Previous Inspection Report.pdf
PDF 3 MB Posted
- Solicitation number
- 36C26224Q1499
About this file
This document is a Previous Inspection Report for a medical gas inspection of the Southern Arizona VA Health Care Systems in Tucson, Arizona. The report covers an inspection that occurred from December 4-6, 2023 and includes detailed information on the medical gas, vacuum, and air systems at the facility.
The key details are:
- The inspection was conducted by Evergreen Medical Services, Inc. and evaluated the medical piping systems against applicable NFPA and CMS codes and standards.
- The report provides an overview of the total number of master alarms (206 total, with 6 deficient), area alarms (117 total, with 6 deficient), and zone valves (200 total, with 13 deficient).
- It also includes the results of testing for system dew point, particulate contaminants, gaseous contaminants, and source equipment, all of which passed inspection.
- The report goes into detail on the specific deficiencies found, including master alarm, area alarm, and zone valve issues, and provides recommendations for corrective actions.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
MEDICAL GAS
TESTING & INSPECTION
Southern Arizona VA Health Care Systems
Tuscon, Arizona
1-800-872-8992 www.evergreenmedical.com
12/04/2023-12/06/2023
MEDICAL GAS INSPECTION REPORT
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
Medical Facility:
City, State:
Medical Gas Contact Name:
Medical Gas Contact Dept:
Inspection Dates(s):
Territory MGR Jason Cates ASSE 6030 CREDENTIAL: 42161
Lead Technician(s) William Sigler ASSE 6030 CREDENTIAL: 12544604
Brandon Scott ASSE 6030 CREDENTIAL: 12615654
Section 7 - Contaminants
REPORT CONTENTS
Section 1 - Systems Overview
Section 2 - Deficiencies
- Master Alarm Deficiencies
- Area Alarm Deficiencies
- Zone Valves Deficiencies . . .
- Outlets & Inlets Deficiencies . . .
Section 3 - Master Alarms
- Master Alarm Inventory
Section 4 - Area Alarms
- Area Alarm Inventory
Evergreen Medical Services, Inc. in accordance with applicable NFPA & CMS codes and standards has evaluated the medical piping systems indicated in this report. These systems have been found to be within applicable guidelines except as noted within this report. Any deficiencies cited are based on the most current applicable edition of NFPA 99.
Conflicts may occur when different editions of NFPA 99 have been adopted by different AHJs, and compliance occurs with one edition but not the other. Changes made to newer editions of NFPA 99 are done so to improve the system's quality, integrity, fire resistance, effectiveness, durability, and safety. Dependent on the opinion of the AHJ(s), the newer code edition is generally acceptable as equivalent or superior to an older version(s).
Per NFPA 99 an existing system that is not in strict compliance with the applicable codes may continue to be used as long as it is not a distinct hazard to safety. Fire/Explosion hazards, labeling, manifold enclosures, medical air usage, & maintenance codes apply to new & existing facilities.
Information recorded on this report represent findings on the indicated evaluation date(s) only. Any action taken in response to recommendations included in this report should be recorded on the appropriate pages. This report should not be copied unless in entirety.
- System Dew Point
- Particulate Contaminants
- Gaseous Contaminants
Section 8 - Source EQPT Summary
- Source Equipment
Section 5 - Zone Valves
- Zone Valve Inventory
Tom Underwood
Engineering
12/04/2023-12/06/2023
Southern Arizona VA Health Care Systems
Tuscon, Arizona
Section 6 - Outlets & Inlets
- Outlets & Inlets Inventory www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
GAS SYSTEMS OVERVIEW
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
MASTER ALARMS/SIGNALS (EACH SIGNAL COUNTED INDIVIDUALLY)
AREA ALARMS (EACH SYSTEM GAS COUNTED AS SEPARATE ALARM)
ZONE VALVES (EACH SYSTEM GAS COUNTED INDIVIDUALLY)
Total Master Alarm Signals
Quantity of Passing Master Alarm Signals
Quantity of Deficient Master Alarm Signals
Total Quantity of Zone Valves has remained the same from previous medical gas systems inspection.
Total Quantity of Master Alarm Signals has remained the same from previous medical gas systems inspection.
Total Quantity of Area Alarms has remained the same from previous medical gas systems inspection.
Total Area Alarms
Quantity of Passing Area Alarms
Quantity of Deficient Area Alarms
Total Zone Valves
Quantity of Passing Zone Valves
Quantity of Deficient Zone Valves
94.9%
5.1% Quantity of Passing Area Alarms
Quantity of Deficient Area Alarms
93.5%
6.5% Quantity of Passing Zone Valves
Quantity of Deficient Zone Valves
97.1%
2.9% Quantity of Passing Master Alarm Signals
Quantity of Deficient Master Alarm Signals www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
GAS SYSTEMS OVERVIEW
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
OUTLETS, INLETS & MANUFACTURED ASSEMBLIES
SYSTEM DEW POINT
PARTICULATE CONTAMINANTS
Total Dew Point Test Sites
Quantity of Passing Test Sites
Quantity of Deficient Test Sites
Total Particulate Contaminant Test Sites
Quantity of Passing Test Sites
Quantity of Deficient Test Sites
The Test Sites are selected in order to represent System Dew Point in both high-use and low-use areas for the systems tested.
The Test Sites are selected in order to represent Particulate Contaminants in both high-use and low-use areas for the systems tested.
Total Quantity of Outlets, Inlets and Manufactured Assemblies has remained the same from previous medical gas systems inspection.
Total Outlets & Inlets
Quantity of Passing Outlets/Inlets
Quantity of Deficient Outlets/Inlets
99.0%
1.0% Quantity of Passing Outlets/Inlets
Quantity of Deficient Outlets/Inlets
85%
15% Quantity of Passing Test Sites
Quantity of Deficient Test Sites
100%
0% Quantity of Passing Test Sites
Quantity of Deficient Test Sites www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
GAS SYSTEMS OVERVIEW
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
GASEOUS CONTAMINANTS
SOURCE EQUIPMENT
Total Source Systems
Quantity of Passing Source Systems
Quantity of Deficient Source Systems
Total Quantity of Source Equipment has remained the same from previous medical gas systems inspection.
