36C26226Q0808.docx

DOCX document 22 KB Posted

Attached to
6515--Maintenance Kits for Draeger Evita V500 Ventilators
Solicitation number
36C26226Q0808
Issued by

About this file

This is a Sources Sought Notice and Request for Information (RFI) issued by the Department of Veterans Affairs Veterans Health Administration, Veterans Integrated Service Network 22, for maintenance kits supporting Draeger Evita V500 Ventilators. The RFI is issued solely for information and planning purposes and does not constitute a binding solicitation. The VA seeks qualified sources capable of providing equivalent or brand Draeger maintenance components.

The requirement encompasses four line items with quantities of 20 units each: Draeger Battery Pack M7 NM2.5 (Part# 8415290 or equal), Draeger Particle Filter (Part# 8415127 or equal), Draeger Kit Air Filter (Part# 8421479 or equal), and Draeger Cable Harness Spirolog Sensor (Part# 8416012 or equal). All components must be compatible with existing Draeger Evita Infinity V500 ventilators, with the battery pack specified as rechargeable NiMH technology. The solicitation applies NAICS code 339113 with a size standard of 800 employees. Responses to this RFI will inform the government's acquisition strategy and may lead to subsequent RFIs and formal solicitations. Responders are responsible for all expenses associated with submitting information, which may be marked as proprietary and will be handled accordingly under FAR 15.201(e).

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice Sources Sought Notice

SUBJECT*
Maintenance Kits for Draeger Evita V500 Ventilators

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
90815
SOLICITATION NUMBER*
36C26226Q0808
RESPONSE DATE/TIME/ZONE
05-27-2026 10:00 PACIFIC TIME, LOS ANGELES, USA
ARCHIVE
30 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
6515
NAICS CODE*
339113
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

Network Contracting Office 22 4811 Airport Plaza Drive Suite 600 Long Beach CA 90815

POINT OF CONTACT*

Contract Specialist Melissa Ramirez Melissa.Ramirez8@va.gov

PLACE OF PERFORMANCE

ADDRESS
Department of Veterans Affairs

VA San Diego Healthcare System

3350 La Jolla Village Drive

San Diego CA

POSTAL CODE
92161

COUNTRY

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

DESCRIPTION

THIS REQUEST FOR INFORMATION (RFI) / SOURCES SOUGHT NOTICE IS ISSUED SOLELY FOR INFORMATION AND PLANNING PURPOSES. THIS IS NOT A SOLICITATION.

SUBMISSION OF INFORMATION ABOUT PRICING, DELIVERY, THE MARKET, AND CAPABILITIES IS HIGHLY ENCOURAGED AND ALLOWED UNDER THIS RFI FOR PLANNING PURPOSES IN ACCORDANCE WITH (IAW) FAR 15.201(e).

DISCLAIMER

This RFI is issued solely for information and planning purposes and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. IAW FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.

SOURCES SOUGHT/RFI DESCRIPTION

This is NOT a solicitation announcement. This is a Sources Sought Notice / RFI only. The purpose of this Sources Sought Notice / RFI is to gain information about potential qualified sources and their size classification relative to NAICS 339113 (size standard of 800 Employees). Responses to this Sources Sought Notice / RFI will be used by the Government to make appropriate acquisition decisions. After review of the responses to this Sources Sought Notice / RFI, further RFIs and/or a solicitation or other announcements may be published.

The Department of Veterans Affairs (VA) is seeking sources capable of providing maintenance kits for Draeger Evita V500 Ventilators equivalent to brand Draeger.

CLIN
DESCRIPTION
QTY
UOM
0001
Draeger Battery Pack M7 NM2.5 Part# 8415290 or equal
20
EA
0002
Draeger Particle Filter Part# 8415127 or equal
20
EA
0003
Draeger Kit Air Filter Part# 8421479 or equal
20
EA
0004
Draeger Cable Harness Spirolog Sensor Part# 8416012 or equal
20
EA

SALIENT CHARACTERISTICS

CLIN 0001 Draeger Battery Pack M7 NM2.5 Part# 8415290 or equal Quantity: 20EA

· Must be compatible with existing Draeger Evita Infinity V500 ventilators.

· Must be made in rechargeable NiMH (Nickel-Metal Hydride) (20 HRX 23/43 (HHR - 250 SCH – L4x5).

· Must have standard charge condition of 0.25A x 16H.

· Must include an expiration date no earlier than two years from the procurement date.

· Must comply with preventative maintenance guidelines and work with the PM software provided by Draeger.

· Must be able to connect to the existing Evita Infinity V500 battery cables and fuses.

· Must have thermal fuse in case of overheating.

· Must have dimensions be equal to L: 7 7/16” x W: 4 1/16” x H: 1 7/64” with curved sides to properly fit into system.

CLIN 0002 Draeger Particle Filter Part# 8415127 or equal- Quantity: 20EA

· Must be compatible with existing Draeger Evita Infinity V500 ventilators.

