36C26226Q1275 AMD 0001.pdf
PDF 193 KB Posted
- Attached to
- 6515--HEMODYNAMICS MONITOR
- Solicitation number
- 36C26226Q1275
About this file
This is Amendment 0001 to a previous combined solicitation for hemodynamic monitoring equipment issued by the Department of Veterans Affairs, Network Contracting Office 22, located in Long Beach, California. The solicitation number is 36C26226Q1275, with a response deadline of August 3, 2026, at 1:00 PM Pacific Time. The set-aside is designated for Service-Disabled Veteran-Owned Small Businesses (SDVOSBC), with a Product Service Code of 6515 and NAICS Code 339112. The place of performance is the VA San Diego Healthcare System at 3350 La Jolla Village Drive, San Diego, California 92161. The point of contact is Contracting Officer Debby Abraham at debby.abraham@va.gov.
Amendment 0001 specifies the salient characteristics and approved brand names/part numbers for the hemodynamic monitor equipment required. The specified products include LiDCO Hemodynamic Monitor kits, Masimo O3 Regional Oximeter and Radical-7 Color Screen Handheld units, Root monitoring systems with accessories, interface cables for Edwards and Philips connections, and fixed-term software licenses for Pleth Vari Index and SpHB/SpOC monitoring capabilities. The hemodynamic monitor must utilize existing arterial lines or non-invasive dual finger cuffs for monitoring, measure blood pressure, and calculate cardiac output (CO), stroke volume (SV), systemic vascular resistance (SVR), stroke volume variation (SVV), and pulse pressure variation (PPV) using PulseCO software for pulse contour analysis. Additional requirements include suitability for hemodynamically unstable patients in operating rooms and ICUs, an intuitive display interface with 3-hour battery life, and capability for both continuous monitoring via arterial line and quick, intermittent, accurate readings. All other terms and conditions of the previous solicitation remain unchanged.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ 36C26226Q1275 HEMODYNAMIC MONITOR.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Amendment to a Previous Combined Solicitation
*= Required Field Amendment to a Previous Combined Solicitation
SUBJECT* HEMODYNAMICS MONITOR
GENERAL INFORMATION
CONTRACTING OFFICE’S ZIP CODE* 90815
SOLICITATION NUMBER* 36C26226Q1275
RESPONSE DATE/TIME/ZONE 08-03-2026 1PM PACIFIC TIME, LOS ANGELES, USA
ARCHIVE 30 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS N
SET-ASIDE SDVOSBC
PRODUCT SERVICE CODE* 6515
NAICS CODE* 339112
CONTRACTING OFFICE ADDRESS Department of Veterans Affairs
Network Contracting Office 22 4811 Airport Plaza Drive Suite 600 Long Beach CA 90815
POINT OF CONTACT*
CONTRACTING OFFICER
DEBBY ABRAHAM
DEBBY.ABRAHAM@VA.GOV
PLACE OF PERFORMANCE
ADDRESS VA San Diego Healthcare System 3350 La Jolla Village Drive
San Diego CA
POSTAL CODE 92161
COUNTRY
A. AMENDMENT 0001 IS HEREBY ISSUED TO INCORPORATE THE SALIENT
CHARACTERISTICS AS FOLLOWS:
BRAND NAME/PART #
Kit, LiDCO Hemodynamic Monitor/# 99026
Kit, O3 Regional Oximeter, Masimo/# 9637
Kit, Radical-7 Color Screen Handheld/#9500
Kit, Root, Masimo/# 9513
LiDCO to Edwards, BP Module Cable, 1/Box/# 5121
LiDCO to Philips, Analog Cable, 1/Box/# 5127
PLETH VARI INDEX, FIXED TERM LICENSE/#2635
RD RNBW SET MD20-12,Patient CB,12FT,1/BX/#4073
Root Roll Stand with Quick Connect/#300238
TOTAL (SpHB) & OXYGEN (SpOC), FIXED TERM/#2633
EQUAL TO PRODUCT
HEMODYNAMIC MONITOR
HEMODYNAMIC MONITOR SALIENT CHARACTERISTICS
Must use existing arterial lines or non-invasive finger cuffs (dual finger cuffs) for monitoring.
Must measures blood pressure and calculates CO, SV, SVR, Stroke Volume Variation (SVV), and Pulse Pressure Variation (PPV).
Must utilizes PulseCO softwave for pulse contour analysis to provide real-time data, often calibrated using Lithium Dilution (LiDCOplus) for accuracy.
Must be suitable for hemodynamically unstable patients in operating rooms and ICUs.
Must feature an intuitive display with a 3-hour battery life
Must provides both continuous monitoring (via arterial line) and quick, intermittent, accurate readings
B. All other terms and conditions remain unchanged.