Quantity of Deficient Test Sites
Total Gaseous Contaminants Test Sites
Quantity of Passing Test Sites
The Test Sites are selected in order to represent Gaseous Contaminants in all separate distribution and source systems. 100%
0% Quantity of Passing Test Sites
Quantity of Deficient Test Sites
100%
0% Quantity of Passing Source Systems
Quantity of Deficient Source Systems www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARM DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
QTY
DEFICIENT
QTY
CORRECTED
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
6 0
0 0
6 0
Rewire Panels Independently to Initiating Device
Rewire Panels Independently to Initiating Device With Correct Number of Conductors
Protect Master Alarm Wiring
Repair/Replace Annunciator Module/Alarm Panel
Repair/Replace Power Supply
O) Labeling: No Condition Monitored
P) Other:
Verify and Relabel Condition Monitored
N/A
TOTALS:
MASTER ALARM DEFICIENCIES OVERVIEW
DEFICIENCY DESCRIPTION
A) Initiating Device: Incorrect Location
B) Initiating Device: Improper Set Point
C) Initiating Device: None in Place
CORRECTIVE ACTION NEEDED
Move Initiating Device to Proper Location and/or Install Demand Check
Adjust Alarm Set Point
Install Initiating Device
D) Wiring: Not Independent to at Least (2) Master Panels
E) Wiring: Shared Common Conductors
F) Wiring: Not Protected
G) Panel: Visual and/or Audible Function Failed
H) Panel: Power Failure
N) Labeling: No Source Location
I) Panel: Incorrect Power Source
J) Panel: Incorrect Location
K) Panel: No Signal in Place (Local Alarm Only)
L) Panel: No Signal in Place (No Initiating Device/Wiring)
M) Labeling: System Gas
Connect Alarm to Life Safety Branch
Move Alarm Panel to a Proper Location
Wire Initiating Device to Master Alarm Panel(s)
Add Signal/Signal Components to Master Alarm System
Verify and Relabel System Gas
Verify and Relabel Source Location www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARM DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation &
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
26 Police O2 Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ O ✓
O) Labeling:
Condition
Monitored-Should be Line Pressure
Low
27 Police O2 Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ O ✓
O) Labeling:
Condition
Monitored-Should be Line Pressure
High
87 Engineering O2 Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ O ✓
O) Labeling:
Condition
Monitored-Should be Line Pressure
Low
88 Engineering O2 Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ O ✓
O) Labeling:
Condition
Monitored-Should be Line Pressure
High
Old Switchboard B57
B102A O2 Beacon Total Alert 2 High Line Pressure
Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ O ✓
O) Labeling:
Condition
Monitored-Should be Line Pressure
Low
DEFICIENT MASTER ALARM LOCATION
DEFICIENCY
CODE
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
/O R
C
O R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
MASTER ALARM DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARM DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation &
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
DEFICIENT MASTER ALARM LOCATION
DEFICIENCY
CODE
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
MASTER ALARM DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
Old Switchboard B57
B102A O2 Beacon Total Alert 2 Low Line Pressure
Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ O ✓
O) Labeling:
Condition
Monitored-Should be Line Pressure
High www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
AREA ALARM DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
L) Labeling: Gas System Monitored Not Labeled
M) Other: 0
J) Module Function: Module Display Blank or Error Message
K) Labeling: Area Monitored Labeling Missing/Incorrect
B) Transducer Location: Incorrect Transducer Location
C) Transducer Location: No Demand Check
D) Transducer Function: Incorrect Gas Pressure Displayed
E) Alarm Location: Improper Alarm Panel Location
F) Alarm Location: No Alarm in Place
G) Panel Function: Visual and/or Audible Alarm Function Failure
H) Panel Function: Power Failure
I) Panel Function: Incorrect Power Source
Install Demand Check
Repair/Recalibrate/Replace Transducer
Move Alarm to Appropriate Location
Install Area Alarm & Transducer
Repair/Replace Annunciator Module / Ribbon Cable / Alarm Panel
Repair/Replace Power Supply
Connect Alarm to Life Safety Branch
Repair/Replace Module or Defective Transducer
Label Area Monitored
Label Gas System Monitored
N/A
AREA ALARM DEFICIENCIES OVERVIEW
ACTION NEEDED
Adjust Set Points to +/-20% Operating Pressure
Move Transducers to Correct Side of Zone Valve
QTY
CORRECTED
DEFICIENCY DESCRIPTION
A) Alarm Set Points: Incorrect High and/or Low Alarm Set Point(s)
QTY
DEFICIENT
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
AREA ALARM DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
A larm
P h ysical L ocation
A rea/Z on e M on itored
S ystem
G as
M an u facturer
M od el
P ressu re R eading
(P ositive in P
SI/N
egative in In/H g)
H igh P ressu re A larm S etting (P ositive in P
SI/N
egative in In/H g)
L ow
P ressure A larm S etting (P ositive in P
SI/N
egative in In/H g)
A larm
S et P oin ts
T ransd u cer L ocation
T ransd u cer F un ction
P an el L ocation
P an el F u nction
M od u le F un ction
L abelin g
O th er
47 OPP Fast Track 10 A2
B38 Infusion Suite A153(Chairs 1-8), Fast
Track 1-10 (A154, A155, A156, A157, A158, A159, A163, A164, O2 Allied Impact 53 12 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored Labeling Missing/Incorrect
