1. 36C26124Q0782 Questions and Answers.xlsx
XLSX spreadsheet 56 KB Posted
- Attached to
- 6640--VA Northern California Health Care System Chemistry & Immunoassay CPRR Equipment and Tests
- Solicitation number
- 36C26124Q0782
About this file
This document contains questions and answers related to a federal solicitation for a chemistry and immunoassay equipment and test contract (36C26124Q0782) for the VA Northern California Health Care System (VANCHCS).
The key details include:
- The solicitation requires pricing for a variety of chemistry, immunoassay, and other clinical laboratory tests, some with zero estimated demand initially.
- The solicitation has been extended to August 26, 2024 at 12PM PDT.
- Drawings of the laboratory spaces at the Mather and Martinez locations have been provided.
- The contract will require the contractor to move equipment to a temporary location during a planned renovation at the Mather laboratory, with revalidation required.
- The government is open to using control vendors other than the current Bio-Rad, and requires the ability to export and upload QC data in real-time.
- Detailed test volumes have been provided for the Mather, Martinez, and Stockton laboratory locations.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| S02 36C26124Q0782 RFQ Price List.xlsx | XLSX spreadsheet | |
| 36C26124Q0782 0004.docx | DOCX document | |
| 36C26124Q0782 0003.docx | DOCX document | |
| 4. Mather Laboratory Sink Removed.docx | DOCX document | |
| 3. Stockton Clinical Laboratory_floor plan.pdf | ||
| 2. Mather Updated Temp Lab Layout.pdf | ||
| 36C26124Q0782 0002.docx | DOCX document | |
| 36C26124Q0782 0001.docx | DOCX document | |
| S02 36C26124Q0782.docx | DOCX document | |
| Redding Floor Plan.pdf | ||
| S02 36C26124Q0418 Price List.xlsx | XLSX spreadsheet | |
| Mather Floor Plan.pdf | ||
| Martinez Floor Plan.docx | DOCX document |
Show all 13
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
General Questions
| VA Northern California Health Care System (VANCHCS) | ||
| Questions | Answers | |
| Q. Document "S02 36C26124Q0418 Price List". We noticed a number of tests are listed with a qty of 0. Can you confirm if you want us to include these test? | ||
| Examples are: BNP and Cystatin C | A. All volumes estimated with a value of one (1) and/or zero (0) require only a best price/base price please. There is no known demand at this time; however, we need a price for future cases should they appear. |
See applicable tab for a breakdown of test volumes
| Request: Due to the ADLM meeting taking place the week of 7/29/24 and various other task related to digesting the information gathered during the 7/22/24 site visits as well as the scope of the clarifications contained within, we request a minimum extension on the due date of at least 3 weeks. | A. Solicitation has been extended to 26 August 2024 at 12PM PDT. |
| Request: We request a CAD for each location and the proposed temporary space for the Mather renovation to best understand the dimensions of each laboratory. | A. Stockton CAD drawing and temporary Mather Floor Plans are attached. All other plans were attached to the original solicitation - 36C26124Q0782. |
| Request: Please indicate how many analyzers are being requested for each location. If backups are needed, must they be the same model? | A. See Statement of Work for analyzer types. |
A. Yes, backup must be same as the primary analyzer.
| Requirement: Page 53, Instructions to Offerors, point 1. Quote submittal requirements, section a. note on solicitation terms and conditions states that Offerors need to provide a price for ALL assays to include open channel assays. | |
| This can be source limiting, please consider a preferred statement inclusion or provide evaluation criteria against this. | A. There are no changes to the instructions at this time. |
| Q. SO2+36C26124Q0418: Tab 1. |
Request: We are requesting more extensive assay definitions for the following assays:
ALBM, ALBX, ALPD, CR-S, CR-T, CREM, CRTS, TP3, PTM, CKNA, LD-P, PHOSM, PHS, TG-B, AMY7, CMBX, HDLC, HDLD, HDLX, M-TP, PAM, T3U, UIBX, UIBXR, CAR, GEN, PHE, PHY, TOB, METD, METQ, PROX, UDR, CHE, CHEX, LIPM, MA, MAB, PAB, CR-S, CR-T, HEMOL, CSA, EVR, MPA.
| Request: for assays listed by group, we request that the test volume for each assay be provided so that we can provide accurate pricing. | A. See "Test Key" tab for a list of Test Names and Acronyms. A. See applicable tab for a breakdown of test volumes |
| Q. Data Management |
Requirement: Page 19, section 2.7, point 4 requires the capability to perform real-time moving averages, etc.
Clarification requested: Based on the current deployment of instrumentation, is it a fair assumption that all rules are to be written in vendor-supplied middleware?
Clarification requested: If there is a split between the VA Data Innovations and the vendor Data Innovations, please specify which party provides which rules.
Clarification requested: Is autoverification to be performed using your instance of Data Innovations Instrument Manager, or are you expecting the vendor to perform this on their middleware?
A. Do not assume rules are to be written in vendor-supplied middleware. We are not part of the current deployment.
A. At this time we do not know until award of contract.
A. 2.7.6: Can not be determined until award of contract. Our current plan is to use data innovations.
Q. Controls:
Requirement: Page 11 of 36C26124Q0782 states ‘the per reportable result price will include cost covering (a) equipment use, (b) all reagents, standards, controls, supplies, consumable/disposable items, etc.
Clarification requested: Please identify which control vendor you currently use.
Clarification requested: Are you open to using other control vendors?
Clarification requested: Are you using QC peer review software, and are we to include the cost of the software license for this?
