This document is a Minimal Essential Characteristics (MEC) specification for the United States Army Medical Research Institute for Infectious Diseases (USAMRIID). The Facilities Management Office is requesting two Aerosol Photometers with specific technical requirements, including precise dimensional constraints (max 33 cm x 25.7 cm x 15.2 cm), weight limitations (max 8.6 kg), electrical specifications (100-250V AC), and comprehensive performance standards. The photometers must meet rigorous technical specifications such as automatic zero reading, selectable alarm types, precise flow control (1 cfm +/- 10%), and compatibility with multiple reagent settings.

The equipment must adhere to multiple industry standards including ISO 14644-3, NSF-49, and various electrical safety certifications (CE, EN 61010-1:2010). The system requires specific technical performance capabilities, including a dynamic range from 0.0001 mg/m3 to 120 mg/m3, 1% reading accuracy, and 0.5% repeatability. Delivery is specified for the USAMRIID facility at 1425 Porter Street, Fort Detrick, MD 21702, with the delivery coordination details to be determined. The opportunity is currently in a pre-solicitation phase under the Defense Health Agency, with the solicitation number HT9426-26-RFI-0001.

View the file

Other files

File Type Posted
HT9426-26-RFI-0001.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

United States Army Medical Research Institute for Infectious Diseases (USAMRIID) Minimal Essential Characteristics (MEC) The Facilities Management Office within USAMRIID requests two (2) Aerosol Photometers meeting the following MECs:

1. The system dimensions height must not exceed 33 cm x 25.7 cm x 15.2 cm (13 in. x 10.1 in. x 6 in.).

2. The system weight must not exceed 8.6 kg (19 lbs).

3. The system electric consumption must be 100 to 250 Volts AC, 50/60 Hz automatic adjustment 0.5 amps @ 120V, 0.25 amps @ 240V.

4. The system must automatically establish zero reading at startup.

5. The system alarm must be r selectable audible, visual, and vibratory.

6. The system flow control must be 1 cfm (28.3 lpm) +/- 10%.

7. The system reagent settings must be PAO-4, DOP, DOS/DEHS, Ondina, PEG, Krydol, Corn Oil, Mineral Oil, Paraffin.

8. The system dynamic range must be set to 0.0001 mg/m3 (µg/l) to 120 mg/m3 (µg/l).

9. The system must have an accuracy of 1% of readings (0.01% to 100%).

10. The system must perform a repeatability of 0.5% of readings > 0.01% to 100%.

11. The system standards must meet the industry ISO 14644-3, NSF-49, IEST-RP-CC034, EN 12469 / Electrical & Safety – CE, EN 61010-1:2010, EN 61326-1:2006, FCC Part 15 Subpart B.

All software licensing, warranty, and services associated with the equipment shall be in accordance with Original Equipment Manufacturer (OEM) terms and conditions.

Delivery Requirements:

Items shall be shipped to:

United States Army Medical Research Institute of Infectious Diseases (USAMRIID) 1425 Porter Street Fort Detrick, MD 21702

Delivery of items shall be coordinated with:

TBD