Attachment 1: Draft Requirements
Key Performance Parameters (KPP) that must be met at the threshold level to be considered as a viable option are noted by a bold KPP in the ID # column of the requirements table. Some requirements have unique sub-requirements for detection as it relates to specific radionuclides; references to these radionuclides can be found in Appendix 1, Tables 1-5.
The requirements are derived from a series of sources. These include:
• “Technical Capability Standard for Handheld Instruments Used for the Detection and Identification of Radionuclides,” (TCS) Document #500-DNDO-117250v0.00, Date Revised: October 2011. The handheld instruments TCS was published in accordance with the SAFE Port Act [H.R. 4954-16, Subtitle C-Port Operations, Section 121(f) Standards] and provides performance requirements for radionuclide detection and identification of selected industrial and special nuclear materials (SNM), both bare and shielded, and detection and identification of SNM under conditions of masking by industrial, medical, or NORM sources. (http://www.dhs.gov/xlibrary/assets/dndo/dndo-technical-capability-standard-for-handheld-final.pdf)
• ANSI N42.34-2015, “American National Standard Performance Criteria for Hand-Held Instruments for the Detection and Identification of Radionuclides”.
• SIGMA Edge API, Release 8.0.0-SNAPSHOT, March 31, 2018.
• ANSI N42.42-2012, “American National Standard Data Format for Radiation Detectors Used for Homeland Security”.
• MIL-STD-810G, MIL-STD-464C, MIL-STF-1472G.
• Certain design and operational characteristics sought by CWMD on the basis of user needs, experience and preferences.
http://www.dhs.gov/xlibrary/assets/dndo/dndo-technical-capability-standard-for-handheld-final.pdf
ID # Requirement Threshold Objective Clarification
Meets (T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
KPP
System shall detect gamma radiation.
Probability of Detection (PD) = 80% for 0.66 µR/hr of Co- 57 @ 1ft/s relative speed and 2 ft.
distance of closest approach & 7.53 µR/hr of Ba-133 @ 1ft/s relative speed and 2 ft. distance of closest approach.
PD = 80% for 0.16 µR/hr of Co-57 @ 2 ft./s relative speed and 4 ft. distance of closest approach & 1.88 µR/hr of Ba-133 @ 2 ft./s relative speed and 4 ft.
distance of closest approach.
Threshold and objective exposure rates are calculated net of background. The alarm signal threshold settings must be the same as those used for the false alarm tests (BHH-OR-03). For testing, the device will be oriented as used in standard scanning operations (held in an operator’s hand). For testing, the exposure rate to be detected will be scaled as the square root of the ratio of the test site background exposure rate to the background exposure rate on which the requirement is based. The requirement is based on a background of 7 µR/hr measured with a Fluke Biomedical 451P pressurized ion chamber survey meter. The device is expected to operate in backgrounds between 0.34 µR/hr and 25 µR/hr.
BHH-FR-
01-01
The device shall detect gamma rays over a relevant range of energies.
25keV - 3000keV 15keV - 8000keV None
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
01-02
The device shall allow modification of the gamma detection alarm settings/thresholds in restricted (password protected) mode.
Ability to modify detection alarm settings/thresholds in restricted mode.
Objective = Threshold
The detection alarm is unique and different from the personal safety alarm described in BHH-OR-10. A gamma detection alarm is the alert to the operator that the detected radiation has exceeded a certain level. That level is set using thresholds or other sensitivity parameters depending on how complex the alarm algorithm is.
01-03
System shall provide notification of gamma detections to the operator.
Notify the operator of gamma detections. Objective = Threshold None
BHH-OR-
KPP
System shall detect neutrons.
PD =80% for a 20,000 (±20%) neutrons/s 252Cf source surrounded by a 4-cm thick sphere of high density polyethylene (HDPE) at 25 cm distance of closest approach (center of source to detector reference point) at 0.5 m/sec relative speed.
PD =80% for a 60,600 (±20%) neutrons/s 252Cf source surrounded by a 4-cm thick sphere of high density polyethylene (HDPE) at 61 cm distance of closest approach (center of source to detector reference point) at 0.5 m/sec relative speed
The alarm signal threshold settings must be the same as those used for the false alarm tests (BHH-OR-04). For testing the device will be oriented as used in standard scanning operations (held in an operator’s hand) with expected amount of moderation provided by the operator included (e.g., torso, arm and hand phantoms).
Expected Range of Neutron background is 0.002-0.013 neutrons/cm2-second (Absorption in the 4 cm of HDPE is neglected.) Both the source and detector will be at a height of 1 meter.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
02-01
The neutron detector shall be insensitive to photon radiation.
Gamma radiation at exposure rates up to 10 mR/h from 137Cs shall not cause the system to indicate presence of neutrons or activate a neutron alarm.
Objective = Threshold
In accordance with ANSI N42.34-2015, section 6.7. This test will be done at the same neutron alarm threshold setting used for OR-2 and OR-4.
02-02
The system shall detect neutrons while exposed to high gamma exposure rates.
Detect presence of neutrons or activate alarm while being exposed to gamma radiation up to 10 mR/hr from 137Cs.
Objective = Threshold
In accordance with ANSI N42.34-2015, section 6.7. This test will be done at the same neutron alarm threshold setting used for OR-2 and OR-4.
02-03
The device shall allow modification of the neutron detection alarm settings/thresholds in restricted (password protected) mode.
Ability to modify detection alarm settings/thresholds in restricted mode.
Objective = Threshold
The detection alarm is unique and different from the personal safety alarm described in BHH-OR-10. A neutron detection alarm is the alert to the operator that the detected radiation has exceeded a certain level. That level is set using thresholds or other sensitivity parameters depending on how complex the alarm algorithm is.
02-04
System shall provide notifications of neutron detections to the operator.
Notify the operator of neutron detections. Objective = Threshold None
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall have a low rate of gamma false detections.
