36C26123Q1120 Amendment 0001.docx
DOCX document 120 KB Posted
- Attached to
- 6540--RFQ 36C26123Q1120 | Haag-Streit IM910 and Reliance Chair Equipment for Stockton CBOC
- Solicitation number
- 36C26123Q1120
About this file
This is an amendment to a request for quotation issued by the Department of Veterans Affairs for ophthalmic equipment including a Haag-Streit IM910 digital video package, Reliance examination chairs and chair glides, and Reliance instrument stands. The equipment will be delivered to the Stockton Community Based Outpatient Clinic. The amendment clarifies delivery addresses and points of contact, adds required attachments, and revises attachment names referenced in the documents. Offerors must acknowledge receipt of the amendment by the response deadline of August 18, 2023. Pricing is to be provided for the equipment items with delivery within 60 days of award, which is expected from September 1, 2023 through October 31, 2023.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment Eye Exam Chair Glide_Product Data.pdf | ||
| Attachment Eye Exam Chair_Product Data.pdf | ||
| Attachment Instrument Stands_Product Data.pdf | ||
| Attachment Chair Stand Locations.pdf | ||
| 36C26123Q1120_1.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
5. PROJECT NUMBER (if applicable)
CODE
7. ADMINISTERED BY
2. AMENDMENT/MODIFICATION NUMBER
CODE
6. ISSUED BY
8. NAME AND ADDRESS OF CONTRACTOR
4. REQUISITION/PURCHASE REQ. NUMBER
3. EFFECTIVE DATE
9A. AMENDMENT OF SOLICITATION NUMBER
9B. DATED
PAGE OF PAGES
10A. MODIFICATION OF CONTRACT/ORDER NUMBER
10B. DATED
BPA NO.
1. CONTRACT ID CODE
FACILITY CODE
CODE
Offers must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended, by one of the following methods:
The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offers
E. IMPORTANT:
is extended,
(a) By completing Items 8 and 15, and returning __________ copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted; or (c) By separate letter or electronic communication which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY is not extended.
12. ACCOUNTING AND APPROPRIATION DATA
(REV. 11/2016)
is required to sign this document and return ___________ copies to the issuing office.
is not, A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE CONTRACT ORDER NO. IN ITEM 10A.
15C. DATE SIGNED
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(b).
RESULT IN REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by letter or electronic communication, provided each letter or electronic communication makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER
Contractor
16C. DATE SIGNED
14. DESCRIPTION OF AMENDMENT/MODIFICATION
16B. UNITED STATES OF AMERICA
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER
16A. NAME AND TITLE OF CONTRACTING OFFICER
15B. CONTRACTOR/OFFEROR
STANDARD FORM 30
PREVIOUS EDITION NOT USABLE
Prescribed by GSA - FAR (48 CFR) 53.243 (Type or print) (Type or print) (Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
(Number, street, county, State and ZIP Code) (If other than Item 6) (Specify type of modification and authority) (such as changes in paying office, appropriation date, etc.)
(If required)
(SEE ITEM 11)
(SEE ITEM 13)
(X)
CHECK
ONE
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS,
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
(Signature of person authorized to sign) (Signature of Contracting Officer)
612-23-3-6214-0264
VA-23-00083579
00261 Department of Veterans Affairs Network Contracting Office 21 VA Southern Nevada Healthcare System 6900 N. Pecos Road, Building 6 North Las Vegas NV 89086
00261 Department of Veterans Affairs Network Contracting Office 21 Southern Nevada Healthcare System 6900 N. Pecos Road, Building 6 North Las Vegas NV 89086 To all Offerors/Bidders
36C26123Q1120 08-18-2023
X
X
X
See CONTINUATION Page
A. The purpose of this Amendment is to correct the following items from the initial solicitation RFQ 36C26123Q1120:
1. Add delivery address and POC to DELIVERY section.
2. Add Attachments mentioned in Salient Characteristic documents.
3. Revise Attachment names on the Salient Characteristic documents.
4. Removed product item 6.
All other terms and conditions are in full force and effect.
