This is a specifications and requirements document issued by Broward County Water and Wastewater Services for the purchase, installation, training, and service of one new Total Organic Carbon (TOC) Analyzer system to be located at the North Regional Wastewater Treatment Plant in Pompano Beach, Florida. The TOC Analyzer must comply with Standard Methods 5310 B (High-Temperature Combustion Method) and operate under FDEP/TNI (NELAP) accreditation for the analysis of drinking water and wastewater samples. The analyzer shall include an integrated autosampler configured for 40 mL vials with a minimum of 68 positions, a controlling personal computer with Windows 11 Professional or newer operating system, instrument control and data processing software with 21 CFR Part 11 capability, accessories including a halogen scrubber and suspended sample kit, and start-up consumables. The scope of work encompasses delivery with inside delivery to the laboratory, installation and start-up by a factory-trained service engineer, on-site operator training for up to four laboratory staff over a minimum of two days, and comprehensive documentation including operator and maintenance manuals. The contract includes a minimum one-year full warranty covering all parts, labor, and travel with a two-business-day repair response requirement, as well as optional annual preventive maintenance service thereafter at the County's discretion.
The document specifies detailed analyzer specifications including measurement capabilities of Total Carbon (TC) from 0 to 30,000 mg/L and Inorganic Carbon (IC) from 0 to 35,000 mg/L with automatic dilution, a detection limit of 4 µg/L or lower, measurement reproducibility with CV of 1.5 percent maximum or 4 µg/L, and automatic sample injection by syringe pump with variable injection volumes of 10 to 2,000 µL. Payment terms are net 30 days following receipt and approval of proper invoices, which must include the Broward County Purchase Order number, description, part number, quantity, list price, discount price, and extended price, with invoices directed to the Accounts Payable email and a copy to the technical contact. The contractor must provide next business day service during normal business hours Monday through Friday from 8:00 AM to 4:00 PM EST, with each additional service visit including a minimum of four service hours and additional time billed in full-hour increments. No set-asides are indicated in this specifications document.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 3-Security Requirements.docx | DOCX document | |
| 1-CAB Sales Tax Exemption Certificate.pdf | ||
| 7-Ins Requirement- Purchase Installation Training and Service of a Total Organic Carbon (TOC) Analyzer.pdf | ||
| 6-Anti-Human Trafficking Affidavit.docx | DOCX document | |
| 5-General Conditions - Quotes Bids.docx | DOCX document | |
| 2-Insurance Requirements - Quotes Bids.docx | DOCX document | |
| 4-Vendor Questionnaire - CABs.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATIONS AND REQUIREMENTS
PURCHASE, INSTALLATION, TRAINING AND SERVICE OF A TOTAL ORGANIC CARBON (TOC) ANALYZER A. Scope of Work Contractor shall furnish, deliver, install, and place into operation one (1) new Total Organic Carbon (TOC) Analyzer with autosampler, controlling computer and software, accessories, start-up consumables, operator training, and warranty for the Broward County Water and Wastewater Services Environmental Laboratory located at the North Regional Wastewater Treatment Plant in Pompano Beach, FL. The analyzer will be used for the analysis of drinking water and wastewater samples by Standard Methods 5310 B (High-Temperature Combustion Method) under FDEP/TNI (NELAP) accreditation. The TOC Analyzer purchase shall include the following:
1.) Minimum Analyzer Specifications:
a. Measurement method: 680 °C combustion catalytic oxidation with non-dispersive infrared (NDIR) detection, suitable for Standard Methods 5310 B
b. Measured parameters: Total Carbon (TC), Inorganic Carbon (IC), TOC (TC − IC), and Non-Purgeable Organic Carbon (NPOC)
c. Measurement range: TC 0 to 30,000 mg/L; IC 0 to 35,000 mg/L (with automatic dilution)
d. Detection limit: 4 µg/L or lower for TC and IC
e. Reproducibility: TC, IC, and NPOC CV 1.5% maximum or ±4 µg/L maximum
f. Measurement time: approximately 3 minutes for TC and 3 minutes for IC
g. Automatic sample injection by syringe pump and slide-type injection mechanism, with variable injection volume of 10 to 2,000 µL
h. Automatic acidification and sparging for IC removal and NPOC measurement
i. Automatic sample dilution of 2x to 50x by syringe pump, with dilution accuracy of ±2% maximum (2x to 20x)
j. Automatic re-measurement with changed dilution and injection volume for samples exceeding the calibration range
k. Storage of up to three (3) calibration curves with automatic selection of the optimal curve for each sample
l. Automatic exclusion of anomalous replicate injections with recalculation of mean, standard deviation, and CV