<1 false gamma alarm in 8 hours when operated in a stable background environment.
<1 false gamma alarm in 12 hours when operated in a stable background environment.
None
BHH-OR-
System shall have a low rate of neutron false detections.
<1 false neutron alarm in 12 hours when operated in a stable background environment.
<1 false neutron alarm in 24 hours when operated in a stable background environment.
None
BHH-OR-
System shall support the operator in localizing a radiation source.
Notify operator if radiation level is rising or falling in real time.
Inform operator of direction to source. None
BHH-FR-
05-01
The device shall display a time series plot to assist in localization.
Device displays time series plot of gamma gross count using time interval ≤ 1second.
Objective = Threshold None
BHH-FR-
05-02
The device shall provide audible indication of changing gamma count levels to assist in localization.
Device provides audible indication (e.g. change in pitch and frequency) of increasing and decreasing gamma count rates.
Objective = Threshold None
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
05-03
The device shall display neutron count rate.
Device displays neutron count rate in counts/second.
Device displays neutron count rate and indication of increase or decrease in the rate.
None
BHH-FR-
05-04
The device shall display gamma count rate.
Device displays gamma count rate in counts per second.
Objective = Threshold None
BHH-FR-
05-05
The device shall provide tactile indication of changing gamma count levels to assist in localization.
There is no threshold.
Device provides tactile indication (e.g. change in intensity and frequency) of increasing and decreasing gamma count rates.
Objective only requirement.
BHH-OR-
06a
KPP
System shall correctly identify present radionuclides that are unshielded.
Probability of Identification (PID) ≥ 80% for each of the radionuclides listed for OR-6a in Radionuclide Table - Appendix 1 Table 1 (KPP Radionuclides).
Objective = Threshold
Gamma-ray exposure rate from each unshielded source shall be 50 µR/hr net of background at the detector’s gamma reference point. Correct identification is defined as when the device identifies the required radionuclide (i.e., the radionuclide listed in Appendix 1 Table 1 KPP Radionuclides). The device may also identify additional acceptable radionuclides that are present in conjunction with the required radionuclide such as daughter products, known contaminants, and NORM from background. It will not be considered correct to identify only the additional acceptable radionuclides and
(T or O)
Or Does Not Meet Reference/explan ation not the required radionuclide. Any radionuclide(s) that are not present, but are identified by the device, will be assessed as a weakness against BHH- OR-07, but will not constitute as an incorrect identification as long as the device identifies the required radionuclide.
BHH-OR-
06b
System shall correctly identify present radionuclides that are unshielded.
Probability of Identification (PID) ≥ 80% for each of the radionuclides listed as Threshold for OR- 6b in Radionuclide Table - Appendix 1 Table 2.
Probability of Identification (PID) ≥ 80% for each of the radionuclides listed as Objective for OR-6b in Radionuclide Table - Appendix 1 Table 2.
Gamma-ray exposure rate from each unshielded source shall be 50 µR/hr net of background at the detector’s gamma reference point with the exception that thorium-232, radium- 226, and potassium-40 be tested at a 20 µR/hr net of background. Medical Radionuclides (see Table 4) shall be surrounded by 8 cm of PMMA to represent in vivo conditions. Correct identification is defined as when the device identifies the required radionuclide (i.e., the radionuclide listed in Appendix 1 Table 2). The device may also identify additional acceptable radionuclides that are present in conjunction with the required radionuclide such as daughter products, known contaminants, and NORM from background. It will not be considered correct to identify only the additional acceptable radionuclides and not the required radionuclide. Any radionuclide(s) that are not present, but are identified by the device, will be assessed as a weakness against BHH-
(T or O)
Or Does Not Meet Reference/explan ation OR-07, but will not constitute as an incorrect identification as long as the device identifies the required radionuclide.
BHH-OR-
06c
System shall correctly identify present radionuclides that are shielded.
Probability of Identification (PID) ≥ 80% for each of the radionuclides listed as Threshold for shielded in Radionuclide Table - Appendix 1 Table 3.
Probability of Identification (PID) ≥ 80% for each of the radionuclides listed as Objective for shielded in Radionuclide Table - Appendix 1 Table 3.
All radionuclides shall be shielded with 5mm steel. HEU (U-235) and WGPu (U-239) radionuclides also shielded with 4 cm of HDPE in addition to the 5 mm steel. Medical Radionuclides (see Table 4) will not be surrounded by 8 cm of PMMA, only the 5 mm of steel.
Shielded configurations will be evaluated at same exposure rates at the detector gamma reference point as in the unshielded configurations.
Correct identification is defined as when the device identifies the required radionuclide (i.e., the radionuclide listed in Appendix 1 Table 3). The device may also identify additional acceptable radionuclides that are present in conjunction with the required radionuclide such as daughter products, known contaminants, and NORM from background. It will not be considered correct to identify only the additional acceptable radionuclides and not the required radionuclide. Any radionuclide(s) that are not present, but are identified by the device, will be assessed as a weakness against BHH- OR-07, but will not constitute as an incorrect identification as long as the
(T or O)
Or Does Not Meet Reference/explan ation device identifies the required radionuclide.
BHH-FR-
06-01
The device shall identify present radionuclides within a certain duration of time from initiation of spectrum collection to display of results.
Within a spectrum acquisition time of 2 minutes and additional analytical time of no more than 15 seconds.
Within a spectrum acquisition time of 30 seconds and additional analytical time of no more than 15 seconds.
None
BHH-FR-
06-02
The device shall allow user to add additional spectrum acquisition time to last spectrum collected and reanalyze results.
Device provides ability for operator to add additional acquisition time to the last spectrum collected.
Objective = Threshold None
BHH-FR-
06-03
The device shall have user adjustable spectrum acquisition duration for identification.
Up to a minimum of 5 minutes.
