36C26224Q0558.docx

DOCX document 74 KB Posted

Attached to
J065--HOOD FUME MAINTENANCE - Base plus 4 (8/1/24 - 7/31/29
Solicitation number
36C26224Q0558
Issued by

About this file

This document is a statement of work for a federal contract solicitation to provide annual preventive maintenance and corrective repair services for laboratory equipment at the Southern Arizona VA Health Care System. The contractor will perform on-site certification testing and calibration of fume hoods, biosafety cabinets, clean rooms, and other laboratory equipment listed in an attachment in accordance with standards such as USP 797 and 800. Services include air flow and filter testing, smoke studies, particle counting, and environmental sampling. The contractor must respond within 24 hours for emergency repairs from 8am to 5pm Monday through Friday. Preventative maintenance will be scheduled monthly with the contracting point of contact. The contractor must document all work, provide certification stickers, and dispose of waste properly. The contract period is five years from August 1, 2024 to July 31, 2029. The solicitation number is 36C26224Q0558 and responses are due by March 4, 2024. The place of performance is the Southern Arizona VA Health Care System facilities.

View the file

Other files

File Type Posted
36C26224Q0584 0001 HOOD FUME MAINTENANCE.docx DOCX document
RFQ_36C26224Q0584.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SUBJECT*

HOOD FUME MAINTENANCE - Base plus 4 (8/1/24 - 7/31/29

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
85723
SOLICITATION NUMBER*
36C26224Q0558
RESPONSE DATE/TIME/ZONE
03-04-2024 4:00 PM MOUNTAIN TIME, DENVER, USA
ARCHIVE
60 DAYS AFTER THE RESPONSE DATE
SET-ASIDE
SDVOSBC
PRODUCT SERVICE CODE*
J065
NAICS CODE*
811210
PLACE OF PERFORMANCE
Southern Arizona Healthcare System

(SAVAHCS)

3601 S. 6th Avenue

Tucson AZ

POSTAL CODE
85723
COUNTRY
USA

CONTACT INFORMATION

CONTRACTING OFFICE ADDRESS

Department of Veterans Affairs NCO22- Network Contracting 3601 S. 6th Avenue Tucson, AZ 85723

POINT OF CONTACT*

Contract Specialist Marie Conforto marie.conforto@va.gov

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

Presolicitation Notice Presolicitation Notice

DESCRIPTION

*=Required Field
Presolicitation Notice

Presolicitation Notice

This is a SOURCES SOUGHT ANNOUNCEMENT ONLY. It is neither a solicitation announcement nor a request for proposals or quotes and does not obligate the Government to award a contract. Requests for a solicitation will not receive a response. Responses to this Sources Sought announcement must be in writing. The purpose of this sources sought announcement is for market research to make appropriate acquisition decisions and to gain knowledge of potential qualified Service-Disabled Veteran Owned Small Businesses, Veteran Owned Small Businesses, 8(a), HUBZone and other Small Businesses interested and capable of providing the services described below. Documentation of technical expertise must be presented in sufficient detail for the Government to determine that your company possesses the necessary functional area expertise and experience to compete for this acquisition. Responses to this notice shall include the following: (a) company name (b) address (c) point of contact (d) phone, fax, and email (e) Unique Entity ID (f) Cage Code (g) Tax ID Number (h) Type of small business, e.g. Services Disabled Veteran Owned small Business, Veteran-owned small business, 8(a), HUBZone, Women Owned Small Business, Small disadvantaged business, or Small Business HUBZone business and (i) must provide a capability statement that addresses the organizations qualifications and ability to perform as a contractor for the work described below.

The Southern Arizona VA Health Care System (SAVAHCS) located at 3601 S. 6th Avenue, Tucson, AZ 85723, is seeking a potential qualified contractor that can provide annual preventive maintenance and corrective repair service on-site for the equipment and systems listed.