Clarification requested: Please also share the quantity and list numbers for your current controls that you acquire today. A. Control vender is currently Bio-RAD
A. We are open to using other control vendors. Vendor QC must meet requirements in PWS.
A. We are using QC peer review software, needed licencing should be provided by contractor. QC must have ability to export and upload QC data in real time to peer group program per SOW..
A. See applicable tab for Bio-Rad/QC information.
Test Key
| Acronym | Test Name |
| ALBM | Ablumin |
| ALBX | Ablumin |
| ALPD | Alkaline Phosphatase |
| AMY7 | Amylase |
| CAR | Carbamazepine |
| CHE | Cholinesterase |
| CHEX | Cholinesterase UDR Assay (CHEX) |
| CKNA | Remove - No longer Needed |
| CMBX | Creatine Kinase, Isoenzyme MB |
| CREM | Creatinine |
| CR-S | Creatinine |
| CR-T | Creatinine |
| CRTS | Creatinine |
| CSA | Cyclosporine |
| EVR | Iron |
| GEN | Gentamicin |
| HDLC | High density lipoprotein- Cholesterol |
| HDLD | High density lipoprotein- Cholesterol |
| HDLX | High density lipoprotein- Cholesterol |
| HEMOL | Remove - No longer Needed |
| LD-P | Lactate Dehydrogenase |
| LIPM | Lipase |
| MA | Microalbumin/ Urine albumin |
| MAB | Microalbumin/ Urine albumin |
| METD | Methadone |
| METQ | Methaqualone |
| MPA | Mycophenolic Acid (MPA) Assay |
| M-TP | Miroprorotein |
| PAB | Prealbumin |
| PAM | P-Amylase Reagent (PAM) |
| PHE | Phenobarbital |
| PHOSM | Phosphorus |
| PHS | Phosphorus |
| PHY | Phenytoin |
| PROX | Propoxyphene |
| PTM | Remove - No longer Needed |
| T3U | T3 uptake |
| TG-B | Triglycerides |
| TOB | Tobramycin |
| TP3 | Total Protein |
| UDR | Remove - No longer Needed |
| UIBX | Remove - No longer Needed |
| UIBXR | Remove - No longer Needed |
Mather
| VANCHCS - Mather | |
| Questions | Answers |
| Q: Do we need to integrate costs for the move to a temporary location and back? | A: Yes, this is mentioned in the Statement of Work, Section Three - Installation/Site Preperation. |
| Q: Can we get the CAD drawing for the temporary space. | A. Yes, please see attached |
| Q: When is the move to the tempary space? | A: So far we have been told CY 2026. |
| Q: Will the current analyzers go away? | A: I did not understand the question so she refrased it to say will the current analyzers be prensent when the new analyzers are put in. The answer is yes, until the new ones are fully validated and signed off. |
| Q: Will there be a temporary location to bring the new analyzers in to validate? | A: I was not sure, we do not have that as a plan at this time and it depends on the analyzers that will be put in with the new contract. |
| Q: Were any existing staff present when the current analyzers were installed and how was that accomplishedt? | A: Yes, we moved the current analyzers to a different location in the lab and installed the new ones in their permenant positions. |
| Q: Middleware to be used for rule building? | A: The plan is that we will build the rules and autovalidation into Data Innovations not the vendor middleware. We use the vendor middleware for the function of the analyzer and track system. |
| Q: How is Remisol being used now? As a pass through? | A: Pass through only. |
| Q: Is QC automatically uploading? | A: Yes, BD. (I found out after they left that they never set it up but they do want it that way - lask of support with computer since we had other priorities for the Laboratory Information Manager/LIM.) |
| Q: Do we want the servers to be virtual or physical servers. | A: We prefer virtual servers. |
| Q: Will out hematology analyzers/Sysmex be attached to the track system? | A: No, the system is not the system by Sysmex that can be attached to a track system. |
| Q: What is the name of the vendor taking care of the water systems attached to the chemistry analyzers. | A: Evoqua Water Technologies. |
| Q: Did we want storage on the track? | A: Yes |
| Q: Where does that door go (from site visit)? | A: Histology and they asked if they needed to provide access, I answered that yes that needs to be accessible. |
| Q: Can the pole be removed (pole with utilities) | A: The drain can be moved but the pole may be acting as a vent/airflow and may not be removed. |
| Q: Can refrigerators be moved? | A: Yes |
| Q: Do we have Autovalidation set up | A: Not yet, but we plan to set it up. |
| Q: Are cabinets modular or fixed? | A: They are modular, not fixed. |
| Q: Do we have dedicated LIS (Laboratory Information System) support? | A: Yes, we have one Laboratory Information Manager/LIM and another one starting soon. |
| Q: Is the pneumatic tube system new? | A: Yes. |
| Q: Any planned upgrades to the LIS? | A: Yes, Cerner but we are not on the schedule yet. |
| Q: Any growth? | A: We did take over Modesto, Sonora and Stockton and the workload will be going to the Stockton laboratory. |
| Q: Do we want to be able to load all the laboratory tests into the track, not just chemistry. | A. We prefer to have the ability |
| Q: Could we switch to a new QC vendor and can we get a list of what QC you are using currently so they know how to meet out requirements. | A: We are open to using other control vendors. Vendor QC must meet requirements in SOW. |
| Q: Does VHA need bracketing (tieing down of analyzers due to earthquate shaking danger)? | A. VHA does require bracing for items 4 feet or taller such as shelving, medical equipment, or storage units, the VHA would require these items to be properly braced or anchored to prevent tipping during an earthquake. |
| Q: Is it the same space and no changes? | A. Correct, no changes |
| Q: When is the plan to move to the temproary space? | A: Last we were told CY 2026. |
| Q: Is the water system owned by Mather? | A: No, by the current vendor. |
| Q: What vendor is maintaing them? | A: Evoqua Water Technologies. |
| Q: The venodor rep asked about Stockton. | A: Stockton will have a lab and will take over the workload for Modesto, Sonora and Stockton from Mather. |
| Q: What is the internal door width at Mather for the purpose of bringing equipment in? | A. Both doors from the Lab to the main hallway are 40.5” |
The rear door is 34.5” SO2+36C26124Q0782: page 19, section 3. Installation/Site Preparation.
