Amendment 36C25723Q0708 0001.pdf
PDF 109 KB Posted
- Attached to
- 6515--Olin E. Teague Veterans Medical Center Medical Service is seeking to purchase Intra-Aortic pumps
- Solicitation number
- 36C25723Q0708
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25723Q0708 0001_1.docx | DOCX document | |
| 36C25723Q0708_1.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Amendment to a Previous Combined Solicitation
*= Required Field Amendment to a Previous Combined Solicitation
SUBJECT* Intra-Aortic pumps
GENERAL INFORMATION
CONTRACTING OFFICE’S ZIP CODE* 78240
SOLICITATION NUMBER* 36C25723Q0708
RESPONSE DATE/TIME/ZONE 04-20-2023 11PM CENTRAL TIME, CHICAGO,
USA
ARCHIVE 5 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS N
SET-ASIDE
PRODUCT SERVICE CODE* 6515
NAICS CODE* 339112
CONTRACTING OFFICE ADDRESS Department Of Veterans Affairs
Network Contracting Office 17
5441 Babcock Road Ste. 302
San Antonio TX 78240
POINT OF CONTACT*
Contracting Specialist
Matt Lee matthew.lee5@va.gov
PLACE OF PERFORMANCE
ADDRESS Olin E. Teague Veterans Medical Center
1901 Veterans Memorial Drive
Attn: Medical Service
Temple TX
POSTAL CODE 76504
COUNTRY US
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
AGENCY CONTACT’S EMAIL ADDRESS matthew.lee5@va.gov
EMAIL DESCRIPTION Business
DESCRIPTION
Amendment 36C25723Q0708 P0001 - Intra-Aortic pumps
1. Adding Technical Characteristics, which are below:
TECHNICAL CHARACTERISTICS
Brand Name Item:
Teleflex AC3 Optimus Intra-Aortic Balloon Pump
DIMENSIONS
Control Module with Monitor
Height: ≤ 11” (27.94 cm)
Width: ≤ 15” (38.1 cm)
Depth: ≤ 2” (5.08 cm)
Pneumatic Drive Unit
Height: ≤ 35” (84.6 cm)
Width: ≤ 13” (33.02 cm)
Depth: ≤ 25” (63.5 cm)
WEIGHT
Control Module: ≤ 6 lbs. (2.7 kg)
Pneumatic Unit: ≥ 100 lbs. (45.36 kg)
Total Weight: ≤ 106 lbs. (48.08 kg)
DISPLAY
• Color touchscreen display
• ≥ 12” (30.48 cm)
• ≥ 3-channel multicolor waveforms
SYSTEM MODES
Autopilot/Operator
PNEUMATICS
• Drive system
• Drive gas
• Pumping volume ≤ 0.5cc increments
CONTINUOUS CONDENSATION REMOVAL WITHOUT COUNTERPULSATION
INTERRUPTION
Essential/significant physical, functional, or performance characteristics.
• Mobility – Must be on a stand (Kiosk) and movable and manageable by a single person.
• Touch Screen Access
• Balloon Volume Adjustment 0.5cc increments
• Alarms
• Alarm Priority distinction
• A visual Alarm with 360 degree view
• Status Reports
• Graphics
• Alarm History Report
• Continuous analysis of 6 or more ECG sources
• Multiple Trigger modes
• Supports Heart rates as high as 200 bpm
• Rechargeable Battery Reserve
• Power – Must be able to run on standard 120v outlet
2. No other changes have been made
File details come from the government source that posted it. Updated .