Brand Name or Equal Worksheet (“Salient Characteristics”) v 2.0
Brand Name Item: California P200DTXICG with color, AF, FA, ICGA OPTOS
1) Essential/significant physical, functional, or performance characteristics.
• Single image must be able to capture 82% of the retina
• Single capture must be able to cover 200 degrees of the internal surface of the retina
• Single capture image must be able to reach beyond the equator showing vortex vessels and ampullae
• Ultra-widefield (200 degree) image must be able to be captured in less than ½ second
• Montage image must be able to show up to 97% of the retina (1166 mm^2)
• Ellipsoidal mirror with virtual point technology must enable maximum imaging field of view
• Must have clinically validated color imaging – must be proven to be equivalent to color fundus photography for pathology detection
• Must provide multiple wavelength color imaging: color, red-free, choroidal, and true-color views with a single capture
• Must present the images in a consistent geometry that accurately represents anatomical features across the retina and that allows for accurate measurements across the retina
• Must have color imaging with red/green lasers to minimize exposure to blue light
• Must have auto capture capabilities
• Must be non-mydriatic, cSLO technology images through 2 mm pupil
• Must include the following light source wavelengths: red 635 nm, green 532 nm, blue 488 nm, IR 802 nm
• Must include DICOM and be DICOM compliant
• Must be on the VistA imaging approved DICOM list
• Must be able to perform fundus autofluorescence (AF) with green laser to minimize exposure to blue light
• Must be able to perform fluorescein angiography (FA)
• Must be able to perform indocyanine green angiography (ICGA)
• Must include the following functions in one system: color photo, red-free photo, choroidal imaging, AF imaging, FA, and ICGA
• Must include alignment software to ensure that pictures are taken in the same place each time and that images in the same place taken w/different modalities and/or on different dates can be accurately overlaid for clinical comparison
• Must include image overlay software for accurate comparison of images from different days and in different modes
• Must be HIPAA compliant
Brand Name or Equal Worksheet (“Salient Characteristics”) v 2.0
• Must include image measurement tools
• Must be capable of combined FA and ICGA capture
• Must be able to connect to 3rd party image management software
• Must include an automatic, wheelchair accessible table
• Must be powered via connection to a standard AC outlet (100 to 240V)
• Maximum weight 75 lbs
• At least two-year warranty
• Must include installation and training
2) Complete generic identification – Fundus Camera
3) Applicable model/make/catalog number – California P200DTXICG with color, AF, FA, ICGA OPTOS, see quote #Q804246
4) Manufacturer name –
OPTOS
Per VAAR 811.104 and 811.105, I certify the following regarding the above information:
• The salient characteristics listed above are essential to the Government’s requirements.
• To my knowledge, the restrictions listed above will not limit the acquisition to an item peculiar to one manufacture.