This document is an installation instructions and templates guide for plumbing and electrical supplies required for the A-2 dental delivery system spotting equipment.
The document specifies detailed requirements for five utility installations: Air requires a 1/2" nominal pipe compression fitting protruding 2" from the finished floor or wall with 80-100 psi at 2.5 scfm minimum (7.5 scfm peak intermittent flow); Water requires a 1/2" nominal pipe with 60±20 psi at 1.5 gpm minimum, not exceeding 40°C, with city water plumbing only needed if the unit lacks a self-contained water system; Electrical requires a 1/2" conduit and box with a Hospital Grade horizontal duplex receptacle positioned no higher than 4.25" above finished floor at 120 volts, 3-wire; Central Vacuum (wet systems) requires a 1/2" O.D. tube protruding 1" perpendicular to the floor at 10-12 in Hg and 9 scfm minimum, while dry/semi-dry systems require 4.5±1 in Hg and 12 scfm minimum; and Gravity Drain requires a 1-1/2" nominal pipe protruding 1" from the finished floor with a trap installed per local codes. The contractor supplies all piping, while the dental dealer supplies shut-off valves and electrical components. All plumbing and utilities must conform to prevailing local codes and be installed by licensed plumbers and electricians. The document includes a floor plan template showing typical utility center locations relative to the A-2 dental chair.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| B.01 Pre-Solicitation Notice (26-Q-2000).docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
WATER
ELECTRICAL
DRAIN
VACUUM
AIR
INSTALLATION INSTRUCTIONS & TEMPLATES FOR
SPOTTING EQUIPMENT PLUMBING AND ELECTRICAL
SUPPLIES FOR THE A2 DELIVERY SYSTEM
AIR: 1/2" nominal pipe compression fitting protruding 2"(51mm_ from finished floor or wall. Supplied by the contractor.Right angle manual shutoff valve supplied by dental dealer to be installed by contractor.Air requirements are 80100 psi (550 to 860 Kpa),2.5 scfm (71 sl/min) minimum,during normal use (7.5 scfm[ 210 sl/min] peak intermittent flow). Air plumbing should be flushed clean before making final connections to dental equipment.
WATER: 1/2" nominal pipe compression fitting protruding 2"(51mm_ from finished floor or wall.Supplied by the contractor.Right angle manual shutoff valve supplied by dental dealer to be installed by contractor.Water requirements are 60 20 psi( 410 140 KPa), 1.5 gpm (5.71/min) minimum,not to exceed 40° C(104° F). Water plumbing should be flushed clean before making final connections to dental equipment.City water plumbing is only required when the unit being installed does not have a self Contained Water System or has a cuspidor( for bowl rinse functions).
ELECTRICAL: 1/2" (13mm) conduit and box with duplex or equal receptacle,supplied by contractor. Wire and use Hospital Grade Duplex receptacle. The top of the duplex box to be no higher than 4.25" (108mm) above finished floor. Duplex receptacle shall be oriented horizontally. Voltage 120volts 3 wire.
CENTRAL
VACUUM:
1) THIS TEMPLATE SHOWS THE TYPICAL METHODS FOR LOCATING THE FLOOR BOX CENTERS WITH
RELATIONSHIP TO DENTAL CHAIR.
2) The REGULATIONS PROVIDE THAT LICENSED PLUMBERS AND ELECTRICIANS SHALL INSTALL UTILITIES.
3) MAKE SURE ALL PLUMBING AND UTILITIES CONFORM TO PREVAILING LOCAL CODES.
FLIGHT DENTAL SYSTEMS MAKES NO WARRANTY OF ANY
KIND WITH REGARD TO THIS DOCUMENT,INCLUDING,BUT
NOT LIMITED TO,THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE.
THE CUSTOMER MUST CONSULT WITH PLANNING AND
INSPECTION AUTHORITIES FOR COMPLIANCE WITH
LOCAL CODES.
THE CUSTOMER ASSUMES ALL RESPONSIBILITY IN
ASSEMBLING AND USING THIS DOCUMENT
GRAVITY
DRAIN :
Plumbing up to floor box utility center should be specified by central vacuum supplier and terminated in utility center.
Wet systems: Plumbing should terminate with 1/2"(16mm) O.D. tube protruding 1" perpendicular to the floor. 10 to 12 in Hg, 9 scfm minimum.
Dry/Semidry systems: Plumbing should terminate with 1/2" (16 mm) O.D. tube protruding 1" perpendicular to the floor. 4.5 ± 1 in Hg,12 scfm minimum.
11/2" (38mm) nominal pipe protruding 1" from finished floor. NOTE: place trap in line to conform with local codes. supplied by contractor. floor mounting only.
pipe size supplied from dental chair is 5/8" ID