Amendment 0002, RFP 75H70523R00002
IV c. Provide MRI-safe wheelchair, emergency crash cart and crash cart supplies, gloves, room disinfectant and disinfectant wipes.
For clarification, the contractor shall: follow safety protocol thereby removing the patient from the MRI area in a patient crisis. The crash cart and emergency personnel will be provided by each site. Contractor shall provide MRI-safe wheelchair. Contractor shall provide gloves, room disinfectant and disinfectant wipes.
IV d. Provide MRI mobile unit with minimum 1.5 Tesla strength, minimum 60 cm bore;
minimum 550 lb table weight limit in both vertical and horizontal movement.
For clarification, the contractor shall: provide MRI unit with the above-referenced capabilities; however, adjustment to the table weight limit being minimum 350 lb. capacity.
IV f. Supply adequate uninterruptable power supply (UPS) for equipment.
For clarification, the contractor shall: provide adequate uninterruptable power supply (UPS) for computer system backup protection in the event of power interruptions, to ensure protection of exam data.
IV r. Have both audible and visual alarms located in the mobile scan room. The audible alarm shall be capable of being heard clearly in the scan room and also in the operator control room. The contractor shall provide a safety system and emergency shutdown, visual and audible alarms, heat sensor shutoff for room temperature problems. The system shall also include warning devises for temperature limits, and scan-in-progress.
For clarification, the contractor shall: ensure both MRI and CT units are equipped with intercom system and that the technologist can maintain visual contact with the patient from the computer work station.
IV s. Include the following components and systems: pulse oximeter, oxygen delivery system, wheelchair, emergency crash cart and crash cart supplies, patient monitor and video camera, patient intercom system, lead shields and positioning sponges.
For clarification, the contractor shall: Provide MRI-safe wheelchair available for patients. Emergency crash cart and supplies have been addressed above (see IV c.). Patient monitoring is addressed above (see IV r.).