Text version
Tel: (800) 457-4117 Web: www.belimed.com
1535 Hobby Street, Charleston, SC 29405
Dwg. No. :
Drawn By :
Rev. :
Scale :
of
Page :
Date :
Beli ed
I n f e c t i o n C o n t r o l Typical installation of a UW565D ultrasonic
1/22/18
N.T.S.
G
A. JONES
CUT-UW565D
UW565D
ULTRASONIC CLEANER
3D VIEW
cleaner.
UTILITY FLEX HOSES TO UNIT
(BY BELIMED)
DRAINPIPE TO DRAIN
(BY OTHERS)
RECEPTACLE
(BY OTHERS)
HOT WATER
(BY OTHERS)
COLD WATER
(BY OTHERS)
MIXING VALVE
(PROVIDED BY BELIMED,
INSTALLED BY OTHERS)
DI WATER
(BY OTHERS)
FRONT
ACCESS DOOR
42"
UTILITY CONNECTIONS
(LOCATED BEHIND UNIT)
CONTROL
PANEL
DRAINPIPE TO DRAIN
(BY OTHERS)
DR
0 0
32"
L E
V E
L I N
G
F
E E
T
DR
A
B
U T
I L I T
Y
C
O N
N E
C T
I O
N
E L
D I
C W
H W
H C
M
M I X
I N
G
V
A L V
E
C
O N
N E
C T
I O
N
DR
C W
H
W
D
I
E L
U T
I L I T
Y
C
O N
N E
C T
I O
N
15"
SIDE ACCESS
PANELS
27"5" MIN.
DRAINPIPE TO DRAIN
(BY OTHERS)
CONTROL
PANEL
L E
V E
L I N
G
F
E E
T
D R
A I N
C O
N N
E C
T I O
N
M I X
I N
G
V
A L V
E
C
O N
N E
C T
I O
N
42"
M
I N
D I
C W
11 1/4"
21 1/4"
25 1/4"
DR
6"
32"
E L
WALL
(BY OTHERS)
DRAINPIPE TO DRAIN
(BY OTHERS)
H W
37"
MANUAL SERVICE VALVE, BY OTHERS.
H C
M
1. REFER TO EQUIPMENT PLUMBING,
MECHANICAL AND ELECTRICAL
PLANS FOR UTILITY LOCATIONS.
2. SEE TECHNICAL SPECIFICATIONS
FOR UTILITY SIZING, INPUT
REQUIREMENTS, AND FINAL
CONNECTIONS.
FINISHED FLOOR
RECOMMENDED MIXING VALVE, PROVIDED
BY BELIMED (IF PURCHASED), INSTALLED
BY OTHERS.
H C
M
C O
L D
W
A T
E R
H O
T
W
A T
E R
CONNECTION POINT
32" A.F.F.
EL
HW CW
3/4" GHT FITTING,
BY OTHERS.
D I
W
A T
E R
MIXING VALVE CONNECTION
52" A.F.F.
1" SWEAT,
BY OTHERS.
Tel: (800) 457-4117 Web: www.belimed.com
1535 Hobby Street, Charleston, SC 29405
Dwg. No. :
Drawn By :
Rev. :
Scale :
of
Page :
Date :
Beli ed
I n f e c t i o n C o n t r o l Typical installation of a UW565D ultrasonic
1/22/18
3/8" = 1'
G
A. JONES
CUT-UW565D
UW565D
ULTRASONIC CLEANER
TECHNICAL DRAWINGS
cleaner.
LEGEND
SERVICE SHUT-OFF VALVE, BY OTHERS
TYPICAL PLAN VIEW
TYPICAL SIDE ELEVATIONTYPICAL FRONT ELEVATION
*MIXING VALVE IS INSTALLED TO ACHIEVE CONTROLLED INCOMING HOT
WATER TEMPERATURE, AS REQUIRED FOR DETERGENTS USED FOR
PROCESSING ROBOTICS OR OTHER APPLICATIONS AS SPECIFIED BY
INSTRUMENT MANUFACTURER IFU.
TYPICAL MIXING VALVE ASSEMBLY (N.T.S.)
UW565D ULTRASONIC CLEANER - TECHNICAL SPECIFICATIONS
REQUIRED INPUTS
Symbol Description
Value
Utility:
Hot Water
Hot water must have a total hardness less than 85 ppm. Temperature range of 100° to
140°F. Maximum consumption per load based on 5 minute rinse cycle at maximum flow rate.