Clarifications requested: Point 1, c Mather Laboratory states:
i. Mather will be renovated during the life of this contract.
ii. The contractor will be required to move equipment to a temporary location, and then to a final location.
iii. Revalidation is required at each movement.
Clarifications requested: We request details on the point i., ii., and iii. Please provide additional details on the expected location of this new laboratory. Also, if available, please provide any details (CAD, diagram, etc.) that depicts the new space and temporary location to assist the vendor with an understanding of the scale of this future move.
Clarifications requested: We ask for a proposed timeline to be defined for these moves.
Clarifications requested: We ask if vendor provided water and ancillary items are to be included in these moves. Provided CAD drawing of new space. Currently planned for the parking lot behind cafeteria adjacent to building 649 and 647. Currently projected for Calendar year 2026. Subject to changes.
Clarification:
Yes. Vendor should provide water and ancillary items are to be included in these moves.
Mather Test Volumes
| Tests | Quarterly QTY | Annual QTY |
| Access 25 OH Vit D DxI - Access 25 OH Vit D for use on DxI 2x50 DET | 6,746 | 26,984 |
| Access TSH (3rd IS) 2x100 Det | 9,804 | 39,216 |
| AccuTnI+3 DxI - AccuTnI+3 for use on DxI | 2,159 | 8,636 |
| ACTM - ACETAMINOPHEN 1X100 TEST KIT | 107 | 428 |
| ALB - CX ALBUMIN REAGENT 2 X 300 | 19,041 | 76,164 |
| ALP - CX/LX ALP KIT 2 X 400 TEST | 18,256 | 73,024 |
| ALT - CX/LX ALT KIT 2 X 400 TEST | 18,920 | 75,680 |
| AMM - CX AMMONIA REAGENT 2X25 TEST | 100 | 400 |
| AMPH - CX AMPHETAMINE KIT 1X250 TESTS | 2,907 | 11,628 |
| AMY7 - G7 AMY 2 x 200 SYN Kit | 193 | 772 |
| AST - CX/LX AST KIT 2 X 400 TEST | 18,910 | 75,640 |
| B12 2X 50 DET - ACCESS B12 2 X 50 DET | 5,942 | 23,768 |
| BARB - BARBITURATE KIT (1X250 TESTS) | 2,117 | 8,468 |
| BNP2 - Access BNP | 1,534 | 6,136 |
| BNZG - Benzodiazepine Reagent | 2,118 | 8,472 |
| BUN - CX BUN REAGENT 2X300 TEST | 28,197 | 112,788 |
| CALC - Calcium | 28,437 | 113,748 |
| CAR - CX/LX CARBAMAZEPINE KIT 200T | 4 | 16 |
| CAR - CX/LX CARBAMAZEPINE KIT 200T REF | 26 | 104 |
| CEA 2X50 TESTS - ACCESS CEA 2X50 TESTS | 158 | 632 |
| CHOL - CX CHOLESTEROL 2 X 300 TEST | 12,227 | 48,908 |
| CK - CX/LX CK KIT 2 X 400 TST | 865 | 3,460 |
| CK-MB REAGENT KIT | 17 | 68 |
| CL - Chloride | 28,316 | 113,264 |
| CO2A - LX C02 ACID REAGENT 2X2L | 28,320 | 113,280 |
| COCM - COCAINE METABOLITE KIT(1X250 TESTS | 2,908 | 11,632 |
| Cortisol 2X50 DET - ACCESS CORTISOL | 397 | 1,588 |
| CRPH - SYNCHRON HIGH SENSTVTY CRP 2X200 TST | 1,433 | 5,732 |
| CR-S - CX4 CREATININE KIT 2X300 | 36,672 | 146,688 |
| DBIL - CX/LX DBIL KIT 2 X 300 TEST | 6,820 | 27,280 |
| Digoxin 2X50 DET - ACCESS DIGOXIN | 79 | 316 |
| ETOH - ETOH KIT 2X150 TEST | 2,766 | 11,064 |
| FE - CX/LX IRON FERROUS KIT (2X200) | 2,577 | 10,308 |
| Ferritin 2 X 50 - ACCESS FERRITIN 2 X 50 DET | 3,072 | 12,288 |
| Folate 2x50 Determination - Access Folate 2x50 Determination | 2,401 | 9,604 |
| Free T4, 2X50 DET - ACCESS FREE T4, 2X50 DET | 2,662 | 10,648 |
| GEN - CX/LX GENTAMICIN 2 X 100 TEST | 16 | 64 |
| Gentian Cystatin C - 1100 Gentian Cystatin C reagent | 31 | 124 |
| GLUCM - LX GLUCOSE 2X2L | 28,230 | 112,920 |
| HDL - *RF*KIT, HDL 2 x 200 DXC | 12,160 | 48,640 |
| Hybritech PSA - ACCESS HYBRITECH PSA RGT | 5,449 | 21,796 |
| IG-A - LX IMMUNOGLOBULIN A (IG-A) RGNT | 197 | 788 |
