This document contains questions and answers related to a federal solicitation for a chemistry and immunoassay equipment and test contract (36C26124Q0782) for the VA Northern California Health Care System (VANCHCS).

The key details include:

  • The solicitation requires pricing for a variety of chemistry, immunoassay, and other clinical laboratory tests, some with zero estimated demand initially.
  • The solicitation has been extended to August 26, 2024 at 12PM PDT.
  • Drawings of the laboratory spaces at the Mather and Martinez locations have been provided.
  • The contract will require the contractor to move equipment to a temporary location during a planned renovation at the Mather laboratory, with revalidation required.
  • The government is open to using control vendors other than the current Bio-Rad, and requires the ability to export and upload QC data in real-time.
  • Detailed test volumes have been provided for the Mather, Martinez, and Stockton laboratory locations.

View the file

Other files

File Type Posted
S02 36C26124Q0782 RFQ Price List.xlsx XLSX spreadsheet
36C26124Q0782 0004.docx DOCX document
36C26124Q0782 0003.docx DOCX document
4. Mather Laboratory Sink Removed.docx DOCX document
3. Stockton Clinical Laboratory_floor plan.pdf PDF
2. Mather Updated Temp Lab Layout.pdf PDF
36C26124Q0782 0002.docx DOCX document
36C26124Q0782 0001.docx DOCX document
S02 36C26124Q0782.docx DOCX document
Redding Floor Plan.pdf PDF
S02 36C26124Q0418 Price List.xlsx XLSX spreadsheet
Mather Floor Plan.pdf PDF
Martinez Floor Plan.docx DOCX document
Show all 13

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

General Questions

VA Northern California Health Care System (VANCHCS)
QuestionsAnswers
Q. Document "S02 36C26124Q0418 Price List". We noticed a number of tests are listed with a qty of 0. Can you confirm if you want us to include these test?
Examples are: BNP and Cystatin CA. All volumes estimated with a value of one (1) and/or zero (0) require only a best price/base price please. There is no known demand at this time; however, we need a price for future cases should they appear.

See applicable tab for a breakdown of test volumes

Request: Due to the ADLM meeting taking place the week of 7/29/24 and various other task related to digesting the information gathered during the 7/22/24 site visits as well as the scope of the clarifications contained within, we request a minimum extension on the due date of at least 3 weeks.A. Solicitation has been extended to 26 August 2024 at 12PM PDT.
Request: We request a CAD for each location and the proposed temporary space for the Mather renovation to best understand the dimensions of each laboratory.A. Stockton CAD drawing and temporary Mather Floor Plans are attached. All other plans were attached to the original solicitation - 36C26124Q0782.
Request: Please indicate how many analyzers are being requested for each location. If backups are needed, must they be the same model?A. See Statement of Work for analyzer types.

A. Yes, backup must be same as the primary analyzer.

Requirement: Page 53, Instructions to Offerors, point 1. Quote submittal requirements, section a. note on solicitation terms and conditions states that Offerors need to provide a price for ALL assays to include open channel assays.
This can be source limiting, please consider a preferred statement inclusion or provide evaluation criteria against this.A. There are no changes to the instructions at this time.
Q. SO2+36C26124Q0418: Tab 1.

Request: We are requesting more extensive assay definitions for the following assays:

ALBM, ALBX, ALPD, CR-S, CR-T, CREM, CRTS, TP3, PTM, CKNA, LD-P, PHOSM, PHS, TG-B, AMY7, CMBX, HDLC, HDLD, HDLX, M-TP, PAM, T3U, UIBX, UIBXR, CAR, GEN, PHE, PHY, TOB, METD, METQ, PROX, UDR, CHE, CHEX, LIPM, MA, MAB, PAB, CR-S, CR-T, HEMOL, CSA, EVR, MPA.

Request: for assays listed by group, we request that the test volume for each assay be provided so that we can provide accurate pricing.A. See "Test Key" tab for a list of Test Names and Acronyms. A. See applicable tab for a breakdown of test volumes
Q. Data Management

Requirement: Page 19, section 2.7, point 4 requires the capability to perform real-time moving averages, etc.

Clarification requested: Based on the current deployment of instrumentation, is it a fair assumption that all rules are to be written in vendor-supplied middleware?

Clarification requested: If there is a split between the VA Data Innovations and the vendor Data Innovations, please specify which party provides which rules.

Clarification requested: Is autoverification to be performed using your instance of Data Innovations Instrument Manager, or are you expecting the vendor to perform this on their middleware?

