Text version

Salient Characteristics S21058

MICROBIOLOGY ANALYZER

SALIENT CHARACTERISTICS:

1 EA, Analyzer:

1. Shall be capable of (1) performing automatic, sample collection and microbiological analysis and (2) sending data and alerts in real-time.

2. Shall be capable of being controlled remotely and wirelessly via a smart phone or a web interface.

3. Shall be field-ruggedized weighing no more than 25lbs.

4. Shall have a response time no less than 1 hour.

5. Shall use reagents.

6. Shall be capable of obtaining E. coli, Total Coliforms, or Enterococci measurements through internal vacuum sampling and optical detection.

7. Shall have measurement range of at least 4 x10 Colony Forming Units per 100 mL.

8. Shall be able to take at least 7 measurements on one charge.

9. Shall use GSM network for programing and data management.

i) Shall allow the user to start measurements via cell phone either manually or through pre-programming.

10. Shall include a Li lon battery.

11. Shall have integrated internal antenna.

1 EA, Annual Web Based Service:

12. Shall be compatible with analyzer specs above.

13. Shall allow for annual subscriptions.

147 EA, Disposable Reagent Cartridges:

14. Shall mix an incubate samples for analyzer, specs above.

1 EA, Extra Battery:

15. Shall be compatible with analyzer specs above.

1 EA, Training:

16. Shall be at least 1 day on site training or online video conference, TBD.

1 EA, Factory Maintenance:

17. Shall be for one (1) annual factory maintenance.

18. Customer can send directly to factory for maintenance and upgrade of all components that may have degraded from normal use.

19. Vendor shall verify device functionalities, the integrity of the mechanical components, the sampling pump, changing all vacuum tubing inside the device, changing all o-rings and seals, and upgrade firmware or hardware if applicable, one (1) year after equipment is received by customer.