36C25718Q0520-001.docx
DOCX document 157 KB Posted
- Solicitation number
- 36C25718Q0520
About this file
36C25718Q0520 36C25718Q0520_1.docx
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25718Q0520-000.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
36C25718Q0520
OF 85
1. REQUISITION NO.
2. CONTRACT NO.
3. AWARD/EFFECTIVE DATE
4. ORDER NO.
5. SOLICITATION NUMBER
6. SOLICITATION ISSUE DATE
a. NAME
b. TELEPHONE NO. (No Collect Calls)
8. OFFER DUE DATE/LOCAL
TIME
9. ISSUED BY
CODE
10. THIS ACQUISITION IS
UNRESTRICTED OR
SET ASIDE:
% FOR:
SMALL BUSINESS
HUBZONE SMALL
BUSINESS
SERVICE-DISABLED
VETERAN-OWNED
SMALL BUSINESS
WOMEN-OWNED SMALL BUSINESS
(WOSB) ELIGIBLE UNDER THE WOMEN-OWNED
SMALL
BUSINESS PROGRAM
EDWOSB
8(A)
NAICS:
SIZE STANDARD:
11. DELIVERY FOR FOB DESTINA-
TION UNLESS BLOCK IS
MARKED
SEE SCHEDULE
12. DISCOUNT TERMS
13a. THIS CONTRACT IS A
RATED ORDER UNDER
DPAS (15 CFR 700)
13b. RATING
14. METHOD OF SOLICITATION
RFQ
IFB
RFP
15. DELIVER TO
CODE
16. ADMINISTERED BY
CODE
17a. CONTRACTOR/OFFEROR
CODE
FACILITY CODE
18a. PAYMENT WILL BE MADE BY
CODE
TELEPHONE NO.
DUNS:
DUNS+4:
PHONE:
FAX:
17b. CHECK IF REMITTANCE IS DIFFERENT AND PUT SUCH ADDRESS IN OFFER 18b. SUBMIT INVOICES TO ADDRESS SHOWN IN BLOCK 18a UNLESS BLOCK BELOW IS CHECKED
SEE ADDENDUM
19.
20.
21.
22.
23.
24.
ITEM NO.
SCHEDULE OF SUPPLIES/SERVICES
QUANTITY
UNIT
UNIT PRICE
AMOUNT
(Use Reverse and/or Attach Additional Sheets as Necessary)
25. ACCOUNTING AND APPROPRIATION DATA
26. TOTAL AWARD AMOUNT (For Govt. Use Only) 27a. SOLICITATION INCORPORATES BY REFERENCE FAR 52.212-1, 52.212-4. FAR 52.212-3 AND 52.212-5 ARE ATTACHED. ADDENDA
ARE
ARE NOT ATTACHED.
27b. CONTRACT/PURCHASE ORDER INCORPORATES BY REFERENCE FAR 52.212-4. FAR 52.212-5 IS ATTACHED. ADDENDA
ARE
ARE NOT ATTACHED
28. CONTRACTOR IS REQUIRED TO SI
GN THIS DOCUMENT AND RETURN _______________
29. AWARD OF CONTRACT: REF. ___________________________________ OFFER
COPIES TO ISSUING OFFICE. CONTRACTOR AGREES TO FURNISH AND
DATED ________________________________. YOUR OFFER ON SOLICITATION
DELIVER ALL ITEMS SET FORTH OR OTHERWISE IDENTIFIED ABOVE AND ON ANY
(BLOCK 5), INCLUDING ANY ADDITIONS OR CHANGES WHICH ARE
ADDITIONAL SHEETS SUBJECT TO THE TERMS AND CONDITIONS SPECIFIED
SET FORTH HEREIN IS ACCEPTED AS TO ITEMS:
30a. SIGN
ATURE OF OFFEROR/CONTRACTOR
31a. UNITED STATES OF AMERICA (SIGNATURE OF CONTRACTING OFFICER) 30b. NAME AND TITLE OF SIGNER (TYPE OR PRINT) 30c. DATE SIGNED 31b. NAME OF CONTRACTING OFFICER (TYPE OR PRINT) 31c. DATE SIGNED
AUTHORIZED FOR
LOCAL REPRODUCTION
(REV. 2/2012)
PREVIOUS EDITION IS NOT USABLE
Prescribed by GSA - FAR (48 CFR) 53.212
7. FOR SOLICITATION
INFORMATION CALL:
STANDARD FORM 1449
OFFEROR TO COMPLETE BLOCKS 12, 17, 23, 24, & 30
SOLICITATION/CONTRACT/ORDER FOR COMMERCIAL ITEMS
674-18-1-5167-0001 36C25718Q0520 03-06-2018 Jennifer L. Salerno-Diciuccio 254-899-6023 03-23-2018 Department of Veterans Affairs Veterans Healthcare Administration Network Contracting Office NCO 17 2002 Scott Boulevard Temple TX 76504 X X 541690 $15 Million X N/A X Department of Veterans Affairs Central Texas Veterans HCS 1901 S. 1st Street Temple TX 76504 Department of Veterans Affairs Veterans Healthcare Administration Network Contracting Office NCO 17 2002 Scott Boulevard Temple TX 76504
Department of Veterans Affairs
FMS-VA-2(101)
Financial Services Center PO Box 149971 Austin TX 78714-9971 The Central Texas Veterans Healthcare System - Radiation Safety Program Office - located at the Temple TX VA Medical Center - 1901 South 1st Street Temple TX 76504 has a requirement for a medical physicist to conduct routine performance, quality assurance surveys, performance evaluations, radiation shieldin g design analysis, post construction shielding integrity analysis, regulatory compliance surveys and site health physics in support of the radiation safety office personnel.
