Opportunties 36C25722Q1055.docx

DOCX document 36 KB Posted

Attached to
J065--674-22-4-995-0203 | NEW | CTVHCS Waco- COE Research -140C Freezer Repair
Solicitation number
36C25722Q1055
Issued by

View the file

Other files

File Type Posted
RFQ 36C25722Q1055.docx DOCX document
RFQ 36C25722Q1055.docx DOCX document
36C25722Q1055.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Presolicitation Notice Presolicitation Notice

SUBJECT*
674-22-4-995-0203 | NEW | CTVHCS Waco- COE Research -140C Freezer Repair

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
78240
SOLICITATION NUMBER*
36C25722Q1055
RESPONSE DATE/TIME/ZONE
08-31-2022 1:00PM CENTRAL TIME, CHICAGO, USA
ARCHIVE
20 DAYS AFTER THE RESPONSE DATE
SET-ASIDE
SBA
PRODUCT SERVICE CODE*
J065
NAICS CODE*
811219
PLACE OF PERFORMANCE
DEPARTMENT OF VETERANS AFFAIRS

Central Texas Veterans Health Care System

Attn: Mental Health

1901 Veterans Memorial Drive

Temple TX

POSTAL CODE
76504

COUNTRY

CONTACT INFORMATION

CONTRACTING OFFICE ADDRESS
Department Of Veterans Affairs

Network Contracting Office 17

5441 Babcock Road Ste. 302 San Antonio TX 78240

POINT OF CONTACT*

Contract Specialist David Simmons david.simmons@va.gov 210-694-6337

ADDITIONAL INFORMATION

AGENCY’S URL
https://www.va.gov
URL DESCRIPTION
Department Of Veterans Affairs
AGENCY CONTACT’S EMAIL ADDRESS
david.simmons@va.gov
EMAIL DESCRIPTION
david.simmons@va.gov

DESCRIPTION

STATEMENT OF WORK (SOW)

GENERAL INFORMATION

Title of Project: One-Time Repair of -140C Ultra-Low Laboratory Freezer; VA Research; Biomarkers and Genetics Core in Waco

Scope of Work: The VISN 17 Center of Excellence, Biomarkers and Genetics Core, located at 4800 Memorial Drive, Building 93, room 2A113, Waco, TX 76711 has requested rebuild of the refrigeration system on -140C Freezer (M#ULT7150-9-D19; SN# 1117367801190921) currently in use in support of multiple ongoing Veteran Research Studies.

Background: The Ultra-Low Laboratory Freezers are a critical piece of equipment used by researchers at the CoE and required by most funded research programs at this location.

SERVICE COVERAGE DESCRIPTIONS: Below are requirements for service and maintenance of the freezers:

1. Obtain warrantied compressor from manufacturer

0. The unit’s scroll compressor is under warranty from ThermoFisher until 2024

0. Must be acquired by ThermoFisher approved repair company

1. Pick up non-working unit from customer site

1. Rebuild the refrigeration system on the unit at service facility

2. Rebuild refrigeration system includes:

0. Replacement of scroll compressor

0. Filter dryer replaced

0. System will be processed with cleaning agent to clean old lines

0. Vacuum pull on the system

0. Replacement of ultra-low temperature coolant

4. Must be performed by highly trained technician as specific refrigerant mixture must be added to OEM specification

0. All parts and labor associated with rebuild

1. Complete system testing to ensure functioning is in line with strict laboratory requirements

1. Deliver unit back to customer and ensure on-site functioning REPORTING PROCEDURES: The contractor will provide at least 24 hours’ notice prior to performing servicing of the instrument. For service calls, repair/service technicians must report to the Center of Excellence and sign-in before proceeding to work location. A copy of the service report will either be directly given or emailed to the research point of contact within two weeks after service.

1. Performance Period:

One-Time Repair: ASAP

B. TRAVEL: Travel cost and transport of the unit are included in pricing.

*=Required Field
Presolicitation Notice

Presolicitation Notice

File details come from the government source that posted it. Updated .