CLIN 001 Chromaster 5110 Pump w/ Auto Purge Valve, Low Pressure Gradient Kit, Degassing Unit and UI Pad

HPLC quaternary operation with flow range of 10ul/min to 9.99 ml/min

HPLC Pump Auto-purge valve

6-channel degassing

Gradient solvent mixing using double-frequency pulsing for low system dwell volumes

Built-in plunger washing mechanism

Built-in leak sensors in each: pump, autosampler, column oven and detector

Dead volume 200ul

All maintenance parts, including light sources, are accessible from front panels

Accuracy +/- 1% or +/- 2 μL/min, whichever is greater

Precision +/- 0.075 % RSD or 0.02 min SD

Operational from approx 0.8 MPa (approximately 120 psi) to maximum pressure 40 MPa (0.001 to 5.000 mL/min), 20 MPa (5.001 to 9.999 mL/min)

GLP Functions: total flow rate, speed error, proportioning valve count, mixer count, auto purge count, plunger wash time

Low Pressure Gradient Kit

Degassing Unit 6-Channel

UI Pad

CLIN 002 Chromaster 5310 Column Oven and UI Pad

Column oven capable of ambient minus 15C to ambient plus 60C

Accuracy: +/- 1 oC

Temp Control Precision: SD 0.2 oC

Ambient and oven temperature recording

Column oven capable of accommodating 3 x 300mm columns

Includes safety vapor/leak sensor

UI Pad

CLIN 003 Chromaster Autosampler-injector Mod and UI Pad

Capacity: 120 vials x1.5 mL (standard), 195 x 1mL, 72 x 4mL, 2 x MTP

Injection volume: 0.1 μL to 50 μL (100uL loop, standard), Sample to sample carryover < 0.003% (cut method)

2 wash solvents for autosampler injector system, Loop Sample Injection (0.003%, cut method)

Injection volume precision: 0.2% RSD (at 10-μL cut injection)

Injection linearity Corr. Coeff. =0.999

Autosampler programmable needle wash sequence

UI Pad

Clear 1.5 ml HPLC Vials, caps and pre-slit PTFE/silicon septa

CLIN 004 Chromaster UV Detector Optics: Double beam ratio photometric system and UI Pad

Built-in Hg lamp for easy and accurate wavelength calibration, Front panel access

Noise: < 0.5 x 10-5 AU

Drift: < 0.1 x 10-4 AU/h

Wavelength Range: 190 to 600 nm

Wavelength accuracy: +/- 1nm

Spectral bandwidth: 6nm

Noise: <0.5 x 10-5 AU at 250nm

Drift: <1.0 x 10-4 AU/h at 250nm

Light Source: D2 lamp, Hg lamp for checking wavelength; Front panel access for source maintenance

Availability of optional temperature controlled flow cell, 4-30°C

GLP functions: Auto calibration using Hg lamp, Lamp energy check, Log for lamp usage, Lamp replacement date, Serial number registration in unit memory

UI Pad

CLIN 005 Chromaster Interface Control Board and Accessories

Control Board integrated for 5210 Autosampler

Cable, USB A to B, M to M 6ft

Chromaster Organizer – Solvent Bottle Organizer

(6) Solvent Bottle – 1 Liter

(6) Cap for HPLC Solvent Bottles

CLIN 006 OpenLab Workstation Software for Hitachi HPLC Systems

Core Software

LC Control License

CLIN 007 Instrument/Software Installation

Includes Instrument/Software Installation and Basic Training

CLIN 008 Shipping/Handling/Freight charges if applicable for delivery of items to:

USDA APHIS, 1920 Dayton Ave., Ames, IA 50010