Amendment to a Previous Combined Solicitation

*= Required Field Amendment to a Previous Combined Solicitation

SUBJECT* Intra-Aortic pumps

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE* 78240

SOLICITATION NUMBER* 36C25723Q0708

RESPONSE DATE/TIME/ZONE 04-20-2023 11PM CENTRAL TIME, CHICAGO,

USA

ARCHIVE 5 DAYS AFTER THE RESPONSE DATE

RECOVERY ACT FUNDS N

SET-ASIDE

PRODUCT SERVICE CODE* 6515

NAICS CODE* 339112

CONTRACTING OFFICE ADDRESS Department Of Veterans Affairs

Network Contracting Office 17

5441 Babcock Road Ste. 302

San Antonio TX 78240

POINT OF CONTACT*

Contracting Specialist

Matt Lee matthew.lee5@va.gov

PLACE OF PERFORMANCE

ADDRESS Olin E. Teague Veterans Medical Center

1901 Veterans Memorial Drive

Attn: Medical Service

Temple TX

POSTAL CODE 76504

COUNTRY US

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS matthew.lee5@va.gov

EMAIL DESCRIPTION Business

DESCRIPTION

Amendment 36C25723Q0708 P0001 - Intra-Aortic pumps

1. Adding Technical Characteristics, which are below:

TECHNICAL CHARACTERISTICS

Brand Name Item:

Teleflex AC3 Optimus Intra-Aortic Balloon Pump

DIMENSIONS

Control Module with Monitor

Height: ≤ 11” (27.94 cm)

Width: ≤ 15” (38.1 cm)

Depth: ≤ 2” (5.08 cm)

Pneumatic Drive Unit

Height: ≤ 35” (84.6 cm)

Width: ≤ 13” (33.02 cm)

Depth: ≤ 25” (63.5 cm)

WEIGHT

Control Module: ≤ 6 lbs. (2.7 kg)

Pneumatic Unit: ≥ 100 lbs. (45.36 kg)

Total Weight: ≤ 106 lbs. (48.08 kg)

DISPLAY

• Color touchscreen display

• ≥ 12” (30.48 cm)

• ≥ 3-channel multicolor waveforms

SYSTEM MODES

Autopilot/Operator

PNEUMATICS

• Drive system

• Drive gas

• Pumping volume ≤ 0.5cc increments

CONTINUOUS CONDENSATION REMOVAL WITHOUT COUNTERPULSATION

INTERRUPTION

Essential/significant physical, functional, or performance characteristics.

• Mobility – Must be on a stand (Kiosk) and movable and manageable by a single person.

• Touch Screen Access

• Balloon Volume Adjustment 0.5cc increments

• Alarms

• Alarm Priority distinction

• A visual Alarm with 360 degree view

• Status Reports

• Graphics

• Alarm History Report

• Continuous analysis of 6 or more ECG sources

• Multiple Trigger modes

• Supports Heart rates as high as 200 bpm

• Rechargeable Battery Reserve

• Power – Must be able to run on standard 120v outlet

2. No other changes have been made