48 OPP Fast Track 10 A2 MA Allied Impact 48 60 40 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored Labeling Missing/Incorrect
49 OPP Fast Track 10 A2 VAC Allied Impact 18 60 40 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored Labeling Missing/Incorrect
O/S E3 Roadrrunner Rd Entrance Middle alarm
Panel
B60 Roadrunner Rd E146- 156+ SE102, Vac Only
O2 Amico A2A 54 60 40 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored Labeling Missing/Incorrect
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
DEFICIENCY CODE
AREA ALARM DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
DEFICIENT AREA ALARM LOCATION
TOTALS: 9 0
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
AREA ALARM DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
A larm
P h ysical L ocation
A rea/Z on e M on itored
S ystem
G as
M an u facturer
M od el
P ressu re R eading
(P ositive in P
SI/N
egative in In/H g)
H igh P ressure A larm
S etting
(P ositive in P
SI/N
egative in In/H g)
L ow
P ressure A larm S etting (P ositive in P
SI/N
egative in In/H g)
A larm
S et P oin ts
T ransd u cer L ocation
T ran sd u cer F un ction
P an el L ocation
P an el F u n ction
M od u le F u n ction
L abelin g
O th er
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
DEFICIENCY CODE
AREA ALARM DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
DEFICIENT AREA ALARM LOCATION
O/S E3 Roadrrunner Rd Entrance Middle alarm
Panel MA Amico A2A 47 12 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored
Labeling Missing/Incorrect
O/S E3 Roadrrunner Rd Entrance Middle alarm
Panel VAC Amico A2A 21 60 40 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored
Labeling Missing/Incorrect
O/S E3 Roadrrunner Rd Entrance Bottom alarm
Panel
B60 E2 Medicine E111- E122
O2 Amico A2A 54 60 40 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored Labeling Missing/Incorrect
O/S E3 Roadrrunner Rd Entrance Bottom alarm
Panel MA Amico A2A 48 12 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored
Labeling Missing/Incorrect
O/S E3 Roadrrunner Rd Entrance Bottom alarm
Panel VAC Amico A2A 22 60 40 ✓ ✓ ✓ ✓ ✓ ✓ K ✓ K) Labeling: Area Monitored
Labeling Missing/Incorrect www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
ZONE VALVE DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
Replace Handle
Replace Window
Verify Rooms/Area Served and Relabel
Verify Rooms/Area Served and Relabel (Repiping or Decommissioning Valve May Be Necessary)
E: General Condition: Missing/Broken Handle
F: General Condition: Missing/Broken Window
G: Labeling: Inaccurate Rooms/Area Served
H: Labeling: Labeled as if in Series
ACTION NEEDED
Replace Gauge
Install Gauge With Correct Scale Range
Install Gauge
Repipe Valve So Gauge is Downstream of Valve
Q: Location: Improper Location
DEFICIENCY DESCRIPTION
A: Gauge: Inaccurate Pressure Reading
B: Gauge: Incorrect Scale Range
C: Gauge: No Gauge in Port
D: Gauge: On Source Side of Valve
N: Leak: Gas Passes Through Valve When Closed
P: Location: Hidden or Not Accessible
M: Leak: External Leak
J: Labeling: Missing System Gas
K: Labeling: Missing Warning Verbiage
L: Labeling: Rooms/Area not Attributed a Zone Valve
O: Leak: Gauge Connection Leaks
I: Labeling: Missing Non-Standard Pressure Verbiage
R: Location: No Valve in Place
N/A
15 0TOTALS:
S: Other: 0
Label All Required Components with Non-Standard Pressure
Update Label to Include System Gas
Update Label to Include Warning Verbiage
Verify if Rooms/Area is Served by an Existing Zone Valve (Installation Necessary if Rooms/Area is not Currently Served by a Zone Valve)
Rebuild if Possible or Replace Leaking Valve
Rebuild if Possible or Replace Leaking Valve
Re-install Gauge Using Oxygen-Safe Sealant (Replacement of Valve May be Necessary)
Move Valve to Visible and Accessible Location
Move Valve to Compliant Location
Install Zone Valve
QTY
DEFICIENT
ZONE VALVE DEFICIENCIES OVERVIEW
QTY
CORRECTED
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
ZONE VALVE DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
L ocation
(O /S = utside & pp = O pposite)
R oom s or A rea Served
System G as
M an ufacturer
B all
G ate/G lob e
G au ge P ressure R eading (P ositive in P
SI / N
egative in "H g)
G au ge
G eneral C ondition
L abeling
L eaks
L ocation
O ther
47 O/S A147 1st Floor B38 BIPAP/CPAP, A144, Pulm Testing A144-146, Home Oxy A148
O2 Allied ✓ 54 ✓ ✓ G ✓ ✓ ✓ G) Labeling: Inaccurate Rooms/Area Served: 1st Floor B38 BIPAP/CPAP, A144, Pulm Testing A144- 146, Home Oxy A148
48 O/S A147 MA Allied ✓ 51 ✓ ✓ G ✓ ✓ ✓ G) Labeling: Inaccurate Rooms/Area Served: 1st Floor B38 BIPAP/CPAP, A144, Pulm Testing A144- 146, Home Oxy A148
49 O/S A147 VAC Allied ✓ 20 ✓ ✓ G ✓ ✓ ✓ G) Labeling: Inaccurate Rooms/Area Served: 1st Floor B38 BIPAP/CPAP, A144, Pulm Testing A144- 146, Home Oxy A148
50 Opp Fast Track
B38 Infusion Suite A153(Chairs 1-8), Fast Track 1-10 (A154, A155, A156, A157, A158, A159, A163, A164, A165, A166)
O2 Allied ✓ 56 ✓ ✓ G ✓ ✓ ✓
G) Labeling: Inaccurate Rooms/Area Served: B38 Infusion Suite A153(Chairs 1-8), Fast Track 1-10 (A154, A155, A156, A157, A158, A159, A163, A164, A165, A166)
51 Opp Fast Track MA Allied ✓ 851 ✓ ✓ G ✓ ✓ ✓
G) Labeling: Inaccurate Rooms/Area Served: B38 Infusion Suite A153(Chairs 1-8), Fast Track 1-10 (A154, A155, A156, A157, A158, A159, A163, A164, A165, A166)
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction Taken, &
R efer to W ork O rder if A pplicable
DEFICIENT VALVE LOCATION
