Attachment 3 - Purchase Description CDC SPIROMETRY LAB.docx
DOCX document 2 MB Posted
- Attached to
- Mobile (1x) and Spirometry (1x) Labs for the CDC
- Solicitation number
- 47QMCA27Q0006
- Issued by
-
GSA Federal Acquisition Service
About this file
This is a Purchase Description document for a Mobile Spirometry Lab required by the Centers for Disease Control and Prevention (CDC). The spirometry lab must be a fully operational, mobile unit capable of conducting pulmonary function testing and respiratory assessments in field settings. The lab should include essential spirometry equipment, calibration devices, quality control mechanisms, and data management systems compliant with American Thoracic Society (ATS) and European Respiratory Society (ERS) standards. The unit must be transportable, self-contained, and equipped with necessary power systems, environmental controls, and safety features to support clinical-grade respiratory testing operations.
The mobile spirometry lab shall accommodate patient testing workflows, staff workstations, and secure storage for sensitive respiratory data and testing materials. Equipment specifications include spirometers, bronchodilator delivery systems, oxygen systems, and supporting diagnostic instruments. The lab must meet infection control standards, maintain temperature and humidity controls suitable for accurate testing, and provide adequate space for patient comfort and assessment privacy. Documentation and record-keeping capabilities must integrate with CDC data systems, and all equipment must comply with applicable FDA regulations, CLIA requirements, and occupational safety standards for laboratory operations.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 1 - RFQ 47QMCA27Q0006 and SF18-95a.pdf | ||
| Attachment 6 - GSA Fleet Systems Account Form.pdf | ||
| Attachment 7 - Order Management User Guide.pdf | ||
| Attachment 2 - Purchase Description CDC X-RAY LAB.docx | DOCX document | |
| Attachment 4 - Technical Exceptions Worksheet.xlsx | XLSX spreadsheet | |
| Attachment 8 - Letter of Supply.docx | DOCX document | |
| Attachment 5 - Brand Name or Equal Worksheet.xlsx | XLSX spreadsheet |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Purchase Description Mobile Spirometry Lab
CDC-NIOSH
Coal Workers Health Surveillance Program
19 October 2026
Note to Supplier: Supplier shall explicitly and completely identify any and all exceptions taken to the requirements of this specification. Drawings, literature, and any other information submitted with an offer do not constitute a stated or implied exception unless specifically identified in writing as an exception and accepted and implemented by an amendment to the solicitation or modification to a contract. Any exception deemed acceptable by the Government shall only apply to the specific item, requirement, etc. cited, and shall not extend to any other requirement.
Supplier offering other than brand name items identified below, when an “or equal” may be acceptable, shall furnish with their offers all technical data information, product descriptions, etc. to ensure that a determination may be made as to the equality of the product(s) offered (see clause titled Brand “Name or Equal” set forth in Section FC 11.204 of the Federal Acquisition Regulation Companion V2.0) The following shall be provided with the quote.
1. Detailed explanation, listed in a separate letter, of any exceptions to the specification, including alternate equipment proposed or a declaration of comprehensive understanding of the requirements with no exceptions taken.
2. The Supplier shall submit with the offer preliminary drawings that show the following:
a. Layout views of the quoted vehicle including all equipment locations and basic dimensions verifying compliance with the requirements herein.
b. Exterior front, rear, and side view with overall dimensions.
c. Estimated curb weight and Gross Vehicle Weight Rating (GVWR)
d. Documented evidence of quality assurance procedures or system
1.0 SCOPE
The purpose of this vehicle is to provide a site for conducting interviews with workers participating in medical examinations and for performing spirometry examinations.
This purchase description is for a custom manufactured vehicle to serve as a mobile medical clinic. The vehicle shall be based on the design solution proposed by the Supplier and approved by GSA + CDC.
2.0 APPLICABLE DOCUMENTS
2.1 FEDERAL STANDARDS AND CODES
· 49 CFR, part 38, subpart B – Americans with Disabilities Act (ADA)
· 42 CFR, part 37, Specifications for Medical examinations of Coal Miners
· 49 CFR, parts:
· 393 (FMCSR)
· 571 (FMVSS)
Application for copies of these publications should reference the code of Federal Regulations, 42 and 49 CFR, and the Federal Register and should be addressed to the Superintendent of Documents, U.S. Government Printing Office, Washington, D.C. 20402. They may also be accessed on the Internet through GPO Access at http://www.access.gpo.gov.
| 2.2 | INDUSTRIAL STANDARDS |
| A. | NFPA 70: NATIONAL ELECTRICAL CODE |
Application for copies of the standard should apply to NFPA, 1 Batterymarch Park, PO Box 9101, Quincy, MA 02269-9101, An International Codes and Standards Organization
B. SOCIETY OF AUTOMOTIVE ENGINEERS, INC.
· J537 - Storage Batteries.
· J551 - Measurement of Electromagnetic Radiation from Motor Vehicles and Devices.
· J553 - Circuit Breakers
· J575 - Tests for Motor Vehicle Lighting Devices & Components.
· J576 - Plastic Materials, For Use In Optical Parts Such As Lenses and Reflectors of Motor Vehicle Lighting Devices.
· J578 - Color Specification for Electric Signal Lighting Devices.
· J682 - Rear Wheel Splash and Stone Throw Protection.
· J702 - Brake and Electrical Connection Locations – Truck-Tractor and Truck-Trailer
· J994 - Backup Alarms, Performance Test and Application.
· J1054 -Warning Lamp, Alternating Flashers.
· J1127 -Battery Cable.
· J1128 -Low Tension Primary Cable.
· J1292 -Automobile, Truck, Truck-Tractor, Trailer, and Motor Coach Wiring.