Mixing Valve Connection Size/Type:
1 Inch / SWEAT
Pressure:
Peak Consumption: 17.4 gal/min
*Consumption per Load:
Utility:
Electric
Connection Type:
Voltage (Phase):
Voltage (Nominal):
Circuit Protection:
Peak Consumption:
Power Consumption per Load:
Medium: Noise Radiation
Maximum Radiation (Net):
80 dBA
Parameter:
Carry-In Dimensions
ESTIMATED EMISSIONS
PHYSICAL SPECIFICATIONS
TS-UW565D_RevG
Provide floor/sink and trap, to accommodate maximum discharge rate. Optional Drain pump assembly with 1/2" GHT connection for areas with wall drain location or floor drain located greater than 12 feet from unit.
Utility:
Cold Water
Pressure:
Peak Consumption: 3.2 gal/min
*Consumption per Load:
Cold water must have a total hardness less than 170 ppm.
Machine alone
(i.e. removed from packaging)
Min: 40 Max: 60 psig
Min: 0 gal / Max: 66 gal
Min: 40 Max: 60 psig
Min: 1.0 Max: 1.0 gal
120/208 VAC (4 Wire)
1 Phase
20 Amps
14 Amps
Min. 0.6 Max 0.7 kW-h
Utility:
Connection Size/Type:
Open Drain
1 1/2 Inch / Indirect Drain
Discharge Flow Rate: 21.0 gal/min
Height:
Width:
Length:
53.5 Inch
42 Inch
27 Inch
Parameter:
Weights
Transport Weight:
Operating Weight:
550 lbs
680 lbs
* Values are for an "Instruments - General Cycle". Values will vary with load size, utility inputs, and cycle setup.
Additional Information:
GHT = Garden Hose Thread
FNPT = Female Pipe Thread
Output connection on mixing valve is 3/4" GHT, by others.
All utility supplies and connections are to be provided by others, per local code.
Operational Wt = Machine + Tray + Water + Instruments
Transport Weight = Machine + Crating
Utility Notes
Final Utility Conn. By:
Others Belimed
Connection
Method
X Flex
X Flex
X
Utility:
Deionized Water
Connection Size/Type:
1/2 Inch / FNPT
Pressure:
Operating Consumption:
*Consumption per Load:
DI Rinse Connection - Maximum consumption per load based on 2 minute rinse cycle at maximum flow rate.
Min: 20 Max: 60 psig
Min: 0.0 Max: 14.8 gal
X Flex
Min: 0.0 Max: 7.4 gal/min
1-1/2" Copper
Pipe
Receptacle
X
Receptacle
Junction Box with Receptacle NEMA L14-20R to NEC local codes.
OSHPD Number: OPM-0219-13
Unit Connection Size/Type:
3/4 Inch / GHT
Mixing Valve Connection Size/Type:
1 Inch / SWEAT
Unit Connection Size/Type:
3/4 Inch / GHT
X
1" Copper Pipe
X
1" Copper Pipe
HW
CW
DI
DR
EL
USE 1/2" DIA. HILTI KB-TZ EXPANSION
ANCHORS (MINIMUM EMBED = 2") .
ANCHOR BRACKET (4-PLACES)
ULTRASONIC
LEVELING FOOT
#14 SHEET METAL SCREW
(4 PER ANCHOR BRACKET)
Tel: (800) 457-4117 Web: www.belimed.com
1535 Hobby Street, Charleston, SC 29405
Dwg. No. :
Drawn By :
Rev. :
Scale :
of
Page :
Date :
Beli ed
I n f e c t i o n C o n t r o l Typical installation of a UW565D ultrasonic
1/22/18
NONE
G
A. JONES
CUT-UW565D
UW565D
ULTRASONIC CLEANER
TECHNICAL SPECIFICATIONS
cleaner.
TYPICAL SEISMIC ANCHORING (N.T.S.)
NOTE: OPTIONAL DRAIN PUMP ASSEMBLY AVAILABLE
FOR AREAS WITH WALL DRAIN LOCATION OR FLOOR
DRAIN LOCATED GREATER THAN 12' FROM UNIT.
DISCHARGE RATE OF 13.3-14.0 GAL/MIN. CONNECTION
SIZE 1/2" GHT.