About this file

This document contains the technical specifications for the ZOLL AED Pro defibrillator device. Key details include:

  • The AED Pro device has 3-lead ECG monitoring, defibrillation waveform, and configurable energy levels for adult and pediatric use. It provides real-time feedback on CPR performance.
  • The device meets or exceeds various safety, environmental, and electromagnetic compatibility standards.
  • The related federal contract opportunity is a 100% Indian Small Business Economic Enterprise (ISBEE) set-aside solicitation for 7 ZOLL AED Pro packages and 7 non-rechargeable battery packs. The solicitation is for the Indian Health Service in Fort Thompson, South Dakota, with a 90-day delivery requirement after award. Award will be made to the lowest-priced technically acceptable offer.

View the file

Other files

File Type Posted
75H70624Q00041.pdf PDF
IHS IEE Representation Form_IHS1499498.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

AED PRO Technical Specifications

ECG Monitoring

Patient Connection: 3-lead ECG cable.

Input Selection: Fully defibrillator protected.

ECG Size: Automatic Gain Control (AGC) gain.

Heart Rate: 30-300 bpm ±5%.

Display

Screen Type: High-resolution LCD display.

Screen Size: 3.02 inches W x 2.27 inches H.

Sweep Speed: 25 mm/sec.

Viewing Time: 3 seconds.

Displayed Information: Heart rate, ECG waveform, text prompts, CPR bar graph, battery gauge, elapsed time, number of shocks delivered.

Event Documentation

Type: Internal non-volatile memory.

Memory Capacity: With audio recording enabled, 20 minutes of ECG, operator and device actions;

without audio recording, 1.45 hours each for 4 patient records or up to 5.8 hours for a single recordof

ECG and event data. Expanded external data storage and transfer available with USB memory stick (not included).

Event Review & Reporting:

RescueNet® Code Review software version 3.30 or higher.

Communications: IrDA infrared wireless for communication with personal computer or external USB memory stick transfer.

Defibrillator

Waveform: ZOLL Rectilinear Biphasic™ waveform.

Energy: 50, 70, 85 joules (pediatric); 120, 150, 200 joules (adult) preconfigured.

Charge Time: Less than 10 seconds with a new fully charged battery. Depleted batteries will result in a longer defibrillator charge time.

Energy Display: Monitor display indicates selected energy in manual mode only.

AED Function: Auto analyze and charge x3 with programmable auto energy level selection, screen prompts, and voice prompts.

Electrodes: One-piece CPR-D-padz®, CPR stat-padz®, stat-padz® II, and Pediatric pedi-padz® II. CPR-D padz one-piece electrode and CPR stat-padz two-piece electrodes provide Real CPR Help for rate anddepth of CPR chest compressions.

Manual: Energy selection user configurable, 120, 150, 200 joules (adult); 50, 70, 85 joules (pediatric).

Semi-Automatic: Evaluates electrode connection and patient ECG to determine if defibrillation is required.

Shockable Rhythms: Ventricular fibrillation with amplitude >100 uV and wide complex ventricular tachycardia with rates greater than 150 bpm (adult) and greater than 200 bpm (pediatric).

Defibrillation Pads Impedance Measurement Range: 10-300 ohms.

Prompts: Audible voice prompts and text messages guide user through complete sequence of operation.

Text prompts are always displayed. With ZOLL Administration Software (ZAS), user can configure voice prompts on/off.

Controls: ON/OFF, Shock, Softkeys.

Configurability

ZOLL Administration Software (ZAS): Allows user to manually adjust the configuration of the AED Pro to desired medical protocol or direction.

Analyze OPTIONS: User can configure the ability to analyze a patient during the CPR period by depressing a softkey.

CPR Duration: User can configure CPR duration from 30 seconds to indefinite, depending on protocol.

User also has the ability to configure the AED Pro to “start with CPR”, which, when turned on, will prompt the rescuer to perform CPR prior to analyzing the patient.

Voice Prompts: User can adjust wording (limited options) and/or turn on/off particular voice prompts.

Note: Text prompts are always displayed.

Shock Energy: Adjust shock energy level for semi-auto mode, for both adult and pediatric modes.

General

Size: 3 in (7.6 cm) x 9.2 in (23.5 cm) x 9.4 in (23.9 cm) deep.

Weight: < 6.5 lbs (2.9 kg) withdisposable battery and electrodes.

Design Standards: Meets or exceeds AAMI DF-80, EN 60601-1, EN 60601-1-2.

Patient Safety: All patient connections are electrically isolated.

Environmental

Operating Temperature: 0° C to 50° C.

Storage and Shipping Temperature: -30° C to 70° C.

Humidity: 10% to 95% relative humidity, non-condensing.

Vibration: MIL Std. 810F, Minimum Helicopter Test.

Shock: IEC 68-2-27; 100G.

Drop Test: 1.5m per IEC 68-2-32.

Material/Water Ingress: IEC 60529, IP55.

Electromagnetic Compatibility (EMC): CISPR; 11 Class B Rated and Conducted Emissions.

Electromagnetic Immunity: AAMI DF-80; IEC 60601.

Electrostatic Discharge: IEC 61000-4-2.

Conducted Susceptibility: IEC 61000-4-6.

Options: Vinyl foam-molded carry case for added protection; external IrDA port; wall sign.