DEFICIENCY
CODE
ZONE VALVE DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
ZONE VALVE DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
L ocation
(O /S =
O pposite)
R oom s or A rea Served
System G as
M an ufacturer
B all
G ate/G lob e
G au ge P ressure R eading (P ositive in P
SI / N
egative in "H g)
G au ge
G eneral C ondition
L abeling
L eaks
L ocation
O ther
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction Taken, &
R efer to W ork O rder if A pplicable
DEFICIENT VALVE LOCATION
DEFICIENCY
CODE
ZONE VALVE DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
52 Opp Fast Track VAC Allied ✓ 19 ✓ ✓ G ✓ ✓ ✓
G) Labeling: Inaccurate Rooms/Area Served: B38 Infusion Suite A153(Chairs 1-8), Fast Track 1-10 (A154, A155, A156, A157, A158, A159, A163, A164, A165, A166)
83 O/S OR 3 VAC Beacon ✓ 14 A ✓ ✓ ✓ ✓ ✓ A) Gauge: Inaccurate Pressure Reading
Bldg 60 O/S E2 A4
Entry B60 E2 Medicine E111-E122 O2 Beacon ✓ 60 A ✓ G ✓ ✓ ✓ A) Gauge: Inaccurate Pressure Reading: G) Labeling:
Inaccurate Rooms/Area Served
Bldg 60 O/S E2 A4
Entry MA Beacon ✓ 42 A ✓ G ✓ ✓ ✓ A) Gauge: Inaccurate Pressure Reading: G) Labeling:
Inaccurate Rooms/Area Served
Bldg 60 O/S E2 A4
Entry VAC Beacon ✓ 20 ✓ ✓ G ✓ ✓ ✓ G) Labeling: Inaccurate Rooms/Area Served www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
ZONE VALVE DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
L ocation
(O /S =
O pposite)
R oom s or A rea Served
System G as
M an ufacturer
B all
G ate/G lob e
G au ge P ressure R eading (P ositive in P
SI / N
egative in "H g)
G au ge
G eneral C ondition
L abeling
L eaks
L ocation
O ther
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction Taken, &
R efer to W ork O rder if A pplicable
DEFICIENT VALVE LOCATION
DEFICIENCY
CODE
ZONE VALVE DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
Bldg 60 O/S E2
A5 Entry B60 Roadrunner Rd E146-156+
SE102, Vac Only O2 Allied ✓ 56 ✓ ✓ G ✓ ✓ ✓ G) Labeling: Inaccurate Rooms/Area Served
Bldg 60 O/S E2
A5 Entry MA Allied ✓ 52 ✓ ✓ G ✓ ✓ ✓ G) Labeling: Inaccurate Rooms/Area Served
Bldg 60 O/S E2
A5 Entry VAC Allied ✓ 20 ✓ ✓ G ✓ ✓ ✓ G) Labeling: Inaccurate Rooms/Area Served www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
OUTLET & INLET DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
QTY
DEFICIENT
QTY
CORRECTED
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
13 0
8 2
0 0
0 0
0 0
0 0
0 0
0 0
0 0
0 0
21 2
S: Static Pressure: Pressure Out of Required Range
T: Location: Insufficient Terminals per Bed
U: Location: Terminals in Non-Compliant Location
N: Leak: Outer Check
O: Low Flow: Depressurized or Disconnected from System
P: Low Flow: Pressure Gas
Q: Low Flow: Undersized Piping/Components
R: Low Flow: Vacuum
Install Outlets/Inlets to Meet Required Minimum
Rebuild Outer Check
Rebuild Outlet / Remove Obstruction
Replace or Convert to Compliant Outlet/Inlet Style
Rebuild Inlet and Flush with Enzyme
Reset System Pressure or Label as Nonstandard
Identify point of disconnection and repair
Replace Damaged Hose Assembly
Rebuild/Replace Damaged Faceplate
Replace with Correct Gas Specific Faceplate
Replace Missing Screws
Replace Damaged Faceplate
ACTION NEEDED
Identify point of cross-connection and separate
Purge system and/or remove contamination source
Purge system and/or service dryer (if Medical Air)
Replace Damaged Faceplate
V: Other
TOTALS:
M: Leak: Inner Check
Replace System Label
Replace or Convert to Compliant Outlet/Inlet Style
Replace or Convert to Compliant Outlet/Inlet Style
Rebuild Inner Check - Temporary Shutdown Required
OUTLET/INLET DEFICIENCIES OVERVIEW
Remove Outlets/Inlets from Non-Compliant Area
N/A
DEFICIENCY DESCRIPTION
A: Gas Concentration: Cross-Connection Found
B: Gas Concentration: Gas Concentration Contaminated
C: Gas Concentration: Water Present
D: General Condition: Damaged Faceplate
E: General Condition: Damaged Hose Assembly
F: General Condition: Failed Adapter Connect and Release
G: General Condition: Incorrect Gas Connection
H: General Condition: Missing Screws
I: Labeling: Missing 'Use No Oil'
J: Labeling: Missing System Label
K: Labeling: No 'Use No Oil'
L: Labeling: No System Label www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
OUTLET & INLET DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
A rea or Z one
R oom or B ed #
(Inventory taken clockw ise around room , from top to bottom and/or left to right)
S ystem
G as
M anufacturer
M odel
P osition
G as C oncentration
G eneral C ondition
L abeling
L eaks
M inim um F low
S tatic P ressure
S ufficient T erm inals/B ed
O ther
2nd Floor Step
Down 232 VAC Beacon DC-DISS Rail ✓ ✓ ✓ N✓ ✓ ✓ ✓ ✓ N) Leak: Outer Check:
Repaired on 12/04/2023
2nd Floor Step
Down 241 VAC Beacon DC-DISS Rail ✓ ✓ ✓ N✓ ✓ ✓ ✓ ✓ N) Leak: Outer Check:
Repaired on 12/04/2023
1219 B50 Procedure 4 350 VAC Allied 560 Recessed ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1381 B57 OR Suite OR 3 Rm 293 VAC Allied 460 Wall
(Recessed)
✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1407 B57 OR Suite OR 4 Rm 299 VAC Allied 560 Recessed ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1409 B57 OR Suite OR 4 Rm 299 VAC Allied HOSE-DISS Hose Drop ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1410 B57 OR Suite OR 4 Rm 299 VAC Allied HOSE-DISS Hose Drop ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
DEFICIENT OUTLET/INLET LOCATION DEFICIENCY CODE
OUTLET/INLET DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
OUTLET & INLET DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
