Advantage Plus Cut Sheet.pdf

PDF 459 KB Posted

Attached to
N065--SL-23-012-NY Medivator Site Prep at Manhattan
Solicitation number
36C24224Q0050
Issued by

About this file

This document contains a technical drawing and specifications for an Advantage Plus Endoscope Reprocessing System. The drawing provides plumbing, electrical, and ventilation details for the system, including connections for water, air, drainage, and optional components such as an air compressor and vapor management system. Related federal contract opportunity N065-SL-23-012-NY seeks site preparation services at a Manhattan Veterans Affairs facility to install an Advantage Plus Endoscope Reprocessing System as specified in the technical drawing. The solicitation was issued by the Department of Veterans Affairs Veterans Health Administration VISN 2 for an unknown response date or award amount.

View the file

Other files

File Type Posted
36C24224Q0050 0002.docx DOCX document
36C24224Q0050 0001.docx DOCX document
36C24224Q0050_3.docx DOCX document
Appendix C - Advantage Site Requirements 50098 1520 EN US Rev C.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Drain connection8

Water connection9

Air connection10

Electrical connection11

Network connection12

Vent connection13

Vapor management (Optional)16

Sanitize connection14

Air Compressor (Optional)15 -Caster or Rubber Feet included

ADVANTAGE PLUS

Endoscope Reprocessing System

TM

For clarification and/or questions please contact Technical Support at 1.800.444.4729.

2019 Revit and CAD file available at www.endoinfectionprevention.com

Water shut off valve

Bypass valve for testing and warming water supply

Temp/Pressure combo gauge

Mixing valve

Pressure gauge

Pressure regulator

Line to main drain

(See Appendix E)

(See Appendix D)

A

B

*1 8"

Plumbing Details per machine

A

CANTEL supplied Pre-Filter Assembly

B

.5

34"

**38"

45"

2” Minimum drain inlet

4"

(Optional)

Important Callouts

(Door Swing)

33"

16"

Back

.5

19.6"

Air Compressor

Caster Rubber

.2

31"17”

***36"

Side (Optional)

31"

38"

Vapor Management

Opti on al

*18” Maximum height for proper drainage (Gravity Drain).

**Wall space needed for Assembly.

***5” Mininum in back for plumbing.

Filtration Unit is 6” in depth.