Ability to start and stop spectrum data collection for an unlimited amount of time
None
BHH-FR-
06-04
The device shall indicate if exposure rate is too high or too low for radionuclide identification.
Provide an indication stating that the radiation level is too low or high to perform an identification.
Provide an indication to move closer or further away to obtain a radiation field conducive to identification.
In accordance with ANSI 42.34-2015 section 6.8
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
06-05
The device shall be able to identify Special Nuclear Material (SNM) while masked by NORM, Industrial or Medical sources.
SNM correctly identified in the masking cases listed in TCS tables 8, 9 and 10 at a 5:1 [NORM or Industrial or Medical]:SNM masking ratio.
SNM correctly identified in the masking cases listed in TCS tables 8, 9 and 10 at a 10:1 [NORM or Industrial or Medical]:SNM masking ratio.
Tested In accordance with the DNDO Technical Capability Standard for Handheld Instruments Used for the Detection and Identification of Radionuclides, October 2011
06-06
The device shall be able to identify three non-NORM radionuclides at the same time.
Correctly identify three non-NORM radionuclides of those listed in Radionuclide Table - Appendix 1 Table 4.
Objective = Threshold Each radionuclide will contribute 50 µR/hr net of background and will be unshielded.
06-07
System shall have high gamma energy resolution.
Detector employed shall have no more than 4% energy resolution as measured on the Cs- 137 662 keV peak.
Objective = Threshold
Resolution greater than that of NaI is needed to resolve spectral features related to nuclear smuggling scenarios.
Resolution (Re) is obtained from the peak full width at half of the maximum height (FWHM) using: Re = FWHM*100 / E, where E is the related energy of the peak.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
06-08
System shall provide a message indicating that it cannot identify SNM because of excessive count rates from masking sources.
For masking ratios greater than 5:1, If the instrument is unable to identify the presence of HEU, WGPu, or DU at the fluence rates for the masking sources due to excessive count rates (at the energy region of interest for HEU, WGPu, and DU), then the instrument shall provide a message indicating that it cannot identify HEU, WGPu or DU.
Objective = Threshold
Masking sources and ratio are defined in Table 8, Table 9 and Table 10 of the DNDO Technical Capability Standard for Handheld Instruments Used for the Detection and Identification of Radionuclides, October, 2011.
BHH-OR-
System shall not display false identifications.
No non-present sources identified and displayed.
Objective = Threshold
False identification refers to the identification of a radionuclide that is not present, regardless of presence of source material. Identification of present source material and additional acceptable radionuclides that are present in conjunction with the source material such as daughter products, known contaminants, and NORM from background do not constitute a false identification. Declaring “unknown” for a present, but unidentified radionuclide does not constitute a false identification.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall report identification results to the operator.
Display identification results as element symbol and mass number.
Display identification results as full element name and mass number.
None
BHH-FR-
08-01
The device shall notify operator if it indicates the presence of a source, but is unable to identify a radionuclide.
Display an indication (e.g. “unknown”) when a radionuclide(s) is present and cannot be identified.
Objective = Threshold
Per ANSI N42.34-2015 section 6.8.1, expose the instrument to a radionuclide that is not in the library and verify that result is unknown. Additionally, When an "unknown" and additional identifiable radionuclides are present, ensure the device correctly displays an indication of "unknown" plus the identification of identified radionuclides.
08-02
The device shall display all radionuclides in a prioritized list, if there are two or more radionuclides identified at the same time.
Radionuclides are displayed in correct priority order based on Radionuclide Table - Appendix 1 Table 5.
Objective = Threshold An indication of "unknown" is a priority 2 for display purposes.
BHH-FR-
08-03
The device shall assign and display a unique identifier/name for each spectrum and associated identification result.
Display unique identifier. Objective = Threshold
A unique identifier/name is required to easily reference and find spectral files and identification results generated and stored by the device. For example, the identifier can be a sequential number (i.e., Event 0001, Event 0002, Event 0003…). A unique identifier must be available for each spectrum up to the maximum number that can be stored in the device (see FR-14-01).
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall categorize identified radionuclides.
Categorize as specified in the Radionuclide Table - Appendix 1 Table 4.
Objective = Threshold None
BHH-FR-
09-01
The device shall display radionuclide category as part of displaying identification results.
Correctly display the category of the identified radionuclides as specified in Radionuclide Table - Appendix 1 Table 4.
Objective = Threshold None
BHH-FR-
09-02
The device shall allow operator to set the categories in restricted (password protected) mode.
Ability to set categories in restricted mode.
Objective = Threshold None
BHH-OR-
System shall provide personal radiation safety information to the operator.
In accordance with
ANSI N42.34-2015,
Section 5.14.
Objective = Threshold None
BHH-FR-
10-01
The device shall display gamma exposure and / or dose rate.
Display exposure and / or dose rate. Objective = Threshold None
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
10-02
The device shall alarm on gamma exposure and / or dose rates above the user-definable safety alarm setting.
Distinct audible, visible, and tactile alarm when the radiation field exceeds alarm setting.
Objective = Threshold
IAW ANSI N42.34-2015, section 5.14.
The intent of this alarm is for personal safety. The safety alarm is a separate and unique alarm to those for detection and identification (OR-11).
10-02a
The device shall provide a user-definable safety alarm setting.
Safety alarm settable in the restricted access mode.
Objective = Threshold None
BHH-FR-
10-03
The device shall measure exposure rates over a wide range.
0.1 µR/hr-25mR/hr 0.1 µR/hr-1R/hr None
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
10-04
The device shall provide an exposure rate over-range indication.
When exposed to radiation fields that are greater than the maximum designed exposure rate, the device shall remain
on. The detector shall display an indication that the instrument is at the maximum of its exposure rate range and a separate over-range indication is to be displayed for the duration of the over-range exposure.
Objective = Threshold None
BHH-OR-
System shall provide notifications of detections and identifications in overt and covert / low-visibility operations.