Applications for copies of SAE publications should be addressed to SAE, Inc., 400 Commonwealth Drive, Warrendale, PA 15096
C. AMERICAN NATIONAL STANDARDS INSTITUTE (ANSI)
ANSI S1.4-1983 (R2006)/ANSI S1.4A-1985 (R2006) Specification for Sound Level Meters ANSI standards may be found at: http://webstore.ansi.org
2.3 OTHER STANDARDS
RP 803-Pre-Service Vehicle Inspection (Published by The Technical and Maintenance Council of the American Trucking Association, 2200 Mill Road, Alexandria, VA 22314)
2.4 ORDER OF PRECEDENCE
In the event of a conflict between the text of this purchase description and the references cited above, the most stringent requirements, as determined by the government, takes precedence. Nothing in this purchase description, however, supersedes applicable state and federal safety laws and regulations and the requirements set forth in the vehicle manufacturer’s Bodybuilders and Incomplete Vehicle Manuals.
3.0 REQUIREMENTS
Materials shall be as specified herein. When materials are not specified, the vehicle and all parts there shall be furnished to provide the intended function, durability, safety, and maintain good long-term appearance. All materials shall be new, shall be suitable for the intended purpose, and shall be free of any characteristics or defects in material and workmanship, which may affect the performance, function, durability, and serviceability of the finished vehicle, or detract from its appearance. The government reserves the right to make the final determination of the suitability of all components and their arrangement on and in the vehicle. The contractor shall not proceed with first production vehicle manufacture until approval to proceed is given by the government.
Supplier shall have documented quality assurance processes to assure compliance with all specification requirements including requirements established by reference, i.e. SAE, FMVSS, etc. Supplier shall submit, with their offer, a copy of their documented processes or procedures.
The Supplier (manufacturers) who respond to this solicitation shall have a minimum of five (5) years of engineering, design, and manufacturing experience in providing similar vehicles with equal or greater complexity to the requirements described herein. Supplier shall provide 24 hour/7days per week access to customer service. Supplier shall provide detailed information describing their customer service access.
The Supplier shall ensure that the application and installation of major subcomponents and systems is compliant with subcomponent vendors’ requirements and recommendations.
3.1 COMPLETED VEHICLE COMPONENT SPECIFICATIONS
The components, assemblies, and accessories to be delivered under the contract shall meet or exceed the requirements of this purchase description. All items shall be as represented in the manufacturer's technical data book. Special or mounted equipment shall be as represented in the equipment manufacturer's technical data. No removal, substitution or alternation of the vehicle manufacturer's standard or optional components shall be made.
3.1.1 MATERIALS
Materials used shall be free from defects which would adversely affect the performance or maintainability of individual components or of the overall assembly. Materials not specified herein shall be of the same quality used for the intended purpose in commercial practice. Unless otherwise specified herein, all equipment, material, and articles incorporated in the work covered by this specification are to be new and fabricated using materials produced from recovered materials to the maximum extent practicable without compromising the intended use. The term “recovered materials” means materials which have been collected or recovered from solid waste and reprocessed to become a source of raw materials, as opposed to virgin raw materials. Unless otherwise specified, none of the above shall be interpreted to mean that the use of used, repaired, or rebuilt products is allowed under this specification.
3.1.2. PROHIBITED MATERIAL
Asbestos materials shall not be used in any form in any part of the vehicle. No item, part, or assembly shall contain radioactive materials in which the specific activity is greater than 0.002 micro curie per gram or activity per item equals or exceeds 0.01 micro curie.
Wood products shall not be used for any interior finishes and cabinetry, unless specified within.
3.1.3 DISSIMILAR METALS
The use of dissimilar metals shall be avoided whenever practicable. When it is necessary to combine dissimilar metals, a permanent dielectric medium compatible to both metals shall be used to prevent corrosion caused by galvanic action.
3.1.4 FLAME RESISTANCE
Flame resistance of all materials shall be in accordance with Federal Motor Vehicle Standard Number 302.
3.1.5 MOLD AND FUNGUS PROTECTION
Materials that can support mold or fungus growth shall not be used in any part of the vehicle.
3.1.6 VOLATILE ORGANIC COMPOUNDS
All materials including but not limited to paints, insulation, glues and adhesives shall have low or no VOC emissions rating. For paint, low VOC levels are 150 g/l or less for non-flat finishes and 100 g/l or less for flat finishes. For other materials including but not limited to insulation, glues and adhesives, manufacturer must state material has a rating of low or no VOC. Contractor shall request and receive approval from the Government before using materials or compounds containing toluene and/or styrene.
3.2 GENERAL DESIGN
3.2.1 ACCESSIBILITY
The design of the vehicle, including installation of equipment, shall permit access for routine servicing and shall permit access for replacement and adjustment of component parts and accessories with minimal disturbance of other components and systems.
3.2.2 RATINGS
The vehicle ratings shall be the manufacturer’s published ratings. Component and chassis ratings shall not be altered to meet the requirements of this purchase description.
3.2.3 RADIO INTERFERENCE SUPPRESSION
The 12-volt electrical system shall be shielded to limit electromagnetic radiation according to SAE J551 or the most current standard.
3.3 WEIGHT AND PAYLOADS
3.3.1 CURB WEIGHT
The curb weight of the vehicle shall include the weight of the completed vehicle, air conditioner, generator, and all other installed and stowed equipment, accessories and attachments specified herein.
3.3.2 GROSS VEHICLE WEIGHT RATING
The Gross Vehicle Weight Rating of the vehicle shall not exceed 26,000 pounds; the drivers will be non-CDL.
3.3.3 WATER AND AIRTIGHTNESS PROPERITIES
The completed vehicle, including the expandable side section, entrance doors and stowage compartments, shall be adequately sealed to prevent the entrance of water. The vehicle shall be tested for water tightness. The vehicle, including the expandable side section and stowage compartments, shall be airtight during transit to prevent dust from entering the interior.