A. Do not assume rules are to be written in vendor-supplied middleware. We are not part of the current deployment.

A. At this time we do not know until award of contract.

A. 2.7.6: Can not be determined until award of contract. Our current plan is to use data innovations.

Q. Controls:

Requirement: Page 11 of 36C26124Q0782 states ‘the per reportable result price will include cost covering (a) equipment use, (b) all reagents, standards, controls, supplies, consumable/disposable items, etc.

Clarification requested: Please identify which control vendor you currently use.

Clarification requested: Are you open to using other control vendors?

Clarification requested: Are you using QC peer review software, and are we to include the cost of the software license for this?

Clarification requested: Please also share the quantity and list numbers for your current controls that you acquire today. A. Control vender is currently Bio-RAD

A. We are open to using other control vendors. Vendor QC must meet requirements in PWS.

A. We are using QC peer review software, needed licencing should be provided by contractor. QC must have ability to export and upload QC data in real time to peer group program per SOW..

A. See applicable tab for Bio-Rad/QC information.

Test Key

AcronymTest Name
ALBMAblumin
ALBXAblumin
ALPDAlkaline Phosphatase
AMY7Amylase
CARCarbamazepine
CHECholinesterase
CHEXCholinesterase UDR Assay (CHEX)
CKNARemove - No longer Needed
CMBXCreatine Kinase, Isoenzyme MB
CREMCreatinine
CR-SCreatinine
CR-TCreatinine
CRTSCreatinine
CSACyclosporine
EVRIron
GENGentamicin
HDLCHigh density lipoprotein- Cholesterol
HDLDHigh density lipoprotein- Cholesterol
HDLXHigh density lipoprotein- Cholesterol
HEMOLRemove - No longer Needed
LD-PLactate Dehydrogenase
LIPMLipase
MAMicroalbumin/ Urine albumin
MABMicroalbumin/ Urine albumin
METDMethadone
METQMethaqualone
MPAMycophenolic Acid (MPA) Assay
M-TPMiroprorotein
PABPrealbumin
PAMP-Amylase Reagent (PAM)
PHEPhenobarbital
PHOSMPhosphorus
PHSPhosphorus
PHYPhenytoin
PROXPropoxyphene
PTMRemove - No longer Needed
T3UT3 uptake
TG-BTriglycerides
TOBTobramycin
TP3Total Protein
UDRRemove - No longer Needed
UIBXRemove - No longer Needed
UIBXRRemove - No longer Needed

Mather

VANCHCS - Mather
QuestionsAnswers
Q: Do we need to integrate costs for the move to a temporary location and back?A: Yes, this is mentioned in the Statement of Work, Section Three - Installation/Site Preperation.
Q: Can we get the CAD drawing for the temporary space.A. Yes, please see attached
Q: When is the move to the tempary space?A: So far we have been told CY 2026.
Q: Will the current analyzers go away?A: I did not understand the question so she refrased it to say will the current analyzers be prensent when the new analyzers are put in. The answer is yes, until the new ones are fully validated and signed off.
Q: Will there be a temporary location to bring the new analyzers in to validate?A: I was not sure, we do not have that as a plan at this time and it depends on the analyzers that will be put in with the new contract.
Q: Were any existing staff present when the current analyzers were installed and how was that accomplishedt?A: Yes, we moved the current analyzers to a different location in the lab and installed the new ones in their permenant positions.
Q: Middleware to be used for rule building?A: The plan is that we will build the rules and autovalidation into Data Innovations not the vendor middleware. We use the vendor middleware for the function of the analyzer and track system.
Q: How is Remisol being used now? As a pass through?A: Pass through only.
Q: Is QC automatically uploading?A: Yes, BD. (I found out after they left that they never set it up but they do want it that way - lask of support with computer since we had other priorities for the Laboratory Information Manager/LIM.)
Q: Do we want the servers to be virtual or physical servers.A: We prefer virtual servers.
Q: Will out hematology analyzers/Sysmex be attached to the track system?A: No, the system is not the system by Sysmex that can be attached to a track system.
Q: What is the name of the vendor taking care of the water systems attached to the chemistry analyzers.A: Evoqua Water Technologies.
Q: Did we want storage on the track?A: Yes
Q: Where does that door go (from site visit)?A: Histology and they asked if they needed to provide access, I answered that yes that needs to be accessible.
Q: Can the pole be removed (pole with utilities)A: The drain can be moved but the pole may be acting as a vent/airflow and may not be removed.
Q: Can refrigerators be moved?A: Yes
Q: Do we have Autovalidation set upA: Not yet, but we plan to set it up.
Q: Are cabinets modular or fixed?A: They are modular, not fixed.
Q: Do we have dedicated LIS (Laboratory Information System) support?A: Yes, we have one Laboratory Information Manager/LIM and another one starting soon.
Q: Is the pneumatic tube system new?A: Yes.
Q: Any planned upgrades to the LIS?A: Yes, Cerner but we are not on the schedule yet.
Q: Any growth?A: We did take over Modesto, Sonora and Stockton and the workload will be going to the Stockton laboratory.
Q: Do we want to be able to load all the laboratory tests into the track, not just chemistry.A. We prefer to have the ability
Q: Could we switch to a new QC vendor and can we get a list of what QC you are using currently so they know how to meet out requirements.A: We are open to using other control vendors. Vendor QC must meet requirements in SOW.
Q: Does VHA need bracketing (tieing down of analyzers due to earthquate shaking danger)?A. VHA does require bracing for items 4 feet or taller such as shelving, medical equipment, or storage units, the VHA would require these items to be properly braced or anchored to prevent tipping during an earthquake.
Q: Is it the same space and no changes?A. Correct, no changes
Q: When is the plan to move to the temproary space?A: Last we were told CY 2026.
Q: Is the water system owned by Mather?A: No, by the current vendor.
Q: What vendor is maintaing them?A: Evoqua Water Technologies.
Q: The venodor rep asked about Stockton.A: Stockton will have a lab and will take over the workload for Modesto, Sonora and Stockton from Mather.
Q: What is the internal door width at Mather for the purpose of bringing equipment in?A. Both doors from the Lab to the main hallway are 40.5”

