This document outlines requirements for a vibration dampening table and EMI shielding system to support a Thermo Fischer FEI Talos F200x TEM system. The vibration table must provide isolation beginning at below 1 Hz and attenuation levels ranging from -5 to -15 dB over frequencies from 1 to 2.5 Hz or higher. It requires a 6 degree of freedom active isolation design, steel laminate platform tops to support the TEM body and electronics cabinet, and an extension to raise the enclosure height. The EMI shielding system must accommodate a room up to 224x253x162 inches, include sensors and controllers for analog feedback and digital gain control to cancel AC and DC fields up to 1 kHz with continuous monitoring. It must be delivered to the US Army Engineer Research and Development Center in Vicksburg, MS by 31 December 2023. The related federal contract opportunity is solicitation number W912HZ23Q7753 from the Department of the Army Corps of Engineers Engineer Research and Development Center to provide the specified vibration dampening table and EMI shielding system.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Description of Requirement
Vibration Dampening Table and EMI Shielding System for
Talos F200x TEM System
SPECIFICATIONS:
The CMB requires a vibration dampening table and a EMI shielding system for a Thermo Fischer FEI Talos F200x TEM system. Both the table and shielding system must be compatabile with the Thermo Fischer FEI Talos F200x TEM.
Specifications for the equipment are:
Vibration table
• Vibration isolation for all floor-generated vibration beginning at below 1 Hz -5 to -10 dB
(40 - 70%) @ 1 Hz - 1.25 Hz, -10 to -15 dB (70 - 85%) @ 2 Hz -20 dB (90%) or more @
2.5 Hz and higher
• Full 6 Degree-of-Freedom active isolation design
• Highly damped steel laminate platform top to support Thermo Fisher FEI Talos TEM main body on top of Tool-Base inside enclosure frame
• Highly damped steel laminate platform top to support Thermo Fisher FEI Talos electronics cabinet inside enclosure frame.
• Maximum load capacity 5000 lbs
• Includes custom designed Acoustic Enclosure Rigid Extension to support and raise Talos enclosure to the same height as inner TOOL-Base Platform. Utilizes TMC Multilayer acoustic panel design
• Active Vibration Isolation Floor Platform with Acoustic Enclosure System for Thermo Fisher FEI Talos F200X TEM main body and extension
• Active Piezoelectric Vibration Isolation 3-Mount System
• Digital controller with Menu-Driven LCD user interface and PC Based GUI via Ethernet
Connection
EMI Shielding
• EMI Shielding room system for rooms up to 224X253X162 inch
• Controller with sensors
• Analog feedback with digital gain control
• Continuous feedback AC and DC field cancellation to 1kHz
• Continuous field monitoring
• Digital feed-forward optional for additional inputs
• PC-based monitoring and control software with graphical interface included
• DC - 3kHz compact fluxgate magnetometer
• 3-axis field cancellation
• Includes Cable and Connector Set, Helmholtz coil pairs and conduit tracks for wall mounting
• Includes aluminum thresholds to cover coils on the floor by doorways if needed
Shipping costs must be included in quote.
DELIVERY LOCATION:
US Army Engineer Research and Development Center 3909 Halls Ferry Road Bldg. 6000 Vicksburg, MS 39180
DELIVERY DATE:
The system must be provided to the ERDC on or before 31 Dec 2023.