3.3.4 SHOCK AND VIBRATION
The completed vehicle shall withstand the road conditions specified herein. Chassis components, fasteners and accessories shall show no loosening, damage, or malfunction when loaded with the rated payload.
3.3.5 LUBRICATION
All parts of the vehicle requiring lubrication, whether considered by the manufacturer to be permanently lubricated or not, shall be provided with ready access for lubrication. Access panels or doors shall be installed as necessary, with hinged access doors used wherever space permits. All parts of the chassis, body and installed equipment that require lubricants shall be capable of being lubricated by one man in not more than two hours, including time to clean lubrication fittings and open and re-close access panels or doors. Any special lubrication equipment or lube gun extensions required to meet these requirements shall be provided with the vehicle and securely stowed in the stowage compartments.
3.5 CHASSIS
The vehicle shall be built on a 2026 or current model year, medium duty, conventional cab, front engine, chassis meeting 2026 EPA emission requirements and furnished with the following:
· Cab: Conventional, flat roof
· Air Conditioning: Driver, heater, defroster
· Diesel Engine: minimum ratings 300 HP and 860 Lb-Ft torque
· Exhaust: Horizontal
· Transmission: Allison 3000HS automatic
· Front Suspension: taper leaf with shock absorbers
· Rear Suspension: Air ride suspension with shock absorbers
· Frame: 120,000 PSI yield strength
· Tow Hooks: frame mounted
· Alternator: Minimum 325 amperes
· Brakes: Cam-type type with slack adjusters and ABS
· Air tanks: heated automatic moisture ejectors
· Auxiliary Braking: engine compression
· Air Compressor: 18 CFM minimum
· Air Dryer: in-line
· Fuel Tank: Minimum 90 gallons
· Fuel Filter: spin on type
· Separator: fuel water filter
· DEF tank: Minimum 5 gallons
· Tires: 295/75R 22.5 14 ply radial
· Wheels: Polished aluminum (outers), hub piloted
· Front Bumper: three-piece chrome with collapsible ends
· Steering: Power with tilt and telescopic controls
· Lights: Daytime Running Lights
· Batteries: two (2) 12V 1900 CCA group 31
· Heater: Electric block 750 watts with receptacle
· Driver seat: air suspension, high back with arm rests
· Passenger seat: 2 persons mid-back vinyl
· Electronics: Backup alarm
· Cruise Control: Electronic
· Horn: air and dual electric
· Door locks: power electric
· Mirrors: dual heated remote with 8” convex
· Windows: power electric with tinted glass
· Wipers: Two-speed with intermittent
· Radio: AM-FM-WB-Bluetooth, clock, aux input
· Outlet: 12V power
· Switch: allows chassis battery to start generator & generator battery to start chassis
· Safety Kit: in-cab, fire extinguisher
· Color: white
3.5.1 CHASSIS TIRE PRESSURE MONITORING AND TIRE FILLING
The vehicle shall be furnished with a tire pressure monitoring system. An aftermarket system may be provided if the system is not released and available from the chassis OEM. A system with in-tire sensors is preferred. Valve stem sensors may only be used if in-tire sensors are not available either from the chassis OEM or aftermarket sources. The system shall include a display located near the driver.
The vehicle shall be equipped provisions to manually fill and maintain tire pressure. The OEM chassis air system may be used, or a dedicated air compressor shall be provided if sufficient pressure is not obtainable from the chassis air compressor. A retractable reel with 3/8-inch minimum air line of adequate length shall be provided to reach all tires. Quick couple connectors shall be provided.
3.5.2 BACK UP CAMERA AND PARK SENSING SYSTEMS
The vehicle shall be furnished with a color backup camera and monitor allowing the driver to view the area to the rear of the vehicle when backing up. The system shall include IR night vision capability, minimum 6-inch monitor, minimum 100-degree rear viewing angle, and one-way audio. The monitor shall be mounted on the instrument panel or dashboard and be easily viewed by the driver. The camera and monitor shall be installed in accordance with all manufacturer installation instructions and recommendations. The camera and camera wiring mounting shall be watertight utilizing gaskets and grommets as required to assure a watertight installation.
The vehicle shall be furnished with a rear parking sensor system. The system shall include sensors mounted in the rear bumper to detect obstacles and a dash-mounted audible indicator.
3.5.3 DOT LIGHTING SYSTEM
All exterior Federal Motor Vehicle Safety Standard (FMVSS) lighting shall be in accordance with Federal Motor Vehicle Safety Standard (FMVSS) 108 (49 CFR 571.108). All lamps shall be LED type and shall be mounted in recessed locations or guarded to provide protection from operational hazards, except lower rear marker lamps and ABS light. The stop, turn and tail lamps shall be recessed into the rear of the body. The vehicle shall be furnished with conspicuity markings in accordance with Federal Motor Vehicle Safety Standard (FMVSS) 108 (49 CFR 571.108).
3.5.4 MUD FLAPS
The vehicle shall be equipped with anti-sail mud flaps on the rear axles
3.5.5 REAR BUMPER
The vehicle shall be equipped with DOT ICC approved rear under-run bumper
3.5.6 LEVELING AND STABLIZATION SYSTEM
A leveling and stabilizing system shall be furnished. The system shall be completely computer controlled (joystick control not acceptable), and able to be deployed and retracted with hydraulics. At no time shall the system allow a single leg to operate independently, but rather two legs shall always operate in tandem. The leveling system shall automatically level the vehicle with a built-in leveling light display. The system shall be capable of side-to-side and fore-and-aft leveling that keeps the vehicle in a common plane to minimize twisting of the chassis. The system shall consist of the following:
· Four frame mounted power up/power down hydraulic leveling jacks (minimum)
· hydraulic pump
· 12-VDC controlled air / hydraulic power unit
· control panel
· reservoir
· hydraulic hoses, valves and fittings
· Labeled Activation light in sight of the driver
Minimum load capacity shall handle the vehicle as built. The jacks and components, accessories, and attachments shall withstand, without damage and deformation, the combined static and dynamic forces due to the portion of the gross weight sustained, and the force resulting from impact during coupling and uncoupling operations. The system shall be installed in accordance with the manufacturer’s instructions and recommendations. The jacks shall be equipped with self-leveling sand pads. The jacks and components, accessories, and attachments shall withstand, without damage and deformation, the combined static and dynamic forces due to the portion of the gross weight sustained, and the force resulting from impact during coupling and uncoupling operations. Each jack shall be mounted to reinforced mounting plates that resist forward, rearward, and side thrust forces on the jacks. When raised to the travel position, the jacks shall remain positively locked. Details of the leveling system and installation plans shall be furnished to the Government for review, verification and approval.
