This is a Sources Sought Notice and Request for Information (RFI) issued by the Department of Veterans Affairs Network Contracting Office for market research purposes only. The government seeks information about qualified sources capable of supplying SUREPOWER rechargeable lithium-ion battery packs and R Series WiFi Cards with Clock Sync and FIPS certification for defibrillators at the Carl T. Hayden Veterans' Medical Center in Phoenix, Arizona. The battery packs must have 5.8 Ah capacity, high-density lithium-ion chemistry, a RunTime Indicator, automatic calibration readiness, and the ability to store use and maintenance history. The R Series WiFi Cards must meet specific IEEE 802.11 a/b/g/n standards with CompactFlash interface, dual-band capability (2.4 GHz and 5 GHz), WPA/WPA2PSK security, FCC and IC certifications, and Windows XP/7 operating system support. The equipment will be used by nursing staff to power defibrillators on crash carts throughout the facility.
Responses must be submitted via email to Contract Specialist Sam Choo at sam.choo@va.gov no later than November 30, 2026 at 10:00 AM Pacific Time. The solicitation number is 36C26227Q0152, with the Product Service Code 6515 and NAICS code 339112 (Medical Surgical Instruments Equipment and Supplies). The SBA small business size standard for this NAICS code is 1,000 employees. Prospective vendors must provide a capabilities statement addressing all salient characteristics, company size status, manufacturer or distributor status, compliance with subcontracting limitations, Commercial Off-the-Shelf (COTS) status, Buy American Act conformance, lead times, product lifespan, pricing, SAM.gov Unique Entity ID/Cage Code, and any existing federal contracts or GSA schedules. This RFI is issued solely for information and planning purposes and does not constitute a solicitation; responses are not offers and cannot form a binding contract.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C26227Q0152_Batteries_Wifi_Card.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sources Sought Notice Sources Sought Notice
| SUBJECT* |
| Battery Packs and Wi-Fi Cards |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 90815 |
| SOLICITATION NUMBER* |
| 36C26227Q0152 |
| RESPONSE DATE/TIME/ZONE |
| 11-30-2026 10AM PACIFIC TIME, LOS ANGELES, USA |
| ARCHIVE |
| 60 DAYS AFTER THE RESPONSE DATE |
| RECOVERY ACT FUNDS |
| N |
SET-ASIDE
| PRODUCT SERVICE CODE* |
| 6515 |
| NAICS CODE* |
| 339112 |
| CONTRACTING OFFICE ADDRESS |
| Department of Veterans Affairs |
Network Contracting Office 22 4811 Airport Plaza Drive Suite 600 Long Beach CA 90815
POINT OF CONTACT*
Contract Specialist Sam Choo Sam.choo@va.gov Sam.choo@va.gov
PLACE OF PERFORMANCE
| ADDRESS |
| Department of Veterans Affairs |
Phoenix VA
Carl T. Hayden Veterans' Medical Center
650 E Indian School Rd
Phoenix AZ
| POSTAL CODE |
| 85012 |
| COUNTRY |
| USA |
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
AGENCY CONTACT’S EMAIL ADDRESS
EMAIL DESCRIPTION
DESCRIPTION
THIS REQUEST FOR INFORMATION (RFI) / SOURCES SOUGHT NOTICE ISSUED SOLELY FOR INFORMATION AND PLANNING PURPOSES. THIS IS NOT A SOLICITATION.
SUBMISSION OF INFORMATION ABOUT PRICING, DELIVERY, THE MARKET, AND CAPABILITIES IS HIGHLY ENCOURAGED AND ALLOWED UNDER THIS RFI FOR PLANNING PURPOSES IN ACCORDANCE WITH (IAW) FAR 15.201(e).
DISCLAIMER
This RFI is issued solely for information and planning purposes and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. IAW FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.