A rea or Z one
R oom or B ed #
(Inventory taken clockw ise around room , from top to bottom and/or left to right)
S ystem
G as
M anufacturer
M odel
P osition
G as C oncentration
G eneral C ondition
L abeling
L eaks
M inim um F low
S tatic P ressure
S ufficient T erm inals/B ed
O ther
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
DEFICIENT OUTLET/INLET LOCATION DEFICIENCY CODE
OUTLET/INLET DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
1426 B57 OR Suite OR 5 Rm 292 VAC Allied 560 Wall
(Recessed)
✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1427 B57 OR Suite OR 5 Rm 292 VAC Allied 560 Wall
(Recessed)
✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1429 B57 OR Suite OR 5 Rm 292 VAC Allied 560 Wall
(Recessed)
✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1500 B57 OR Suite OR 10 Rm 2209 VAC Allied HOSE-DISS Hose Drop ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1501 B57 OR Suite OR 10 Rm 2209 VAC Allied 560 Wall
(Recessed)
✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1514 B57 OR Suite OR 11 Rm 2205 VAC Allied HOSE-DISS Hose Drop ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1528 B57 OR Suite OR 12 Rm 2203 VAC Allied HOSE-DISS Hose Drop ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
OUTLET & INLET DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
A rea or Z one
R oom or B ed #
(Inventory taken clockw ise around room , from top to bottom and/or left to right)
S ystem
G as
M anufacturer
M odel
P osition
G as C oncentration
G eneral C ondition
L abeling
L eaks
M inim um F low
S tatic P ressure
S ufficient T erm inals/B ed
O ther
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
DEFICIENT OUTLET/INLET LOCATION DEFICIENCY CODE
OUTLET/INLET DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
1542 B57 OR Suite OR 13 Rm 2202 VAC Allied HOSE-DISS Hose Drop ✓ ✓ ✓ M ✓ ✓ ✓ ✓ M) Leak: Inner Check
1554 B57 PRE-OP Bay 2 O2 Allied 460 Wall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check
1565 B57 PRE-OP Bay 4 MA Allied 460 Wall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check
1569 B57 PRE-OP Bay 5 O2 Allied 460 Wall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check
1574 B57 PRE-OP Bay 6 O2 Allied 460 Wall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check
1580 B57 PRE-OP Bay 7 O2 Allied 460 Wall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check
1585 B57 PRE-OP Bay 8 O2 Allied 460 Wall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
OUTLET & INLET DEFICIENCIES
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
A rea or Z one
R oom or B ed #
(Inventory taken clockw ise around room , from top to bottom and/or left to right)
S ystem
G as
M anufacturer
M odel
P osition
G as C oncentration
G eneral C ondition
L abeling
L eaks
M inim um F low
S tatic P ressure
S ufficient T erm inals/B ed
O ther
D E
F
IC
IE
N
C Y
D E
T A
IL
A
N D
R
R E
C T
IV
E
A C
T
IO
N
*P lease D ate, D escribe A ction T aken, & R efer to W ork
O rder if A pplicable
DEFICIENT OUTLET/INLET LOCATION DEFICIENCY CODE
OUTLET/INLET DEFICIENCIES BREAKDOWN (BY ITEM NUMBER)
1606 B57 PACU Bay 2 MA Allied 460 Headwall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check
1669 B57 PACU Bay 11 MA Allied 460 Headwall
(Recessed)
✓ ✓ ✓ N ✓ ✓ ✓ ✓ N) Leak: Outer Check www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
1 Police MA Beacon Total Alert 2 Low Line Pressure Bldg. 57 Basement BA046 Mech Rm
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
2 Police MA Beacon Total Alert 2 High Line Pressure Bldg. 57 Basement BA046 Mech Rm
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
3 Police MA Beacon Total Alert 2 Inlet Filter Dirty Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
4 Police MA Beacon Total Alert 2 Dental-Dry Malfunction Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
5 Police MA Beacon Total Alert 2 Backup Compressor On Bldg. 57 Basement BA046 Mech Rm
75 PSI ✓ ✓ ✓ ✓ ✓ PASS
6 Police MA Beacon Total Alert 2 Aftercooler Hi Temp Bldg. 57 Basement BA046 Mech Rm
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
7 Police MA Beacon Total Alert 2 Dew Point High Bldg. 57 Basement BA046 Mech Rm
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
8 Police MA Beacon Total Alert 2 Carbon Monoxide High Bldg. 57 Basement BA046 Mech Rm
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
9 Police MA Beacon Total Alert 2 Dry Malfunction Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
10 Police MA Beacon Total Alert 2 Post Filter Dirty Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
11 Police VAC Beacon Total Alert 2 Filter Back Pressure Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
12 Police VAC Beacon Total Alert 2 Low Bldg. 57 Basement BA046 Mech Rm
12 inHg ✓ ✓ ✓ ✓ ✓ PASS
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
13 Police VAC Beacon Total Alert 2 Backup Pump On Bldg. 57 Basement BA046 Mech Rm
18 inHg ✓ ✓ ✓ ✓ ✓ PASS
14 Police MA Beacon Total Alert 2 Dental-Low Line Pressure Bldg. 57 Basement BA046 Mech Rm
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
15 Police MA Beacon Total Alert 2 Dental-High Line Pressure Bldg. 57 Basement BA046 Mech Rm
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
16 Police MA Beacon Total Alert 2 Dental-Thermal Shut Down Bldg. 57 Basement BA046 Mech Rm
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
17 Police MA Beacon Total Alert 2 Dental-Dew Point High Bldg. 57 Basement BA046 Mech Rm
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