Important information: The Government is not obligated to nor will it pay for or reimburse any costs associated with responding to this sources sought synopsis request. This notice shall not be construed as a commitment by the Government to issue a solicitation or ultimately award a contract, nor does it restrict the Government to a particular acquisition approach. The Government will in no way be bound to this information if any solicitation is issued. The VA is mandated by Public Law 109-461 to consider a total set-aside for Service-Disabled Veteran Owned Small Business set aside. However, if response by Service-Disabled Veteran Owned Small Business firms proves inadequate, an alternate set-aside or full and open competition may be determined. No sub-contracting opportunity is anticipated. The North American Classification System (NAICS) code for this acquisition is 811210 ($34 Million Dollars).

Notice to potential offerors: All offerors who provide goods or services to the United States Federal Government must be registered in the System for Award Management (SAM) at www.sam.gov. Additionally, all Service-Disabled Veteran Owned Businesses or Veteran Owned Businesses who respond to a solicitation on this project must be registered with the SBA Veteran Small Business Certification (VetCert) Registry located at https://veterans.certify.sba.gov. All interested Offerors should submit information by e-mail to marie.conforto@va.gov. All information submissions to be marked Attn: Marie Conforto, Contract Specialist and should be received no later than 4:00 pm PT on March 4, 2024.

STATEMENT OF WORK

DRAFT

The contractor shall provide all labor, personnel, equipment, transport of materials to certified lab, reporting of results, disposal of hazardous/nonhazardous waste materials, proof of licensing, documentation of required service and permits when required to provide testing, tools, materials, supervision, and other items necessary to provide annual preventive maintenance and corrective repair service on-site for the equipment and systems listed in Section 3 below for the Southern Arizona VA Health Care System (SAVAHCS) and as defined in this Statement of Work (SOW).

1. SERVICE/WORK TO BE PERFORMED:

· The contractor will perform on-site certification testing on equipment/systems as provided for in the contract agreement. Since there are multiple types of hoods and certification requirements, the standards for each device and room are listed in the equipment table below. Service engineer/technicians must be Controlled Environment Testing Association (CETA) registered cleanroom certification professionals.

· Travel and labor included.

· All test equipment requiring calibration is traceable through the National Bureau of Standards.

· Certification:

i. All primary engineering controls (PEC) and secondary engineering controls (SEC) and passthrough certification are completed no less than every 6 months or whenever changes to the area such as redesign, construction, replacement or relocation of any PEC, or alteration in the configuration of the room that could affect airflow or air quality. PEC, SEC and passthroughs will be retested if they fail to meet standards of USP 797 and USP 800. Certification will occur under dynamic operating conditions. Testing for certification includes:

a. Airflow testing (PEC and SEC)

b. HEPA filter integrity testing (PEC and SEC)

c. Total particle count testing (PEC, SEC and passthroughs)

d. Dynamic airflow smoke pattern testing (PEC and SEC)

e. Viable air sampling (PEC, SEC and passthroughs)

ii. Viable surface sampling (SS) of all PEC and SEC are performed no less than monthly and after every personnel medial fill (MF) testing. Compounders and personnel with direct observation of compounding will need to be retested if they fail to meet standards of USP 797. IV room, anteroom, hazardous drug (HD) room and segregated compounding area (SCA) PEC, all SEC and passthroughs will need to be retested if viable SS testing fails to meet standards of USP 797.

iii. Personnel gloved fingertip (GFT) testing shall occur no less than every 6 months for compounders and no less than annually for direct observation of compounding. Personnel GFT testing is performed after donning of garb and MF testing. Personnel GFT testing is performed three times after each donning of garb for initial personnel certification and once for recertification. Compounding and personnel with direct observation of compounding will need to be retested if they fail to meet USP 797 standards.

iv. Environmental wipe sampling for hazardous drug residue is performed no less than every 6 months or more frequently as needed to verify containment per the standards of USP 800. The kits used to complete wipe sampling must be verified prior to use to ensure the sensitivity and validity of the methods and reagents used to recover a specific percentage of known marker drugs. A minimum of 6 surface samples will be completed per certification. Facilities can elect to expand beyond the minimum number of samples, but coordination with the contractor will be required prior to the certification visit. Surface sampling should include the following areas:

1. Interior of containment primary engineering control (C-PEC) and any equipment contained inside the C-PEC

2. Pass-through chambers

3. Surfaces in staging or work areas near C-PEC

4. Areas adjacent to C-PECs (e.g., floors directly under C-PEC, staging, and dispensing area)

5. Areas immediately outside the hazardous drug buffer room or C-SCA

6. Patient administration areas

ii. No less than 5 commonly used chemotherapy drugs will be tested at your facility at the time of testing.

