This is a Sources Sought Notice issued by the Department of Veterans Affairs, Veterans Integrated Service Network 20, in support of the Mann-Grandstaff VA Medical Center in Spokane, Washington. The VA is conducting market research to identify qualified sources capable of providing a Captus 4000e or equivalent thyroid uptake system and to assess the availability of small business concerns eligible for set-aside consideration, including Service-Disabled Veteran-Owned Small Businesses, Veteran-Owned Small Businesses, HUBZone 8(a), Women-Owned Small Businesses, and Small Disadvantaged Businesses. The NAICS code is 334516 (Analytical Laboratory Instrument Manufacturing) with a small business size standard of 1,000 employees. Responses are due by August 13, 2026, at 2:00 PM Pacific Time and should be submitted via email to Krystal Weeks (Contracting Officer) at krystal.weeks@va.gov and nco20supplyteam2@va.gov. This notice does not constitute a request for proposal, quote, or pricing information, and responding does not preclude future participation in formal solicitations.
The required thyroid uptake system must be a comprehensive nuclear medicine measurement system with specific technical capabilities including a NaI(Tl) scintillation detector (2-inch diameter), a multichannel analyzer with minimum 1024 channels, a color touch screen interface (minimum 20 inches), and Windows 10 or 11 operating system compatibility. The system must support thyroid uptake measurements, wipe tests, bioassay monitoring, automated quality assurance functions, DICOM connectivity for hospital information systems integration, and reporting with spectrum graphics and data archival in standard formats. Optional features include a well counter with NaI(Tl) detection and 1-inch lead shielding. Physical requirements include a mobile ergonomic stand with adjustable height, articulating detector arm, storage for phantoms and quality control sources, and hospital-grade power configuration. Respondents must indicate small business eligibility status, original equipment manufacturer or authorized distributor status, GSA/FSS or GWAC contract availability, Buy American Act compliance, and estimated delivery date. This notice is for market research purposes only, with any formal solicitation to be posted on sam.gov.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sources Sought Notice Sources Sought Notice
| SUBJECT* |
| Thyroid Uptake System |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 98661 |
| SOLICITATION NUMBER* |
| 36C26026Q0928 |
| RESPONSE DATE/TIME/ZONE |
| 08-13-2026 14:00 PACIFIC TIME, LOS ANGELES, USA |
| ARCHIVE |
| 15 DAYS AFTER THE RESPONSE DATE |
| RECOVERY ACT FUNDS |
| N |
SET-ASIDE
| PRODUCT SERVICE CODE* |
| 6515 |
| NAICS CODE* |
| 334516 |
| CONTRACTING OFFICE ADDRESS |
| NCO 20 Network Contracting Office |
ATTN: V4CONT
1601 E. Fourth Plain Blvd Bldg.17, Suite B428 Vancouver WA 98661
POINT OF CONTACT*
Contracting Officer Krystal Weeks Krystal.weeks@va.gov nco20supplyteam2@va.gov
PLACE OF PERFORMANCE
| ADDRESS |
| Deptartment of Veterans Affairs |
Mann-Granstaff VA Medical Center
4815 North Assembly Street
Spokane WA
| POSTAL CODE |
| 99205-6485 |
| COUNTRY |
| United States |
ADDITIONAL INFORMATION
| AGENCY’S URL |
| https://www.va.gov/spokane-health-care/ |
| URL DESCRIPTION |
| VA Spokane health care |
| AGENCY CONTACT’S EMAIL ADDRESS |
| See Point of Contact Above |
| EMAIL DESCRIPTION |
| See Point of Contact Above |
DESCRIPTION
SOURCES SOUGHT NOTICE
THIS IS A SOURCES SOUGHT NOTICE ONLY. THIS IS NOT A REQUEST FOR PROPOSAL OR QUOTE; THE VA IS NOT SEEKING PRICING OR OFFERS.
The Department of Veterans Affairs, VISN 20, in support of the Mann-Grandstaff VA Medical Center, is performing market research to compile data regarding: (1) The availability and capability of qualified small business concerns; (2) Identify if there are sufficient number of qualified Service-Disabled Veteran-Owned Small Business, Veteran Owned Small Business, HUB Zone 8(a), Women-Owned Small Business, or Small Disadvantaged Business, Small Business concerns relative to the subject North American Industry Classification System (NAICS) code. Responses to the information requested will assist the Government in determining the appropriate procurement method, including any set-aside decision. Respondents should clearly describe their capabilities to include management experience, equipment and personnel or the ability to obtain them. Furthermore, interested sources must also meet all requirements of Federal, State, and Local Municipal codes regarding operations of this type of service.
The NAICS code assigned is: 334516 – Analytical Laboratory Instrument Manufacturing The small business size standard is 1,000 employees.
Not responding to this Sources Sought does not preclude participation in any future or potential solicitation, or Request for Quotation. It is the intent of the Mann-Grandstaff VA Medical Center to use the information gathered for market research purposes only. If a formal solicitation is released, it will be posted via the Contract Opportunies on Sam.gov.
DESCRIPTION OF THE REQUIREMENT:
The Department of Veterans Affairs, Network Contracting Office 20, is seeking sources that can provide a Captus 4000e or equal thyroid uptake system. Equal product will be considered if it has the same characteristics as listed below:
1. General System Functionality
· Must be a comprehensive nuclear medicine measurement system capable of performing thyroid uptake measurements, wipe tests, bioassay measurements, and automated quality assurance functions.