Visible, audible, and tactile selectable alarm notification options.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
11-01
The device shall have a visible display
Incorporate a display and operational controls which are part of the handheld unit.
Objective = Threshold
Clarification: Sole display configurations are only acceptable if the display and detector are a physically integrated single system.
Use of a separate system, such as a cell phone or tablet, which acts as the system’s only display and operations control is unacceptable.
11-02
The device shall be equipped with an adjustable audio volume function.
Volume adjustable from muted/off up to a minimum of 85 dB(A), maximum of 100 dB(A) at 30 cm.
Objective = Threshold Audio volume measurements will be taken 30 cm normal to the device’s primary speaker.
11-03
The device shall have a selectable tactile alarm feature.
Device shall provide option for vibration alerts to operator for detection.
Objective = Threshold None
BHH-FR-
11-04
The device’s visible indications shall be readable in ambient indoor and outdoor lighting conditions.
Visible in low light (<150 lx) and bright light (˃10,000 lx).
Objective = Threshold None
BHH-FR-
11-05
Selectable alarm notification options shall be available without use of a password.
Ability to turn on and off visible, audible, and tactile alarms in a routine (no password) mode.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
12 KPP
System shall support reachback operations.
Transmit gamma spectral data (background, unknown, and a reference spectrum) from detection device to reachback via existing communications methods.
Transmit gamma spectral data (background, unknown, and a reference spectrum) directly from detection device to reachback.
Existing communications include (1) personal or laptop computer, windows tablet and email; (2) satellite phone; (3) Cell phone.
12-01
The device shall be able to transfer/download data to component computers (i.e., PC/laptops/tablets).
Connect with and download stored data via direct USB connection from device to computers using, Windows 7 and higher (to include Windows 10) and Win Server 2012.
Wireless connection to computers.
The ability to conduct data transfer without installation of additional software application is desired.
BHH –
FR-12-02
The device shall be able to transfer data via the component's approved satellite phone communications system.
Direct connection [without use of an intermediate device] and data transmission via Iridium 9575 satellite phones.
Direct connection [without use of an intermediate device] and data transmission via Iridium 9575 satellite phones.
None
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
12-03
The device shall be able to transfer data via component deployed cellular phones.
Connect with and download stored data to smart phones using Android 6.0 and higher and iOS 9.3 and higher.
Objective = Threshold None
BHH-FR-
12-04
The system shall encrypt data when transferred to DHS information systems using “wireless” means (i.e., wireless, Bluetooth, infrared).
Use encryption that is
FIPS 197 (AES‐256)
compliant and has received
FIPS 140‐2
validation.
Objective=Threshold
Wireless communications to PC, laptop, cell phone shall implement strong identification, authentication and encryption. Systems requiring encryption shall comply with at least one of the following methods:
• Products using FIPS 197 Advanced Encryption Standard (AES) algorithms with at least 256 bit encryption that have been validated under FIPS 140-2
• National Security Agency (NSA) Type 2 or Type 1 encryption (Note: The use of triple Data Encryption Standard [3DES] and FIPS 140-1 is no longer permitted.)
12-05
The device shall allow the operator to choose which communication modes are active.
Ability to enable and disable the communication modes (e.g.
Bluetooth, wireless WiFi, cellular, GPS) capability in only the restricted (password protected) mode.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
12-06
The ability to conduct any available reachback method shall be readily accessible to the operator.
Ability to view and transfer identification data and results in the routine mode (no password required).
Objective = Threshold None
BHH-FR-
12-07
The device shall support voice-only reachback.
Ability for an operator to access, in the routine operating mode, spectral features of the measured gamma energy spectrum that were used to make the identification determination, e.g.
tables of peak properties.
A “Manual mode” where the operator can accumulate a spectrum of desired duration and access spectral features with a pointer-like interface.
Spectral features of interest: Peak energy centroids in keV, peak amplitudes in sigma-units, peak FWHM in %, peak total counts.
12-08
Device shall transmit data in recognizable format.
Transmit data in
ANSI N42.42-2012
format.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall continuously upload radiation scan and alarm data to a remote repository.
Through an intermediate device (e.g., cell phone, web enabled periphery) send every 1 second an N42.42 formatted 1 second real-time gamma spectral file, neutron count rate, alarm information (if in alarm state), time and geographical location.
Not using an intermediate device, send every 1 second an N42.42 formatted 1 second real-time gamma spectral file, neutron count rate, alarm information (if in alarm state), time, and geographical location.
Reference SIGMA Edge API, Release 8.0.0-SNAPSHOT, March 31, 2018.
BHH-FR-
13-01
Maintain a log of sequential radiation scan and alarm data.
Ability to maintain 12 hours of information.
Ability to maintain 24 hours of information.
Minimum data to log includes: 1 second real-time gamma spectral files, neutron count rate readings, associated time and geographical location.
13-02
The device shall link to a periphery device.
Using Bluetooth 4.0 or better and/or WiFi, link to cell phone and/or web enabled periphery.
Objective = Threshold None
BHH-OR-
System shall provide data storage.
Store and transfer detection, identification, and status of system health information.
Objective = Threshold System health data such as cumulative operating time, Built in Test (BIT), etc.
stored in non-volatile memory.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
14-01
The device shall have the ability to store complete spectral files.
Device shall be capable of storing 1000 5-minute spectral acquisition time files with associated identification results.
Objective = Threshold Data format is defined in ANSI N42.42- 2012.
BHH-FR-
14-02
The device shall automatically save spectral files to non-volatile memory upon display of identification results.
Automatically save all spectral files. Objective = Threshold None
BHH-FR-
14-03
The device shall notify operator when storage capacity is almost full.
There is no threshold.
Notify operator when storage capacity is 95% full.
Objective only requirement.
BHH-FR-
14-04
The device shall allow the user in routine (no password) mode to be able to set time and date in any specified time zone.
