Mg, K, Na, Sr, As, Se, V, Ti, Mo.

THC < 0.5 ppm O2 < 4.0 ppm H2O <3.5 ppm

7. Atomic Absorption Acetylene

Purity: 99.6% Total Sulfurs <15 ppm Phosphine <10 ppm

8. Research Carbon Dioxide

Purity: 99.999% Ar+O2+CO <1 ppm H2O <3 ppm THC <1 ppm N2 < 5 ppm

Specialty/Liquid Gas

9. Liquid Argon (Dewar=s)

Purity - 99.997% Cylinder contents: 180 L high-pressure Dewar Cylinder valve outlet: CGA 580-valve outlet for high-pressure service, pressure regulator set at 135 psi with a 235 psi pressure relieve valve.

CO2 < 1 ppm N2 <20 ppm O2 <5 ppm THC <1 ppm H2O <10.5 ppm Dew Point -76 degrees F

10. Liquid Nitrogen

Purity - 99.998% Cylinder size: 180 L low pressure (below 30 psi) Dewar on casters.

Cylinder valve outlet: 1 / 2" ODTx45 flair fitting with 22 psi liquid control valve for liquid service.

11. 5% UHP Methane/ 95% UHP Argon (P5)

Cylinder contents: 223 cubic feet Cylinder valve outlet: CGA 350

Analytical requirements: Must be suitable for ECD analysis.

12. Liquid Nitrogen 10L Transfill.

Industrial Gas

13. CO2 Industrial Grade - Dip Tube

14. CO2 Industrial Grade- No Dip Tube

15. Industrial Nitrogen

Purity: 99.998%

O2 <1 ppm H2O <4 ppm Dew Point -90 degrees F

16. Industrial Argon

Purity: 99.997% H2O < 10.5 ppm CO2 < 1 ppm N2 < 20 ppm O2 < 5 ppm THC < 1 ppm Dew Point -76 degrees F