SOURCES SOUGHT/RFI DESCRIPTION
This is NOT a solicitation announcement. This is a Sources Sought Notice / RFI only. The purpose of this Sources Sought Notice / RFI is to gain information about potential qualified sources and their size classification relative to NAICS code is 339112 -Medical surgical instruments equipment and supplies, with an SBA small business size standard of 1,000 employees. Responses to this Sources Sought Notice / RFI will be used by the Government to make appropriate acquisition decisions. After review of the responses to this Sources Sought Notice / RFI, further RFIs and/or a solicitation or other announcements may be published.
The purpose of this procurement is to acquire SUREPOWER rechargeable lithium-ion battery packs and R Series WiFi Cards with Clock Sync, FIPS, for defibrillators used throughout the facility by Nursing staff.
The batteries shall power the defibrillators used on crash carts. The R Series WiFi Cards shall provide network connectivity, with clock synchronization performed automatically through DXS after scheduled readiness tests.
The anticipated requirement includes battery packs with 5.8 Ah capacity, high-density lithium-ion chemistry, a RunTime Indicator, automatic calibration readiness, and the ability to store use and maintenance history. The WiFi Cards shall meet the physical, wireless, security, environmental, certification, and operating-system requirements identified in the Statement of Work below.
The contractor shall comply with the complete VA cybersecurity and privacy requirements included in the Statement of Work, including the requirements applicable to medical devices, specialized systems, wireless connectivity, FIPS-validated cryptography, patching, product integrity, incident reporting, and VA information-system access.
STATEMENT OF WORK
1) Description of Equipment Requested:
a) SUREPOWER RECHARGEABLE LITHIUM-ION BATTERY PACK and R Series WiFi Card with Clock Sync, FIPS for defibrillators
2) How Equipment Works:
a) Batteries power the defibrillators to enable them to deliver a controlled electrical shock that depolarizes all heart muscle cells at once, stopping chaotic rhythms like ventricular fibrillation so the heart’s natural pacemaker can restart a normal rhythm. This process is performed by devices such as AEDs, which analyze heart rhythms and shock only when needed. The R Series’ Wi‑Fi card enables network connectivity, while clock synchronization is performed automatically through DXS after scheduled readiness tests.
3) Who Will Use This Equipment:
a) Nursing staff
4) Where it will be used:
a) Throughout the facility
5) Salient Characteristics:
a) Battery
i) Must have 5.8 Ah Capacity
ii) Must have High density lithium-ion chemistry
iii) Must have RunTime™ Indicator
iv) Must have Automatic calibration ready
v) Must be able to store history of use and maintenance
b) R Series Wifi Card
i) Must have Standard: IEEE 802.11 a/b/g/n (single‑stream)
ii) Must have Interface: CompactFlash (CF) — self‑contained CF‑to‑Wi‑Fi functionality (no host CPU load)
iii) Must have Bands: 2.4 GHz (2.412–2.462 GHz) & 5 GHz (5.180–5.825 GHz)
iv) Must have Data rates: Up to 65 Mbps (includes 54/58.5/52/39/26/19.5/13/6/5/5.5/2/1 Mbps)
v) Must have Modulation: OFDM with BPSK, QPSK, 16QAM, 64QAM, CCK
vi) Must have Security: WPA/WPA2‑PSK, WEP (64/128 bits), TKIP
vii) Must have Antenna: Embedded PCB printed, 1.30 dBi @ 2.4 GHz / 1.97 dBi @ 5 GHz
viii) Must have TX power: 9 dBm @ 802.11a, 14 dBm @ 802.11b, 11 dBm @ 802.11g, 10 dBm @ 802.11n, 8 dBm @ 802.11an
ix) Must have Sensitivity: −70 dBm @ 802.11a, −75 dBm @ 802.11b, −70 dBm @ 802.11g, −69 dBm @ 802.11n, −67 dBm @ 802.11an
x) Must have Dimensions: 60.5 × 41.7 × 6.45 mm
xi) Must have Operating temp: 0 – 55 °C
xii) Must have Humidity: 20 % – 95 %
xiii) Must have Certifications: FCC, IC
xiv) Must have OS support: Windows XP / 7
6) Additional Equipment Or Replacement Equipment (EE number):
i) Replacement Equipment – No EE#s - MVO
7) Justification:
i) Defibrillators are required for crash carts
VA CYBERSECURITY AND PRIVACY LANGUAGE FOR INCLUSION IN ACQUISITIONS. – See attached PDF file.