When placed in the travel position, the jacks shall remain positively locked. Clearance under the rear jacks shall not reduce the normal rear departure angle of the vehicle.
3.5.7 LEVELING BLOCKS
A set of eight (8) polypropylene leveling blocks, each a minimum 8 X 8 X 1 inch thick, with carrying bag, shall be furnished.
3.6 BODY
The body shall be furnished with insulated attic storage above the chassis cab. A boundary layer material providing galvanic protection shall be placed between aluminum and steel components.
The completed vehicle, resting on the landing legs on level ground, and with all specified equipment, shall conform to the dimensions specified in Table I.
TABLE I. Dimensions and clearances
| Requirement |
| Value |
| Overall maximum length (less bolt-on accessories) |
| 40 feet |
| Overall maximum height from ground |
| 162 inches (13 feet, 6 inches) |
| Overall maximum width |
| 102 inches |
3.6.1 BODY MOUNTING
The body shall be attached to the chassis in accordance to the OEM instructions and shall be suitable for severe service duty as described within these requirements. Fasteners shall be a minimum of Grade 8. The interior finished floor height shall be as low to grade as possible, with no wheel wells protruding above interior finished floor, to ease loading and maximize interior height.
3.6.2 BODY FRAMING
Structural members shall be extruded tubing or structural extruded channel or hat sections spaced 16 inches or less on center for sides, front, and rear posts. Corners and corner posts shall be rounded cast, extruded, or formed aluminum.
3.6.3 BODY EXTERIOR WALLS
The body exterior front, sides, and rear skin panels shall be a minimum .050inch aluminum attached to wall studs with buck rivets or Huck bolt type fasteners. The exterior of the body shall be furnished with symmetrically radius corners and edges. An extruded aluminum base rail shall be furnished connecting the front, side, and rear side panels to the lower longitudinal channels connecting the transverse cross-members.
All door openings shall have structural headers and vertical uprights around the perimeter of each opening. Drip moldings (rain gutters) shall be furnished immediately above any entry doors, and all exterior compartments.
3.6.4 BODY ROOF
The roof of the vehicle shall be of one-piece pre-tensioned aluminum alloy sheet (fiberglass reinforced polyester material not acceptable). The roof shall be supported and secured with equally spaced roof bows on not more than 24-inch centers. Extruded aluminum top rails shall be furnished riveted to and connecting the front, rear and side panels to the roof bows and roof panels. The roof shall be secured to the top rails. The roof shall have aerodynamic front top corners. All joints/seams shall be sealed and weatherproofed.
3.6.5 BODY FLOOR
A single level floor, without wheel housings, shall be provided the full length and width of the vehicle’s body. The floor shall be constructed of a minimum 1-inch thick structural closed-cell urethane composite. The floor shall be glued and screwed to flanges of the floor cross members. The body floor structural cross-members shall be on maximum 16-inch center spacing. All joints and voids in the floor shall be filled and sealed with body filler prior to floor tile installation.
3.6.6 SIDE (MAIN) ENTRANCE DOOR & STAIRS
A side entrance door shall be located on the passenger side of the body. The door shall provide a minimum clear opening of 32 inches wide and 72 inches high. A 31 percent maximum light transmission gray tinted, rectangular, tempered safety glass window, shall be installed vertically in the upper half of the door. The window shall be a minimum of 8 inches wide and 22 inches tall, centered in the door.
The door shall open out to the right, towards the front of the vehicle, from outside of the body. The door shall open a minimum of 90 degrees. A spring-loaded door support shall be installed to retain the door in the open position. The door shall be furnished with an automotive style slam latch. An electronic programmable digital keypad combination lock with key override shall be provided, Alarm Lock Trilogy series or equal. A deadbolt lock, capable of locking from inside shall be provided. All door hardware shall have stainless steel finish. The door shall be made with aluminum frame and sheet and include full insulation and a finished interior panel to match the interior walls. A rain gutter shall be furnished above the door, extending past the edges of the door.
A stairway, deploying perpendicular to the body, shall be provided for the body entrance door to facilitate ingress/egress into the vehicle. The stairway shall be aluminum construction. The tread surface shall comply with ADA, NFPA, and OSHA requirements. The stairway shall be constructed using aluminum channels, angles, and plate to form the frame. The frame components shall be welded or bolted. The stairways tread depth and riser height shall both be a minimum 7 inches. The stairs shall be constructed such that they do not interfere with the operation of the door when the entrance door is in the full-open position, 90 degrees from the body. When stowed they shall not hinder use of the entrance when not deployed.
3.6.7 WINDOWS
A minimum of two (2) windows shall be provided in the body section. One window shall be provided on the street side and shall allow for emergency escape.