v. EMERGENCY LINE ITEM:

a. Service calls with on-site response within 24 hours (Monday to Friday 8:00 AM to 5:00 PM). Emergency Designated Response time 24 hours unless otherwise agreed upon by COR or designee: Completed as needed and price added separately.

b. Response: All requests for repair service will be considered requests for emergency service unless a statement to the contrary is made by the COR, POC or designee. A competent service representative adequately equipped with tools, test instruments, service manuals, and repair parts will arrive at the SAVAHCS within the designated response time. The service representative may contact VA staff by phone before arrival at the SAVAHCS. Such contacts, however, will not serve to alter the requirements for arrival on-site except when the request for service is withdrawn by designee.

vi. EQUIPMENT:

vii. PERFORMANCE:

a. Hours: Normal duty hours are 8:00 a.m. through 4:30 p.m., Monday through Friday, not including federal holidays. Whenever contact with Operations Supervisor is required outside of normal duty hours, the Utility System Repair Operator (USRO) may be contacted as an alternate. The hospital operator at extension "0" can contact the USRO via radio.

b. Preventive maintenance will be scheduled with a contact person as assigned by the Contracting Officer’s Representative (COR), Point of Contact (POC) or designee. This PM work must be completed in the month specified during the request unless another date is agreed upon by COR or designee. The Operations supervisor can be contacted at (520) 792-1450 ext. 15020. All contractor personnel (without exception) must report to the Maintenance & Operations Department, Bldg. 2-room B17, to sign in and receive a contractor identification badge before any work is performed and return to the same location to sign out and turn-in said badge along with the documented service report (as described in section 4, DOCUMENTATION) for the service performed.

c. Noncontract Charges: The contractor shall not perform any service that will result in additional charges without prior approval from the Contracting Officer.

d. Payment: Payment will be made in Tungsten, and upon receipt of a properly prepared invoice which references the period for which payment is due, the contract number and/or assigned purchase order number. The invoice must include the correct payment address.

e. Waste Removal: The contractor is responsible to remove and properly dispose of all waste related to the services provide on-site at SAVAHCS.

f. Failure to carry out the above procedures will be considered as a failure to document the service visit and will constitute a failure to perform.

g. All shipping for parts will be included in the cost of the contract.

viii. DOCUMENTATION:

a. Service reports will be provided for all service activities including preventive maintenance, emergency repair, and safety inspections, and must contain the following information:

a) Date(s) and time of service.

b) Complete description of equipment serviced including model number and serial number.

c) Complete description of equipment services performed including upgrades, updates and software changes made.

d) Complete list of all materials, parts, or assemblies installed or replaced including serial numbers of parts where applicable.

e) A statement of the complaint that initiated the service call.

f) Total standard labor hours, total overtime/premium hours, and total travel hours expended during the repair.

g) Last calibration date of service equipment.

· For each PEC passing the required certification tests, the contractor will supply each unit with a certification sticker with the following information:

· Company name and contact information.

· Unit make, model and serial number.

· Report number

· Location (e.g., building, room number)

· Certification date

· Recertification date

· Technician signature

ix. REMOVAL OF EQUIPMENT:

a. Whenever possible, all repairs and preventive maintenance will be performed at the SAVAHCS. When that is not possible, the COR, POC or designee may give approval for service work to be done at another location. All transportation charges will be borne by the contractor.

b. VA property may not be removed from the SAVAHCS prior to execution, and approval of all documents as required by Logistics Service. Computer hard drives and other storage media items may not leave VA property.

c. Requirements for service reports for designated equipment not at this location are the same as for work done on-site. Service reports for incoming equipment will be presented for signature at the time of delivery. If equipment is returned via commercial carrier, the service report may be included with the shipment.

x. GOVERNMENT REPAIRS: Government employees will work on contract equipment only with authorization of the contractor or following the failure of the contractor to respond. The government is not obligated to perform any work on contract equipment.