If your company is interested in and capable of providing the required supplies/services, you MUST provide the information indicated below OR your intent, interest to participation, and information provided will not be considered as valid for research purposes.
Response to this notice should include company name, address, point of contact, size of business pursuant to the following questions:
(1) Submit your capabilities statement illustrating how your organization can/cannot meet the list of DRAFT/SALIENT CHARACTERISTICS/SOW requirements. For instances where your company cannot meet the DRAFT/SALIENT CHARACTERISTICS/SOW requirement(s), please explain. For instances where your company can meet the DRAFT/SALIENT CHARACTERISTICS/SOW requirement(s), please clearly show how your company meets/exceeds each requirement otherwise your response will be recorded as “interest only”.
(2) Please review the list of DRAFT/SALIENT CHARACTERISTICS/SOW requirements and provide any additional feedback or suggestions. If none, please reply as N/A.
(3) Please indicate the size status and representations of your business, such as but not limited to: Service-Disabled Veteran-Owned Small Business (SDVOSB), Veteran-Owned Small Business (VOSB), HUBZone, Woman Owned Small Business (WOSB), Large Business, etc.
(4) Is your company considered small under the NAICS code identified in this RFI?
(5) Are you the manufacturer, authorized distributor, and/or can your company provide a solution to the required supplies/services described in the list of DRAFT/SALIENT CHARACTERISTICS/SOW?
(6) If you are a large business, do you have any designated/authorized distributors? If so, please provide their company name, telephone, point of contact and size status (if available).
(7) If you’re a small business and you are an authorized distributor/reseller for the items identified above, do you alter; assemble; modify; the items requested in any way? If you do, state how and what is altered; assembled; modified.
(8) Limitations on Subcontracting: How does your business ensure compliance with the limitations on subcontracting as outlined in 13 CFR § 125.6?
(9) Are the items you are identifying/providing considered Commercial of the Shelf (COTS) items as defined in FAR Part 2.101 under commercial items?
(10) Non-Manufacturer Rule: If applicable, can you confirm your business complies with the Non-Manufacturer rule? Specifically, does your company: Provide a product from a small business manufacturer or processor? Not exceed 500 employees? Primarily engage in the retail or wholesale trade and normally sell the type of item being supplied? Take ownership or possession of the item(s) with its personnel, equipment or facilities in a manner consistent with industry practice?
(11) Please indicate whether your product conforms to the requirements of the Buy American Act? Please indicate if manufacturer is/will be US domestic Small Business / US Domestic Large Business / Foreign.
(12) What is your lead time to deliver a single unit with all components? Is there scale in lead time with greater quantities? Please elaborate.
(13) What is estimated life span of your solution? What support/services does that entail?
(14) Does your organization offer a leasing solution? Please elaborate.
(15) Does your company have a Federal Supply Schedule (FSS) GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract? If so, please provide the contract number(s).
(16) If you are an FSS GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract holder, are all the items/solutions you are providing information about available on your schedule/contract? Please state if all or some items are available on the contract.
(17) General pricing of your products/solution is encouraged. Pricing will be used for the purpose of market research only. It will not be used to evaluate for any type of award.
(18) Please provide your SAM.gov Unique Entity ID/Cage Code number.
Responses to this notice shall be submitted via email to sam.choo@va.gov. Telephone responses will not be accepted. Responses must be received no later than Monday, November 30, 2026 by 10:00 AM PST.
All responses to this Sources Sought/RFI will be used for planning purposes only. Responses to this Sources Sought Notice / RFI are not considered a request to be added to a prospective bidders list or to receive a copy of the solicitation. If further RFIs and/or a solicitation or other announcement is issued as a result of the information provided from this RFI, all interested parties must respond to the specific posting separately IAW the specifications of that announcement.
See attached document: 36C26227Q0152_Batteries_Wifi_Card.
| *= Required Field |
| Sources Sought Notice |
Sources Sought Notice
File details come from the government source that posted it. Updated .