· Must be designed to filter incoming oxygen before it gets pushed to the patient.

CLIN 0003 Draeger Kit Air Filter Part# 8421479 or equal- Quantity: 20EA

· Must be compatible with existing Draeger Evita Infinity V500 ventilators.

· Must be designed to filter incoming air before it gets pushed to patient.

CLIN 0004 Draeger Cable Harness Spirolog Sensor Part# 8416012 or equal - Quantity: 20EA

· Must be compatible with existing Draeger Evita Infinity V500 ventilators.

· Must be low-voltage medical sensor connector (flow/pressure sensor), not a standard DIN or aviation connector.

· Must have circular panel-mount receptacle, 4 small signal pins in a single circular pattern, keyed plastic shroud and molded PA12 housing; no substitutions will be accepted.

· Must be marked "PA CB30" indicating a nylon-12 based molded housing.

· Must comply with preventative maintenance guidelines and work with the PM software provided by Draeger.

· Must be compatible with the disposable hot-wire flow sensor (Spirolog) for Draeger Evita Infinity V500 ventilator breathing system.

The list of maintenance kits characteristics is intended to be descriptive, not restrictive, of the supplies/services that are required.

If your company is interested and capable of providing the required supplies/services, please provide the information indicated below. Incomplete responses to the questions below this notice will not be accepted. Response to this notice should include company name, address, point of contact, size of business pursuant to the following questions:

(1) Submit your capabilities statement illustrating how your organization can/cannot meet the SALIENT CHARACTERISTICS. For instances where your company cannot meet the SALIENT CHARACTERISTICS, please explain. For instances where your company can meet the SALIENT CHARACTERISTICS requirement(s), please show how your company meets/exceeds each requirement.

(2) Please review the SALIENT CHARACTERISTICS and provide any additional feedback or suggestions. If none, please reply as N/A.

(3) Please indicate the size status and representations of your business, such as but not limited to: Service-Disabled Veteran-Owned Small Business (SDVOSB), Veteran-Owned Small Business (VOSB), HUBZone, Woman Owned Small Business (WOSB), Large Business, etc.

(4) Is your company considered small under the NAICS code identified in this RFI?

(5) Are you the manufacturer, authorized distributor, and/or can your company provide a solution to the required supplies/services described in the list of SALIENT CHARACTERISTICS?

(6) If you are a large business, do you have any designated/authorized distributors? If so, please provide their company name, telephone, point of contact and size status (if available).

(7) If you’re a small business and you are an authorized distributor/reseller for the items identified above, do you alter; assemble; modify; the items requested in any way? If you do, state how and what is altered; assembled; modified.

(8) Limitations on Subcontracting: How does your business ensure compliance with the limitations on subcontracting as outlined in 13 CFR § 125.6?

(9) Are the items you are identifying/providing considered Commercial of the Shelf (COTS) items as defined in FAR Part 2.101 under commercial items?

(10) Non-Manufacturer Rule: If applicable, can you confirm your business complies with the Non-Manufacturer rule? Specifically, does your company: Provide a product from a small business manufacturer or processor? Not exceed 500 employees? Primarily engage in the retail or wholesale trade and normally sell the type of item being supplied? Take ownership or possession of the item(s) with its personnel, equipment or facilities in a manner consistent with industry practice?

(11) Please indicate whether your product conforms to the requirements of the Buy American Act?

(12) What is your lead time to deliver a single unit with all components? Is there scale in lead time with greater quantities? Please elaborate.

(13) What is estimated life span of your solution? What support/services does that entail?

(14) Does your proposed equipment have FDA clearance? Please specify what FDA clearance(s) have been obtained.

(15) Does your organization offer a leasing solution? Please elaborate.

(16) Does your company have a Federal Supply Schedule (FSS) GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract? If so, please provide the contract number(s).

(17) If you are an FSS GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract holder, are all the items/solutions you are providing information about available on your schedule/contract? Please state if all or some items are available on the contract.

(18) General pricing of your products/solution is encouraged. Pricing will be used for the purpose of market research only. It will not be used to evaluate for any type of award.

(19) Please provide your SAM.gov Unique Entity ID/Cage Code number.

Responses to this notice shall be submitted via email to Melissa.Ramirez8@va.gov. Telephone responses will not be accepted. Responses must be received no later than Wednesday, May 27th, 2026 by 10:00 AM Pacific Local Time.

All responses to this Sources Sought/RFI will be used for planning purposes only. Responses to this Sources Sought Notice / RFI are not considered a request to be added to a prospective bidders list or to receive a copy of the solicitation. If further RFIs and/or a solicitation or other announcement is issued as a result of the information provided from this RFI, all interested parties must respond to the specific posting separately IAW the specifications of that announcement.

*= Required Field
Sources Sought Notice

Sources Sought Notice

File details come from the government source that posted it. Updated .