xi. PRIOR INSPECTION OF EQUIPMENT: Potential contractors may, at their own expense, inspect equipment covered by this solicitation (ref 52.237-1 SITE VISIT). Arrangements must be made in advance with Rodolfo Guzman and Marie Conforto.

xii. GUARANTEE: The contractor will guarantee workmanship and all parts furnished by the contractor for a period of not less than 90 days.

image1.emf

Inventory.xlsx Research

INV# LOCATION MAKE MODEL SN Certification Standards EQU. TYPE Area Column1 Column2

36272 181L-57 PCI G32 10915 Smoke Containment/Face Velocity Fume Hood SPS IN USE

55283 181L-57 PCI G14KA 29590 Smoke Containment/Face Velocity Fume Hood SPS IN USE

76270 107-10 NUAIRE NU-437-400 163682061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76272 107-10 NUAIRE NU-437-400 163684061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53702 110-10 BAKER SG403A 94853 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76256 110-10 NUAIRE NU-437-500 163685061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76269 110-10 NUAIRE NU-430-600 163698061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76261 112-78 NUAIRE NU-437-600 163690061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76263 113-10 NUAIRE NU-435-400 163692061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76259 113-78 NUAIRE NU-437-500 163688061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76260 114A-78 NUAIRE NU-437-500 163689061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76262 115-78 NUAIRE NU-437-600 163691061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53149 116-78 Fisher Scientific 115V 80383682 Smoke Containment/Face Velocity Fume Hood Research IN USE ON COUNTER TOP

76264 116-78 NUAIRE NU-435-400 163693061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

56724 138A-78 Thermo Electron C8004055 68013 AIHA/ANSI Lab Ventilation Requirements Downdraft Table Research IN USE GROSSING TABLE

30029 111-78 LABCONCO 6900000 255087 Ductless-Smoke Containment/Face Velocity Fume Hood Research IN USE ON COUNTER TOP

68100 0BA085-57 IV Room Per CETA CAG 003 per USP 797; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm IN USE

68100 0BA086-57 Ante Room Per CETA CAG 003 per USP 797; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm ROOM

68100 0BA088-57 HD Room Per CETA CAG 003 per USP 797 and USP 800; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm ROOM

92627 BA080-57 BAKER EDGEGARD VF401 128826 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm ADD

48628 BA085-57 BAKER CS-600 87631 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Negative Isolator Pharm TURNED IN 11/2/18

48629 BA088-57 BAKER SS-500 87450 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm TURNED IN 9/19/19

48630 BA088-57 BAKER SS-800 88782 Left Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

48631 BA088-57 BAKER SS-800 88782 Right Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

92626 BA088-57 BAKER EDGEGARD VF401 128825 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm ADD

76657 227-57 Mopec BF100 BF100H14005 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

75491 2756-80 AIR SCIENCE P-30 P70084 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

66199 A109-38 COMECER FHR1 50 LAR 40C1C\0111\V11246 Per Manufacturer's Specifications;Per IES-RP-CC-002 Laminar Air Flow Workstation Other IN USE

76658 0B146-57 MOPEC MB600 BF100H14006 Per Manufacturer's Specifications;Directional Smoke/Face Velocity Fume Hood Lab IN USE

86323 0B147-57 NUAIRE NU-435-400 176331092816 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

54704 0BA018-57 Air Clean Sys AC632DB AC032DB3501 Per Manufacturer's Specifications;Per IES-RP-CC-001 Class I Biological Safety Cabinet Lab ADD

66616 B128-57 Air Science USA Purair P5-36 P67929 Ductless-Smoke Containment/Face Velocity Fume Hood Lab TURNED IN 2/21/18