-------------------------------------------------END OF SF 30----------------------------------------------------------
Mandy Thompson Contracting Officer
CONTINUATION PAGE
A.1 PRICE/COST SCHEDULE
ITEM INFORMATION
| ITEM NUMBER |
| DESCRIPTION OF SUPPLIES/SERVICES |
| QUANTITY |
| UNIT |
| UNIT PRICE |
| AMOUNT |
| 1.00 |
| EA |
| __________________ |
| __________________ |
SERVICE/LABOR CAL COAST INSTALLATION & SETUP OF EQUIPMENT. FREIGHT IS ADDITIONAL.
Contract Period: Base POP Begin: 09-01-2023 POP End: 10-31-2023 LOCAL STOCK NUMBER: Labor
| 3.00 |
| EA |
| __________________ |
| __________________ |
HAAG STREIT IMAGING MODULE IM910,RM-03,PC
POP Begin: 09-01-2023
LOCAL STOCK NUMBER: SLHS02-501
| 10.00 |
| EA |
| __________________ |
| __________________ |
RELIANCE 980L CHAIR- COLOR: 634 PUMICE
POP Begin: 09-01-2023
LOCAL STOCK NUMBER: CSRE05-001SC
| 10.00 |
| EA |
| __________________ |
| __________________ |
940 RELIANCE CHAIR GLIDE
POP Begin: 09-01-2023
LOCAL STOCK NUMBER: CSRE15-000
| 10.00 |
| EA |
| __________________ |
| __________________ |
7900IC-WC RELIANCE STAND
POP Begin: 09-01-2023
LOCAL STOCK NUMBER: CSRE20-001
| GRAND TOTAL |
| __________________ |
A.2 DELIVERY SCHEDULE
| ITEM NUMBER |
| QUANTITY |
| DELIVERY DATE |
| 0001 |
| SHIP TO: |
| Department of Veteran Affairs Stockton CBOC |
6505 S. Manthey Rd French Camp, CA 95231
USA
| 1.00 |
| 60 days ARO |
| MARK FOR: |
| Jennifer Ledward |
916-843-7550 jennifer.ledward@va.gov
| 0002 |
| SHIP TO: |
| Department of Veteran Affairs Stockton CBOC |
6505 S. Manthey Rd French Camp, CA 95231
USA
| 3.00 |
| 60 Days ARO |
| MARK FOR: |
| Jennifer Ledward |
| 0003 |
| SHIP TO: |
| Department of Veteran Affairs Stockton CBOC |
6505 S. Manthey Rd French Camp, CA 95231
USA
| 10.00 |
| 60 days ARO |
| MARK FOR: |
| Jennifer Ledward |
| 0004 |
| SHIP TO: |
| Department of Veteran Affairs Stockton CBOC |
6505 S. Manthey Rd French Camp, CA 95231
USA
| 10.00 |
| 60 days ARO |
| MARK FOR: |
| Jennifer Ledward |
| 0005 |
| SHIP TO: |
| Department of Veteran Affairs Stockton CBOC |
6505 S. Manthey Rd French Camp, CA 95231
USA
| 10.00 |
| 60 days ARO |
| MARK FOR: |
| Jennifer Ledward |
Haag-Streit IM910 Digital Video Package for Haag-Streit BQ900 LED Slit Lamp or Equal
1. General The IM910 is a digital video package directly connected to the Haag-Streit BQ900 slit lamp. It encompasses a camera, a computer, and software which allows capturing optic images for storage.
| 2. | Warranty Requirements |
| a. | Manufacturer must offer a one year warranty that product is free from defects in materials and workmanship. All defects will be replaced or repaired by the manufacturer at no cost to the buyer. |
3. Distinct Features
a. Can store eye images directly into the EyeSuite software.
b. By the turn of a knob, IM910 can share eye view within microscope directly onto a monitor screen instantly.
c. Capture eye images with the camera trigger button.
d. It has built-in sensor for the camera.
e. A performant auto-exposure mode in combination with the automatic aperture control.