3.6.8 WHEELCHAIR LIFT
A wheelchair lift shall be furnished on curb side. The useable platform width shall be 30 inches minimum. The useable platform length shall be 45 inches minimum. Lifting capacity shall be at 700 pounds minimum. The wheelchair lift shall be hydraulic powered using a 12-volt direct current motor. A manual hydraulic pump back-up system shall be provided in case of electrical failure. The wheelchair lift shall have dual hand rails. Safety interlocks shall be provided to prevent operation of the wheelchair lift, when the wheelchair door is closed. The wheelchair lift shall be furnished with a button pendant control. The wheelchair lift shall comply with the Americans with Disability Act requirements.
3.6.9 EQUIPMENT MOUNTING PROVISIONS
Unless otherwise specified herein, aluminum tapping plates shall be furnished to provide reinforced attachment of all equipment, systems, cabinets and equipment mounts. Sidewall tapping plates shall be at least 1/4 inch thick by at least 4 inches wide and welded to multiple sidewall structural members as required. Plates to mount all other features to the floor shall be at least 3/8 inch thick by at least 4 inches wide and welded to multiple floor structural members.
The mounting plates for the specified medical equipment shall be provided and installed in accordance to the manufacturer’s instructions.
3.6.10 STOWAGE COMPARTMENTS
Exterior stowage compartments shall be furnished under both sides of the body to the maximum extent possible. Drip rails, or similar provisions, shall be provided over, or on all stowage compartment doors. Drip rail ends shall be rounded. Drainage with plugs shall be provided in the storage compartments.
3.6.10.1 COMPARTMENT DOOR CONSTRUCTION
Compartment doors provide access to all portions of each compartment. Each door shall be box pan constructed or furnished with a full perimeter channel to assure rigidity. The doors shall be hinged at the top. Means shall be provided automatically (action of opening doors shall set mechanism to hold doors open) and positively retain the doors in the full-open position, manual release of hold open mechanism is acceptable. Separate manual latches to hold compartment doors open are not acceptable. Compartments shall be weatherproof incorporating a protected tubular neoprene type automotive sealing system. The mating edges of the ends of neoprene tubular seals shall be on the bottom of the doors. Each door shall be equipped with recessed latches and key operated locks. Latches and locks shall be stainless steel. All compartments shall be keyed alike.
The storage compartments interiors shall be continuously coated with “Gator Kote”, “Line-X “or “Rhino Lining” or equal to seal and protect the compartments
3.6.11 UNDERCOATING
The vehicle shall be undercoated for sound deadening, corrosion and stone damage protection. A commercial, fire resistant, petroleum base or rubberized undercoating, or other material providing equivalent protection, shall be spray applied to the underbody and under chassis sheet metal surfaces, except to the drive shafts, drain holes, lubrication points, engine/transmission oil pans, fuel tanks, heavy castings, suspension components, heat shields and areas within 12 inches or less from exhaust system components, as well as other areas specifically excluded by the chassis manufacturer. These areas shall be kept free of undercoating. The chassis frame, underside of the engine compartment hood, and underbody surfaces that are inaccessible without removing vehicle fuel tank(s) or other major components, shall not require undercoating.
3.6.12 WHEEL CHOCKS
A set of wheel chocks shall be provided for the rear axles, sized for the tires provided. The chocks shall be stowed in one of the underbody storage compartments close to the rear of the vehicle. Final size and storage location shall be approved by the Government.
3.7 INTERIOR FINISH
3.7.1 INTERIOR WALL CONSTRUCTION
The interior walls shall be constructed of metal studs or tubing. The walls shall be insulated. The insulation material shall be hard board insulation. All insulation materials shall not support combustion or sustain flame, mold, rot, support vermin, corrode metals, absorb orders, and shall be essentially odorless and of low VOC materials. The insulation shall not disintegrate under exposed operational or storage temperatures, or from vibration or settle and leave voids.
3.7.2 INSULATION
The vehicle body shall be completely insulated. Insulation shall be installed to fill the cavities between the inner walls and body exterior panels, ceiling, and under floor areas. It is acceptable for the insulation to be furnished with hollow chase-ways for electrical conduit. All insulation materials shall not support combustion or sustain flame, mold, rot, support vermin, corrode metals, absorb orders, and shall be essentially odorless. The insulation shall not disintegrate under exposed operational or storage temperatures, or from vibration or settle and leave voids. Fiberglass batt insulation shall not be used.
Insulation shall be installed under the entire floor in areas between the cross members. Insulation in under-floor areas shall be spray-in-place CFC-free urethane foam having a minimum thickness of 2 inches. Spray-in-place urethane foam insulation shall conform to ASTM E-84 for Class A fire retardant.
3.7.3 FINISH INTERIOR FLOORING
Fiberglass, aluminum, structural polyester foam, or other non-hydroscopic composites shall be used as the sub-floor, the minimum thickness shall be ½ inch. Plywood, particleboard, or equivalent wood materials are not acceptable
Finish flooring shall be fire retardant, elastomer profiled rubber composition. The color shall be determined at the VDR meeting and approved by the Government.
3.7.4 CEILING AND WALL FINISH
Fiberglass, aluminum, or other non-hydroscopic composites shall be used as underlayment on the walls and ceiling, the minimum thickness shall be 1/2 inches. Plywood, particleboard, or equivalent wood materials are not acceptable.
Panels shall be constructed of either fiberglass reinforced plastic laminated to both sides of a closed cell, expanded polystyrene foam substrate or ABS plastic. The sills and division bars may be fastened to the floor, ceiling, or wall stud with fasteners. Such fasteners shall be covered by vertical trim molding that match the matte wall finish. A cove base shall be furnished throughout the vehicle to protect the floor to wall transition. The wall panels shall have a smooth finish.
Finished ceiling height, in the main body area, shall be a minimum of 90 inches. A fiberglass or aluminum channel ceiling grid system shall be suspended beneath the roof of the vehicle. The ceiling grid system shall be designed so that no rivets, screws, or fasteners shall be visible. The ceiling panels shall be Class C fire-rated in accordance with UL. The ceiling shall be an acoustical ceiling. The ceiling tiles shall be architectural grade embossed panels. There shall be no ductwork or other equipment intruding below the ceiling. Lay-in type ceiling panels shall be furnished throughout the entire vehicle having a nominal size of 24 inches x 24 inches. The ceiling panels shall be constructed of an acoustical vinyl textured face.
