This is a Questions and Answers document for Solicitation 36C26126Q0595, an optometry eye lane maintenance contract with the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 21. The document addresses a single inquiry regarding preventive maintenance requirements for the Reichert 11636 Phoroptor, a diagnostic instrument included in the optometry eye lane inventory. The contractor asked whether calibration verification confirming the unit operates within manufacturer specifications would be acceptable as an alternative to factory calibration, noting that Reichert does not publish field calibration procedures and factory calibration requires specialized manufacturer equipment. The Government confirmed acceptance of calibration verification confirming the unit operates within manufacturer specifications as part of the required preventive maintenance, eliminating the necessity for factory calibration under this contract.

View the file

Other files

File Type Posted
36C26126Q0595 0001.pdf PDF
36C26126Q0595_4.docx DOCX document
P09-SOW OPTOMETRY EYE LANE MAINTENANCE.pdf PDF
WD 2015-5593.pdf PDF
RFQ 36C26126Q0595_2.pdf PDF
36C26126Q0595_3.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Questions and Answers Solicitation 36C26126Q0595 Optometry Eye Lane Maintenance

Question: Regarding the preventive maintenance of the Reichert 11636 Phoroptor included in the optometry eye lane inventory. Because Reichert does not publish field calibration procedures and factory calibration requires specialized manufacturer equipment, would the Government consider calibration verification confirming the unit is operating within manufacturer specifications acceptable as part of the required preventive maintenance, or is factory calibration required under this contract?”

Answer: Yes, we will consider calibration verification confirming the unit is operating within manufacturer specifications acceptable as part of the required preventive maintenance.