4. Technical Specifications (minimum requirements)
a. Compatibility and fully integrated with: Haag-Streit BQ900 slit lamps.
| b. | Beam path: right side. |
| c. | Control panel: RM03. |
| d. | There is a depth of focus control. |
| e. | Camera interface: CoaXPress. |
| f. | Shutter system: Global Shutter. |
| g. | Pixel size: 3.45 µm. |
| h. | Sensor size: 2/3”. |
| i. | Resolution: 2448 x 2048. |
| j. | Frame rate: 30 (frames per second). |
| k. | Software: EyeSuite Imaging (optional). |
| l. | EMR interfaces: EyeSuite script language, DICOM (SCU) GDT, Samba Share for direct import into patient files. |
| m. | There is a connector box. |
| n. | Connector box connectivity: |
| a. | HDMI out. |
| b. | DisplayPort out. |
| c. | Ethernet. |
| 5. | Delivery Requirements |
| a. | All the items shall be delivered and installed at their respected locations considering existing constraints. Install per approved installation shop drawings. |
| b. | Deliver all parts/ unit at once from factory to project site. No split deliveries will be accepted. |
| 6. | Care and Maintenance Requirements |
| a. | Follow manufacturer’s care and maintenance instructions. |
| 7. | Locations |
| a. | Consult SP 3A112S |
| b. | Exam/ Treatment 4A114B |
| c. | Eye Procedure 4A114E |
See attached Attachment Chair Stand Locations plan showing the locations of Haag Streit IM910 Digital Video Package for Haag-Streit BQ900 LED slit lamps in the CBOC building.
Page 1 of
Reliance/ Haag-Streit 940 Chair Glide or Equal
1. General Chair Glide with Tapered Track is especially designed for Reliance/ Haag-Streit 980 Premium Full Power Exam / Procedure Chair. It is an equipment gear consisting of a base plate, a stainless steel track, and a carriage. The exam chair sits on this chair glide, allowing it to move back and forth thereby accommodating the movement of a wheelchair.
| 2. | Warranty Requirements |
| a. | Manufacturer must offer a one year warranty that product is free from defects in materials and workmanship. All defects will be replaced or repaired by the manufacture at no cost to the buyer. |
3. Distinct Features
f. Heavy-duty stops on base plate limit distance of travel.
g. Carriage to be heavy-duty, cast aluminum.
h. Tapered track improves access to patient.
i. Locking pedal secures chair in any position along its 27” path.
j. Case-hardened steel rollers allow to easily slide chair out of the way for wheelchair access.
4. Technical Specifications (minimum requirements)
b. Base plate length: 52” overall; wide section: 28”; narrow: 24” (max. dimensions).
| b. | The base plate is centered on the plate. A track guides the carriage and prevents lateral movement. |
| c. | The carriage is made from heavy-duty cast aluminum. It fits all Reliance/ Haag Streit exam chairs. |
| d. | Base plate width: wide section: 20”; narrow: 6” (max. dimensions). |
| e. | Track: ¾” wide by 3/16” high by 52” long (max. dimensions). |
| f. | Overall carriage length: 29”; width: 26”. |
| g. | Chair Glide must be compatible with Reliance/ Haag Streit 980 exam chair. |
| h. | Complies with the mandatory requirements of the Americans with Disabilities Act. |
| 5. | Delivery Requirements |
| a. | All the items shall be delivered and installed at their respected locations considering existing constraints. Install per approved installation shop drawings. |
| b. | Deliver all parts/ unit at once from factory to project site. No split deliveries will be accepted. |
| 6. | Care and Maintenance Requirements |
| a. | Follow manufacturer’s care and maintenance instructions. |
| 7. | Locations |
| a. | Consult SP 3A112S |
| b. | Consult/ Exam SP 3A112R |
| c. | Consult/ Exam SP 3A112N |
| d. | Consult/ Exam SP 3A112M |
| e. | Consult/ Exam SP 3A112L |
| f. | Exam/ Treatment 4A114B |
| g. | Exam/ Treatment 4A114C |
| h. | Exam/ Treatment 4A114L |
| i. | Eye Procedure 4A114E |
| j. | Low Vision Exam 4A114F |
See attached Attachment Chair Stand Locations plan showing the locations of Reliance/ Haag Streit 940 chair glides in the CBOC building.