The rear door is 34.5” SO2+36C26124Q0782: page 19, section 3. Installation/Site Preparation.

Clarifications requested: Point 1, c Mather Laboratory states:

i. Mather will be renovated during the life of this contract.

ii. The contractor will be required to move equipment to a temporary location, and then to a final location.

iii. Revalidation is required at each movement.

Clarifications requested: We request details on the point i., ii., and iii. Please provide additional details on the expected location of this new laboratory. Also, if available, please provide any details (CAD, diagram, etc.) that depicts the new space and temporary location to assist the vendor with an understanding of the scale of this future move.

Clarifications requested: We ask for a proposed timeline to be defined for these moves.

Clarifications requested: We ask if vendor provided water and ancillary items are to be included in these moves. Provided CAD drawing of new space. Currently planned for the parking lot behind cafeteria adjacent to building 649 and 647. Currently projected for Calendar year 2026. Subject to changes.

Clarification:

Yes. Vendor should provide water and ancillary items are to be included in these moves.

Mather Test Volumes

TestsQuarterly QTYAnnual QTY
Access 25 OH Vit D DxI - Access 25 OH Vit D for use on DxI 2x50 DET6,74626,984
Access TSH (3rd IS) 2x100 Det9,80439,216
AccuTnI+3 DxI - AccuTnI+3 for use on DxI2,1598,636
ACTM - ACETAMINOPHEN 1X100 TEST KIT107428
ALB - CX ALBUMIN REAGENT 2 X 30019,04176,164
ALP - CX/LX ALP KIT 2 X 400 TEST18,25673,024
ALT - CX/LX ALT KIT 2 X 400 TEST18,92075,680
AMM - CX AMMONIA REAGENT 2X25 TEST100400
AMPH - CX AMPHETAMINE KIT 1X250 TESTS2,90711,628
AMY7 - G7 AMY 2 x 200 SYN Kit193772
AST - CX/LX AST KIT 2 X 400 TEST18,91075,640
B12 2X 50 DET - ACCESS B12 2 X 50 DET5,94223,768
BARB - BARBITURATE KIT (1X250 TESTS)2,1178,468
BNP2 - Access BNP1,5346,136
BNZG - Benzodiazepine Reagent2,1188,472
BUN - CX BUN REAGENT 2X300 TEST28,197112,788
CALC - Calcium28,437113,748
CAR - CX/LX CARBAMAZEPINE KIT 200T416
CAR - CX/LX CARBAMAZEPINE KIT 200T REF26104
CEA 2X50 TESTS - ACCESS CEA 2X50 TESTS158632
CHOL - CX CHOLESTEROL 2 X 300 TEST12,22748,908
CK - CX/LX CK KIT 2 X 400 TST8653,460
CK-MB REAGENT KIT1768
CL - Chloride28,316113,264
CO2A - LX C02 ACID REAGENT 2X2L28,320113,280
COCM - COCAINE METABOLITE KIT(1X250 TESTS2,90811,632
Cortisol 2X50 DET - ACCESS CORTISOL3971,588
CRPH - SYNCHRON HIGH SENSTVTY CRP 2X200 TST1,4335,732