34107 B138-57 NUAIRE Series 26 77388 AFR Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

34546 B138-57 NUAIRE Series 26 82675 AGR Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

57903 B138A-57 BSL III Hood Room Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab ROOM

55366 B139-57 Ante Room Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE

B140-57 BSL III Lab Room Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE

62974 B146-57 Air Science P5-36 P66066 Ductless-Smoke Containment/Face Velocity Fume Hood Lab IN USE

53189 B146-57 MOPEC MB600 MB600D08004 Smoke Containment/Face Velocity Fume Hood Lab IN USE

62136 B146-57 MOPEC MB600 MB600A11004 Smoke Containment/Face Velocity Fume Hood Lab IN USE

66932 B147-57 TBS SC-DD 21202SCDD0009-02 Smoke Containment/Face Velocity Fume Hood Lab IN USE

57903 BA015-57 Extraction Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE

73235 BA015-57 Labconco 3440001 130171290 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Lab IN USE

39668 BA018-57 Air Clean Sys AC648LFUVC 2356 Per Manufacturer's Specifications;Per IES-RP-CC-001 Class I Biological Safety Cabinet Lab IN USE

615990B133-57NUAIRE425-600140865102610Per Manufacturer's Specifications
Per NSF Standard No. 49Class II A Direct Inflow MethodIN USE

86409 138-78 SAFETY LAB 112537002 171150096 Walk-In Walk-in Research IN USE

81099 107-10 LABCONCO 3955200 160728749B Smoke Containment/Face Velocity Fume Hood Research IN USE ON COUNTER TOP

76271 111-78 NUAIRE NU-3437-400 163683061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II, Type A2 Biological Safety Cabinet Research IN USE

Pharmacy

INV# LOCATION MAKE MODEL SN Certification Standards EQU. TYPE Area Column1 Column2

36272 181L-57 PCI G32 10915 Smoke Containment/Face Velocity Fume Hood SPS IN USE

55283 181L-57 PCI G14KA 29590 Smoke Containment/Face Velocity Fume Hood SPS IN USE

76257 Warehouse (15) NUAIRE NU-437-500 163686061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research STORAGE

76270 107-10 NUAIRE NU-437-400 163682061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76272 107-10 NUAIRE NU-437-400 163684061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53702 110-10 BAKER SG403A 94853 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76256 110-10 NUAIRE NU-437-500 163685061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76269 110-10 NUAIRE NU-430-600 163698061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76268 111-78 NUAIRE NU-435-600 163697061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76261 112-78 NUAIRE NU-437-600 163690061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76263 113-10 NUAIRE NU-435-400 163692061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76259 113-78 NUAIRE NU-437-500 163688061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76260 114A-78 NUAIRE NU-437-500 163689061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76262 115-78 NUAIRE NU-437-600 163691061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53149 116-78 Fisher Scientific 115V 80383682 Smoke Containment/Face Velocity Fume Hood Research IN USE

76264 116-78 NUAIRE NU-435-400 163693061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76266 135-56 NUAIRE NU-430-400 163695061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research STORAGE 8/13/2019/ TAKE OFF SOW

76271 135-56 NUAIRE NU-437-400 163683061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research STORAGE TAKE OFF SOW

56724 138A-78 Thermo Electron C8004055 68013 AIHA/ANSI Lab Ventilation Requirements Downdraft Table Research IN USE

30029 111-78 LABCONCO 6900000 255087 Ductless-Smoke Containment/Face Velocity Fume Hood Research IN USE ON COUNTER TOP/ Add to SOW. Needs cert

76258 221-6 NUAIRE NU-437-500 163687061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research LOANED OUT REMOVE FROM SOW. AT UA

92627 BA080-57 Baker Edgegard VF401 128826 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

93003 BA080-57 Baker Edgegard VF601 130072 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

88030 BA080-57 Baker SG-304 126553 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Negative Isolator Pharm IN USE

98084 BA080-57 Baker CS-500 125183 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Negative Isolator Pharm IN USE

48630 BA093-57 Baker SS-800 88783 Left Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

48631 BA093-57 Baker SS-800 88782 Right Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

92626 BA080-57 Baker Edgegard VF401 128825 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

76657 227-57 Mopec BF100 BF100H14005 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

75491 2756-80 AIR SCIENCE P-30 P70084 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

66199 A109-38 COMECER FHR1 50 LAR 40C1C\0111\V11246 Per Manufacturer's Specifications;Per IES-RP-CC-002 Laminar Air Flow Workstation Other IN USE

76658 0B146-57 MOPEC MB600 BF100H14006 Per Manufacturer's Specifications;Directional Smoke/Face Velocity Fume Hood Lab IN USE

86323 0B147-57 NUAIRE NU-435-400 176331092816 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