| 8. | Attachment |
| a. | Attachment Chair Stand Locations |
| b. | Attachment Eye Exam Chair Glide Product Data |
Reliance/ Haag-Streit 980 Premium Full-Power Exam/ Procedure Chair or Equal
1. General Full-power articulating round-base examination chair.
| 2. | Warranty Requirements |
| a. | Manufacturer must offer a one year warranty that product is free from defects in materials and workmanship. All defects will be replaced or repaired by the manufacturer at no cost to the buyer. |
3. Distinct Features
k. Memory programmable function automatically positions chair to preferred exam position.
l. Programmable safety function prevents children from operating chair.
m. Dual side controls.
n. Rotation locks allow you to swivel chair 330 degrees from both sides.
o. Armrests lock in place and release for easy patient access and security.
p. Auto return button returns chair to upright and lowered position.
q. Large hinged footrest supports patient’s full weight for safe ingress and egress.
r. Detached footswitch allows for hands-free patient positioning.
s. Headrest adjusts with a single hand and locks into place.
t. Finger switches and a corded footswitch that raises, lowers or reclines the chair and auto return.
u. Backlit membrane switches.
v. Long lasting hand-sewn upholstery with high-density foam.
w. Removable cushions for easy cleaning or replacement.
x. Cast iron bases and hinges.
4. Technical Specifications (minimum requirements)
c. Depth of chair (without headrest) table position: 60”, Seat: 18¾” wide, 18¼”deep (max. dimension).
| b. | Two base heights. HIGH: 23-5/8” lowered, 34 ¼” raised. LOW 21-1/8” lowered, 29-5/8” raised. |
| c. | Diameter: 24 ¼”. |
| d. | Distance between armrests: 19”. |
| e. | Shipping Weight: 325 lbs. |
| f. | Standard electric: 120 volts, 50/60Hz, 6.3 amperes total. Optional electric: 230 volts, 50 Hz, 4.0 amperes. |
| g. | Color: TBD during technical evaluation. |
| h. | Canadian Standards Association (CSA, C/US) certificate compliance. |
| i. | Chair must be compatible with Reliance/ Haag-Streit 940 Chair Glide for mandatory requirements of the Americans with Disabilities Act. |
| 5. | Delivery Requirements |
| a. | All the items shall be delivered and installed at their respected locations considering existing constraints. Install per approved installation shop drawings. |
| b. | Deliver all parts/ unit at once from factory to project site. No split deliveries will be accepted. |
| 6. | Care and Maintenance Requirements |
| a. | Follow manufacturer’s care and maintenance instructions. |
| 7. | Locations |
| a. | Consult SP 3A112S |
| b. | Consult/ Exam SP 3A112R |
| c. | Consult/ Exam SP 3A112N |
| d. | Consult/ Exam SP 3A112M |
| e. | Consult/ Exam SP 3A112L |
| f. | Exam/ Treatment 4A114B |
| g. | Exam/ Treatment 4A114C |
| h. | Exam/ Treatment 4A114L |
| i. | Eye Procedure 4A114E |
| j. | Low Vision Exam 4A114F |
See attached Attachment Chair Stand Locations plan showing the locations of Reliance/ Haag Streit 980 premium full-power exam/ procedure chairs in the CBOC building.
| 8. | Attachment |
| a. | Attachment Chair Stand Locations |
| b. | Attachment Eye Exam Chair Product Data |
Reliance/ Haag-Streit 7900IC Ophthalmic Instrument Stand or Equal
1. General The Reliance 7900IC is an ophthalmic instrument stand with rechargeable wells for handheld instruments and multiple arms for mounting slit lamps and phoropters.
| 2. | Warranty Requirements |
| a. | Manufacturer must offer a one year warranty that product is free from defects in materials and workmanship. All defects will be replaced or repaired by the manufacturer at no cost to the buyer. |
3. Distinct Features
y. The stand has exam chair controls to adjust chair’s heights located on the slit lamp arm.