Time and Date shall be in local time with operator ability to change to "GMT" in routine mode.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
The device shall have two different operating modes:
routine mode (no password) and restricted mode (password protected).
In accordance with
ANSI N42.34-2015,
Section 5.3 (Access Modes). Also include ability to localize, save and transfer data in routine mode.
Objective = Threshold None
BHH-OR-
System shall have a long service life. 10 years 15 years
Intent is for the system to be remain in use and be supported over the entire service life period by warranty, maintenance contracts, and/or software upgrades and not require exchange of limited life components (i.e., calibration/stabilization sources) during the service life period.
System shall become fully functional from a dead start or from standby mode.
< 2 minutes Objective = Threshold ANSI 42.34-2015 5.10 (Warm-up time) provides test method.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall be reliable.
Mean Time Between Failure [MTBF] rate of 8,760 operating hours. (Device shall record operating hours)
Mean Time Between Failure [MTBF] rate of 17,520 operating hours. (Device shall record operating hours)
MTBF is defined for a particular interval, as the total operating time (time that the system is operating or capable of operating) of the population of Basic Handheld devices divided by the total number of failures (requiring corrective maintenance actions).
MTBF can be calculated, but acceptance of stated values is based on an explanation of the methodology used and supporting data. Field data is preferred and, if available, should be provided as justification to any specified MTBF.
System shall support continuous operations.
8 hours 12 hours
Continuous operation refers to: a single shift exclusively using battery power;
remaining responsive without the need to be re-booted after initial start-up.
The device will be continuously measuring radiation, and periodically conducting identifications (for planning purposes assume 6 identifications per hour).
19-01
The device shall have sufficient battery life to support operations at standard ambient temperatures.
8 hours 12 hours
Ambient temperature 18 - 25 degrees C (ANSI 42-34-2015 Table 1) The device will be continuously measuring radiation, and periodically conducting identifications (for planning purposes assume 6 identifications per hour).
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
19-02
If device uses a stand-by power/sleep mode to extend battery life, the device shall maintain functionality.
Continue to detect radiation, alarm, and finish identifications while in stand-by mode.
Objective = Threshold None
BHH-FR-
19-03
The device shall allow for replacement of batteries by operators.
No special tools required. No tools required.
Special tools are tools not readily available or found in a basic, standard tool kit (e.g., of standard tools are pliers, cross point or flathead screw drivers).
19-04
The device allows batteries to be replaced without loss of stored data.
No data loss during battery exchange.
Device cannot be placed on A/C power.
Objective = Threshold None
BHH-FR-
19-05
The device shall display the expected remaining operating battery life.
Display the expected remaining operating battery life (time or percent of life remaining).
Objective = Threshold None
BHH-FR-
19-06
The device shall be operable on A/C Power.
Device operates on A/C. Objective = Threshold None
BBH –
FR-19-07
The device shall have ability to switch between A/C and battery power.
Change between A/C and battery power without having to power down device.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
19-08
While on A/C power the device’s batteries shall be capable of being changed without having to shut down or reboot device.
Objective = Threshold None
BHH-FR-
19-09
The device shall be capable of using rechargeable batteries.
Device is capable of using rechargeable batteries.
Objective = Threshold None
BHH-FR-
19-10
The device shall use replaceable batteries that are commonly available.
Device uses batteries that are commonly and commercially available, non-proprietary and not exclusive to that detector.
Objective = Threshold None
BHH-FR-
19-11
The device shall include a battery charger.
110 volt AC and 12 volt DC charging capability and adapters to connect to standard outlets and power receptacles in commercial vehicles
Objective = Threshold None.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System reliability shall support a high fleet Operational Availability [AO].
0.99 0.994
Operational Availability [AO] indicates the percentage of time that a system is operationally capable of performing an assigned mission and can be expressed as (uptime/ (uptime + downtime)). High fleet operational availability, given that the device requires no scheduled maintenance, is across the entire fleet including those devices unavailable because they are in maintenance or being shipped to and from the maintenance repair facility.
System shall operate without scheduled maintenance.
Maintenance limited to battery replacement and external cleaning in the field without any special tools, procedures, or test equipment.
Objective = Threshold None
BHH-FR-
21-01
The device shall display the software and firmware versions in the device.
Software and Firmware versions displayed upon user command.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
21-02
The device shall not require the use of special tools or test equipment.
No special tools or equipment. Objective = Threshold
Special tools are tools not readily available or found in a standard tool kit (e.g., of standard tools are pliers, cross point or flathead screw drivers). Check sources are those sources required to calibrate/stabilize the device for normal operations. Check sources are not considered special tools or equipment.
System shall notify the operator of degraded operations.
System shall perform self diagnostics at start-up and upon user request and notify operator of any system problems.
Continuous self-check diagnostics during operations and notify operator of any system problems.
Inform operator of system problems and initiate error notification when problems arise. Users shall be able to request the unit perform diagnostics while in a routine (no password) mode.
22-01
The device shall have the ability to self-calibrate.
Built-in calibration capacity. Objective = Threshold None
BHH-FR-
22-02
The device shall maintain a diagnostics status file.
Record diagnostics status, malfunctions, associated date and time in data file.
Objective = Threshold None
BHH-FR-
22-03
The device shall monitor mechanical shock and impact experienced by the device.
Measure, record, and output shock and impact data.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall be ruggedized.
In accordance with
ANSI N42.34-2015
Section 9 [Vibration, Mechanical Shock, Impact, and drop] except the drop will be from a height of
1.5 meters onto each surface of the devices.
In accordance with
ANSI N42.34-2015
Section 9 [Vibration, Mechanical Shock, Impact, and drop] except the drop will be from a height of 3 meters onto each surface of the devices.
Use MIL-STD-810G Method 516.6 Procedure IV modified for drop from specified heights for guidance.
