Text version
A09-144: Microfabricated Mass Spectrometer for Near Real-Time Toxic Chemical Detection
PROJECTED CMMC LEVEL REQUIREMENT:
N/A
TECHNOLOGY AREAS:
Chem Bio Defense | Human Systems
MODERNIZATION PRIORITIES:
KEYWORDS:
mass spectrometry; microfabrication; detection; chemical threats; toxic industrial chemicals; personal protective equipment; air
ADDITIONAL INFORMATION
OBJECTIVE:
Develop a microfabricated mass spectrometer (MS) for use in (1) determining permeation rates through personal protective garments and gear and for (2) hand-portable field detection of toxic volatile and semi-volatile organic chemical threats.
DESCRIPTION:
This Small Business Innovation Research (SBIR) project is aimed at a creative novel approach to develop a small mass spectrometer (MS) based on microfabrication techniques. There is a need to develop an MS that can be easily and widely applied to unique chemical detection problems, such as swatch testing and field operations.
Such a system must be small, inexpensive, robust, and high performing. In order to provide these desirable characteristics, the fabrication methods for the MS analyzer must be simple, highly precise, and low cost. Planar microfabrication techniques offer the potential for satisfying this need. Development and miniaturization of integrated circuits for advanced computation was one of the most significant scientific advances of all time. There are three fundamental design objectives that guided this advance (reference 1): (1) use thin film processes; (2) keep active components like transistors on one level, at the surface of the substrate; and (3) use the most robust material system available. Analytical chemical instrumentation has benefitted greatly by this advance, mostly at the electronic and computer component level. Additional advanced capabilities are possible by using this approach to design and fabricate the measurement components of analytical instrumentation, and the trend has begun by developments in microfluidics such as microchip capillary electrophoresis, microchip gas chromatography, and microchip liquid chromatography (reference 2). Preliminary efforts in microfabrication of MS analyzers have recently been reported, however, significant innovation and effort is needed for these analyzers to reach their future potential (reference 3). By using microfabrication techniques, it should be possible to design a miniature MS which would operate in the tandem MS (MS/MS) mode in a single device and operate at higher pressures than possible in larger mass spectrometers. Since fouling of electrodes with use is common to all MS designs, and usually determines the frequency of service, the development of throw-away analyzers, which is not unreasonable when based on microfabrication, is extremely attractive. There are many applications in which instrument size is critical. Obviously, small size is critical for portability, but it is also important when instrumentation must be operated in confined spaces, such as in hoods, and when multiple instruments must be stacked for parallel operation. Furthermore, certain stationary operations require frequent movement of the equipment from one area to another for interval testing. All of these applications could benefit from a new MS analyzer design that is small and inexpensive to fabricate, robust during movement, and easy to remove for servicing and replacement, while still maintaining excellent MS and MS/MS performance. New microfabrication technology should minimize adsorptive surfaces to which analytes could be exposed, greatly reducing the chance for ion-molecule reactions and lengthening the time between cleanings when used as a detector in test equipment, such as the Swatch, Emergency Responder Toxic Industrial Chemicals (ERTIC), Mask, and Glove & Boot test fixtures and when used for detection of chemical threats in the field. This SBIR topic seeks a novel MS design based on microfabrication techniques to produce an inexpensive, but high performance mass analyzer. The technology should provide simpler, less labor-intensive requirements for cleaning and parts replacement, leading to significantly reduced costs. While feasible, there is significant risk involved in developing a high performance microfabricated MS analyzer. The electrical fields for ion manipulation (e.g., trapping, focusing, and scanning) must be carefully designed and implemented; electrical shielding is especially important for miniaturize analyzers.
Materials that are robust under continuous use for extended periods of time must be selected for use as substrates and coatings for patterning of thin-film electrodes and feed-throughs. Novel approaches for ionizing samples and detecting ions with high efficiency are also needed. In addition, low power consumption is critical for field operations.
PHASE I:
Perform a feasibility study to develop and demonstrate key performance features of novel microfabricated MS instrumentation for accurate, real-time (or near real-time) detection and identification of volatile and semi-volatile toxic organic chemicals in swatch test fixtures. Generate a detailed design of the new MS utilizing microfabrication methods. Report study findings which will include feasibilities of performance features, detailed design, and microfabrication methods.
PHASE II:
Construct a microfabricated MS prototype as designed in Phase I and demonstrate its applicability for detecting target analytes according to the specifications outlined in this topic description. The new MS prototype should demonstrate detection limits better than 100 pg (S/N = 3), and be able to continuously monitor from 0.4 mg/m3 to 200 mg/m3 (direct without split) of target analytes in an air stream. The new MS prototype should also be capable of monitoring an air stream in real-time. This sensitivity and unit mass resolution up to 450 m/z should be maintained for at least 1000 hours of continuous operation.
PHASE III DUAL USE APPLICATIONS:
Deliver a working system to DPG for testing on Swatch, ERTIC, Mask, and Glove & Boot test fixtures. Assist in integrating the new MS in Swatch, ERTIC, Mask, and Glove & Boot fixtures for determining permeation rates of toxic chemical threats. Furthermore, demonstrate the application of the new MS in a hand-portable system for field detection of chemical threats. Provide training for personnel in both application areas. PRIVATE SECTOR COMMERCIAL POTENTIAL/DUAL-USE APPLICATIONS: Private sector applications of the developed technology include (1) use for stationary or mobile routine monitoring of workplace and field environments and (2) use by first responders for field detection of toxic chemicals, solvents involved in arson, drugs of abuse, and explosives, as well as detection of volatile and semi-volatile chemicals associated with agricultural, food quality, biomedical, and environmental issues.
REFERENCES:
Janasek, D.; Franzke, J.; Manz, A. Nature 2006, 442, 374-380.1.
Vilkner, T.; Janasek, D.; Manz, A. Anal. Chem. 2004, 76, 3373-3386.2.
May, G.S.; Sze, S,M. Fundamentals of Semiconductor Fabrication, Wiley, New York, 2004.3.
TECHNICAL POINT OF CONTACT (TPOC):
None
File details come from the government source that posted it. Updated .