This document is a Solicitation (W9114F-25-Q-A086) for a Dental Sterilizer for MEDDAC B Ansbach, detailing comprehensive requirements for a medical sterilization system. The procurement includes a single-door steam sterilizer with 300-liter chamber capacity, an electric steam generator, inverter, base tray, transport cart, and associated accessories. Key technical specifications require FDA approval, European-compatible power supply, a microcomputer-controlled cabinet with integrated printer-recorder, and sterilization parameters ranging from 250-275°F with various cycle durations. Additional requirements include a stand-alone steam generator, custom and standard cycle programming capabilities, and a minimum 2-year parts and labor warranty. The system must have precise environmental monitoring, safety features, and comprehensive testing protocols including installation, operational, performance, and functional testing.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Solicitation - W9114F25QA086-Final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
• 1. Steam Sterilizer single door: x 1ea.
• Capacity: 4 StU at 300 x 300 x 600 mm I 300 Liter
• Type: MSTV (09.02)
• Outer dimensions: Chamber dimensions:
o W = 990mm W = 660mm o D = 1095mm D = 700mm o H = 1970mm H = 660mm
• Loading height: 790mm
• Machine Base Configuration:
o 208 VAC / 60 Hz, according ASME
• ASME Certification Mark with ""U'"' designator
• Vertical sliding door
• User interface type CP-Top Crystal Protected-Touch Operating Panel, 5.7 inch display integrated in the operation panel made of glass and Hi-Macs
• Machine operation via touch sensor keys with acoustic feedback
• Outer material made of stainless steel with hygienic 4N finish
• Chamber made of stainless steel, 1.4404 (AISI 316L)
• built in batch protocol printer
• Programmable controller type Saia PCD -2
• Fully automatic PLC controlled
• Temperatures are measured with 3 circuit Pt 1000 sensors, Class AA according to standard DIN EN 60751
• Chamber and double jacket pressure is measured by 2 independent pressure transducers
• Steam pressure is measured via pressure transducers
• Fractionated pre-vacuum procedures
• Status information on loading and unloading site
2. Electric Steam Generator (30 kW) for generating clean steam according to EN 285 x 1ea.
• Type: ELD (16) / Steam output = 48 kg/hour Machine Base Configuration:
• 400 VAC 50/60Hz, (DGRL 2014/68/EU) according EN 14222, EN 13445
• Steam Generator
• Feed Water Tank with pre-heating
• Feed Water Pump
• Control Cabinet with Touchscreen and Main Switch
3. Inverter 50Hz to 60 Hz x 1ea.
• Converter for electrical power supply 400V/50Hz to 208V/60Hz
4. Base Tray x 1ea.
5. Transport cart x 1ea.
• Height adjustable transport plate (+/- 20 mm)2 swivel castors with total lock 0125mm
• 2 swivel castors with directional lock 0125mm
• Centering and locking device for docking
• Stainless steel 1.4301/304
- Height: 1005 mm
- Width: 637 mm
- Depth: 890 mm
W9114F-25-Q-A086- Dental Sterilizer MEDDAC B Ansbach
Salient Characteristics: Steam Sterilizers Single door
6. Shelf for batch cart x 1ea.
Consisting of:
• Shelf with 30x30x2.5mm grid
• 2 shelf holders ;
• Stainless steel 1.4301/304, electro polished
7. Side panel x 1ea.
• 960x1970 mm
• Material: stainless steel
• 1.4301 for use on left or right
• Side casing for the sterilizer can be installed on left and right. If the ELD switch cabinet is installed at the side, an aperture for the main switch is installed to the left on the side wall. The side position is defined from the point of view of the servicing access (side of the electric switch cabinet).
Consists of:
• Side wall
• Assembly screws
8. Additional paneling, per m2 1 ea.
9. Assembly / per unit 1 ea.
10. Acceptance and functional testing per unit 1ea
• Installation test
• Operational test
• Performance test
• Tests and inspections of safety devices
• Technical functional testing
• Functional testing according to manufacturer's specifications
• Verification of the calibration values of the plant's own measuring systems
• Air leakage testing (vacuum leak test)
• Vapor penetration testing
• Thermometric testing
• Testing with hollow body loading
• Drying tests of the loading
W9114F-25-Q-A086- Dental Sterilizer MEDDAC B Ansbach
11. Removal and Disposal Old Installation 1 ea.
12. Additional Requirements:
• Must have FDA approval but European compatible power supply.
• Dimensions are not to exceed: h(mm) x d(mm) x w(mm) = 1970x1095x990
• Must have a minimum chamber volume of 300 Liters
• Should conform with average optimum loading height of ca. 800 mm.
• Should process instruments at temperatures ranging from 250 F through 275 F via Pre-VAC.
• Should be pre-programmed with standard cycles and testing protocols but allow for custom programs.
• Must be cabinet mounted and controlled by microcomputer that monitors all operations and provides audible and visual warnings of any deviation.
• Must have integrated printer-recorder for cycle documentation.
• Each unit must have stand-alone steam generator, inverters (50Hz to 60Hz), the necessary trays, carts, and other miscellaneous items for standard operations.
• Maximum sterilization parameters to be met: 250 F – 30 min / 270 F – 15 min / 273 F – 12 min / 274 F – 12 min / 275 F – 12 min.
• Manufacturer must certify operation and proper function after installation and commissioning.
• Must have a minimum of a 2-year parts and labor warranty