Specifications for the purchase of a MagNA Pure 96 System are as follows:

1) Shall be a benchtop standalone instrument with a separate control unit.

2) Must be able to support from 1 to 96 reactions per run.

3) Must have a sample input volume of 50 μL to 1 mL and up to 4 mL for a plasma sample.

4) Must have an elution volume of 50 to 200 μL.

5) Must have kits that are able to be re-used up to 28 days after first use.

6) Must be able to support the following sample types: whole blood, plasma, serum, fresh-frozen tissue, FFPE tissue, cultured cells, urine, swabs, sputum, CSF, BAL, stool.

7) Shall have kits, as well as the consumables, that are barcoded and read by the MagNA Pure onboard scanner for inventory check, loading errors, and process documentation.

8) Must be able to isolate DNA, RNA, and viral nucleic acids from a wide range of starting materials.

9) Must have the ability to support pre-programmed protocols tailored to different sample requirements.

10) Must offer regent kits that are prefilled and ready-to-use trays that can be easily loaded into racks.

11) Must provide a typical run time of less than 1hour for 96 standard volume samples.

12) Must have MagNA Pure 96 Software that provides all instructions to operate the system, such as sample loading to run start.

13) Must have MagNA Pure 96 Software that is able to detect loading errors and provide information to the user for the appropriate corrective action.

14) Must have MagNA Pure 96 software with audit capabilities.

15) Must be able to process parallel handling of all samples with 2 robotic arms.

16) Must be able to produce qPCR results from MagNA Pure 96 extracted samples that are highly reproducible (CV <2%) for intra-run and inter-run, and show reagents' lot-to-lot consistency.

17) Must be able to support Simplified sample preparation for dramatic reduction of handling errors.

18) The MagNA Pure 96 system must have preloaded protocols for a broad range of sample types.

19) Delivery Charges included.