9084 - Questions and Answers.pdf

PDF 106 KB Posted

Attached to
X-Ray Diffractometer System
Solicitation number
ITB-08224
Issued by
Alachua County, Glencoe CDP, Florida

About this file

This document contains the Questions and Answers related to Bid No. DOT-RFP-24-9084-CA issued by the Florida Department of Transportation for the procurement of an X-Ray Diffractometer System. The bid opening date is March 27, 2024 at 10:00 am (local time). The bid requires a maximum angular resolution between 0.021 - 0.03 FWHM on LaB6 and a maximum angular speed of 12-15 deg/s. The contract opportunity is open to vendors registered in the State of Florida's MyFloridaMarketplace system, and out-of-state companies must be authorized to do business in Florida. The contract is subject to the MFMP Transaction Fee and the Scrutinized Companies Lists requirements under Florida Statutes.

Regarding pricing, the solicitation does not provide specific details on pricing terms, set asides, incumbents, or funding requirements. However, it notes that all payment(s) to the vendor(s) resulting from the contract(s) will be subject to the MFMP Transaction Fee, unless the transaction is exempt pursuant to the Florida Administrative Code.

View the file

Other files

File Type Posted
9084 Solicitation Document.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

FLORIDA DEPARTMENT OF TRANSPORTATION

Procurement Office

605 Suwannee Street, MS 20 Tallahassee, Florida 32399-0450

Phone: (850) 414-4100

QUESTIONS AND ANSWERS

DATE: March 21, 2024

BID NO.: DOT-RFP-24-9084-CA

BID TITLE: X-Ray Diffractometer System

OPENING DATE/TIME: March 27, 2024 at 10:00 am (LOCAL TIME)

Notice is hereby given of the following changes to the above-referenced SOLICITATION:

• Questions and Answers provided below:

QUESTION 1

After studying and reviewing the technical specifications addressed in ITB DOT- ITB-24-9084-CA, our engineers have a question regarding the Technical Specification Section “3. Goniometer”.

The bid requires a maximum angular resolution of 0.021° FWHM on LaB6. We note that this specification is made without a matching requirement for peak intensity and/or measurement speed. We believe that only one system on the market (not ours) meets this resolution specification, so we request the specification be changed to 0.03° FWHM on LaB6.

We suggest that the specification could be changed to require minimum performance in both peak FWHM and intensity, because a practical diffractometer must balance good resolution and speed. A specification that should be achievable by multiple instruments on the market would be “when measuring the (111) peak of LaB6, must be able to provide 0.045° FWHM resolution and 15,000 counts per second peak height”. This would ensure that the quoted systems deliver the total performance needed to analyze complex mixtures.

Would you please consider the above comments and provide us the response?

ANSWER 1

We have a correction. It should be a range for the resolution as we did for the speed.

Maximum Angular Resolution: 0.021°- 0.03° (FWHM on LaB₆).

Maximum Angular Speed: 12-15 deg/s.