CR-S - CX4 CREATININE KIT 2X30036,672146,688
DBIL - CX/LX DBIL KIT 2 X 300 TEST6,82027,280
Digoxin 2X50 DET - ACCESS DIGOXIN79316
ETOH - ETOH KIT 2X150 TEST2,76611,064
FE - CX/LX IRON FERROUS KIT (2X200)2,57710,308
Ferritin 2 X 50 - ACCESS FERRITIN 2 X 50 DET3,07212,288
Folate 2x50 Determination - Access Folate 2x50 Determination2,4019,604
Free T4, 2X50 DET - ACCESS FREE T4, 2X50 DET2,66210,648
GEN - CX/LX GENTAMICIN 2 X 100 TEST1664
Gentian Cystatin C - 1100 Gentian Cystatin C reagent31124
GLUCM - LX GLUCOSE 2X2L28,230112,920
HDL - *RF*KIT, HDL 2 x 200 DXC12,16048,640
Hybritech PSA - ACCESS HYBRITECH PSA RGT5,44921,796
IG-A - LX IMMUNOGLOBULIN A (IG-A) RGNT197788
IG-G - LX IMMUNOGLOBULIN G (IG-G) RGNT231924
IG-M - LX IMMUNOGLOBULIN M (IG-M) REAGENT173692
K - Potassium28,516114,064
LACT - LACT 2 X 50 DXC KIT1,0034,012
LD-L - VR* CX/LX LD 2 X 300 TEST2911,164
LDLD - KIT, LDLD 2X100 CX/LX3,00112,004
LIP - CX/LX LIP KIT 2 X 60 TEST1,2765,104
Lithium - Lithium 2 x 65 Test Kit140560
MAB - MICROALBUMIN 2X100 TEST KIT3,74614,984
METD - METD KIT (1X250 TESTS)2,1178,468
MG - CX MAGNESIUM 2X100 TEST7,58630,344
M-TP - CX MICROPROTEIN REAGENT 2X50 TEST8383,352
Myoglobin - ACCESS MYOGLOBIN, 2X50 DET312
NA - Sodium28,574114,296
OP - OPIATE KIT (1X250 TESTS)2,90711,628
Oxycodone UDR - 10012653 Oxycodone UDR reagent2,87411,496
PAB - PREALBUMIN 2X100 TEST KIT48192
PHE - CX/LX PHENOBARBITAL KIT 200TST1560
PHS - Phosphorous Assay3,30913,236
PHY - CX/LX PHENYTOIN KIT 200 TEST26104
PTH 2 x 50 DET - ACCESS PTH 2 x 50 DET1,1794,716
SALY - KIT, SALY 2 X 45 CX/LX90360
TBIL - CX/LX TBIL KIT 2 X 400 TEST18,26373,052
Testosterone Reagent1,0084,032
TG - CX TRIGLYCERIDE REAGENT 2X300 TEST12,18448,736
THC5 - THC5 KIT (1X250 TESTS)2,90711,628
THE - CX/LX THEOPHYLLINE KIT, 200T1872
TOTAL B- hCG - ACCESS TOTAL B- hCG 2x50 Det93372
TP - CX TOTAL PROTEIN REAGENT 2X30019,03376,132
TRFN - LX TRANSFERRIN REAGENT2,56210,248
URIC - CX URIC ACID REAGENT 2 X 3008283,312
VANC - KIT, VANC 200 CX/LX2591,036
VPA - CX/LX VALPROIC ACID KT (2X100)152608
Hepatitis C2,2008,800
HIV 1/2 combo screen (4th Generation)2,2008,800
Hepatitis B2,1608,640
Hepatits B core antigen2,1608,640
Procalcitonin4201,680
Syphilis1,0204,080
Hepatitis B surface antigen1,9507,800
Tacrolimus5402,160
Hepatitis A5222,088
Hepatitis A IgM832
Troponin I2,4009,600
Sex binding Hormone92368
Alpha Feta protein8843,537
Prostate-Specific Antigen, free32126
Grand Total488,5412,020,515