54704 0BA018-57 Air Clean Sys AC632DB AC032DB3501 Per Manufacturer's Specifications;Per IES-RP-CC-001 Class I Biological Safety Cabinet Lab ADD

66616 B128-57 Air Science USA Purair P5-36 P67929 Ductless-Smoke Containment/Face Velocity Fume Hood Lab TURNED IN 2/21/18

34107 B138-57 NUAIRE Series 26 77388 AFR Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

34546 B138-57 NUAIRE Series 26 82675 AGR Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

57903 B138A-57 BSL III Hood Room Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab ROOM

55366 B139-57 Ante Room Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE

B140-57 BSL III Lab Room Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE

62974 B146-57 Air Science P5-36 P66066 Ductless-Smoke Containment/Face Velocity Fume Hood Lab IN USE

53189 B146-57 MOPEC MB600 MB600D08004 Smoke Containment/Face Velocity Fume Hood Lab IN USE

62136 B146-57 MOPEC MB600 MB600A11004 Smoke Containment/Face Velocity Fume Hood Lab IN USE

66932 B147-57 TBS SC-DD 21202SCDD0009-02 Smoke Containment/Face Velocity Fume Hood Lab IN USE

57903 BA015-57 Extraction Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE

73235 BA015-57 Labconco 3440001 130171290 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Lab IN USE

39668 BA018-57 Air Clean Sys AC648LFUVC 2356 Per Manufacturer's Specifications;Per IES-RP-CC-001 Class I Biological Safety Cabinet Lab IN USE

615990B133-57NUAIRE425-600140865102610Per Manufacturer's Specifications
Per NSF Standard No. 49Class II A Direct Inflow MethodIN USE

86409 138-78 SAFETY LAB 112537002 171150096 Walk-In Walk-in Research IN USE NEEDS CERT

DO not Have yet BA086 BAKER BCG 401 Do Not Have yet Per USP 797/800 BSC Biochemgard Pharmacy Yet to be installed Will be located in the HD room Inpatient pharmacy

DO not Have yet BA084 BAKER BCG 401 Do Not Have yet Per USP 797/800 BCG Biochemgard Pharmacy Yet to be installed Will be located in the HD room Inpatient pharmacy

Lab

INV# LOCATION MAKE MODEL SN Certification Standards EQU. TYPE Area Column1 Column2

76257 Warehouse (15) NUAIRE NU-437-500 163686061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research STORAGE

76270 107-10 NUAIRE NU-437-400 163682061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76272 107-10 NUAIRE NU-437-400 163684061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53702 110-10 BAKER SG403A 94853 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76256 110-10 NUAIRE NU-437-500 163685061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76269 110-10 NUAIRE NU-430-600 163698061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76268 111-78 NUAIRE NU-435-600 163697061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76261 112-78 NUAIRE NU-437-600 163690061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76263 113-10 NUAIRE NU-435-400 163692061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76259 113-78 NUAIRE NU-437-500 163688061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76260 114A-78 NUAIRE NU-437-500 163689061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76262 115-78 NUAIRE NU-437-600 163691061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53149 116-78 Fisher Scientific 115V 80383682 Smoke Containment/Face Velocity Fume Hood Research IN USE

76264 116-78 NUAIRE NU-435-400 163693061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76266 135-56 NUAIRE NU-430-400 163695061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research STORAGE 8/13/2019/ TAKE OFF SOW

76271 135-56 NUAIRE NU-437-400 163683061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research STORAGE TAKE OFF SOW

56724 138A-78 Thermo Electron C8004055 68013 AIHA/ANSI Lab Ventilation Requirements Downdraft Table Research IN USE

30029 111-78 LABCONCO 6900000 255087 Ductless-Smoke Containment/Face Velocity Fume Hood Research IN USE ON COUNTER TOP/ Add to SOW. Needs cert

76258 221-6 NUAIRE NU-437-500 163687061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research LOANED OUT REMOVE FROM SOW. AT UA

68100 0BA085-57 IV Room Per CETA CAG 003 per USP 797; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm IN USE

68100 0BA086-57 Ante Room Per CETA CAG 003 per USP 797; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm ROOM

