About this file

36C26118Q9212 Salient Characteristics.docx

View the file

Other files

File Type Posted
36C26118Q9212-000.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Brand Name or Equal Worksheet (“Salient Characteristics”) Brand Name Item: Helmer iB125 Blood Blank Refrigerator

1) Essential/significant physical, functional, or performance characteristics.

· Must be monitored with an information center

· Must constantly monitor the chamber temperature and critical refrigerator functions, safeguarding the products inside and providing valuable information to the users

· Must have a monitor for easy access to all functions and features

· Must have a color touchscreen

· Requires Automatic high/low alarm testing (Peltier based) for quality control

· For operation safety, it must be Password protected configuration

· Must have a lockable door

· Must have dual alarm probes and dual temperature probes

· Must have event log records door openings and alarm conditions with date and time for inspection requirement

· Dry contact alarm connection

· Must have USB ports for download/upload/update and event log storage

· Must have a rechargeable battery back up

· Must be able to mute alarm with adjustable ring-back interval

· Displays:

a) Graph display

b) Chart alert display

c) Temperature graph display and digital display of upper and lower chamber temperatures

d) Date and time display

e) Displays compressor temperature performance in compressor logs

· Must have a visual/audible alarm system that alarms for:

a) Low battery

b) High and low temperatures

c) Door ajar

d) Power failure

e) Condenser temperature

· Alarm volume must be adjustable with different tones

· Must be able to hold 504 blood bags

· Interior Dimensions (w x h x d)- 24.75’’ x 58.25’’ x 30.25’’

· Exterior Dimensions (w x h x d)– 29.5”x 79.5’’x 35.5’’

· Interior Volume – 25.5 cu ft / 714 liters

· Weight – 559 LBS.

· Electrical Voltage: 115V 60Hz 7.5FLA

230V50Hz 4.2FLA
230V60Hz 4.2FLA

· Drawers (w x d): 7 solid stainless steel (22in x 27in)

· Door: One door with scratch-guard edge protectors

· Temperature range: 2-10 °C

· Interior, exterior and door handles: Bacteria resistant powder coated

· Drawer handles: Sure-Grip handler

· Lighting: Energy-efficient LED lighting

· Must have access port on top of cabinet for external monitoring probes (checkpoint)

· Must meet AABB, ARC, FDA, EU Blood Directive 2004/33/EC, and Part 4 of the Blood Safety and Quality regulation 2005 standards for blood storage at 4 °C

· Warranty: 7 years compressor, 2 years parts, and 1 year labor

2) Complete generic identification – Blood bank refrigerator

3) Applicable Model/Make - i.Series iB125/Blood Bank Refrigerator

4) Manufacturer name – Helmer Scientific

Regarding the above information:

•The salient characteristics listed above are essential to the Government’s requirements.
•To my knowledge, the restrictions listed above will not limit the acquisition to an item peculiar to onemanufacture.