| IG-G - LX IMMUNOGLOBULIN G (IG-G) RGNT | 231 | 924 |
| IG-M - LX IMMUNOGLOBULIN M (IG-M) REAGENT | 173 | 692 |
| K - Potassium | 28,516 | 114,064 |
| LACT - LACT 2 X 50 DXC KIT | 1,003 | 4,012 |
| LD-L - VR* CX/LX LD 2 X 300 TEST | 291 | 1,164 |
| LDLD - KIT, LDLD 2X100 CX/LX | 3,001 | 12,004 |
| LIP - CX/LX LIP KIT 2 X 60 TEST | 1,276 | 5,104 |
| Lithium - Lithium 2 x 65 Test Kit | 140 | 560 |
| MAB - MICROALBUMIN 2X100 TEST KIT | 3,746 | 14,984 |
| METD - METD KIT (1X250 TESTS) | 2,117 | 8,468 |
| MG - CX MAGNESIUM 2X100 TEST | 7,586 | 30,344 |
| M-TP - CX MICROPROTEIN REAGENT 2X50 TEST | 838 | 3,352 |
| Myoglobin - ACCESS MYOGLOBIN, 2X50 DET | 3 | 12 |
| NA - Sodium | 28,574 | 114,296 |
| OP - OPIATE KIT (1X250 TESTS) | 2,907 | 11,628 |
| Oxycodone UDR - 10012653 Oxycodone UDR reagent | 2,874 | 11,496 |
| PAB - PREALBUMIN 2X100 TEST KIT | 48 | 192 |
| PHE - CX/LX PHENOBARBITAL KIT 200TST | 15 | 60 |
| PHS - Phosphorous Assay | 3,309 | 13,236 |
| PHY - CX/LX PHENYTOIN KIT 200 TEST | 26 | 104 |
| PTH 2 x 50 DET - ACCESS PTH 2 x 50 DET | 1,179 | 4,716 |
| SALY - KIT, SALY 2 X 45 CX/LX | 90 | 360 |
| TBIL - CX/LX TBIL KIT 2 X 400 TEST | 18,263 | 73,052 |
| Testosterone Reagent | 1,008 | 4,032 |
| TG - CX TRIGLYCERIDE REAGENT 2X300 TEST | 12,184 | 48,736 |
| THC5 - THC5 KIT (1X250 TESTS) | 2,907 | 11,628 |
| THE - CX/LX THEOPHYLLINE KIT, 200T | 18 | 72 |
| TOTAL B- hCG - ACCESS TOTAL B- hCG 2x50 Det | 93 | 372 |
| TP - CX TOTAL PROTEIN REAGENT 2X300 | 19,033 | 76,132 |
| TRFN - LX TRANSFERRIN REAGENT | 2,562 | 10,248 |
| URIC - CX URIC ACID REAGENT 2 X 300 | 828 | 3,312 |
| VANC - KIT, VANC 200 CX/LX | 259 | 1,036 |
| VPA - CX/LX VALPROIC ACID KT (2X100) | 152 | 608 |
| Hepatitis C | 2,200 | 8,800 |
| HIV 1/2 combo screen (4th Generation) | 2,200 | 8,800 |
| Hepatitis B | 2,160 | 8,640 |
| Hepatits B core antigen | 2,160 | 8,640 |
| Procalcitonin | 420 | 1,680 |
| Syphilis | 1,020 | 4,080 |
| Hepatitis B surface antigen | 1,950 | 7,800 |
| Tacrolimus | 540 | 2,160 |
| Hepatitis A | 522 | 2,088 |
| Hepatitis A IgM | 8 | 32 |
| Troponin I | 2,400 | 9,600 |
| Sex binding Hormone | 92 | 368 |
| Alpha Feta protein | 884 | 3,537 |
| Prostate-Specific Antigen, free | 32 | 126 |
| Grand Total | 488,541 | 2,020,515 |
Mather Bio-Rad QC
| VA Mather | (YTD=Year to Date) |
| QC Orders through Bio-Rad | |
| as of 8/13/2024 |
| 2024 | ||||
| Part# | Description | YTD | 2023 | 2022 |
| 146 | CARD MKR LT PLUS 1 LIQ 6X3ML | 16 | 20 | 16 |
| 148 | CARD MKR LT PLUS 3 LIQ 6X3ML | 16 | 20 | 16 |
| 361 | IA PLUS LIQ 1 12X5ML | 10 | 10 | 10 |
| 362 | IA PLUS LIQ 2 12X5ML | 10 | 10 | 10 |
| 363 | IA PLUS LIQ 3 12X5ML | 10 | 10 | 10 |
| 364 | SPECIALTY IA LIQ 1 6X5ML | 5 | 15 | 7 |
| 365 | SPECIALTY IA LIQ 2 6X5ML | 5 | 15 | 7 |
| 366 | SPECIALTY IA LIQ 3 6X5ML | 5 | 15 | 7 |
| 397 | URINE CHEMISTRY 1 LIQ 12X10ML | 10 | 4 | |
| 398 | URINE CHEMISTRY 2 LIQ 12X10ML | 10 | 4 | |
| 697 | MULTIQUAL UNASSY 1 LIQ 12X10ML | 10 | 20 | 10 |
| 699 | MULTIQUAL UNASSY 3 LIQ 12X10ML | 10 | 20 | 10 |
| 740 | DIABETES BI LYPH 6X0.5ML | 8 | 12 | 11 |
| 740X | DIABETES LYPH SP | 1 | ||
| 751 | SPINAL FLUID 1 LIQ 6X3ML | 8 | 2 | |
| 752 | SPINAL FLUID 2 LIQ 6X3ML | 8 | 2 |
Martinez
| VANCHCS - Martinez | |
| Questions | Answers |
| Q: The vendor representatives asked for us to breakdown the test list to individual tests not groups. | A. See applicable tab for a breakdown of test volumes |