3.8 INTERIOR LAYOUT
The object of this section is to describe the interior room layout. The Supplier is responsible for the final detailed layout of the vehicle interior which shall be approved by the Government at the VDR meeting.
The vehicle shall be configured into multiple areas. Each area shall be sized to accommodate the equipment, cabinets, or seating as specified herein. All surfaces shall withstand continuous cleanings with disinfectants and cleaning solutions. The Supplier is responsible for the final detailed lay out of the vehicle interior and the final design shall be approved by the Government at the VDR meeting.
The interior noise level shall be measured along the centerline of the body at a height of 3 feet, no reading shall exceed 65 dB (A), with the generator operating at full load, and all air conditioner on and set to coldest setting. Doors between the areas shall be open for one test and closed for the second test. Testing shall be witnessed by the government.
3.8.1 LOFT STORAGE AREA
There shall be an elevated and air-conditioned storage loft over the chassis cabin. The loft ceiling shall be as high as practicable. Access doors and LED lighting shall be provided in the loft.
3.8.2 INTERVIEW AREAS
Two (2) interview work areas shall be provided. Four (4) desks shall be provided with countertop and base cabinets. The countertops shall be provided and shall allow mounting of computer equipment and provide a writing surface. Size, capacity, and configuration of the cabinets and countertops shall be determined and approved by the CDC at the VDR meeting. The layout of the interview areas shall support private/confidential discussions between the Doctor and Miner.
3.8.3 SEATING
The seating shall be suitable for the healthcare environment, commercial grade vinyl, contoured back with lumbar support. Four (4) roll-around chairs with pneumatic seat height and heavy-duty casters shall be provided. Four (4) seating positions with arm-rests shall be provided for the Miners.
3.8.4 ENTRY/INTAKE
An entry area shall be provided. A desk shall be provided with countertop with base and overhead cabinets. A 24-inch diagonal flat screen television with Wi-Fi and PC inputs shall be provided. The configuration of the entry/intake area shall be determined and approved by the CDC at the VDR meeting.
3.8.5 CABINETS AND COUNTERTOPS
All cabinets shall comply with Joint Commission standards for sanitation and workmanship. The CDC shall select the materials and colors for cabinets, from items currently manufactured. Counter surfaces shall be constructed from solid-surface acrylic resin composite. Countertops shall be Class 1 (A) fire-rated, hospital-grade. Size, capacity, and configuration of the cabinets and countertops shall be determined and approved by the CDC at the VDR meeting.
3.8.6 GALLEY
A dual voltage AC/DC refrigerator with freezer tray shall be furnished and installed. The refrigerator/freezer shall be a built-in style and furnished with a means to lock the door for securing contents inside. Separate outlets on separate breakers shall be provided for the appliances.
3.8.7 BUILT-IN SINK
A hand-washing sink built into a fixed cabinet shall be provided. The sink shall include cold and hot water, 30-gallon minimum supply reservoir, 30-gallon waste reservoir, ADA compliant handles, on demand pump, water heater, and a soap dispenser. The sink basin shall be constructed of stainless steel. The sink shall be NSF certified and provide an easy method to legally dispose of the waste water at a dump station.
3.8.8 SPIROMETRY
Desks shall be provided with countertop and clear dividers, with base and overhead cabinets. The spirometry areas shall allow use of a laptop computer where blood pressure is obtained and examination materials for the Miner may be printed on label printer. The configuration of the intake area shall be determined and approved by the CDC at the VDR meeting.
3.8.8.1 SPIROMETRY EXAMINATION ROOM
Two (2) rooms shall be provided for use spirometry equipment.
3.8.8.2 SPIROMETRY TESTING EQUIPMENT
The government (CDC) will acquire and install all spirometry equipment after vehicle delivery.
3.8.9 ULTRA VIOLET GERMICIDAL IRRADIATION EQUIPMENT
UVGI equipment shall be provided, installed, and integrated into all the common areas of the vehicle and within its ventilation system.
3.9 INTERIOR SECURITY ALARM SYSTEM
An intrusion alarm system shall be furnished to protect the interior of the vehicle.
3.10 CABLE RACEWAYS AND DATA CABLES
The vehicle shall be provided with two front-to-rear communications / computer raceways. The raceways shall be installed above the ceiling. One raceway shall be located on the curbside and one on the street side. The cable raceways shall be separate from the raceways used to carry power cables. Drops shall be furnished from these raceways to the work areas where the computers will be placed.
All cable and electrical wiring shall be labeled at each end for ease of maintenance and shown on an as-built vehicle wiring schematic supplied at vehicle delivery. All cable and electrical wiring shall be labeled at each end for ease of maintenance and shown on an as-built vehicle wiring schematic supplied at vehicle delivery. Voice and data wiring products shall comply to Category 6 specifications.
3.11 SMOKE DETECTORS
120VAC with 10-year lithium battery backup combination smoke and CO detectors shall be furnished and installed. One in each section of the body, final location shall be approved by the government. The units shall be tied together to sound the alarm in all areas.
3.12 FIRE EXTINGUISHER / SAFETY TRIANGLES
A CO2 fire extinguisher and a ten-pound ABC extinguisher shall be furnished and mounted in the body, location to be determined at the VDR meeting. A set of DOT approved flares (6), three reflective safety triangles and four orange safety cones shall be provided and shall be shipped loose with the vehicle.