18 Police MA Beacon Total Alert 2 Dental-Carbon Monoxide
High Bldg. 57 Basement BA046 Mech Rm
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
19 Police CO2 Beacon Total Alert 2 Low Line Pressure Manifold Rm 40 PSI ✓ ✓ ✓ ✓ ✓ PASS
20 Police CO2 Beacon Total Alert 2 High Line Pressure Manifold Rm 60 PSI ✓ ✓ ✓ ✓ ✓ PASS
21 Police CO2 Beacon Total Alert 2 Reserve Supply In Use Manifold Rm 120 PSI ✓ ✓ ✓ ✓ ✓ PASS
22 Police O2 Beacon Total Alert 2 Reserve Supply Low Bulk Site 40 inH2O ✓ ✓ ✓ ✓ ✓ PASS
23 Police O2 Beacon Total Alert 2 Liquid Level Low Bulk Site 45 inH2O ✓ ✓ ✓ ✓ ✓ PASS
24 Police O2 Beacon Total Alert 2 Reserve Supply In Use Bulk Site 90 PSI ✓ ✓ ✓ ✓ ✓ PASS www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
25 Police O2 Beacon Total Alert 2 Reserve Pressure Low Bulk Site 90 PSI ✓ ✓ ✓ ✓ ✓ PASS
26 Police O2 Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ O ✓ FAIL
27 Police O2 Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ O ✓ FAIL
28 Police O2 Beacon Total Alert 2 High Line Pressure Bldg. 60 Mech Rm
Near Grill Patio
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
29 Police O2 Beacon Total Alert 2 Low Line Pressure Bldg. 60 Mech Rm
Near Grill Patio
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
30 Police N2O Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
31 Police N2O Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
32 Police N2O Beacon Total Alert 2 Reserve Supply In Use Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
100 PSI ✓ ✓ ✓ ✓ ✓ PASS
33 Police N2 Beacon Total Alert 2 Reserve Supply In Use Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
250 PSI ✓ ✓ ✓ ✓ ✓ PASS
34 Police N2 Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
190 PSI ✓ ✓ ✓ ✓ ✓ PASS
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
35 Police N2 Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
140 PSI ✓ ✓ ✓ ✓ ✓ PASS
36 Police MA Beacon Total Alert 2 Carbon Monoxide High Bldg. 38 Basement A41 Mech Room
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
37 Police MA Beacon Total Alert 2 Dew Point High Bldg. 38 Basement A41 Mech Room
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
38 Police MA Beacon Total Alert 2 Aftercooler Hi Temp Bldg. 38 Basement A41 Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
39 Police MA Beacon Total Alert 2 Backup Compressor On Bldg. 38 Basement A41 Mech Room
75 PSI ✓ ✓ ✓ ✓ ✓ PASS
40 Police MA Beacon Total Alert 2 High Line Pressure Bldg. 38 Basement A41 Mech Room
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
41 Police MA Beacon Total Alert 2 Low Line Pressure Bldg. 38 Basement A41 Mech Room
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
42 Police MA Beacon Total Alert 2 Post Filter Dirty Bldg. 38 Basement A41 Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
43 Police MA Beacon Total Alert 2 Dry Malfunction Bldg. 38 Basement A41 Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
44 Police MA Beacon Total Alert 2 Inlet Filter Dirty Bldg. 38 Basement A41 Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
45 Police MA Beacon Total Alert 2 Backup Compressor On Bldg. 60 Penthouse
Mech Room
75 PSI ✓ ✓ ✓ ✓ ✓ PASS
46 Police MA Beacon Total Alert 2 Dew Point High Bldg. 60 Penthouse
Mech Room
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
47 Police MA Beacon Total Alert 2 Aftercooler Hi Temp Bldg. 60 Penthouse
Mech Room
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
48 Police MA Beacon Total Alert 2 Carbon Monoxide High Bldg. 60 Penthouse
Mech Room
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
49 Police MA Beacon Total Alert 2 High Line Pressure Bldg. 60 Penthouse
Mech Room
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
50 Police MA Beacon Total Alert 2 Low Line Pressure Bldg. 60 Penthouse
Mech Room
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
51 Police MA Beacon Total Alert 2 Post Filter Dirty Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
52 Police MA Beacon Total Alert 2 Dry Malfunction Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
53 Police MA Beacon Total Alert 2 Inlet Filter Dirty Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
54 Police VAC Beacon Total Alert 2 Low Line Pressure Bldg. 38 Basement A41 Mech Room
12 inHg ✓ ✓ ✓ ✓ ✓ PASS
55 Police VAC Beacon Total Alert 2 Filter Back Pressure Bldg. 38 Basement A41 Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
56 Police VAC Beacon Total Alert 2 Backup Pump On Bldg. 60 Penthouse
Mech Room 18 inHg ✓ ✓ ✓ ✓ ✓ PASS
57 Police VAC Beacon Total Alert 2 Filter Back Pressure Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
58 Police VAC Beacon Total Alert 2 Low Line Pressure Bldg. 60 Penthouse
Mech Room 12 inHg ✓ ✓ ✓ ✓ ✓ PASS www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
59 Police O2 Beacon Total Alert 2 High Line Pressure Bldg. 38 Basement A41 Mech Room
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
60 Police O2 Beacon Total Alert 2 Low Line Pressure Bldg. 38 Basement A41 Mech Room
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
61 Police VAC Beacon Total Alert 2 Backup Pump On Bldg. 38 Basement A41 Mech Room
18 inHg ✓ ✓ ✓ ✓ ✓ PASS
62 Engineering MA Beacon Total Alert 2 Low Line Pressure Bldg. 57 Basement BA046 Mech Rm
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
63 Engineering MA Beacon Total Alert 2 High Line Pressure Bldg. 57 Basement BA046 Mech Rm
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
64 Engineering MA Beacon Total Alert 2 Inlet Filter Dirty Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
