80TECH23Q0018 SOW Automated Humidity Generator.pdf

PDF 135 KB Posted

Attached to
Automated Humidity Generator Federal contract opportunity
Solicitation number
80TECH23Q0018
Issued by
National Aeronautics and Space Administration

About this file

This document contains a Statement of Work and Sources Sought notice for an Automated Humidity Generator. The National Aeronautics and Space Administration Kennedy Space Center is seeking a Brand Name or equal Thunder Scientific Model 9500 series 9000 Automated Humidity Generator to replace an end-of-life unit. The equipment must be capable of generating humidity and dew point profiles with uncertainties no greater than 0.17% of the target value plus 0.016 of the target value, and have a high flow capability of 10 to 150 L/min. Responses to the Sources Sought notice are due by May 18th, 2023, and questions by May 16th, 2023. The notice will be used to determine the appropriateness of any small business set-asides. The government anticipates awarding a competitive firm fixed price contract for equipment, training, delivery, and a two-year warranty period.

View the file

Other files for this federal contract opportunity

Other files attached to Automated Humidity Generator, newest first.
File Type Posted
RFI Posting Page - Automated Humidity Generator.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

Automated Humidity Generator

1. Objective

Kennedy Space Center has a requirement for a Brand Name or equal product of one (1) Thunder Scientific Model 9500 (Series 9000) Automated Humidity Generator to replace the current existing Thunder Scientific Model 9000 that is end of life. This system will be utilized to calibrate KSC humidity devices across the center.

2. Characteristics, Scope, and Specs Thunder Scientific Model 9500 (series) 9000 Automated Humidity Generator or equal.

• The equipment shall be capable of generating relative humidity, dew point, frost point, and parts per million water vapor multipoint profiles

• The equipment shall have uncertainty values no greater than 0.17% * target value + 0.016 target value.

• The equipment shall have a high flow capability of 10 to 150 L/min

• The equipment shall have computerized internal transducer calibration

• This standard must be traceable to National Institute of Standards and Technology

(NIST) and generate humidity based on NIST’ proven "two-pressure" principle

• The contractor shall have National Voluntary Laboratory Accreditation Program

(NVLAP) Accredited Certification of Calibration as per International Organization for Standardization/International Electrotechnical Commission

ISO/IEC 17025

3. Warranty

The Contractor shall provide standard manufacturer warranty for the equipment.

4. Training

The contractor shall provide a one-time two (2) day on-site training on the Thunder Scientific Model 9500 Humidity (or equivalent) calibration standard, review operations, maintenance procedures & calibration performance.

The contractor shall provide the Operational Software & Manual.

5. Place of Performance Kennedy Space Center, KSC Calibration Laboratory, Bldg K6-1696, FL 32899

6. Period of Performance

The Period of Performance shall be 52 Weeks after receipt of award (ARO)