3.13 MUSIC SYSTEM / PA SYSTEM
An AM/FM/CD music and PA system shall be furnished for use externally under the awning area. The system shall be available to be used inside the vehicle for staff use. The system shall be located to be controlled by staff only. This radio system shall have sufficient output to handle one external set of speakers, under the awning area and the three sets located in all interior rooms. A wireless microphone shall be provided for the staff inside to make announcements over the outside PA system. Each set of speakers shall have individual volume control and be capable of being turned off independently. Final location of speakers shall be approved by the government.
3.14 HVAC SYSTEM
3.14.1 AIR CONDITIONER
Two (2) wall mount commercial duty HVAC systems shall be provided. The units shall provide a minimum of 36000 BTU and electric heating element rated at 5KW. All the components are to be provided to deliver the unit as a system. The unit shall be installed on the rear wall of the body. The units shall be wired on a dedicated 208/230VAC, 1 Phase, 60 HZ circuits with an appropriately rated circuit breaker at panel, as required by the manufacturer. Drains from the unit shall be able to be hooked to a hose to direct drain water away from the body. The system shall include HEPA filtration (to include at least 1 wall mounted air intake vents in the exam areas) and internal UV disinfection lighting system. Final configuration of the HVAC system and outlets shall be approved by the government.
3.14.2 DUCTS AND DIFUSERS.
Supply and return air ducts and adjustable diffusers shall be provided to meet the air conditioning requirements specified herein. All ducts shall be insulated. Supply air ducts shall be located above the finished ceiling and run in a straight line from the front to rear of the vehicle. Return air ducts shall run in a straight line from the front to rear of the vehicle and be located below the finished ceiling closer to the floor. A 3 H x 2 W foot wall mount exhaust hood type return will be located in the wall of the spirometry testing area. UV disinfection lighting will be located on a wall adjacent to the aforementioned return vent. All ducts shall be held securely by bracing. Supply ducts shall be furnished with adjustable diffusers. Diffusers shall be mounted on the ceiling and match the color of the ceiling panels. Diffusers shall be furnished as follows for optimum distribution of heating and cooling. Diffuser(s) shall be located so as not to interfere with the proper operation of the thermostat. Diffusers(s) shall not be placed directly above the thermostat. Return grilles shall be furnished as follows to provide optimum air flow in the body. The return grilles shall be in either the ceiling. The supply and return ducts shall be sized to allow high volume low velocity, low noise and even distribution of air.
3.14.3 THERMOSTATS
The vehicle shall be furnished with remote thermostat(s) to control the air conditioning and heating. Heating/cooling shall be controlled through the electronic thermostat(s). The thermostat(s) shall control the 5KW heat. The thermostat(s) shall provide the following features: on/off, fan/ air conditioning, heating. The thermostat shall be located such that it can only be accessed by the staff and shall be installed in accordance with the manufacturers’ instructions. Thermostat shall be mounted in a government approved location.
3.15 DC ELECTRICAL SYSTEM
An electrical load analysis shall be performed on the DC circuits. The analysis shall indicate the total current draw on each circuit. The total current draw of each circuit shall not exceed 80 percent of the current protection for that circuit. The load analysis shall be made available to the government for quality assurance purposes. The 12-volt electrical system shall include the following: wiring, main distribution panel board, battery conditioner, batteries, battery isolator, power converter, branch circuits, and lighting as specified herein. The batteries associated shall be charged by shore power or when the generator is running.
3.15.1 WIRING
All 12-volt wiring shall be in accordance with the applicable portion of SAE J1292. All 12-volt wiring mounted to the chassis or outside the body shall be stranded and have SXL type insulation. All conductors shall be stranded copper. All 12-volt wire shall be routed in flexible loom or conduit. Wiring terminals used in all 12-volt wiring shall grip the wire insulation as well as the conductor. All fasteners, conduit, conduit fittings, shall not exhibit electrolysis with adjacent materials. All exterior lighting shall be wired within the body to the exterior lights. Wire shall be accessible for replacement without removing any component of the vehicle. Wire shall be capable of being replaced and serviced through the conduit and junction boxes. Grommets or electrical bulkhead fittings shall be employed for any wire passing through structural member(s). All 12-volt wire shall be color-coded as specified in SAE J560. The 12 VDC panel shall include a digital voltmeter and low voltage alarm, which activates at 12.5 volts.
Wire and circuit breaker size shall be as follows:
Wire size (minimum) Circuit breaker (maximum) 18 AWG 5 amperes 16 AWG 10 amperes 14 AWG 15 amperes 12 AWG 20 amperes 10 AWG 30 amperes 8 AWG 50 amperes
3.15.2 LIGHTING INTERIOR
A minimum of eight (8) 12-volt LED dome lights shall be provided and shall be located throughout the rooms in the ceiling of the body. A switch shall be provided on the interior wall near the rear entrance door to control the interior dome lights in each room. The switches shall be flush or surface mounted, white in color, and shall have a brushed stainless steel cover plate. The switch shall be labeled “12-VOLT INTERIOR LIGHTS”. Final location and number of lights provided shall be approved by the Government.
3.15.3 LIGHTING EXTERIOR
12VDC courtesy lights shall be located under the entry doors to illuminate the stairs/wheel chair lift and shall be controlled by a switch inside the entrance doors. Lighting level for the wheelchair lift shall comply to ADA requirements.
3.15.4 LIGHTING FLOOD EXTERIOR
Five (5) weatherproof LED type flood lights shall be installed on the body exterior. The flood lights shall be mounted as follows: two (2) on the street side the vehicle; two (2) on the curbside of the vehicle and one (1) on the rear body. The flood lights shall be able to be switched on/off manually individually by side. Switches to control the lights shall be approved by the Government at the VDR meeting.
3.15.5 ELECTRICAL OUTLETS
12VDC electrical outlets shall be provided to all appliances that require it in the body and the lower exterior compartments. Locations shall be approved by the Government at the VDR meeting.