Mather Bio-Rad QC

VA Mather(YTD=Year to Date)
QC Orders through Bio-Rad
as of 8/13/2024
2024
Part#DescriptionYTD20232022
146CARD MKR LT PLUS 1 LIQ 6X3ML162016
148CARD MKR LT PLUS 3 LIQ 6X3ML162016
361IA PLUS LIQ 1 12X5ML101010
362IA PLUS LIQ 2 12X5ML101010
363IA PLUS LIQ 3 12X5ML101010
364SPECIALTY IA LIQ 1 6X5ML5157
365SPECIALTY IA LIQ 2 6X5ML5157
366SPECIALTY IA LIQ 3 6X5ML5157
397URINE CHEMISTRY 1 LIQ 12X10ML104
398URINE CHEMISTRY 2 LIQ 12X10ML104
697MULTIQUAL UNASSY 1 LIQ 12X10ML102010
699MULTIQUAL UNASSY 3 LIQ 12X10ML102010
740DIABETES BI LYPH 6X0.5ML81211
740XDIABETES LYPH SP1
751SPINAL FLUID 1 LIQ 6X3ML82
752SPINAL FLUID 2 LIQ 6X3ML82

Martinez

VANCHCS - Martinez
QuestionsAnswers
Q: The vendor representatives asked for us to breakdown the test list to individual tests not groups.A. See applicable tab for a breakdown of test volumes
Q: Do we need bracking?VHA does require bracing for items 4 feet or taller such as shelving, medical equipment, or storage units, the VHA would require these items to be properly braced or anchored to prevent tipping during an earthquake.
Q: Need the Stockton floor planA: Please see attached document
Q: Any upcoming changes to the existing floor plan?A: None planned at this site.
Q: Where is the water system?A. Currently installed in adjacent storage room.
Q: Will we have a staging location for the new analyzers during validation?A: No, we do not have a staging area.
Q: Please provide door width the analyzers will be moved through?A: 42 inches
Q: What is our middleware/interface?A: Data Innovations
Q: Will we do autovalidation - workflow and designA: Yes, not sure when

Martinez Test Volumes

TestsQuarterly QTYAnnual QTY
Access 25 OH Vit D DxI2,73610,944
Access hsTNI 2X50 - ACCESS hsTNI30120
Access TSH (3rd IS) 2x100 Det3,30013,200
AccuTnI+3 DxI - AccuTnI+361244
ACTM - ACETAMINOPHEN 1X100 TEST KIT1040
ALB - CX ALBUMIN REAGENT 2 X 3005,37021,480
ALP - CX/LX ALP KIT 2 X 400 TEST5,17620,704
ALT - CX/LX ALT KIT 2 X 400 TEST5,87423,496
AMM - CX AMMONIA REAGENT 2X25 TEST62248
AMPH - CX AMPHETAMINE KIT 1X250 TESTS7162,864
AMY7 - G7 AMY 2 x 200 SYN Kit189756
AST - CX/LX AST KIT 2 X 400 TEST5,69422,776
B12 2 X 50 DET - ACCESS B12 2 X 50 DET2,56910,276
BARB - BARBITURATE KIT (1X250 TESTS)6322,528
BNP2 - Access131524
BNZG - Benzodiazepine Reagent6322,528
BUN - CX BUN REAGENT 2X300 TEST6,82327,292
CALC - Calcium6,84127,364
CAR - CX/LX CARBAMAZEPINE KIT 200T1248
CHOL - CX CHOLESTEROL 2 X 300 TEST4,22616,904
CK - CX/LX CK KIT 2 X 400 TST3101,240
CL - Chloride6,85227,408
CO2A - LX C02 ACID REAGENT 2X2L6,86127,444
COCM - COCAINE METABOLITE KIT7162,864
CRPH - SYNCHRON HIGH SENSTVTY CRP3131,252
CR-S - CX4 CREATININE KIT 2X3009,95239,808
DBIL - CX/LX DBIL KIT 2 X 300 TEST9873,948
ETOH - ETOH KIT 2X150 TEST7132,852
FE - CX/LX IRON FERROUS KIT (2X200)9813,924
Ferritin 2 X 50 - ACCESS FERRITIN1,0284,112
Folate 2x504171,668
Free T4, 2X50 DET - ACCESS FREE T48903,560
GLUCM - LX GLUCOSE 2X2L6,83827,352
HDL - *RF*KIT, HDL 2 x 200 DXC4,20016,800
HFSH 2 X 50 DET - ACCESS HFSH 2 X 50 D2621,048
HLH 2 X 50 DET - ACCESS HLH 2 X 50 DET3261,304
Hybritech PSA - ACCESS HYBRITECH PSA RGT1,8397,356
K - Potassium6,93827,752
LACT - LACT 2 X 50 DXC KIT74296
LD-L - VR* CX/LX LD 2 X 300 TEST137548
LDLD - KIT, LDLD 2X100 CX/LX6652,660
LIP - CX/LX LIP KIT 2 X 60 TEST140560
MAB - MICROALBUMIN 2X100 TEST KIT1,8237,292
METD - METD KIT (1X250 TESTS)6322,528
MG - CX MAGNESIUM 2X100 TEST1,2795,116
M-TP - CX MICROPROTEIN REAGENT 2X502621,048
NA - Sodium6,92327,692
OP - OPIATE KIT (1X250 TESTS)7162,864
Oxycodone UDR7342,936
PCP - PCP KIT (1X250 TESTS)1040
PHE - CX/LX PHENOBARBITAL KIT 200TST1144
PHS - Phosphorous Assay8153,260
PHY - CX/LX PHENYTOIN KIT 200 TEST1560
Prolactin - ACCESS PROLACTIN 2X50 DET3321,328
PROX - PROX KIT (1X250 TESTS)1040
TBIL - CX/LX TBIL KIT 2 X 400 TEST5,17520,700
Testosterone Reagent3471,388
TG - CX TRIGLYCERIDE REAGENT 2X300 TEST4,20316,812
THC5 - THC5 KIT (1X250 TESTS)7162,864
THE - CX/LX THEOPHYLLINE KIT, 200T1040
TP - CX TOTAL PROTEIN REAGENT 2X3005,17020,680
TRFN - LX TRANSFERRIN REAGENT9773,908
URIC - CX URIC ACID REAGENT 2 X 3006772,708
VANC - KIT, VANC 200 CX/LX2288
VPA - CX/LX VALPROIC ACID KT (2X100)43172
Grand Total133,425533,700