m. Automated system blank check function
n. Stat or priority samples can be added to the run schedule without interrupting an autosampler sequence
o. Electronic carrier gas flow control; carrier gas: high-purity air (CO, CO₂, and hydrocarbons each 1 ppm maximum)
p. Power supply: 100 to 240 V AC, 600 VA; operating ambient temperature 5 to 35 °C
q. Benchtop footprint not to exceed approximately W340 × D660 × H480 mm (analyzer only)
2.) Autosampler:
a. Autosampler is fully integrated with the analyzer and software
b. Configured for 40 mL vials (minimum 68 positions); configuration for 24 mL vials (93 positions) acceptable upon County approval
c. Sample sparging within autosampler vials (external sparging kit with ASI parts)
d. Magnetic stirrer to keep suspended solids in wastewater samples mixed prior to injection
e. Initial supply of autosampler vials, septa, and caps (minimum 100 of each)
3.) Computer, Software, and Data Management:
a. Controlling personal computer and monitor, with a currently supported operating system (Windows 11 Professional or newer), supplied and configured by the Contractor
b. Instrument control and data processing software with user authentication (ID and password) and operational history/audit trail logging (21 CFR Part 11 capable)
c. Creation of run schedules and sample tables, including automatic insertion of quality control samples with automatic re-measurement when results fall outside configured acceptance limits
d. Export of measurement results as text/CSV files for transfer to the Laboratory Information Management System (LabLynx LIMS)
e. Batch and runtime report output for all samples or selected samples
4.) Accessories and Start-Up Consumables:
a. B-type halogen scrubber for samples containing chloride
b. Suspended sample kit with ASI parts, for wastewater samples containing particulates
c. Air supply pipe set and carrier gas regulator/connection fittings
d. Start-up kit including combustion tube, catalyst, IC reaction reagent supplies, and syringe, plus one (1) complete set of spare consumables
5.) Delivery, Installation, and Acceptance:
a. Delivery F.O.B. destination, inside delivery to the laboratory, including uncrating and removal of all packing materials
b. Installation and start-up by a factory-trained service engineer
c. Performance verification demonstrating the system meets manufacturer specifications, with Installation Qualification/Operational Qualification (IQ/OQ) documentation provided to the County
d. Final acceptance will occur after successful installation and performance verification, as approved in writing by the Laboratory Manager
6.) Training and Documentation:
a. On-site operator training for up to four (4) laboratory staff, minimum two (2) days, covering operation, calibration, software, data export, routine maintenance, and troubleshooting
b. Operator and maintenance manuals (printed and electronic) and software documentation
7.) Warranty:
a. Minimum one (1) year full warranty on the analyzer, autosampler, computer, and software, starting at final acceptance, covering all parts, labor, and travel
b. Warranty repair response by a factory-trained service engineer within two (2) business days of the County’s service request
8.) Annual Preventive Maintenance Service (after warranty period, at the County’s option):
a. One (1) preventive maintenance visit per year by a factory-trained service engineer, including labor, travel, and the manufacturer’s preventive maintenance kit
b. Replacement of wear parts and consumables per manufacturer recommendations (combustion tube, catalyst, syringe, seals, filters, and scrubbers as needed)
c. Performance verification after service, with a written service report provided to the County
d. Software updates and patches released by the manufacturer for the installed version
B. Additional Services Contractor may provide additional service visits by request of the County. Contractor must provide next business day service during normal business hours, Monday through Friday, 8:00 AM to 4:00 PM (EST). Hourly labor rates must include travel time to and from the system location. Each additional service visit includes a minimum of four (4) service hours. Additional time on site above the minimum will be billed in increments of full hours only.
C. Additional Material/Equipment Additional consumables, replacement parts, and optional accessories beyond those listed in the Scope of Work shall be billed according to the contract pricing found in the bid line items.
D. Invoices and Payments
1. Payments will be made in net 30 days, after receipt and approval of proper invoices.
2. Invoices shall include Broward County Purchase Order number, Description, Part Number, Quantity, List Price, Discount Price, and Extended Price.
3. To ensure prompt payment, all invoices and backup documentation must be emailed to AccountsPayable@Broward.org and a copy to the technical contact listed on the purchase order.