Text version
CLIN 001 Chromaster 5110 Pump w/ Auto Purge Valve, Low Pressure Gradient Kit, Degassing Unit and UI Pad
HPLC quaternary operation with flow range of 10ul/min to 9.99 ml/min
HPLC Pump Auto-purge valve
6-channel degassing
Gradient solvent mixing using double-frequency pulsing for low system dwell volumes
Built-in plunger washing mechanism
Built-in leak sensors in each: pump, autosampler, column oven and detector
Dead volume 200ul
All maintenance parts, including light sources, are accessible from front panels
Accuracy +/- 1% or +/- 2 μL/min, whichever is greater
Precision +/- 0.075 % RSD or 0.02 min SD
Operational from approx 0.8 MPa (approximately 120 psi) to maximum pressure 40 MPa (0.001 to 5.000 mL/min), 20 MPa (5.001 to 9.999 mL/min)
GLP Functions: total flow rate, speed error, proportioning valve count, mixer count, auto purge count, plunger wash time
Low Pressure Gradient Kit
Degassing Unit 6-Channel
UI Pad
CLIN 002 Chromaster 5310 Column Oven and UI Pad
Column oven capable of ambient minus 15C to ambient plus 60C
Accuracy: +/- 1 oC
Temp Control Precision: SD 0.2 oC
Ambient and oven temperature recording
Column oven capable of accommodating 3 x 300mm columns
Includes safety vapor/leak sensor
UI Pad
CLIN 003 Chromaster Autosampler-injector Mod and UI Pad
Capacity: 120 vials x1.5 mL (standard), 195 x 1mL, 72 x 4mL, 2 x MTP
Injection volume: 0.1 μL to 50 μL (100uL loop, standard), Sample to sample carryover < 0.003% (cut method)
2 wash solvents for autosampler injector system, Loop Sample Injection (0.003%, cut method)
Injection volume precision: 0.2% RSD (at 10-μL cut injection)
Injection linearity Corr. Coeff. =0.999
Autosampler programmable needle wash sequence
UI Pad
Clear 1.5 ml HPLC Vials, caps and pre-slit PTFE/silicon septa
CLIN 004 Chromaster UV Detector Optics: Double beam ratio photometric system and UI Pad
Built-in Hg lamp for easy and accurate wavelength calibration, Front panel access
Noise: < 0.5 x 10-5 AU
Drift: < 0.1 x 10-4 AU/h
Wavelength Range: 190 to 600 nm
Wavelength accuracy: +/- 1nm
Spectral bandwidth: 6nm
Noise: <0.5 x 10-5 AU at 250nm
Drift: <1.0 x 10-4 AU/h at 250nm
Light Source: D2 lamp, Hg lamp for checking wavelength; Front panel access for source maintenance
Availability of optional temperature controlled flow cell, 4-30°C
GLP functions: Auto calibration using Hg lamp, Lamp energy check, Log for lamp usage, Lamp replacement date, Serial number registration in unit memory
UI Pad
CLIN 005 Chromaster Interface Control Board and Accessories
Control Board integrated for 5210 Autosampler
Cable, USB A to B, M to M 6ft
Chromaster Organizer – Solvent Bottle Organizer
(6) Solvent Bottle – 1 Liter
(6) Cap for HPLC Solvent Bottles
CLIN 006 OpenLab Workstation Software for Hitachi HPLC Systems
Core Software
LC Control License
CLIN 007 Instrument/Software Installation
Includes Instrument/Software Installation and Basic Training
CLIN 008 Shipping/Handling/Freight charges if applicable for delivery of items to:
USDA APHIS, 1920 Dayton Ave., Ames, IA 50010