Martinez Bio-Rad QC

VA Martinez(YTD=Year to Date)
QC Orders through Bio-Rad
as of 8/13/2024
2024
Part#DescriptionYTD20232022
12007028qUAntify Advance Liq 6X12ML6818
12016447CARDIAC ADVANCE LIQ 1 6X3ML4
12016448CARDIAC ADVANCE LIQ 2 6X3ML4
12016449CARDIAC ADVANCE LIQ 3 6X3ML4
12016450CARDIAC ADVANCE LIQ 4 6X3ML4
146CARD MKR LT PLUS 1 LIQ 6X3ML9149
148CARD MKR LT PLUS 3 LIQ 6X3ML9149
361IA PLUS LIQ 1 12X5ML599
362IA PLUS LIQ 2 12X5ML599
363IA PLUS LIQ 3 12X5ML599
364SPECIALTY IA LIQ 1 6X5ML561
365SPECIALTY IA LIQ 2 6X5ML561
366SPECIALTY IA LIQ 3 6X5ML561
397URINE CHEMISTRY 1 LIQ 12X10ML36
398URINE CHEMISTRY 2 LIQ 12X10ML36
697MULTIQUAL UNASSY 1 LIQ 12X10ML756
698MULTIQUAL UNASSY 2 LIQ 12X10ML3
699MULTIQUAL UNASSY 3 LIQ 12X10ML756
751SPINAL FLUID 1 LIQ 6X3ML64
752SPINAL FLUID 2 LIQ 6X3ML64

Redding

VANCHCS - Redding
QuestionsAnswers
Q: How will we handle the transition from the old vendor to the new- is this a temporary space?A: We will bring them in the lab - we can position the new ones and move the current analyzers or the reverse.
Q: Would it be 2 Alinity Immunassay analyzers and 2 Alinity chemistry analyzers?A: Yes
Q: What is our Interface/Middleware?A: Data Innovations
Q: Any changes in the SOW for Redding?A: No. The wall cannot be removed and the egress route is stated in the SOW.
Q: Can we get the volumes by test not group?A. See applicable tab for a breakdown of test volumes
Q: Is the Vendor middleware a pass through?A: Yes
Q: Who maintains the current water systems?A: There is a sticker on the outside of the water system that says Evoqua Water Technologies