65 Engineering MA Beacon Total Alert 2 Dental-Dryer Malfunction Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
66 Engineering MA Beacon Total Alert 2 Backup Compressor On Bldg. 57 Basement BA046 Mech Rm
75 PSI ✓ ✓ ✓ ✓ ✓ PASS
67 Engineering MA Beacon Total Alert 2 Aftercooler Hi Temp Bldg. 57 Basement BA046 Mech Rm
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
68 Engineering MA Beacon Total Alert 2 Dew Point High Bldg. 57 Basement BA046 Mech Rm
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
69 Engineering MA Beacon Total Alert 2 Carbon Monoxide High Bldg. 57 Basement BA046 Mech Rm
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
70 Engineering MA Beacon Total Alert 2 Dryer Malfunction Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
71 Engineering MA Beacon Total Alert 2 Post Filter Dirty Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
72 Engineering VAC Beacon Total Alert 2 Filter Back Pressure Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
73 Engineering VAC Beacon Total Alert 2 Low Bldg. 57 Basement BA046 Mech Rm
12 inHg ✓ ✓ ✓ ✓ ✓ PASS
74 Engineering VAC Beacon Total Alert 2 Backup Pump On Bldg. 57 Basement BA046 Mech Rm
18 inHg ✓ ✓ ✓ ✓ ✓ PASS
75 Engineering MA Beacon Total Alert 2 Dental-Low Line Pressure Bldg. 57 Basement BA046 Mech Rm
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
76 Engineering MA Beacon Total Alert 2 Dental-High Line Pressure Bldg. 57 Basement BA046 Mech Rm
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
77 Engineering MA Beacon Total Alert 2 Dental-Thermal Shut Down Bldg. 57 Basement BA046 Mech Rm
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
78 Engineering MA Beacon Total Alert 2 Dental-Dew Point High Bldg. 57 Basement BA046 Mech Rm
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
79 Engineering MA Beacon Total Alert 2 Dental-Carbon Monoxide
High Bldg. 57 Basement BA046 Mech Rm
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
80 Engineering CO2 Beacon Total Alert 2 Low Line Pressure Manifold Rm 40 PSI ✓ ✓ ✓ ✓ ✓ PASS
81 Engineering CO2 Beacon Total Alert 2 High Line Pressure Manifold Rm 60 PSI ✓ ✓ ✓ ✓ ✓ PASS
82 Engineering CO2 Beacon Total Alert 2 Reserve Supply In Use Manifold Rm 120 PSI ✓ ✓ ✓ ✓ ✓ PASS www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
83 Engineering O2 Beacon Total Alert 2 Reserve Supply Low Bulk Site 40 inH2O ✓ ✓ ✓ ✓ ✓ PASS
84 Engineering O2 Beacon Total Alert 2 Liquid Level Low Bulk Site 45 inH2O ✓ ✓ ✓ ✓ ✓ PASS
85 Engineering O2 Beacon Total Alert 2 Reserve Supply In Use Bulk Site 90 PSI ✓ ✓ ✓ ✓ ✓ PASS
86 Engineering O2 Beacon Total Alert 2 Reserve Pressure Low Bulk Site 90 PSI ✓ ✓ ✓ ✓ ✓ PASS
87 Engineering O2 Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ O ✓ FAIL
88 Engineering O2 Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ O ✓ FAIL
89 Engineering O2 Beacon Total Alert 2 High Line Pressure Bldg. 60 Penthouse
Mech Room
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
90 Engineering O2 Beacon Total Alert 2 Low Line Pressure Bldg. 60 Penthouse
Mech Room
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
91 Engineering N2O Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
92 Engineering N2O Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
93 Engineering N2O Beacon Total Alert 2 Reserve Supply In Use Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
100 PSI ✓ ✓ ✓ ✓ ✓ PASS
94 Engineering N2 Beacon Total Alert 2 Reserve Supply In Use Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
250 PSI ✓ ✓ ✓ ✓ ✓ PASS
95 Engineering N2 Beacon Total Alert 2 High Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
190 PSI ✓ ✓ ✓ ✓ ✓ PASS
96 Engineering N2 Beacon Total Alert 2 Low Line Pressure Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
140 PSI ✓ ✓ ✓ ✓ ✓ PASS
97 Engineering MA Beacon Total Alert 2 Carbon Monoxide High Bldg. 38 Basement A41 Mech Room
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
98 Engineering MA Beacon Total Alert 2 Dew Point High Bldg. 38 Basement A41 Mech Room
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
99 Engineering MA Beacon Total Alert 2 Aftercooler Hi Temp Bldg. 38 Basement A41 Mech Room
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
100 Engineering MA Beacon Total Alert 2 Backup Compressor On Bldg. 38 Basement A41 Mech Room
75 PSI ✓ ✓ ✓ ✓ ✓ PASS
101 Engineering MA Beacon Total Alert 2 High Line Pressure Bldg. 38 Basement A41 Mech Room
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
102 Engineering MA Beacon Total Alert 2 Low Line Pressure Bldg. 38 Basement A41 Mech Room
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
103 Engineering MA Beacon Total Alert 2 Post Filter Dirty Bldg. 38 Basement A41 Mech Room www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
104 Engineering MA Beacon Total Alert 2 Dry Malfunction Bldg. 38 Basement A41 Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
105 Engineering MA Beacon Total Alert 2 Inlet Filter Dirty Bldg. 38 Basement A41 Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
106 Engineering MA Beacon Total Alert 2 Backup Compressor On Bldg. 60 Penthouse
Mech Room
75 PSI ✓ ✓ ✓ ✓ ✓ PASS
107 Engineering MA Beacon Total Alert 2 Dew Point High Bldg. 60 Penthouse
Mech Room
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
108 Engineering MA Beacon Total Alert 2 Aftercooler Hi Temp Bldg. 60 Penthouse
Mech Room
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
109 Engineering MA Beacon Total Alert 2 Carbon Monoxide High Bldg. 60 Penthouse
