A01-MEC-PendoTech Sensor Data Acquisition System.pdf

PDF 141 KB Posted

Attached to
PendoTech Sensor Data Acquisition System, WRAIR Federal contract opportunity
Solicitation number
HT942523Q0065
Issued by
Defense Health Agency

View the file

Other files for this federal contract opportunity

Other files attached to PendoTech Sensor Data Acquisition System, WRAIR, newest first.
File Type Posted
A02 Sources Sought Pendo Tech.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PENDOTECH MINIMAL ESSENTIAL CHARACTERITICS

1. PENDOTECH DATA ACQUISITION SYSTEM (PDKT-PCS-DAQ-T)

INPUTS

• 4 pressure measurements with PendoTECH Pressure Sensors which can be re-used extensively or other pressure sensors/transducers optional pH (traditional probe) conductivity electrode

• 2 flowmeters with pulse/freq signal (each with standard calibration entry or manual calibration wizard)

• Air Detector (for air or liquid detection or can be used as an alarm)

• 2 scales (by serial RS232)

• 2 generic analog 4-20mA inputs (configured in software for range, units of measure, and digits of precision)

OTHER FEATURES

• PC software for graphical interface

• Critical alarms on pressures, pH, conductivity, and analog inputs

• Pre-alarm condition notifications with option to send email notifications

• Dry contact relay alarm signal (could be used to shut off a pump)

• Trending up to 8 parameters

• All data logged to a CSV file that is opened with Excel

• Manufactured in stainless steel enclosure

2. PENDOTECH BENCHTOP PHOTOMETER (SPEC-L-1-SU-280)

SPECS

• L = Lab

• 1 = Single wavelength unit

• SU = Optical couplers for use with single use flow cell

• XXX = Wavelength # 1, i.e. 280 nm)

OTHER FEATURES

• Optical Configuration — Absorbance, with LED Light Source and UV Silicone

Photodiode: Internal Source Reference (wavelengths available from 240- 1100) LED Lifetime > 5 years

• Transmitter Box o Optical connectivity via SMA905 o Power by 12 mm plug (bench mount)

• Dimensions (WxDxH) — L Version: 3.25 x 4.51 x 2.69 inch (82.55 x 114.55 x

68.33 mm)

• Power Requirement — 20-30 Volts DC (100-250VAC to 24VDC supply included with bench top unit)

• Output Signal — 4-20mA sourcing with 400ohm maximum at 24VDC via field wireable connectors (bench top)

3. PENDOTECH CABLE FROM CHANNEL 1 OF A SINGLE OR DUAL CHANNEL

PHOTOMETER TO PENDOTECH DAQ CONTROL SYSTEM (PDKT-UV1-PCS)

• Connects the Benchtop photometer to the Data Acquisition system

• mA, 6ft

4. PENDOTECH SINGLE USE UV FLOW CELL

• Several path lengths available for different turbidity ranges from 1/2 inch hosebarb fitting; 1cm path length to 1/8 inch hosebarb fitting; 2mm path length

• Unobstructed flow path & no dead legs

• Qualified for use to 75psi (5bar)

• Compatible with gamma irradiation & autoclavable

• All polymeric materials meet USP Class VI

• Each sensor is 100% tested for quality during manufacturing, including leak test

• Manufactured in a clean room in an FDA registered, ISO 13485 facility

• Caustic and chemical resistant polysulfone

• Lot Certificate of Quality included with each sensor

5. PENDOTECH SINGLE USE TEMPERATURE SENSOR

• Sizes range from 1/8inch hose barb to 3/4inch hose barb

• Compatible with gamma irradiation

• High sensitivity to temperature change

• Better than +/- 0.2°C accuracy (typ. better than 0.1) in the specified range of

0-70°C

• Polycarbonate (luer) or polysulfone (hosebarb sensors) material of construction

• NO calibration required, each is tested during manufacturing to be within specification

• All fluid path materials meet USP Class VI requirements

• Certificate of quality included

• Manufactured in a FDA Registered, ISO 13485 certified facility

File details come from the government source that posted it. Updated .