Redding Test Volumes

TestSum of QTY (quarter)Annual QTY
Access 25 OH Vit D D2,1418,564
Access TSH2,65810,632
ALB - CX ALBUMIN REAGENT 2 X 3004,48317,932
ALP - CX/LX ALP KIT 2 X 400 TEST4,41817,672
ALT - CX/LX ALT KIT 2 X 400 TEST4,57218,288
AMM - CX AMMONIA REAGENT 2X25 TEST38152
AMPH - CX AMPHETAMINE KIT 1X250 TESTS8423,368
AMY7 - G7 AMY 2 x 200 SYN Kit132528
AST - CX/LX AST KIT 2 X 400 TEST4,53118,124
B12 2 X 50 DET - ACCESS B12 2 X 50 DET1,4765,904
BARB - BARBITURATE KIT (1X250 TESTS)6982,792
BNP2 - Access199796
BNZG - Benzodiazepine Reagent6982,792
BUN - CX BUN REAGENT 2X300 TEST5,25221,008
CALC - Calcium5,38521,540
CHOL - CX CHOLESTEROL 2 X 300 TEST3,78815,152
CK - CX/LX CK KIT 2 X 400 TST193772
CL - Chloride5,34721,388
CO2A - LX C02 ACID REAGENT 2X2L5,33221,328
COCM - COCAINE METABOLITE KIT8433,372
CRPH - SYNCHRON HIGH SENSTVTY2871,148
CR-S - CX4 CREATININE KIT 2X3007,82231,288
DBIL - CX/LX DBIL KIT 2 X 300 TEST6142,456
ETOH - ETOH KIT 2X150 TEST7472,988
FE - CX/LX IRON FERROUS KIT (2X200)4971,988
Ferritin 2 X 50 - ACCESS FERRITIN4341,736
Folate 2x503181,272
Free T4, 2X50 DET - ACCESS FREE5362,144
GLUCM - LX GLUCOSE 2X2L7,13128,524
HDL - *RF*KIT, HDL 2 x 200 DXC3,77015,080
Hybritech PSA1,1744,696
K - Potassium7,31529,260
LACT - LACT 2 X 50 DXC KIT70280
LD-L - VR* CX/LX LD 2 X 300 TEST72288
LDLD - KIT, LDLD 2X100 CX/LX4121,648
LIP - CX/LX LIP KIT 2 X 60 TEST167668
MAB - MICROALBUMIN 2X100 TEST KIT1,4145,656
METD - METD KIT (1X250 TESTS)6982,792
MG - CX MAGNESIUM 2X100 TEST4921,968
M-TP - CX MICROPROTEIN REAGENT153612
NA - Sodium5,36621,464
OP - OPIATE KIT (1X250 TESTS)8433,372
Oxycodone UDR - 10012653 Oxycodone UDR8533,412
PHS - Phosphorous Assay247988
TBIL - CX/LX TBIL KIT 2 X 400 TEST4,41917,676
Testosterone Reagent3531,412
TG - CX TRIGLYCERIDE REAGENT 2X3003,78115,124
THC5 - THC5 KIT (1X250 TESTS)8433,372
TP - CX TOTAL PROTEIN REAGENT 2X3004,42517,700
TRFN - LX TRANSFERRIN REAGENT4931,972
URIC CX URIC ACID REAGENT228912
Grand Total109,000436,000

Redding Bio-Rad QC

VA Redding(YTD=Year to Date)
QC Orders through Bio-Rad
as of 8/13/2024
2024
Part#DescriptionYTD20232022
12007028qUAntify Advance Liq 6X12ML268
12016447CARDIAC ADVANCE LIQ 1 6X3ML10
12016448CARDIAC ADVANCE LIQ 2 6X3ML10
12016449CARDIAC ADVANCE LIQ 3 6X3ML10
12016450CARDIAC ADVANCE LIQ 4 6X3ML10
146CARD MKR LT PLUS 1 LIQ 6X3ML11108
148CARD MKR LT PLUS 3 LIQ 6X3ML11108
361IA PLUS LIQ 1 12X5ML844
362IA PLUS LIQ 2 12X5ML844
363IA PLUS LIQ 3 12X5ML844
364SPECIALTY IA LIQ 1 6X5ML57
365SPECIALTY IA LIQ 2 6X5ML57
366SPECIALTY IA LIQ 3 6X5ML57
397URINE CHEMISTRY 1 LIQ 12X10ML43
398URINE CHEMISTRY 2 LIQ 12X10ML43
697MULTIQUAL UNASSY 1 LIQ 12X10ML1224
699MULTIQUAL UNASSY 3 LIQ 12X10ML1224

Stockton

VANCHCS - Stockton
QuestionsAnswers
Q: Can the sink near the pass through be removed?A: Yes, the sink can be removed if needed.
Q: Could the neutralizer be removed? It may be in the way for the service techs for the analyzers.A: The neutralizer cannot be removed.
Q: Not sure if the plug for the 220 power is correct?A. Activation teams can support power requirments.
Q: Who will be responsible for the connection of the water system to the analyzer through the ceiling?A: Engineering can support putting the tubing through the ceiling from the water systems where it will be required and is the current set up (Stockton, Mather, and Martinez).
Q: Vendor representatives asked for a breakdown of the test list to individual tests not groups.A. Stockton is a new facility. We are expecting the same volume as VAMC Martinez. Please See applicable tab for a breakdown of test volumes.
Q: They ask if the placement is side by side.A: Yes
Can a CAD drawing and floor plan for Stockton be provided?A. Yes, see Attached