23-01
System shall survive exposure to vibrations
Physical condition and functionality not affected during or after exposure to random vibration at
0.01 g2/Hz (spectral density) using 5 Hz and 500 Hz as the frequency endpoints for a period of 1 h in each of three orthogonal directions.
Objective = Threshold Tested IAW ANSI N42.34-2015, Section 9.1
BHH-FR-
23-02
System shall survive mechanical shock.
Physical condition and functionality not affected after exposure to shock pulses of 50g peak acceleration in three orthogonal orientations.
Objective = Threshold Tested IAW ANSI N42.34-2015, Section 9.2
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
23-03
System shall survive impacts.
Physical condition and functionality not affected during or after being subjected to three 0.2 Joule intensity impacts on each side, top and bottom of the device.
Objective = Threshold Tested IAW ANSI N42.34-2015, Section 9.3
BHH-FR-
23-04
System shall survive drops.
Physical condition and functionality not affected after drops from 1.5 meters onto concrete, on each side, top and bottom of the device.
Physical condition and functionality not affected after drops from 3 meters onto concrete, on each side, top and bottom of the device.
Tested IAW ANSI N42.34-2015, Section 9.4 as amended by the requirement
23-05
The device shall be provided with an impact resistant case for shipping and storage of the device and all required components.
Provide a single impact resistant case for protection of the device and all required components during shipping and storage.
Objective = Threshold
Should consider need to remove battery from device for shipment.
Required components may include spare battery and external battery packs, manuals, charging adapters, and cables.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
23-06
System shall minimize use of external shock absorption.
External shock absorption restricted to a sleeve that fits over the unit, with no external snagging hazards.
No external shock absorption.
Intent is to have a system that minimizes external shock absorption features. External shock absorption features can add bulk and create a snagging hazard. Preclude creating external snagging hazards in design of any external shock absorption methods used (e.g., a rubber sleeve that fits over the unit) . External bumpers which fit only over a small part of the unit do not meet this requirement. Best case is that no external shock absorption is needed to meet the drop and impact requirements; all mitigation is internal to the device.
System shall operate without degradation in all operational environments.
In accordance with
ANSI N42.34-2015,
Sections 7.1-7.5 and
IEC 60529
Temperature, Temperature Shock, Humidity, Dust and Moisture Protection, and Extreme Temperature Startup.
Objective = Threshold Operational environments include indoor and outdoor (all-weather) conditions.
24-01
The device shall operate without degradation in the operational environment in extreme temperature.
In accordance with
ANSI N42.34-2015,
Section 7.1.except lower temperature changed to -30 degrees C.
In accordance with
ANSI N42.34-2015,
Section 7. 1. except lower temperature changed to -40 degrees C.
None
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
24-02
The device shall operate without degradation in the operational environment during temperature shocks.
In accordance with
ANSI N42.34-2015,
Section 7.2. except lower temperature changed to -30 degrees C.
In accordance with
ANSI N42.34-2015,
Section 7.2. except lower temperature changed to -40 degrees C.
None
BHH-FR-
24-03
The device shall operate in the operational environment without degradation under humidity conditions.
In accordance with
ANSI N42.34-2015,
Section 7.3.
Objective = Threshold None
BHH-FR-
24-04
The device shall be able to operate when switched on at extreme temperature limits.
In accordance with
ANSI N42.34-2015,
Section 7.5. (−20 degrees C and 50 degrees C).
In accordance with
ANSI N42.34-2015,
Section 7.5 except cold temperature extreme is -30degress C.
None
BHH-FR-
24-05
The device shall operate without degradation in the operational environment in accordance with International Electrotechnical Commission (IEC) standard IEC 60529 for Ingress Protection (IP) for dust.
Coded for tight dust protected (IP-65). Objective = Threshold
(T or O)
Or Does Not Meet Reference/explan ation
BHH-FR-
24-06
The device shall operate without degradation in the operational environment in accordance with International Electrotechnical Commission (IEC) standard IEC 60529 for Ingress Protection (IP) for water.
Coded for spraying water (IP-65). Objective = Threshold
When the device is in the operational orientation, low pressure water jets from any direction shall have no harmful effects.
System shall be safe to operate.
Pass Underwriters Laboratories (UL) Certification 61010- 1V3 certification or equivalent.
Objective = Threshold None
BHH-OR-
Systems used shall be able to operate in hazardous/explosive environment in support of maritime operations [Maritime Only].
N/A. Objective only.
System shall be certified to operate in UL913 (Class II) hazardous locations.
If certification is claimed, documentation is to be provided.
Certification is to be based on UL-913- 2004. In accordance with reference:
"Naval Ships' Technical Manual Chapter 074 Volume 3".
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall operate without degradation in a Radio Frequency Interference and Electro Magnetic Interference environment.
In accordance with
ANSI N42.34-2015,
Section 8.1, 8.3, and 8.4.
Objective = Threshold Electrostatic Discharge (ESD), RF Susceptibility, Magnetic Fields
BHH-FR-
27-01
System shall not be affected by radio-frequency fields over the frequency range of 80 MHz to 2500 MHz
IAW ANSI N42.34-
2015, Section 8.1 Objective = Threshold
BHH-FR-
27-02
System shall be fully functional when exposed to a 100 A/m (1.3 G) 60 Hz magnetic field
IAW ANSI N42.34-
2015, Section 8.3 Objective = Threshold
BHH-FR-
27-03
System shall not be affected by exposure to electrostatic discharges (ESDs) at intensities of up to ±6 kV
IAW ANSI N42.34-
2015, Section 8.4 Objective = Threshold
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall not interfere with or degrade operations of other DHS or component systems due to radiated electromagnetic emissions.
In accordance with
ANSI N42.34-2015,
Section 8.2.
Objective = Threshold None
BHH-OR-
System shall be light weight in operational configuration.
