PERFORMANCE WORK STATEMENT (PWS)
HPGE Detector Repair/Reconfiguration and ISOCS/LabSOCS Characterization, Radiological Controls Office Code 105.5 Radiation Health
JML 0107-1540
1. GENERAL: This is a non-personal services contract to provide Service Level 4 – DETSERV-4 HPGE detector repair and reconfiguration. The Government will not exercise any supervision or control over the contract service providers performing the services herein. Such contract service providers shall be accountable solely to the contractor who, in turn, is responsible to the Government.
1.1. Description of Services/Introduction: The contractor shall provide all personnel, equipment, supplies, facilities, transportation, tools, materials, supervision, and other items and non-personal services necessary to perform Service Level 4 – DETSERV-4 HPGE detector repair and reconfiguration as defined in this PWS except for those items specified as Government furnished items and support, if any. The contractor shall perform to the standards in this contract.
1.2. Background: PSNS & IMF uses High Purity Germanium (HPGE) detectors to support sample and item analyses for radioactivity. Detectors require periodic maintenance and repair to keep its reliability and accuracy. Its maintenance and repair can only be performed by the manufacturer hence the requirement for contractual support.
| 1.3. Scope: Contractor will provide Service Level 4 – DETSERV-4 HPGE detector repair and reconfiguration including a new slimline cryostat assembly, LMR chamber assembly, detector end cap, iPA-SL 10 preamp, and RC Cold Preamp assembly. Additionally, the contractor will convert the detector(s) to electrical cooling (from liquid nitrogen cooling) by incorporating the Cryo-Pulse 5 Plus controller. After the repair and reconfiguration is complete, the contractor will perform an ISOCS/LabSOCS Characterization – ISOXCAL to two each HPGE Detector(s), Serial | Number(s) 01069076 and 11059052 to bring them up to manufacturer’s optimum operating parameters. Finally, the contractor will provide two detector lift(s) for using the reconfigured detector(s) in a standard counting cave. | |
| 1.4. Objective: To bring up two each HPGE Detector(s), Serial | Number(s) 01069076 and 11059052 to thier optimum manufacturer’s operating parameters and convert to electrical cooling. |
1.5. Period of performance: 90 days after contract award date.
1.6. Place of performance: Contractor Facility
Mirion Technologies (Canberra) Inc.
800 Research Parkway Meriden, Connecticut 06450 Phone: (800) 255-6370
2. DEFINITIONS:
Not Applicable
3. GOVERNMENT FURNISHED ITEMS AND SERVICES
3.1 Government-furnished property: Government will ship two each HPGE Detector(s), Serial Number(s) 01069076 and 11059052 to manufacturer. The detector(s) are currently located at Bldg. 78 Code 105.55 Environmental Lab, once contract is in place; contact David Olson to ship detector(s) to vendor.
4. CONTRACTOR FURNISHED ITEMS AND RESPONSIBILITIES
4.1 Delivery: Contractor shall return two each HPGE Detector(s), Serial Number(s) 01069076 and 11059052 that are fully functional (i.e. performing at OEM optimum operating parameters and electrically cooled) to following address after completion of tasks described in this PWS:
| PSNS & IMF |
| RECEIVING OFFICER, BLDG 514 |
| 1400 FARRAGUT AVENUE |
| BREMERTON, WA 98314-5001 |
| ATTN: JEFF FOCHT C/2380.4R |
| PHONE: 360-476-3520, CELL 360-340-2766 |
| **ADDITIONAL REQUIREMENTS.** | ||||
| PRIOR TO SHIPPING, VENDOR WILL MAKE VOICE CONTACT VIA | PHONE WITH JEFF FOCHT OR HIS REPRESENTATIVE AT (360)476-3520 | PSNS&IMF CODE 2380.4R WITH DATE OF SHIPMENT AND SHIPMENT | TRACKING INFORMATION (I.E. UPS/FEDEX TRACKING NUMBERS). A | PHONE MESSAGE WILL NOT BE CONSIDERED VOICE CONTACT. |
5. SPECIFIC TASKS:
5.1 Service Level 4 – DETSERV-4- Detector Repair and Reconfiguration:
5.1.1 Diagnosis and restoration;
5.1.2 Element repair (as required);
5.2 Replacement of Detector Components
5.2.1 New end cap – Std Al 3.0” x 5.75” Lg
5.2.2 Slimline Cryostat Assembly
5.2.3 LMR Chamber Assembly
5.2.4 iPA-SL 10 Preamp
5.2.5 RC Cold Preamp Assembly
5.2.6 Cryo-Pulse 5 Plus
5.3 ISOCS Characterization
5.3.1 ISOXCAL
5.3.2 ISOCS Check source.
5.4 Detector Lift
5.4.1 Detector Lift to allow use in a standard cave with CP-5 Unit