Mech Room
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
110 Engineering MA Beacon Total Alert 2 High Line Pressure Bldg. 60 Penthouse
Mech Room
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
111 Engineering MA Beacon Total Alert 2 Low Line Pressure Bldg. 60 Penthouse
Mech Room
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
112 Engineering MA Beacon Total Alert 2 Post Filter Dirty Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
113 Engineering MA Beacon Total Alert 2 Dryer Malfunction Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
114 Engineering MA Beacon Total Alert 2 Inlet Filter Dirty Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
115 Engineering VAC Beacon Total Alert 2 Low Line Pressure Bldg. 38 Basement
Mech Room 12 inHg ✓ ✓ ✓ ✓ ✓ PASS www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
116 Engineering VAC Beacon Total Alert 2 Filter Back Pressure Bldg. 38 Basement
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
117 Engineering VAC Beacon Total Alert 2 Backup Pump On Bldg. 60 Penthouse
Mech Room 18 inHg ✓ ✓ ✓ ✓ ✓ PASS
118 Engineering VAC Beacon Total Alert 2 Filter Back Pressure Bldg. 60 Penthouse
Mech Room
N/A ✓ ✓ ✓ ✓ ✓ PASS
119 Engineering VAC Beacon Total Alert 2 Low Line Pressure Bldg. 60 Penthouse
Mech Room 12 inHg ✓ ✓ ✓ ✓ ✓ PASS
120 Engineering O2 Beacon Total Alert 2 High Line Pressure Bldg. 38 Basement A41 Mech Room
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
121 Engineering O2 Beacon Total Alert 2 Low Line Pressure Bldg. 38 Basement A41 Mech Room
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
122 Engineering VAC Beacon Total Alert 2 Backup Pump On Bldg. 38 Basement A41 Mech Room
18 inHg ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Low Line Pressure
Bldg. 57 Basement BA046 Mech Rm
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 High Line Pressure
Bldg. 57 Basement BA046 Mech Rm
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Inlet Filter Dirty
Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Dental-Dry Malfunction
Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Backup Compressor On
Bldg. 57 Basement BA046 Mech Rm
75 PSI ✓ ✓ ✓ ✓ ✓ PASS
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
Old Switchboard B57
B102A MA Beacon Total Alert 2 Aftercooler Hi Temp
Bldg. 57 Basement BA046 Mech Rm
ºF ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Dew Point High
Bldg. 57 Basement BA046 Mech Rm
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Carbon Monoxide High
Bldg. 57 Basement BA046 Mech Rm
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Dry Malfunction
Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Post Filter Dirty
Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A VAC Beacon Total Alert 2 Filter Back Pressure
Bldg. 57 Basement BA046 Mech Rm
N/A ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A VAC Beacon Total Alert 2 Low Line Pressure
Bldg. 57 Basement BA046 Mech Rm
12 inHg ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A VAC Beacon Total Alert 2 Backup Pump On
Bldg. 57 Basement BA046 Mech Rm
18 inHg ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Dental-Low Line Pressure
Bldg. 57 Basement BA046 Mech Rm
40 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Dental-High Line Pressure
Bldg. 57 Basement BA046 Mech Rm
60 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Dental-Thermal Shut Down
Bldg. 57 Basement BA046 Mech Rm
N/A ºF ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A MA Beacon Total Alert 2 Dental-Dew Point High
Bldg. 57 Basement BA046 Mech Rm
35 ºF ✓ ✓ ✓ ✓ ✓ PASS
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch or S ou rce L ocation
S et P oin t
U nit of M easu re
In itiatin g D evice
W irin g
P an el
L abelin g
O th er
C om p on en t C on d ition
MASTER ALARMS: LOCATION & SPECIFICATION FUNCTION & CONDITION
Old Switchboard B57
B102A MA Beacon Total Alert 2
Dental-Carbon Monoxide High
Bldg. 57 Basement BA046 Mech Rm
10 PPM ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A CO2 Beacon Total Alert 2 Low Line Pressure Manifold Rm 40 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A CO2 Beacon Total Alert 2 High Line Pressure Manifold Rm 60 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A CO2 Beacon Total Alert 2 Reserve Supply In Use Manifold Rm 120 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A O2 Beacon Total Alert 2 Reserve Supply Low Bulk Site 40 inH2O ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A O2 Beacon Total Alert 2 Liquid Level Low Bulk Site 45 inH2O ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A O2 Beacon Total Alert 2 Reserve Supply In Use Bulk Site 90 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A O2 Beacon Total Alert 2 Reserve Pressure Low Bulk Site 90 PSI ✓ ✓ ✓ ✓ ✓ PASS
Old Switchboard B57
B102A O2 Beacon Total Alert 2 High Line Pressure
Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
60 PSI ✓ ✓ ✓ O ✓ FAIL
Old Switchboard B57
B102A O2 Beacon Total Alert 2 Low Line Pressure
Bldg. 57 Outside Mech Rm Across
O2 Bulk Pad
40 PSI ✓ ✓ ✓ O ✓ FAIL
www.evergreenmedical.com © 2021 Evergreen Medical Services, Inc. - VER 1.3 Corporate Office: 1-800-872-8992
MASTER ALARMS
Medical Gas and Vacuum Systems Inspection/Testing
Southern AZ VA Tucson, AZ
M aster A larm P an el
L ocation
If M ultiple P anels
P er Location
(e.g. "E ngineering P anel #2")
S ystem
G as
M an u facturer
M od el
C on d ition
M on itored
S w itch…
This is the start of the file's text. The full file is on GovTribe.