≤5lbs ≤3lbs
Operational configuration is the configuration required for normal use under battery power, including any external shock absorption. Threshold & Objective set in accordance with MIL- STD-1472G, Section 5.9.17.3.4 "Weight and Dimensions" for portable test equipment.
System shall be operable with one hand.
Operator can perform detection, localization, and identification functions using one hand.
Objective = Threshold Examples include menu-driven or icon-based interfaces.
BHH-OR-
System shall be operable by a user wearing gloves.
Operator can perform detection, localization, and identification functions while wearing gloves.
Objective = Threshold Gloves are those typical worn for thermal protection and nuclear, biological, chemical protection.
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System’s display screen shall be readable by a user wearing polarized sunglasses.
User able to read display screen. Objective = Threshold None
BHH-OR-
System shall permit maneuverability in handheld scanning missions.
Operator movement is not obstructed while conducting scanning missions.
Objective = Threshold
Human factors and ergonomic considerations to minimize operator injury and fatigue, such as device shape, weight balance and handle design will be evaluated as part of this requirement. Example missions include ship boarding (from helicopter or a component vessel), scanning cargo containers, screening of personal vehicles, and scanning onboard small vessels, airplanes, buses, and trains.
33-01
The device shall have small physical dimensions.
< 13in x 9in x 6in < 10in x 4 in x 3in None
BHH-OR-
System shall be transportable by a single operator during horizontal missions.
Hands free transportation in operational configuration during horizontal missions.
Objective = Threshold
Not to interfere with the operator's movement or activities. Horizontal missions are those where an operator is walking, riding or moving around and over small obstacles not requiring use of hands for stability or propulsion (e.g., walking patrols, getting on and off buses and trains).
(T or O)
Or Does Not Meet Reference/explan ation
BHH-OR-
System shall be transportable during extended vertical missions.
Transport in operational configuration during vertical missions that exceed 8.5 feet.
"Hands Free" single operator vertical transport during missions that exceed
8.5 feet.
Vertical missions are those involving ascent or descent onto a platform or vessel where the operator must use hands for stabilization or propulsion (e.g., Jacob’s ladders, fast roping).
Appendix 1: Tables referenced in the requirements
Table 1. Radionuclides for Requirement BHH-OR-06a
Radionuclide Threshold (T) or Objective (O) for Unshielded Sources
Americium-241 [Am-241] T
Cesium-137 [Cs-137] T
Cobalt-60 [Co-60] T
Iridium-192 [Ir-192] T
Neptunium-237 [Np-237] T
Plutonium - 239 [Pu-239] T
Selenium-75 [Se-75] T
Uranium-235 [U-235] T
Uranium-238 [U-238] T
Table 1 Note: Requirement BHH-OR-06a is a Key Performance Parameter (KPP). All radionuclides within the requirement are considered KPP.
Table 2. Radionuclides for Requirement BHH-OR-06b
Radionuclide Threshold (T) or Objective
(O) for Unshielded Sources
Barium-133 [Ba-133] T
Cesium-131 [Cs-131] T
Cobalt-57 [Co-57] T
Fluorine-18 [F-18] T
Gallium-67 [Ga-67] T
Iodine-123 [I-123] T
Iodine-125 [I-125] T
Iodine-131 [I-131] T
Lathanum-138 [La-138] T
Lutetium-177 [Lu-177] T
Palladium-103 [Pd-103] T
Plutonium-238 [Pu-238] T
Potassium-40 [K-40] (a) T
Radium-223 [Ra-223] T
Radium-226 [Ra-226 + Daughters(a)] T
Selenium-75 [Se-75] T
Strontium-90/Yittrium-90 [Sr-90/Y-90] (b, c) T
Technetium-99m [Tc-99m] T
Thallium-201 [Tl-201] T
Thorium-232 [Th-232 + Daughters (a)] T
Uranium-232/Uranium-233 [U-232/U-233] (b) T
Scandium-46 [Sc-46] O
Tungsten-187 [W-187] O
Antimony-124 [Sb-124] O
Cromium-51 [Cr-51] O
Europium-152 [Eu-152] O
Hydrogen Neutron Capture O
Indium-111 [In-111] O
Lutetium-176 [Lu-176] O
Manganese-54 [Mn-54] O
Manganese-56 [Mn-56] O
Molybdenum-99 [Mo-99] O
Niobium-95 [Nb-95] O
Samarium-153 [Sm-153] O
Strontium-89 [Sr-89] (c) O
Thallium-202 [Tl-202] O
Thallium-204 [Tl-204] O
Yittrium-88 [Y-88] O
Table 2 Notes:
(a) When found at background levels, these radionuclides should be saved in background spectral files but do not need to be identified on the screen or separately identified in radioisotope ID spectral files.
(b) Any of the Identification Names listed are acceptable for screen display.
(c) “Beta Emitter” may be displayed in lieu of the name of the radionuclide.
