Statement of Work.pdf
PDF 135 KB Posted
- Attached to
- Non-Invasive Blood Pressure (NIBP) Simulation Testing for the Gallup Indian Medical Center
- Solicitation number
- 75H71024Q00012
About this file
This statement of work outlines requirements for a non-invasive blood pressure patient simulator to be procured by the Biomedical Engineering Department at Gallup Indian Medical Center. The simulator must be able to simulate adult, neonatal, hypertensive, and hypotensive blood pressure readings at specified levels. It must also simulate electrocardiograms at various heart rates and arrhythmias, respiration rates, oxygen saturation levels detectable by Masimo Rainbow SET, and invasive blood pressure readings across eight dynamics and seven statics. Additional required capabilities include temperature simulation up to 40 degrees Celsius, compatibility with a YSI temperature probe, and inclusion of six non-invasive blood pressure adapters and a cuff jacket. The simulator and all necessary supplies must be delivered to the Biomedical Engineering Department at Gallup Indian Medical Center. Quotes submitted in response to solicitation number 75H71024Q00012 from the Department of Health and Human Services Indian Health Service should include any applicable shipping and freight charges.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| IHS IEE Representation Form_Jan 2022_508 Compliant.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
GALLUP INDIAN MEDICAL CENTER: NIBP Simulation
PURPOSE OF CONTRACT: Gallup Indian Medical Center (GIMC) is part of the Navajo Area Indian
Health Services (NAIHS) located in Gallup, New Mexico is soliciting for a patient simulation system for
Non‐Invasive Blood Pressure medical device testing for the Biomedical Engineering Department.
GENERAL: Contractor shall furnish all necessary equipment and supplies for the successful delivery, performance of new equipment required by the Biomedical Engineering Department.
Background: IHS is an agency within the U.S. Department of Health and Human Services and is responsible for providing federal health services to American Indians and Alaska Natives. The provision of health services to members of federally recognized tribes grew out of the special government to government relationship between the federal government and Indian Tribes.
The goal is to assure that comprehensive, culturally acceptable personal and public health services are available and accessible to American Indians and Alaska Natives. The IHS currently provides health services to approximately 1.5 million American Indians and Alaska Natives who belong to more than 557 federally recognized tribes in 34 states.
Navajo Area Indian Health Service (NAIHS) administers health centers and hospitals providing health care to approximately 201,583 members of the Navajo Nation. The Navajo Nation is the largest Indian tribe in the United States and has the largest reservation, which encompasses more than 25,516 square miles in northern Arizona, western New Mexico, and southern Utah, with three satellite communities in central New Mexico. The NAIHS is the primary provider of inpatient, ambulatory care, preventive and community health, and environmental health services for members of the Navajo Nation and the San Juan Southern Paiute Tribe.
Gallup Indian Medical Center (GIMC) is a 75‐bed hospital in Gallup, New Mexico, on the border of the Navajo Reservation. Clinical specialties at GIMC includes Internal Medicine, Cardiology, Anesthesia, OB/GYN, General Surgery, Orthopedics, Ophthalmology, ENT, Radiology, Pathology, Pediatrics, Psychiatry, Emergency Medicine, and Urology. The workload at Gallup is one of the largest in the Indian Health Service with 380,000 outpatient encounters and 3,300 inpatient admissions annually. GIMC has the largest staff of all Navajo Area IHS facilities.
STATEMENT OF WORK: The Biomedical Engineering Department is procuring for a non‐invasive blood pressure patient simulator that must consist of the following:
These specifications (Salient Characteristics) shall be meet:
NIBP simulations must be able to provide Adult at 120/80 mmHg, Neonatal at 70/40 mmHg, Hypertensive at 190/120 mmHg and Hypotensive at 80/40 mmHg.
Should include a built in digital manometer.
Over pressure test mode and a leak rate test mode
ECG part of the simulation shall be able to connect up 12 leads at selectable heart rates of 30, 60, 90, 120, 140, 160, 200 and 240 beats per minute
Selectable arrhythmias shall include NSR w/PVCs, VTAH, VFIB, Asys, Bigeminy, Pacer, Trigeminy, ST+ and 2nd degree with square wave
Respiration rates include 15, 30, 45, 60, 70, 80, 100, 120 beats per minute
ECG plug in must be a banana type connection
SpO2 saturations must be in 1% increments between the ranges of 10% ‐ 100%
Pulse rates increment changes of 1 beat per minute from 25 – 240
Must be able to customize the preset selections
SpO2 simulator must be Masimo Rainbow SET compatible
Simulator shall automatically detect SpO2 manufacturer
Invasive blood pressure simulations must have 4 channels with 8 dynamics of 120/80, 60/30, 30/12, 20/12, 125/15, 60/5, 8/2 and 4/1 mmHg
7 Static with 0, 25, 50, 100, 150, 200 and 250 mmHg outputs
Cardiac output simulations with 2, 5 and 7 l/m with no adapter cable required
Temperature simulations must be able to reach up to 25, 37 and 40 degrees Celsius
A YSI temperature probe with 400 and 700 options
6 NIBP adapters with a cuff jacket
Shipping and freight charges must be included in the overall quote and costs.
PLACE OF PERFORMANCE:
Ground East Floor – Biomedical Engineering Department
Gallup Indian Medical Center, 516 E. Nizhoni Blvd. Gallup, NM 87301
Contractor shall provide medical equipment & delivery to the following location: Biomedical
Engineering. Ground East of the Gallup Indian Medical Center.
File details come from the government source that posted it. Updated .