| Q: Do we need bracking? | VHA does require bracing for items 4 feet or taller such as shelving, medical equipment, or storage units, the VHA would require these items to be properly braced or anchored to prevent tipping during an earthquake. |
| Q: Need the Stockton floor plan | A: Please see attached document |
| Q: Any upcoming changes to the existing floor plan? | A: None planned at this site. |
| Q: Where is the water system? | A. Currently installed in adjacent storage room. |
| Q: Will we have a staging location for the new analyzers during validation? | A: No, we do not have a staging area. |
| Q: Please provide door width the analyzers will be moved through? | A: 42 inches |
| Q: What is our middleware/interface? | A: Data Innovations |
| Q: Will we do autovalidation - workflow and design | A: Yes, not sure when |
Martinez Test Volumes
| Tests | Quarterly QTY | Annual QTY |
| Access 25 OH Vit D DxI | 2,736 | 10,944 |
| Access hsTNI 2X50 - ACCESS hsTNI | 30 | 120 |
| Access TSH (3rd IS) 2x100 Det | 3,300 | 13,200 |
| AccuTnI+3 DxI - AccuTnI+3 | 61 | 244 |
| ACTM - ACETAMINOPHEN 1X100 TEST KIT | 10 | 40 |
| ALB - CX ALBUMIN REAGENT 2 X 300 | 5,370 | 21,480 |
| ALP - CX/LX ALP KIT 2 X 400 TEST | 5,176 | 20,704 |
| ALT - CX/LX ALT KIT 2 X 400 TEST | 5,874 | 23,496 |
| AMM - CX AMMONIA REAGENT 2X25 TEST | 62 | 248 |
| AMPH - CX AMPHETAMINE KIT 1X250 TESTS | 716 | 2,864 |
| AMY7 - G7 AMY 2 x 200 SYN Kit | 189 | 756 |
| AST - CX/LX AST KIT 2 X 400 TEST | 5,694 | 22,776 |
| B12 2 X 50 DET - ACCESS B12 2 X 50 DET | 2,569 | 10,276 |
| BARB - BARBITURATE KIT (1X250 TESTS) | 632 | 2,528 |
| BNP2 - Access | 131 | 524 |
| BNZG - Benzodiazepine Reagent | 632 | 2,528 |
| BUN - CX BUN REAGENT 2X300 TEST | 6,823 | 27,292 |
| CALC - Calcium | 6,841 | 27,364 |
| CAR - CX/LX CARBAMAZEPINE KIT 200T | 12 | 48 |
| CHOL - CX CHOLESTEROL 2 X 300 TEST | 4,226 | 16,904 |
| CK - CX/LX CK KIT 2 X 400 TST | 310 | 1,240 |
| CL - Chloride | 6,852 | 27,408 |
| CO2A - LX C02 ACID REAGENT 2X2L | 6,861 | 27,444 |
| COCM - COCAINE METABOLITE KIT | 716 | 2,864 |
| CRPH - SYNCHRON HIGH SENSTVTY CRP | 313 | 1,252 |
| CR-S - CX4 CREATININE KIT 2X300 | 9,952 | 39,808 |
| DBIL - CX/LX DBIL KIT 2 X 300 TEST | 987 | 3,948 |
| ETOH - ETOH KIT 2X150 TEST | 713 | 2,852 |
| FE - CX/LX IRON FERROUS KIT (2X200) | 981 | 3,924 |
| Ferritin 2 X 50 - ACCESS FERRITIN | 1,028 | 4,112 |
| Folate 2x50 | 417 | 1,668 |
| Free T4, 2X50 DET - ACCESS FREE T4 | 890 | 3,560 |
| GLUCM - LX GLUCOSE 2X2L | 6,838 | 27,352 |
| HDL - *RF*KIT, HDL 2 x 200 DXC | 4,200 | 16,800 |
| HFSH 2 X 50 DET - ACCESS HFSH 2 X 50 D | 262 | 1,048 |
| HLH 2 X 50 DET - ACCESS HLH 2 X 50 DET | 326 | 1,304 |
| Hybritech PSA - ACCESS HYBRITECH PSA RGT | 1,839 | 7,356 |
| K - Potassium | 6,938 | 27,752 |
| LACT - LACT 2 X 50 DXC KIT | 74 | 296 |
| LD-L - VR* CX/LX LD 2 X 300 TEST | 137 | 548 |
| LDLD - KIT, LDLD 2X100 CX/LX | 665 | 2,660 |
| LIP - CX/LX LIP KIT 2 X 60 TEST | 140 | 560 |
| MAB - MICROALBUMIN 2X100 TEST KIT | 1,823 | 7,292 |
| METD - METD KIT (1X250 TESTS) | 632 | 2,528 |
| MG - CX MAGNESIUM 2X100 TEST | 1,279 | 5,116 |
| M-TP - CX MICROPROTEIN REAGENT 2X50 | 262 | 1,048 |
| NA - Sodium | 6,923 | 27,692 |
| OP - OPIATE KIT (1X250 TESTS) | 716 | 2,864 |
| Oxycodone UDR | 734 | 2,936 |
| PCP - PCP KIT (1X250 TESTS) | 10 | 40 |
| PHE - CX/LX PHENOBARBITAL KIT 200TST | 11 | 44 |
| PHS - Phosphorous Assay | 815 | 3,260 |
| PHY - CX/LX PHENYTOIN KIT 200 TEST | 15 | 60 |
| Prolactin - ACCESS PROLACTIN 2X50 DET | 332 | 1,328 |
