SOW_Hemosphere.revised.2.pdf
PDF 126 KB Posted
- Solicitation number
- 36C25723Q1135
About this file
This document is a scope of work outlining requirements for the purchase of a Hemosphere Cardiac Monitoring System by the Amarillo VA Health Care System. The scope of work specifies that the system must provide hemodynamic monitoring capabilities using noninvasive and catheter methods, with color-based indicators to communicate patient status. Key requirements for the Hemosphere Advanced Monitoring System include compatibility with a variety of cuffs, sensors and catheters, hypotension predictive software, and decision support tools. The scope of work provides details on equipment specifications, electrical requirements, and connectivity to hospital information systems. It also outlines a service and maintenance plan through the Hemosphere 360 protection program covering replacement of failed equipment, software updates, and technology upgrades for one year at no additional cost. The related federal contract opportunity is solicitation number 36C25723Q1135 from the Department of Veterans Affairs for purchase of the Hemosphere Cardiac Monitoring System by the Amarillo VA Health Care System nursing service.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25723Q1135 0001.docx | DOCX document | |
| 36C25723Q1135.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
The Amarillo VA Health Care System is requesting purchase of the Hemosphere Advanced Monitoring System. This technology provides capability of hemodynamic monitoring utilizing a variety of methods including a noninvasive finger cuff. The Hemosphere Advanced Monitoring System presents hemodynamic information using color-based indicators which communicate patient status at a glance and allow for immediate recognition of clinical status and improving clinical decision making. All equipment must adhere to federal guidelines for network security.
Place of Performance:
Amarillo VA Health Care System, Amarillo TX
General Requirements of equipment:
The contractor will provide the type of hemodynamic monitoring systems to meet or exceed the government’s minimum needs.
Hemosphere Advanced Monitoring System offers a full-range cuff, sensor and catheter compatibility hypotension predictive decision support software allowing proactive individualized patient management -
Features
• HEMOSPHERE ADVANCED MONITORING SYSTEM – SMART RECOVERY WITH CLEARSIGHT,
PRESSURE CABLE FOR ACUMEN IQ / FLOTRAC, HPI, AND ACUMEN ANALYTICS
Kit Contains: HEM1, HEMBAT10, HEMCSM10, PC2K, EVHRS, HEMPSC100, HEMTOM10, HEMDPT1000, 194371001, Acumen Analytics.
• Hemosphere Roll stand
• Hemosphere 360 program 24 MO plan. Covers modules under total protection program highlights.
• Hemosphere Foresight tissue oximetry cable.
• Pressure controller cover – PCCVR Included with HEMAQSR2, HEMCSMUPG, EV1000CS, and EVNIUPG-US kits
Safety Specifications:
Weight: 10 +/- 0.2 lbs (4.5 +/- 0.1 kg) Width: 12.4 in (315 mm) Height: 11.7 in (297 mm) Depth: 5.56 in (141 mm)
Display/Resolution/OS: Active Area 12.1 in (307 mm)/1024x768 LCD/Windows 10 loT
HemoSphere modules
HemoSphere Tissue Oximetry module Weight: approx 1.0 lb (.45 kg); Height: 1.4” (3.5 cm) Width: 3.5 in (8.9 cm); Depth: 5.4 in (13.6 cm) HemoSphere Clear Sight module Weight: approx 2.0 lb (.9 kg); Height: 5.1” (13 cm) Width: 5.6 in (14cm); Depth: 3.9 in (10 cm)
Electrical
Rated Supply Voltage: 100-240 Vac, 50/60 Hz Rated input: 1.5-2.0 Amps
Input/Output
Touch Screen: Projective capacitive touch RS-232 Serial Port (1): Edwards proprietary protocol: maximum data rate = 57.6 kilo baud USB Ports (2): 1 USB 2.0, 1 USB 3.0 Ports: 1 RJ-45 Ehternet, 1 HDMI Hospital information system connectivity: Health Level 7 (HL7); integration services are subject to local availability DPT Pressure: DPT pressure out ECG monitor: Sync line conversion: 1V/mV; Input voltage range: +/- full scale; Resolution: +/- BPM; Range: =30-200 BPM; /4 in. stereo jack: analog cable
HemoSphere battery pack
HemoSphere battery pack Weight: 0.9 lb (0.4 kg); Height: 1.38 in (35 mm); Width: 3.15 in (80 mm); Depth: 5.0 in (126 mm); Cells: 4 x LifePO4 (lithium iron phosphate)
DELIVERY/INSTALLATION
1. Throughout the life of the monitors, the Vendor shall provide ongoing training, education and support.
SERVICE/MAINTENANCE
1. The HemoSphere 360 protection Service provides:
a. Failure of associated models will be replaced at no charge
b. Includes software updates and technology expansion upgrades to the equipment’s software
c. Covers physical and liquid damage to the equipment
d. Guaranteed next day replacement of covered equipment
e. All parts, labor and shipping costs covered
f. One year of manufacturer’s warranty upgraded to total protection at no additional cost.
2. 24/7/365 technical and product phone support
CONTRACTING OFFICER REPRESENATIVE
Thomas E. Creek VAHCS 6010 Amarillo Boulevard West Amarillo, TX 79106-1990 Attention: TBD Email: TBD
File details come from the government source that posted it. Updated .