This file is a Statement of Work (SOW) for a Sources Sought market research notice issued by the Department of Veterans Affairs (VA) Sierra Nevada Health Care System (VASNHCS). The VA is seeking information from vendors regarding the availability of Automated Blood Culture Detection Analyzers (brand name or equal) to support laboratory operations at the facility. This is a pre-solicitation notice only and does not constitute an actual solicitation or commitment to award a contract. The NAICS code is 325413 with a small business size standard of 1,250 employees. Interested vendors must submit capability statements and company information by email to david.alvarez4@va.gov no later than 10:00 AM Pacific Standard Time on December 2, 2026, with no telephone requests accepted and no questions to be answered during this phase.

Vendors must provide their company name, point of contact information (name, phone, email), SAM UEI number, socioeconomic status (such as SDVOSB, VOSB, or Women-Owned Small Business status, which will be verified through VetBiz), capability statements, an Authorized Distributor Letter if applicable, and an estimated price. The Government will not compensate vendors for submitting information in response to this sources sought notice, retains the right to cancel the requirement at its discretion, and will handle all information marked proprietary in accordance with applicable regulations. Interested vendors should review the attached Statement of Work and Salient Characteristics documents for additional technical and operational requirements.

View the file

Other files

File Type Posted
Salient Characteristics Blood Culture Analyzer.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK (SOW)

Automated Blood Culture (Bottles) Analyzer — Incubation & Detection System VA Sierra Nevada Health Care System (VASNHCS) Pathology & Laboratory Medicine Service (PALMS), Reno, NV

1. Scope Contractor shall supply an FDA‑cleared automated blood culture incubation and detection system that provides continuous monitoring of blood culture bottles. The system must include automatic detection, temperature control, continuous loading, automated bottle loading and re‑loading, middleware, QC materials, blood‑volume auditing, training, validation, and service support, including automatic LIS/middleware transmission of results to VistA/DI.

2. Capability Requirements

· Continuous, automated microbial growth detection using validated optical/fluorometric/colorimetric sensors.

· Compatible with adult aerobic, anaerobic, pediatric, and specialty/resin bottles.

· Automated bottle loading mechanism capable of accepting bottles without requiring manual placement into individual cells.

· Automated bottle re‑loading capability, allowing:

· Automated repositioning or re-entry of bottles after a temporary removal,

· Re-loading into the system for continued incubation,

· Automatic reassignment of bottle location with barcode confirmation,

· No interruption of monitoring.

· Real-time barcode tracking for all bottles, operators, and lot numbers.

· Automated temperature and mixing/agitation.

· Critical alarm prioritization and immediate notifications.

· Built‑in blood volume audit capability.

· Middleware supporting automatic result transmission for all bottle events.

3. Quality & Traceability

· Automated QC processes for temperature, alarms, sensors, and barcode integrity.

· Automated traceability including load time, unload time, automatic reload, alarm timestamps, and operator login.

· Blood volume audit logs stored with full audit trails.

4. Software & Data Management

· Intuitive dashboard showing bottle status, TTD metrics, alarm queues, and location maps.

· Role-based security, LDAP/AD, 21 CFR Part 11–ready audit trails.

· Automatic LIS/middleware transmission, including:

· Positive bottle alarms

· Negative results at completion

· QC transmissions

· Blood volume measurements

· Automatic re-load status

· Operator and lot/expiry metadata

· Middleware with retry logic, full auditing, and VA‑compliant network security.

5. Productivity & Workflow Efficiency

· Automated bottle loading supports bulk loading for high-volume shifts.

· Automated re-loading capability allowing bottles to be removed (for Gram stain, subculture) and inserted without manual reassignment.

· Continuous loading and unloading without interrupting active monitoring.

· Real-time volume display for phlebotomy feedback.

· Visual lane mapping for rapid retrieval of positive bottles.

6. System Specifications

· Modular or benchtop system sized to PALMS floor plan.

· Capacity of XX–XXX bottles (finalized at needs assessment).

· Automated temperature control with real-time monitoring.

· Bottles loaded through an automated loading hatch or tray, eliminating the need to open incubator chambers.

· Automatic re-loading mechanism:

· Bottle barcode scanned on re-entry,

· Instrument assigns an open location without technologist input,

· Monitoring resumes without workflow interruption.

· Integrated barcode scanners and redundant detection sensors.

· Integrated middleware engine supporting automatic HL7/ASTM messaging.

7. Installation, Validation & Training Validation must include:

· Temperature stability

· Bottle compatibility

· Alarm verification

· Automated loading/re-loading function testing

· LIS/middleware automatic result transmission

· Blood volume measurement checks

· Barcode traceability workflow Training includes super-user training and workflow optimization for automated loading/reloading procedures.

8. Maintenance & Technical Support

· 24/7 secure remote support.

· ≤ 24‑hour on‑site response.

· Annual PM.

· Monitoring of automated loading system, sensors, alignment, and software.

9. Performance Acceptance Criteria

· Automated bottle loading and re-loading functions perform successfully with ≥ 99% bottle acceptance and no location assignment errors.

· Bottle positions automatically tracked with no mismatches.

· 100% automatic transmission of positive, negative, QC, and blood‑volume audit results.

· Temperature stability and growth-detection performance meet IFU.

· ≥ 95% uptime.

10. Deliverables

· Installed analyzer with automated loading/re-loading functionality enabled.

· Full middleware documentation covering automatic result transmission.

· Validation summary including automated loading tests.

· QC plan, competency documents, and training records.

· Service logs and maintenance reporting.