| PROX - PROX KIT (1X250 TESTS) | 10 | 40 |
| TBIL - CX/LX TBIL KIT 2 X 400 TEST | 5,175 | 20,700 |
| Testosterone Reagent | 347 | 1,388 |
| TG - CX TRIGLYCERIDE REAGENT 2X300 TEST | 4,203 | 16,812 |
| THC5 - THC5 KIT (1X250 TESTS) | 716 | 2,864 |
| THE - CX/LX THEOPHYLLINE KIT, 200T | 10 | 40 |
| TP - CX TOTAL PROTEIN REAGENT 2X300 | 5,170 | 20,680 |
| TRFN - LX TRANSFERRIN REAGENT | 977 | 3,908 |
| URIC - CX URIC ACID REAGENT 2 X 300 | 677 | 2,708 |
| VANC - KIT, VANC 200 CX/LX | 22 | 88 |
| VPA - CX/LX VALPROIC ACID KT (2X100) | 43 | 172 |
| Grand Total | 133,425 | 533,700 |
Martinez Bio-Rad QC
| VA Martinez | (YTD=Year to Date) |
| QC Orders through Bio-Rad | |
| as of 8/13/2024 |
| 2024 | ||||
| Part# | Description | YTD | 2023 | 2022 |
| 12007028 | qUAntify Advance Liq 6X12ML | 6 | 8 | 18 |
| 12016447 | CARDIAC ADVANCE LIQ 1 6X3ML | 4 | ||
| 12016448 | CARDIAC ADVANCE LIQ 2 6X3ML | 4 | ||
| 12016449 | CARDIAC ADVANCE LIQ 3 6X3ML | 4 | ||
| 12016450 | CARDIAC ADVANCE LIQ 4 6X3ML | 4 | ||
| 146 | CARD MKR LT PLUS 1 LIQ 6X3ML | 9 | 14 | 9 |
| 148 | CARD MKR LT PLUS 3 LIQ 6X3ML | 9 | 14 | 9 |
| 361 | IA PLUS LIQ 1 12X5ML | 5 | 9 | 9 |
| 362 | IA PLUS LIQ 2 12X5ML | 5 | 9 | 9 |
| 363 | IA PLUS LIQ 3 12X5ML | 5 | 9 | 9 |
| 364 | SPECIALTY IA LIQ 1 6X5ML | 5 | 6 | 1 |
| 365 | SPECIALTY IA LIQ 2 6X5ML | 5 | 6 | 1 |
| 366 | SPECIALTY IA LIQ 3 6X5ML | 5 | 6 | 1 |
| 397 | URINE CHEMISTRY 1 LIQ 12X10ML | 3 | 6 | |
| 398 | URINE CHEMISTRY 2 LIQ 12X10ML | 3 | 6 | |
| 697 | MULTIQUAL UNASSY 1 LIQ 12X10ML | 7 | 5 | 6 |
| 698 | MULTIQUAL UNASSY 2 LIQ 12X10ML | 3 | ||
| 699 | MULTIQUAL UNASSY 3 LIQ 12X10ML | 7 | 5 | 6 |
| 751 | SPINAL FLUID 1 LIQ 6X3ML | 6 | 4 | |
| 752 | SPINAL FLUID 2 LIQ 6X3ML | 6 | 4 |
Redding
| VANCHCS - Redding | |
| Questions | Answers |
| Q: How will we handle the transition from the old vendor to the new- is this a temporary space? | A: We will bring them in the lab - we can position the new ones and move the current analyzers or the reverse. |
| Q: Would it be 2 Alinity Immunassay analyzers and 2 Alinity chemistry analyzers? | A: Yes |
| Q: What is our Interface/Middleware? | A: Data Innovations |
| Q: Any changes in the SOW for Redding? | A: No. The wall cannot be removed and the egress route is stated in the SOW. |
| Q: Can we get the volumes by test not group? | A. See applicable tab for a breakdown of test volumes |
| Q: Is the Vendor middleware a pass through? | A: Yes |
| Q: Who maintains the current water systems? | A: There is a sticker on the outside of the water system that says Evoqua Water Technologies |
Redding Test Volumes
| Test | Sum of QTY (quarter) | Annual QTY |
| Access 25 OH Vit D D | 2,141 | 8,564 |
| Access TSH | 2,658 | 10,632 |
| ALB - CX ALBUMIN REAGENT 2 X 300 | 4,483 | 17,932 |
| ALP - CX/LX ALP KIT 2 X 400 TEST | 4,418 | 17,672 |
| ALT - CX/LX ALT KIT 2 X 400 TEST | 4,572 | 18,288 |
| AMM - CX AMMONIA REAGENT 2X25 TEST | 38 | 152 |
| AMPH - CX AMPHETAMINE KIT 1X250 TESTS | 842 | 3,368 |
| AMY7 - G7 AMY 2 x 200 SYN Kit | 132 | 528 |
| AST - CX/LX AST KIT 2 X 400 TEST | 4,531 | 18,124 |
| B12 2 X 50 DET - ACCESS B12 2 X 50 DET | 1,476 | 5,904 |
| BARB - BARBITURATE KIT (1X250 TESTS) | 698 | 2,792 |
| BNP2 - Access | 199 | 796 |
| BNZG - Benzodiazepine Reagent | 698 | 2,792 |
| BUN - CX BUN REAGENT 2X300 TEST | 5,252 | 21,008 |
| CALC - Calcium | 5,385 | 21,540 |
| CHOL - CX CHOLESTEROL 2 X 300 TEST | 3,788 | 15,152 |
| CK - CX/LX CK KIT 2 X 400 TST | 193 | 772 |
| CL - Chloride | 5,347 | 21,388 |