· Must include a software-driven workflow that supports predefined protocols and customizable measurement sequences.
2. User Interface & Computer Requirements
· Must utilize a color touch screen user interface with a large-format display (minimum 20 inches for visibility during procedures).
· Must operate on a current, widely supported Windows OS platform. (VA current platform Windows 10 or 11)
· Must support generation of reports containing graphics, isotope spectra, and archived data, with ability to save or print reports in PDF format.
3. Detector & Measurement Specifications Thyroid Uptake Detector
· Must utilize a NaI(Tl) scintillation detector capable of precise uptake measurements. (Spec: 2‑inch diameter NaI(Tl).)
· Must include a collimator meeting recognized nuclear medicine performance standards (e.g., ANSI or IAEA).
· Must include a physical measurement/alignment mechanism to ensure consistent detector‑to‑thyroid distance.
Multichannel Analyzer (MCA)
· Must include an MCA with a minimum of 1024 channels.
· Must provide automated and manual calibration capabilities.
· Must support user‑defined and preset ROIs, live time, real time, and total count presets.
· Must support accurate count rate linearity (≤2% deviation up to at least 150,000 cps).
Optional Well Counter
· Must support well counting for capsule/liquid dose formats utilizing a NaI(Tl) well detector. (Spec: 2‑inch diameter NaI(Tl).)
· Must provide shielding sufficient for radiation safety and measurement accuracy (minimum 1‑inch lead shielding).
4. Quality Assurance Capabilities
· Must provide fully automated QA tests including energy calibration, zero adjust, gain adjust, detector efficiency, resolution, chi‑square, constancy, and minimum detectable activity (MDA).
· Must include both automated and manual calibration options.
5. Thyroid Uptake Measurement Requirements
· Must support capsule and liquid dose formats.
· Must allow pre-dose and pre‑count measurement.
· Must include multiple predefined thyroid uptake protocols and allow custom protocol creation.
6. Wipe Test & Bioassay Capabilities
· Must support multi‑group wipe test setup and isotope identification with spectral display.
· Must flag wipe test results exceeding trigger levels.
· Must support bioassay monitoring for common isotopes (e.g., I‑123, I‑125, I‑131).
· Must allow user‑defined action levels and produce individual and summary reports.
7. Connectivity / System Integration Requirements
· Must provide optional or integrated DICOM capability that can retrieve patient data and upload reports to hospital information systems.
8. Physical / Form Factor Requirements
· Must include a mobile, ergonomic stand with adjustable height and articulating detector arm capable of wide range motion (vertical travel and rotation).
· Must include secure storage for neck phantom and quality control sources.
· Must include cable management to maintain safe, clinical‑grade setup.
· Must include hospital-grade power configuration with isolation transformer.
9. Reporting & Data Archival Requirements
· Must allow reporting that includes full spectrum graphics and thyroid uptake time‑vs‑value graphs.
· Must permit archiving of measurement data and export in standard formats.
The information above is intended to be descriptive, not restrictive, to indicate the quality of equipment that will be satisfactory. It is the responsibility of the interested source to demonstrate to the Government that the interested source can provide equipment that fulfills the required specifications.
POTENTIAL SOURCES SHALL PROVIDE THE FOLLOWING INFORMATION IN THEIR RESPONSE:
Mark if your firm is eligible for participation in one of the following small business programs. If so, please indicate the program:
[ ] yes [ ] no - Small Business (SB) [ ] yes [ ] no - HUBZone [ ] yes [ ] no - Small Business 8(a) [ ] yes [ ] no - Small Disadvantaged Business (SDB) [ ] yes [ ] no - Women-Owned (WO) Small Business [ ] yes [ ] no - Service Disabled Veteran Owned Small Business (SDVOSB) [ ] yes [ ] no - Veteran Owned Small Business (VOSB) [ ] yes [ ] no - Large Business [ ] yes [ ] no - Other (please specify) [ ] yes [ ] no Are you the Original Equipment Manufacturer (OEM)?
[ ] yes [ ] no If not the OEM, are you an authorized distributor, and can produce a letter from the OEM?
[ ] yes [ ] no Is this product available on your GSA/FSS or other GWAC Contract?
If so, what is your contract number?_____________________ [ ] yes [ ] no Does your equipment conform to the Buy American Act and RFO 52.225-1?
52.225-1 Buy American-Supplies | Acquisition.GOV What is the estimated Delivery Date After Receipt of Order (ARO)? _____________
Please include the following information:
Company Name:
POC:
UEI:
Address:
Email:
In addition to providing the information requested above, responding companies are encouraged to include any relevant information (specifications, cut sheets, brochures, capability statement, experience etc.) to confirm the company s ability to meet the requirements outlined in this request.
Disclaimer: This notice does not obligate the Government to award a contract or otherwise pay for the information provided in response. The Government reserves the right to use information provided by respondents for any purpose deemed necessary and legally appropriate. After reviewing the responses received, a pre-solicitation synopsis and solicitation may be published in Contract Opportunities (https://sam.gov/).
Submit Capability Statements to: Department of Veterans Affairs, VISN 20 Attn: Krystal Weeks, no later than 2:00 PM (PDT), August 13, 2026. Responses must be submitted via e-mail to: krystal.weeks@va.gov and nco20supplyteam2@va.gov .
Note: Do not include Proprietary, classified, confidential, or sensitive information in responses.
| *= Required Field |
| Sources Sought Notice |
Sources Sought Notice
File details come from the government source that posted it. Updated .