68100 0BA088-57 HD Room Per CETA CAG 003 per USP 797 and USP 800; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm ROOM

92627 BA080-57 BAKER EDGEGARD VF401 128826 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm ADD

48628 BA085-57 BAKER CS-600 87631 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Negative Isolator Pharm TURNED IN 11/2/18

48629 BA088-57 BAKER SS-500 87450 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm TURNED IN 9/19/19

48630 BA088-57 BAKER SS-800 88782 Left Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

48631 BA088-57 BAKER SS-800 88782 Right Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

92626 BA088-57 BAKER EDGEGARD VF401 128825 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm ADD

76657 227-57 Mopec BF100 BF100H14005 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

75491 2756-80 AIR SCIENCE P-30 P70084 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

66199 A109-38 COMECER FHR1 50 LAR 40C1C\0111\V11246 Per Manufacturer's Specifications;Per IES-RP-CC-002 Laminar Air Flow Workstation Other IN USE

76658 0B146-57 MOPEC BF100 BF100H14006 Per Manufacturer's Specifications;Directional Smoke/Face Velocity Fume Hood Lab IN USE Next to Electrical Panel in 146

86323 0B147-57 NUAIRE NU-435-400 176331092816 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

34107 0B138-57 NUAIRE NU-427-400 77388AFR Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

61599 0B133-57 NUAIRE NU-425-600 140865102610 BSL III Lab IN USE

34546 0B138-57 NUAIRE NU-427-400 82675AGR Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Lab IN USE

96063 B146-57 MOPEC MB1060R MB1060R0135661001 Smoke Containment/Face Velocity Fume Hood Lab IN USE

96064 B146-57 MOPEC MB1060R MB1060R0135661002 Smoke Containment/Face Velocity Fume Hood Lab IN USE

62974 0B146-57 Air Science P5-36 P66066 Ductless-Smoke Containment/Face Velocity Fume Hood Lab IN USE

73235 0BA028A-38 Labconco 3440001 130171290C Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Lab IN USE

102996 0B133-57 Thermofisher Scientific 1300 300429464 Per Manufacturer's Specifications;Per IES-RP-CC-001 Class I Biological Safety Cabinet Lab IN USE

103551 0BA045-57 Thermofisher Scientific 1100-2-C 210514 Fume Hood Lab IN USE On Counter/ Next to Electrical Panel

39668 0BA028B-38 Air Clean Sys AC648LFUVC AC648LFUVC2356 Per Manufacturer's Specifications;Per IES-RP-CC-001 Class I Biological Safety Cabinet Lab IN USE

75491 2756-80 Air Science P-30 P70084 Smoke Containment/Face Velocity Fume Hood Derm IN USE

615990B133-57NUAIRE425-600140865102610Per Manufacturer's Specifications
Per NSF Standard No. 49Class II A Direct Inflow MethodIN USE

86409 138-78 SAFETY LAB 112537002 171150096 Walk-In Walk-in Research IN USE NEEDS CERT

81099 107-10 LABCONCO 3955200 160728749B Research IN USE ADD TO SOW

Derm, Rad, SPS & OR

INV# LOCATION MAKE MODEL SN Certification Standards EQU. TYPE Area Column1 Column2

76257 Warehouse (15) NUAIRE NU-437-500 163686061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research STORAGE

76270 107-10 NUAIRE NU-437-400 163682061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76272 107-10 NUAIRE NU-437-400 163684061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53702 110-10 BAKER SG403A 94853 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76256 110-10 NUAIRE NU-437-500 163685061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76269 110-10 NUAIRE NU-430-600 163698061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76268 111-78 NUAIRE NU-435-600 163697061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76261 112-78 NUAIRE NU-437-600 163690061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76263 113-10 NUAIRE NU-435-400 163692061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76259 113-78 NUAIRE NU-437-500 163688061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76260 114A-78 NUAIRE NU-437-500 163689061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

76262 115-78 NUAIRE NU-437-600 163691061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research IN USE

53149 116-78 Fisher Scientific 115V 80383682 Smoke Containment/Face Velocity Fume Hood Research IN USE

76264 116-78 NUAIRE NU-435-400 163693061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research IN USE