| CO2A - LX C02 ACID REAGENT 2X2L | 5,332 | 21,328 |
| COCM - COCAINE METABOLITE KIT | 843 | 3,372 |
| CRPH - SYNCHRON HIGH SENSTVTY | 287 | 1,148 |
| CR-S - CX4 CREATININE KIT 2X300 | 7,822 | 31,288 |
| DBIL - CX/LX DBIL KIT 2 X 300 TEST | 614 | 2,456 |
| ETOH - ETOH KIT 2X150 TEST | 747 | 2,988 |
| FE - CX/LX IRON FERROUS KIT (2X200) | 497 | 1,988 |
| Ferritin 2 X 50 - ACCESS FERRITIN | 434 | 1,736 |
| Folate 2x50 | 318 | 1,272 |
| Free T4, 2X50 DET - ACCESS FREE | 536 | 2,144 |
| GLUCM - LX GLUCOSE 2X2L | 7,131 | 28,524 |
| HDL - *RF*KIT, HDL 2 x 200 DXC | 3,770 | 15,080 |
| Hybritech PSA | 1,174 | 4,696 |
| K - Potassium | 7,315 | 29,260 |
| LACT - LACT 2 X 50 DXC KIT | 70 | 280 |
| LD-L - VR* CX/LX LD 2 X 300 TEST | 72 | 288 |
| LDLD - KIT, LDLD 2X100 CX/LX | 412 | 1,648 |
| LIP - CX/LX LIP KIT 2 X 60 TEST | 167 | 668 |
| MAB - MICROALBUMIN 2X100 TEST KIT | 1,414 | 5,656 |
| METD - METD KIT (1X250 TESTS) | 698 | 2,792 |
| MG - CX MAGNESIUM 2X100 TEST | 492 | 1,968 |
| M-TP - CX MICROPROTEIN REAGENT | 153 | 612 |
| NA - Sodium | 5,366 | 21,464 |
| OP - OPIATE KIT (1X250 TESTS) | 843 | 3,372 |
| Oxycodone UDR - 10012653 Oxycodone UDR | 853 | 3,412 |
| PHS - Phosphorous Assay | 247 | 988 |
| TBIL - CX/LX TBIL KIT 2 X 400 TEST | 4,419 | 17,676 |
| Testosterone Reagent | 353 | 1,412 |
| TG - CX TRIGLYCERIDE REAGENT 2X300 | 3,781 | 15,124 |
| THC5 - THC5 KIT (1X250 TESTS) | 843 | 3,372 |
| TP - CX TOTAL PROTEIN REAGENT 2X300 | 4,425 | 17,700 |
| TRFN - LX TRANSFERRIN REAGENT | 493 | 1,972 |
| URIC CX URIC ACID REAGENT | 228 | 912 |
| Grand Total | 109,000 | 436,000 |
Redding Bio-Rad QC
| VA Redding | (YTD=Year to Date) |
| QC Orders through Bio-Rad | |
| as of 8/13/2024 |
| 2024 | ||||
| Part# | Description | YTD | 2023 | 2022 |
| 12007028 | qUAntify Advance Liq 6X12ML | 26 | 8 | |
| 12016447 | CARDIAC ADVANCE LIQ 1 6X3ML | 10 | ||
| 12016448 | CARDIAC ADVANCE LIQ 2 6X3ML | 10 | ||
| 12016449 | CARDIAC ADVANCE LIQ 3 6X3ML | 10 | ||
| 12016450 | CARDIAC ADVANCE LIQ 4 6X3ML | 10 | ||
| 146 | CARD MKR LT PLUS 1 LIQ 6X3ML | 11 | 10 | 8 |
| 148 | CARD MKR LT PLUS 3 LIQ 6X3ML | 11 | 10 | 8 |
| 361 | IA PLUS LIQ 1 12X5ML | 8 | 4 | 4 |
| 362 | IA PLUS LIQ 2 12X5ML | 8 | 4 | 4 |
| 363 | IA PLUS LIQ 3 12X5ML | 8 | 4 | 4 |
| 364 | SPECIALTY IA LIQ 1 6X5ML | 5 | 7 | |
| 365 | SPECIALTY IA LIQ 2 6X5ML | 5 | 7 | |
| 366 | SPECIALTY IA LIQ 3 6X5ML | 5 | 7 | |
| 397 | URINE CHEMISTRY 1 LIQ 12X10ML | 4 | 3 | |
| 398 | URINE CHEMISTRY 2 LIQ 12X10ML | 4 | 3 | |
| 697 | MULTIQUAL UNASSY 1 LIQ 12X10ML | 12 | 2 | 4 |
| 699 | MULTIQUAL UNASSY 3 LIQ 12X10ML | 12 | 2 | 4 |
Stockton
| VANCHCS - Stockton | |
| Questions | Answers |
| Q: Can the sink near the pass through be removed? | A: Yes, the sink can be removed if needed. |
| Q: Could the neutralizer be removed? It may be in the way for the service techs for the analyzers. | A: The neutralizer cannot be removed. |
| Q: Not sure if the plug for the 220 power is correct? | A. Activation teams can support power requirments. |
| Q: Who will be responsible for the connection of the water system to the analyzer through the ceiling? | A: Engineering can support putting the tubing through the ceiling from the water systems where it will be required and is the current set up (Stockton, Mather, and Martinez). |
| Q: Vendor representatives asked for a breakdown of the test list to individual tests not groups. | A. Stockton is a new facility. We are expecting the same volume as VAMC Martinez. Please See applicable tab for a breakdown of test volumes. |