3.15.6 STOWAGE COMPARTMENT LIGHTING
12-VDC LED lighting fixtures shall be installed in all underbelly storage bays. Lighting fixtures shall be protected, waterproof, and readily accessible for servicing. The lights shall be powered by the deep cycle batteries. They shall also be controlled by a switch mounted in each bay that automatically shuts off the lights when the bay door is closed. Number and location of lights shall be determined at the VDR meeting and approved by the Government.
3.16 AC ELECTRICAL SYSTEM
An electrical load analysis shall be performed on the AC circuits. The analysis shall indicate the total current draw on each circuit. The total current draw of each circuit shall not exceed 80 percent of the current protection for that circuit. The load analysis shall be made available to the government for quality assurance purposes. The vehicle body AC electrical system shall be in accordance with the National Electrical Code (NEC). The vehicle shall be equipped with a 120-volt, single phase, 60 Hz, 3-wire and a 240-volt, single phase, 60-Hz, 3-wire electrical system. The electrical system supply shall be from the on-board diesel engine generator and shore power provisions. The system shall consist of the following:
· Wiring
· 120/240-volt 3-wire, single phase AC main distribution panel sized to the output of the generator
· Branch circuits
· Breakers for all equipment with a minimum of 15% spare breakers in the main panel
· Receptacles
· Lighting and switches as specified herein
The panel shall include reverse polarity indication/warning, a digital combination voltage, frequency, and current meter, and a high current alarm. A 3-position manual meter line switch shall be furnished. The line switch shall be a spring biased switch and installed such that the switch is normally biased to the vocational branch of the 120 VAC circuit. The meters shall be capable of measuring the input feed (either leg individually (120VAC) or across both legs (240VAC) from either the generator or shore power. A power panel courtesy light shall be installed at the power panel and shall illuminate the power breakers
3.16.1 AC WIRING
The vehicle shall be wired from the main distribution panel board (located to be determined at the VDR). All wiring and cable conductors shall be stranded copper and THHN insulation. Wiring shall be color coded at the junction boxes and at the main distribution panel board for identification purposes. All electrical and electronic wires shall be run in rigid electrical conduit, ELECTRI-FLEX Conduit - Jacketed Metallic Flexible or plastic conduit for ease of repair. Wiring shall be of sufficient size and gauge to carry 125% of the current imposed in operating the electrical system without loss of dielectric strength.
3.16.2 BRANCH CIRCUITS
The vehicle shall be provided with 20-ampere branch circuits, as required by the NEC, throughout. Loads shall be evenly proportioned among the legs and branch circuits. All lighting shall be grouped together on specific branch circuits. Each piece of equipment and each appliance shall have dedicated circuits when required by the manufacturer or when specified herein.
3.16.3 INTERIOR OUTLETS
A minimum of twelve (12) duplex outlets shall be provided as required to each of the work stations, the galley area, examination areas, and throughout the vehicle as required. All receptacles shall be (NEMA 5-20R) white duplex outlet 120VAC, 20 amp duplex electrical receptacles. Hospital grade outlet shall be used in the interior. Receptacles shall be covered with a brushed stainless steel cover plate. All receptacles shall be of the GFCI protected type. Final locations shall be approved by the Government at the VDR meeting.
3.16.4 EXTERIOR OUTLETS
A total of four (4) GFI protected exterior duplex outlets shall be provided on the exterior or in the underbelly. All receptacles shall be (NEMA 5-20R) white duplex outlet 120VAC, 20 amp duplex electrical receptacles.
3.16.5 LIGHTING INTERIOR
LED lighting shall be provided throughout. The light fixtures shall be switched at easily accessible locations. All fixtures shall be securely anchored to prevent loosening or damage due to vibrations caused by rough transit. The lighting fixtures shall be positioned as required to provide adequate work lighting. A minimum light intensity of 50 foot-candles shall be provided, measured at 30 inches above the floor or work-space level.
3.16.6 WALL SWITCHES
Rocker design (touch on and off operation), switches shall be provided for all room lighting. The switches shall be single pole and rated at 20 ampere/120 volts, and UL listed. Each switch shall be recessed into the wall and enclosed in a galvanized steel box. The color of the switches shall be white. Each switch shall be provided with a brushed stainless steel cover plate. Switches shall be provided for all interior lighting at the front of the vehicle body and by the rear doors. Final location shall be reviewed and approved by the government at the vehicle design review.
3.16.7 TASK LIGHTING
Task lighting shall be provided at each countertop location where overhead cabinets allow. Each built-in, flush mounted, 115-volt task light shall have an integrated switch. A minimum light intensity of 50 foot-candles shall be provided at each location
3.16.8 EMERGENCY LIGHTING EXIT SIGN
An emergency light with exit sign shall be provided located near the main door.
3.17 POWER GENERATOR
3.17.1 GENERATOR COMPARTMENT
The generator shall be mounted in dedicated under-belly compartment. The engine-generator compartment shall be designed and lined with sufficient sound deadening material to provide the required sound level in the vehicle when the generator is at full load without hampering the use of the generator.
3.17.2 DIESEL GENERATOR
A minimum rated, 20 KW, 120/240 volt, 60 Hertz, single phase diesel powered liquid-cooled engine-generator shall be furnished. Final sizing of the generator shall be reviewed after the electrical analysis of the system provided by the Supplier along with their recommendation for generator size required. The final size shall be approved by the Government at the VDR meeting. The engine-generator shall be mounted on shock/vibration isolation mounts. The engine-generator shall be capable of operating during transit. An original equipment manufacturer engine-generator control panel shall be furnished and located near the engine-generator with clearly labeled operating instructions posted. The control panel shall include at a minimum a start/stop switch, hour meter for the generator, and glow plug control switch. A remote start switch and control panel shall be provided. The location of the start switch shall be near the electrical panel.
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it. Updated .