76266 135-56 NUAIRE NU-430-400 163695061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II B Direct Inflow Method Research STORAGE 8/13/2019/ TAKE OFF SOW

76271 135-56 NUAIRE NU-437-400 163683061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research STORAGE TAKE OFF SOW

56724 138A-78 Thermo Electron C8004055 68013 AIHA/ANSI Lab Ventilation Requirements Downdraft Table Research IN USE

30029 111-78 LABCONCO 6900000 255087 Ductless-Smoke Containment/Face Velocity Fume Hood Research IN USE ON COUNTER TOP/ Add to SOW. Needs cert

76258 221-6 NUAIRE NU-437-500 163687061014 Per Manufacturer's Specifications;Per NSF Standard No. 49 Class II A Direct Inflow Method Research LOANED OUT REMOVE FROM SOW. AT UA

68100 0BA085-57 IV Room Per CETA CAG 003 per USP 797; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm IN USE

68100 0BA086-57 Ante Room Per CETA CAG 003 per USP 797; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm ROOM

68100 0BA088-57 HD Room Per CETA CAG 003 per USP 797 and USP 800; ISO 14644-1 for ISO Class 7; IEST RP-CC-034; Testing includs particle counts, air volume, air exchange rates, room pressurization, HEPA filter leak testing and viable environmental sampling. Clean Room Pharm ROOM

92627 BA080-57 BAKER EDGEGARD VF401 128826 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm ADD

48628 BA085-57 BAKER CS-600 87631 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Negative Isolator Pharm TURNED IN 11/2/18

48629 BA088-57 BAKER SS-500 87450 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm TURNED IN 9/19/19

48630 BA088-57 BAKER SS-800 88782 Left Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

48631 BA088-57 BAKER SS-800 88782 Right Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm IN USE

92626 BA088-57 BAKER EDGEGARD VF401 128825 Per Manufacturer's Specifications; CETA Applications Guide 002-2006; Per ISO 14644-1 Positive Isolator Pharm ADD

76657 227-57 Mopec BF100 BF100H14005 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

75491 2756-80 AIR SCIENCE P-30 P70084 Ductless-Smoke Containment/Face Velocity Fume Hood Other IN USE

66199 A109-38 COMECER FHR1 50 LAR 40C1C\0111\V11246 Per Manufacturer's Specifications;Per IES-RP-CC-002 Laminar Air Flow Workstation Other IN USE

104536 181U-57 PCI G18-E 121-0019 Smoke Containment/Face Velocity Fume Hood SPS IN USE

55283 181U-57 PCI G14KA 29590 Smoke Containment/Face Velocity Fume Hood SPS IN USE

75491 2756-80 Air Science P30 P70084 Chemical Fume Fume Hood Derm IN USE

76657 227-57 Mopec BF100 BF100H14005 Per Manufacturer's Specifications;Directional Smoke/Face Velocity Fume Hood OR IN USE **Located in Inner OR Core**

66199 A109-38 Comecer FHR1 50 LAR 40C1C\0111\V11246 Chemical Fume Fume Hood Radiology IN USE

75491 2756-80 Air Science P-30 P70084 Smoke Containment/Face Velocity Fume Hood Derm IN USE

615990B133-57NUAIRE425-600140865102610Per Manufacturer's Specifications
Per NSF Standard No. 49Class II A Direct Inflow MethodIN USE

86409 138-78 SAFETY LAB 112537002 171150096 Walk-In Walk-in Research IN USE NEEDS CERT

81099 107-10 LABCONCO 3955200 160728749B Research IN USE ADD TO SOW

Filters

57903 0B138A-57 BSL III Hood Room Filters HEPA Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab ROOM **This is not equipment. EE number is used to track semi-annual certification.

55366 0B139-57 Ante Room Filters HEPA Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE **This is not equipment. EE number is used to track semi-annual certification.

0B140-57 BSL III Lab Room Filters HEPA Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE **This is not equipment. EE number is used to track semi-annual certification.

57903 0BA015-57 Extraction Filters HEPA Testing includs air volume, air exchange rates, room pressurization, HEPA filter leak testing BSL III Lab IN USE **This is not equipment. EE number is used to track semi-annual certification.