| Q: They ask if the placement is side by side. | A: Yes |
| Can a CAD drawing and floor plan for Stockton be provided? | A. Yes, see Attached |
Stockton Est. Test Volumes
| Tests | Quarterly QTY | Annual QTY |
| Access 25 OH Vit D DxI | 2,736 | 10,944 |
| Access hsTNI 2X50 - ACCESS hsTNI | 30 | 120 |
| Access TSH (3rd IS) 2x100 Det | 3,300 | 13,200 |
| AccuTnI+3 DxI - AccuTnI+3 | 61 | 244 |
| ACTM - ACETAMINOPHEN 1X100 TEST KIT | 10 | 40 |
| ALB - CX ALBUMIN REAGENT 2 X 300 | 5,370 | 21,480 |
| ALP - CX/LX ALP KIT 2 X 400 TEST | 5,176 | 20,704 |
| ALT - CX/LX ALT KIT 2 X 400 TEST | 5,874 | 23,496 |
| AMM - CX AMMONIA REAGENT 2X25 TEST | 62 | 248 |
| AMPH - CX AMPHETAMINE KIT 1X250 TESTS | 716 | 2,864 |
| AMY7 - G7 AMY 2 x 200 SYN Kit | 189 | 756 |
| AST - CX/LX AST KIT 2 X 400 TEST | 5,694 | 22,776 |
| B12 2 X 50 DET - ACCESS B12 2 X 50 DET | 2,569 | 10,276 |
| BARB - BARBITURATE KIT (1X250 TESTS) | 632 | 2,528 |
| BNP2 - Access | 131 | 524 |
| BNZG - Benzodiazepine Reagent | 632 | 2,528 |
| BUN - CX BUN REAGENT 2X300 TEST | 6,823 | 27,292 |
| CALC - Calcium | 6,841 | 27,364 |
| CAR - CX/LX CARBAMAZEPINE KIT 200T | 12 | 48 |
| CHOL - CX CHOLESTEROL 2 X 300 TEST | 4,226 | 16,904 |
| CK - CX/LX CK KIT 2 X 400 TST | 310 | 1,240 |
| CL - Chloride | 6,852 | 27,408 |
| CO2A - LX C02 ACID REAGENT 2X2L | 6,861 | 27,444 |
| COCM - COCAINE METABOLITE KIT | 716 | 2,864 |
| CRPH - SYNCHRON HIGH SENSTVTY CRP | 313 | 1,252 |
| CR-S - CX4 CREATININE KIT 2X300 | 9,952 | 39,808 |
| DBIL - CX/LX DBIL KIT 2 X 300 TEST | 987 | 3,948 |
| ETOH - ETOH KIT 2X150 TEST | 713 | 2,852 |
| FE - CX/LX IRON FERROUS KIT (2X200) | 981 | 3,924 |
| Ferritin 2 X 50 - ACCESS FERRITIN | 1,028 | 4,112 |
| Folate 2x50 | 417 | 1,668 |
| Free T4, 2X50 DET - ACCESS FREE T4 | 890 | 3,560 |
| GLUCM - LX GLUCOSE 2X2L | 6,838 | 27,352 |
| HDL - *RF*KIT, HDL 2 x 200 DXC | 4,200 | 16,800 |
| HFSH 2 X 50 DET - ACCESS HFSH 2 X 50 D | 262 | 1,048 |
| HLH 2 X 50 DET - ACCESS HLH 2 X 50 DET | 326 | 1,304 |
| Hybritech PSA - ACCESS HYBRITECH PSA RGT | 1,839 | 7,356 |
| K - Potassium | 6,938 | 27,752 |
| LACT - LACT 2 X 50 DXC KIT | 74 | 296 |
| LD-L - VR* CX/LX LD 2 X 300 TEST | 137 | 548 |
| LDLD - KIT, LDLD 2X100 CX/LX | 665 | 2,660 |
| LIP - CX/LX LIP KIT 2 X 60 TEST | 140 | 560 |
| MAB - MICROALBUMIN 2X100 TEST KIT | 1,823 | 7,292 |
| METD - METD KIT (1X250 TESTS) | 632 | 2,528 |
| MG - CX MAGNESIUM 2X100 TEST | 1,279 | 5,116 |
| M-TP - CX MICROPROTEIN REAGENT 2X50 | 262 | 1,048 |
| NA - Sodium | 6,923 | 27,692 |
| OP - OPIATE KIT (1X250 TESTS) | 716 | 2,864 |
| Oxycodone UDR | 734 | 2,936 |
| PCP - PCP KIT (1X250 TESTS) | 10 | 40 |
| PHE - CX/LX PHENOBARBITAL KIT 200TST | 11 | 44 |
| PHS - Phosphorous Assay | 815 | 3,260 |
| PHY - CX/LX PHENYTOIN KIT 200 TEST | 15 | 60 |
| Prolactin - ACCESS PROLACTIN 2X50 DET | 332 | 1,328 |
| PROX - PROX KIT (1X250 TESTS) | 10 | 40 |
| TBIL - CX/LX TBIL KIT 2 X 400 TEST | 5,175 | 20,700 |
| Testosterone Reagent | 347 | 1,388 |
| TG - CX TRIGLYCERIDE REAGENT 2X300 TEST | 4,203 | 16,812 |
| THC5 - THC5 KIT (1X250 TESTS) | 716 | 2,864 |
| THE - CX/LX THEOPHYLLINE KIT, 200T | 10 | 40 |
| TP - CX TOTAL PROTEIN REAGENT 2X300 | 5,170 | 20,680 |
| TRFN - LX TRANSFERRIN REAGENT | 977 | 3,908 |
| URIC - CX URIC ACID REAGENT 2 X 300 | 677 | 2,708 |
| VANC - KIT, VANC 200 CX/LX | 22 | 88 |
| VPA - CX/LX VALPROIC ACID KT (2X100) | 43 | 172 |
| Grand Total | 133,425 | 533,700 |