Stockton Est. Test Volumes

TestsQuarterly QTYAnnual QTY
Access 25 OH Vit D DxI2,73610,944
Access hsTNI 2X50 - ACCESS hsTNI30120
Access TSH (3rd IS) 2x100 Det3,30013,200
AccuTnI+3 DxI - AccuTnI+361244
ACTM - ACETAMINOPHEN 1X100 TEST KIT1040
ALB - CX ALBUMIN REAGENT 2 X 3005,37021,480
ALP - CX/LX ALP KIT 2 X 400 TEST5,17620,704
ALT - CX/LX ALT KIT 2 X 400 TEST5,87423,496
AMM - CX AMMONIA REAGENT 2X25 TEST62248
AMPH - CX AMPHETAMINE KIT 1X250 TESTS7162,864
AMY7 - G7 AMY 2 x 200 SYN Kit189756
AST - CX/LX AST KIT 2 X 400 TEST5,69422,776
B12 2 X 50 DET - ACCESS B12 2 X 50 DET2,56910,276
BARB - BARBITURATE KIT (1X250 TESTS)6322,528
BNP2 - Access131524
BNZG - Benzodiazepine Reagent6322,528
BUN - CX BUN REAGENT 2X300 TEST6,82327,292
CALC - Calcium6,84127,364
CAR - CX/LX CARBAMAZEPINE KIT 200T1248
CHOL - CX CHOLESTEROL 2 X 300 TEST4,22616,904
CK - CX/LX CK KIT 2 X 400 TST3101,240
CL - Chloride6,85227,408
CO2A - LX C02 ACID REAGENT 2X2L6,86127,444
COCM - COCAINE METABOLITE KIT7162,864
CRPH - SYNCHRON HIGH SENSTVTY CRP3131,252
CR-S - CX4 CREATININE KIT 2X3009,95239,808
DBIL - CX/LX DBIL KIT 2 X 300 TEST9873,948
ETOH - ETOH KIT 2X150 TEST7132,852
FE - CX/LX IRON FERROUS KIT (2X200)9813,924
Ferritin 2 X 50 - ACCESS FERRITIN1,0284,112
Folate 2x504171,668
Free T4, 2X50 DET - ACCESS FREE T48903,560
GLUCM - LX GLUCOSE 2X2L6,83827,352
HDL - *RF*KIT, HDL 2 x 200 DXC4,20016,800
HFSH 2 X 50 DET - ACCESS HFSH 2 X 50 D2621,048
HLH 2 X 50 DET - ACCESS HLH 2 X 50 DET3261,304
Hybritech PSA - ACCESS HYBRITECH PSA RGT1,8397,356
K - Potassium6,93827,752
LACT - LACT 2 X 50 DXC KIT74296
LD-L - VR* CX/LX LD 2 X 300 TEST137548
LDLD - KIT, LDLD 2X100 CX/LX6652,660
LIP - CX/LX LIP KIT 2 X 60 TEST140560
MAB - MICROALBUMIN 2X100 TEST KIT1,8237,292
METD - METD KIT (1X250 TESTS)6322,528
MG - CX MAGNESIUM 2X100 TEST1,2795,116
M-TP - CX MICROPROTEIN REAGENT 2X502621,048
NA - Sodium6,92327,692
OP - OPIATE KIT (1X250 TESTS)7162,864
Oxycodone UDR7342,936
PCP - PCP KIT (1X250 TESTS)1040
PHE - CX/LX PHENOBARBITAL KIT 200TST1144
PHS - Phosphorous Assay8153,260
PHY - CX/LX PHENYTOIN KIT 200 TEST1560
Prolactin - ACCESS PROLACTIN 2X50 DET3321,328
PROX - PROX KIT (1X250 TESTS)1040
TBIL - CX/LX TBIL KIT 2 X 400 TEST5,17520,700
Testosterone Reagent3471,388
TG - CX TRIGLYCERIDE REAGENT 2X300 TEST4,20316,812
THC5 - THC5 KIT (1X250 TESTS)7162,864
THE - CX/LX THEOPHYLLINE KIT, 200T1040
TP - CX TOTAL PROTEIN REAGENT 2X3005,17020,680
TRFN - LX TRANSFERRIN REAGENT9773,908
URIC - CX URIC ACID REAGENT 2 X 3006772,708
VANC - KIT, VANC 200 CX/LX2288
VPA - CX/LX VALPROIC ACID KT (2X100)43172
Grand Total133,425533,700

File details come from the government source that posted it. Updated .