Table 3. Radionuclides for Requirement BHH-OR-06c
Radionuclide Threshold (T) or Objective
(O) for Shielded Sources
Americium-241 [Am-241] T
Barium-133 [Ba-133] T
Cesium-137 [Cs-137] T
Cobalt-57 [Co-57] T
Cobalt-60 [Co-60] T
Iridium-192 [Ir-192] T Neptunium-237 [Np-237] T Plutonium - 239 [Pu-239] T
Plutonium-238 [Pu-238] T
Potassium-40 [K-40] (a) T
Radium-226 [Ra-226 + Daughters(a)] T
Selenium-75 [Se-75] T
Thorium-232 [Th-232 + Daughters (a)] T
Uranium-232/Uranium-233 [U-232/U-233] (b) T
Uranium-235 [U-235] T
Uranium-238 [U-238] T
Scandium-46 [Sc-46] O
Tungsten-187 [W-187] O
Antimony-124 [Sb-124] O
Cesium-131 [Cs-131] O
Cromium-51 [Cr-51] O
Europium-152 [Eu-152] O
Fluorine-18 [F-18] O
Gallium-67 [Ga-67] O
Hydrogen Neutron Capture O
Indium-111 [In-111] O
Iodine-123 [I-123] O
Iodine-125 [I-125] O
Iodine-131 [I-131] O
Lathanum-138 [La-138] O
Lutetium-176 [Lu-176] O
Lutetium-177 [Lu-177] O
Manganese-54 [Mn-54] O
Manganese-56 [Mn-56] O
Molybdenum-99 [Mo-99] O
Niobium-95 [Nb-95] O
Palladium-103 [Pd-103] O
Radium-223 [Ra-223] O
Samarium-153 [Sm-153] O
Strontium-89 [Sr-89] (c) O
Strontium-90/Yittrium-90 [Sr-90/Y-90] (b, c) O
Technetium-99m [Tc-99m] O
Thallium-201 [Tl-201] O
Thallium-202 [Tl-202] O
Thallium-204 [Tl-204] O
Yittrium-88 [Y-88] O
Table 3 Notes:
(a) When found at background levels, these radionuclides should be saved in background spectral files but do not need to be identified on the screen or separately identified in radioisotope ID spectral files.
(b) Any of the Identification Names listed are acceptable for screen display.
(c) “Beta Emitter” may be displayed in lieu of the name of the radionuclide.
Table 4. Radionuclides for Requirements BHH-FR-06-06, BHH-OR-09, and BHH-FR-09-01
Radionuclide Category
Neptunium-237 [Np-237] SNM
Plutonium - 239 [Pu-239] SNM
Uranium-232/Uranium-233 [U-232/U-233] SNM
Uranium-235 [U-235] SNM
Americium-241 [Am-241] SNM-Related (a)
Hydrogen Neutron Capture SNM-Related (a)
Plutonium-238 [Pu-238] SNM-Related (a)
Uranium-238 [U-238] SNM-Related (a)
Scandium-46 [Sc-46] IND
Tungsten-187 [W-187] IND
Antimony-124 [Sb-124] IND
Barium-133 [Ba-133] IND
Cesium-137 [Cs-137] IND
Cobalt-57 [Co-57] IND
Cobalt-60 [Co-60] IND
Europium-152 [Eu-152] IND
Iridium-192 [Ir-192] IND
Manganese-54 [Mn-54] IND
Manganese-56 [Mn-56] IND
Molybdenum-99 [Mo-99] IND
Niobium-95 [Nb-95] IND
Strontium-90/Yittrium-90 [Sr-90/Y-90] IND
Thallium-204 [Tl-204] IND
Yittrium-88 [Y-88] IND
Selenium-75 [Se-75] IND
Cesium-131 [Cs-131] MED
Cromium-51 [Cr-51] MED
Fluorine-18 [F-18] MED
Gallium-67 [Ga-67] MED
Indium-111 [In-111] MED
Iodine-123 [I-123] MED
Iodine-125 [I-125] MED
Iodine-131 [I-131] MED
Lutetium-177 [Lu-177] MED
Molybdenum-99 [Mo-99] IND
Palladium-103 [Pd-103] MED
Radium-223 [Ra-223] MED
Samarium-153 [Sm-153] MED
Strontium-89 [Sr-89] MED
Technetium-99m [Tc-99m] MED
Thallium-201 [Tl-201] MED
Thallium-202 [Tl-202] MED
Lathanum-138 [La-138] NORM
Lutetium-176 [Lu-176] NORM
Potassium-40 [K-40] NORM
Radium-226 [Ra-226 + Daughters] NORM
Thorium-232 [Th-232 + Daughters] NORM
Table 4 Notes:
(a) ‘SNM-Related’ is the desired Category name for screen display, but a similar designation is also acceptable.
Table 5. Radionuclides for Requirement BHH-FR-08-02
Radionuclide Display Priority
Americium-241 [Am-241] 1
Cobalt-60 [Co-60] 1
Hydrogen Neutron Capture 1
Neptunium-237 [Np-237] 1
Plutonium - 239 [Pu-239] 1
Plutonium-238 [Pu-238] 1
Uranium-232/Uranium-233 [U-232/U-233] 1
Uranium-235 [U-235] 1
Uranium-238 [U-238] 1
Cesium-137 [Cs-137] 2
Iridium-192 [Ir-192] 2
Selenium-75 [Se-75] 2
Strontium-90/Yittrium-90 [Sr-90/Y-90] 2
Scandium-46 [Sc-46] 3
Tungsten-187 [W-187] 3
Antimony-124 [Sb-124] 3
Barium-133 [Ba-133] 3
Cesium-131 [Cs-131] 3
Cobalt-57 [Co-57] 3
Cromium-51 [Cr-51] 3
Europium-152 [Eu-152] 3
Fluorine-18 [F-18] 3
Gallium-67 [Ga-67] 3
Indium-111 [In-111] 3
Iodine-123 [I-123] 3
Iodine-125 [I-125] 3
Iodine-131 [I-131] 3
Lathanum-138 [La-138] 3
Lutetium-176 [Lu-176] 3
Lutetium-177 [Lu-177] 3
Manganese-54 [Mn-54] 3
Manganese-56 [Mn-56] 3
Molybdenum-99 [Mo-99] 3
Niobium-95 [Nb-95] 3
Palladium-103 [Pd-103] 3
Radium-223 [Ra-223] 3
Samarium-153 [Sm-153] 3
Strontium-89 [Sr-89] 3
Technetium-99m [Tc-99m] 3
Thallium-201 [Tl-201] 3
Thallium-202 [Tl-202] 3
Thallium-204 [Tl-204] 3
Yittrium-88 [Y-88] 3
Potassium-40 [K-40] 4 Radium-226 [Ra-226 + Daughters] 4 Thorium-232 [Th-232 + Daughters] 4