z. The stand has (3) rechargeable wells for handheld instruments.
aa. The stand has following arms (at a minimum):
a. Counterbalanced slit-lamp arm: allow to lock and adjust the arm with a single hand. Arm can travel 12’’ in vertical direction, and can be counterbalanced for instruments weighing12 to 60 lbs.
b. Upper refractor arm: offers 10’’ range of vertical movement, and can support instruments up to 20 lbs. This arm must be able to support and compatible with Reichert Technologies Ophthalmic Instruments or equal’s phoropter that will be attached to this arm.
c. Third instrument arm available as an option: it has capacity to support 10- to 50 lb. instruments, and has the capability to travel 10’’ in vertical direction.
d. Monitor Arm: Instrument stand supports an additional monitor arm.
ab. Durability features:
a. Made from durable base materials, including polycarbonate, to be able to resist scrapes, scuffs and scratches.
b. The stand has backlit membrane switches. They control the chart projector, overhead lamp, and binding post for an indirect ophthalamoscope, and raise and lower the exam chair.
c. The stand has halogen lamp on the top of the stand that provides bright, direct light for minor procedures. The halogen lamp is supported on an arm that can easily be adjusted with hand without the use of any tools. The halogen lamp has low voltage for safety.
ac. Options:
a. An optional lower slit lamp arm to serve wheelchair patients.
b. Has an eclipse option to activate all room lighting from the stand. Lighting adjusts based on instrument moves or quick manual adjustments.
i. Verify with the end users if this feature needs to be included during technical evaluation period.
4. Technical Specifications (minimum requirements)
d. Height to top console: 36”.
| b. | Height to top pole: 71”. |
| c. | Base: 21”x 21’’ (max dimension). |
| d. | Shipping weight: 292 lbs. (approximate) |
| e. | Electrical (Standard UL Listed), 115 volts, 60 Hz, 3 amperes. |
| f. | Complies with following safety standard: |
| a. | Conforms to CE standard. |
| b. | CSA, C/US tested, equivalent to UL 60601-01. |
| g. | Must be compatible with Reliance/ Haag Streit 980 exam chairs so the control switch features on the stand can function as they are supposed to. |
| h. | The rechargeable wells must be compatible with Hillrom-Welch Allyn or equal’s hand-held instruments. |
| 5. | Delivery Requirements |
| a. | All the items shall be delivered and installed at their respected locations considering existing constraints. Install per approved installation shop drawings. |
| b. | Deliver all parts/ unit at once from factory to project site. No split deliveries will be accepted. |
| 6. | Care and Maintenance Requirements |
| a. | Follow manufacturer’s care and maintenance instructions. |
| 7. | Locations |
| a. | Consult SP 3A112S |
| b. | Consult/ Exam SP 3A112R |
| c. | Consult/ Exam SP 3A112N |
| d. | Consult/ Exam SP 3A112M |
| e. | Consult/ Exam SP 3A112L |
| f. | Exam/ Treatment 4A114B |
| g. | Exam/ Treatment 4A114C |
| h. | Exam/ Treatment 4A114L |
| i. | Eye Procedure 4A114E |
| j. | Low Vision Exam 4A114F |
See attached Attachment Chair Stand Locations plan showing the locations of Reliance/ Haag Streit 7900IC Ophthalmic Instrument Stands in the CBOC building.
| 8. | Attachment |
| a. | Attachment Chair Stand Locations |
| b. | Attachment Instrument Stands Product Data |
ATTACHMENT LIST
See attached document: Attachment Chair Stand Locations.
See attached document: Attachment Eye Exam Chair Glide_Product Data.
See attached document: Attachment Eye Exam Chair_Product Data.
See attached document: Attachment Instrument Stands_Product Data.
The Amendment to the solicitation RFQ 36C26123Q1120 corrects the following items:
1. Add delivery address and POC to DELIVERY section.
2. Add Attachments mentioned in Salient Characteristic documents.
3. Revise Attachment names on the Salient Characteristic documents.
4. Removed product item 6.
All other terms and conditions are in full force and effect.
File details come from the government source that posted it. Updated .