This document is a detailed Request for Quote (RFQ) for two steam sterilizers for MEDDAC B Stuttgart. The solicitation requires two single-door steam sterilizers with specific technical specifications, including a 300-liter chamber capacity, MSTV type (09.02), vertical sliding door, and a user interface with a 5.7-inch touch display. Additional required components include two 30 kW electric steam generators, inverters, base trays, transport carts, shelves, side panels, and associated accessories. Key requirements mandate FDA approval, European-compatible power supply, microcomputer control with integrated documentation, and sterilization parameters ranging from 250°F to 275°F. The procurement includes comprehensive testing and acceptance procedures, removal of old installation, and requires a minimum two-year parts and labor warranty.

View the file

Other files for this federal contract opportunity

Other files attached to Steam Sterilizer (Quantity 2), newest first.
File Type Posted
Solicitation - W9114F25QA080.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics: Two Steam Sterilizers Single door

1. Steam Sterilizer single door: 2 ea.

• Capacity: 4 StU at 300 x 300 x 600 mm I 300 Liter

• Type: MSTV (09.02)

• Outer dimensions: Chamber dimensions:

o W = 990mm W = 660mm o D = 1095mm D = 700mm o H = 1970mm H = 660mm

• Loading height: 790mm

• Machine Base Configuration:

o 208 VAC / 60 Hz, according ASME

• ASME Certification Mark with ""U'"' designator

• Vertical sliding door

• User interface type CP-Top Crystal Protected-Touch Operating Panel, 5.7 inch display integrated in the operation panel made of glass and Hi-Macs

• Machine operation via touch sensor keys with acoustic feedback

• Outer material made of stainless steel with hygienic 4N finish

• Chamber made of stainless steel, 1.4404 (AISI 316L)

• built in batch protocol printer

• Programmable controller type Saia PCD -2

• Fully automatic PLC controlled

• Temperatures are measured with 3 circuit Pt 1000 sensors, Class AA according to standard DIN EN 60751

• Chamber and double jacket pressure is measured by 2 independent pressure transducers

• Steam pressure is measured via pressure transducers

• Fractionated pre-vacuum procedures

• Status information on loading and unloading site

2. Electric Steam Generator (30 kW) for generating clean steam according to EN 285 2 ea.

• Type: ELD (16) / Steam output = 48 kg/hour Machine Base Configuration:

• 400 VAC 50/60Hz, (DGRL 2014/68/EU) according EN 14222, EN 13445

• Steam Generator

• Feed Water Tank with pre-heating

• Feed Water Pump

• Control Cabinet with Touchscreen and Main Switch

3. Inverter 50Hz to 60 Hz 2 ea.

• Converter for electrical power supply 400V/50Hz to 208V/60Hz

4. Base Tray 2 ea.

5. Transport cart 2 ea.

• Height adjustable transport plate (+/- 20 mm)2 swivel castors with total lock 0125mm

• 2 swivel castors with directional lock 0125mm

• Centering and locking device for docking

• Stainless steel 1.4301/304

- Height: 1005 mm

- Width: 637 mm

- Depth: 890 mm

W9114F-25-Q-A080- Dental Sterilizer, MEDDAC B Stuttgart

6. Shelf for batch cart 2 ea.

Consisting of:

• Shelf with 30x30x2.5mm grid

• 2 shelf holders ;

• Stainless steel 1.4301/304, electro polished

7. Side panel 2 ea.

• 960x1970 mm

• Material: stainless steel

• 1.4301 for use on left or right

• Side casing for the sterilizer can be installed on left and right. If the ELD switch cabinet is installed at the side, an aperture for the main switch is installed to the left on the side wall. The side position is defined from the point of view of the servicing access (side of the electric switch cabinet).

Consists of:

• Side wall

• Assembly screws

8. Additional paneling, per m2 1 ea.

9. Assembly / per unit 1 ea.

10. Acceptance and functional testing per unit 1ea

• Installation test

• Operational test

• Performance test

• Tests and inspections of safety devices

• Technical functional testing

• Functional testing according to manufacturer's specifications

• Verification of the calibration values of the plant's own measuring systems

• Air leakage testing (vacuum leak test)

• Vapor penetration testing

• Thermometric testing

• Testing with hollow body loading

• Drying tests of the loading

W9114F-25-Q-A080- Dental Sterilizer, MEDDAC B Stuttgart

11. Removal and Disposal Old Installation 1 ea.

12. Additional Requirements:

• Must have FDA approval but European compatible power supply.

• Dimensions are not to exceed: h(mm) x d(mm) x w(mm) = 1970x1095x990

• Must have a minimum chamber volume of 300 Liters

• Should conform with average optimum loading height of ca. 800 mm.

• Should process instruments at temperatures ranging from 250 F through 275 F via Pre-VAC.

• Should be pre-programmed with standard cycles and testing protocols but allow for custom programs.

• Must be cabinet mounted and controlled by microcomputer that monitors all operations and provides audible and visual warnings of any deviation.

• Must have integrated printer-recorder for cycle documentation.

• Each unit must have stand-alone steam generator, inverters (50Hz to 60Hz), the necessary trays, carts, and other miscellaneous items for standard operations.

• Maximum sterilization parameters to be met: 250 F – 30 min / 270 F – 15 min / 273 F – 12 min / 274 F – 12 min / 275 F – 12 min.

• Manufacturer must certify operation and proper function after installation and commissioning.

• Must have a minimum of a 2-year parts and labor warranty

Aleksandr.Gutman Highlight