Please refer to the statement of work and schedule of services below for specific contract requirements.
X X x Jennifer L. Salerno-Diciuccio
NCO1715L2-0401
Table of Contents
| SECTION A | 1 |
| A.1 SF 1449 SOLICITATION/CONTRACT/ORDER FOR COMMERCIAL ITEMS | 1 |
| SECTION B - CONTINUATION OF SF 1449 BLOCKS | 3 |
| B.1 CONTRACT ADMINISTRATION DATA | 3 |
| B.2 SCHEDULE OF SERVICES | 4 |
| B.3 STATEMENT OF WORK | 17 |
| B.4 IT CONTRACT SECURITY | 31 |
| SECTION C - CONTRACT CLAUSES | 41 |
| C.1 52.212-4 CONTRACT TERMS AND CONDITIONS—COMMERCIAL ITEMS (JAN 2017) | 41 |
| C.2 52.217-9 OPTION TO EXTEND THE TERM OF THE CONTRACT (MAR 2000) | 47 |
| C.3 VAAR 852.203-70 COMMERCIAL ADVERTISING (JAN 2008) | 47 |
| C.4 VAAR 852.203-71 DISPLAY OF DEPARTMENT OF VETERAN AFFAIRS HOTLINE POSTER (DEC 1992) | 47 |
| C.5 VAAR 852.219-10 VA NOTICE OF TOTAL SERVICE-DISABLED VETERAN-OWNED SMALL BUSINESS SET-ASIDE (JUL 2016)(DEVIATION) | 48 |
| C.6 VAAR 852.232-72 ELECTRONIC SUBMISSION OF PAYMENT REQUESTS (NOV 2012) | 49 |
| C.7 VAAR 852.237-70 CONTRACTOR RESPONSIBILITIES (APR 1984) | 50 |
| C.8 LIMITATIONS ON SUBCONTRACTING-- MONITORING AND COMPLIANCE (JUN 2011) | 50 |
| C.9 52.252-2 CLAUSES INCORPORATED BY REFERENCE (FEB 1998) | 51 |
| C.10 52.212-5 CONTRACT TERMS AND CONDITIONS REQUIRED TO IMPLEMENT STATUTES OR EXECUTIVE ORDERS—COMMERCIAL ITEMS (JAN 2018) | 51 |
| SECTION D - CONTRACT DOCUMENTS, EXHIBITS, OR ATTACHMENTS | 59 |
| SECTION E - SOLICITATION PROVISIONS | 60 |
| E.1 52.212-1 INSTRUCTIONS TO OFFERORS—COMMERCIAL ITEMS (JAN 2017) | 60 |
| E.2 52.209-7 INFORMATION REGARDING RESPONSIBILITY MATTERS (JUL 2013) | 64 |
| E.3 52.216-1 TYPE OF CONTRACT (APR 1984) | 65 |
| E.4 52.233-2 SERVICE OF PROTEST (SEP 2006) | 65 |
| E.5 VAAR 852.209-70 ORGANIZATIONAL CONFLICTS OF INTEREST (JAN 2008) | 66 |
| E.6 VAAR 852.233-70 PROTEST CONTENT/ALTERNATIVE DISPUTE RESOLUTION (JAN 2008) | 66 |
| E.7 VAAR 852.233-71 ALTERNATE PROTEST PROCEDURE (JAN 1998) | 67 |
| E.8 VAAR 852.270-1 REPRESENTATIVES OF CONTRACTING OFFICERS (JAN 2008) | 67 |
| E.9 52.212-2 EVALUATION—COMMERCIAL ITEMS (OCT 2014) | 68 |
| E.10 52.212-3 OFFEROR REPRESENTATIONS AND CERTIFICATIONS—COMMERCIAL ITEMS (NOV 2017) | 69 |
SECTION B - CONTINUATION OF SF 1449 BLOCKS
B.1 CONTRACT ADMINISTRATION DATA
(continuation from Standard Form 1449, block 18A.)
1. Contract Administration: All contract administration matters will be handled by the following individuals:
a. CONTRACTOR:
b. GOVERNMENT: Contracting Officer 36C257 Department of Veterans Affairs Veterans Healthcare Administration Network Contracting Office NCO 17 2002 Scott Boulevard Temple TX 76504
2. CONTRACTOR REMITTANCE ADDRESS: All payments by the Government to the contractor will be made in accordance with:
| [] |
| 52.232-33, Payment by Electronic Funds Transfer—System For Award Management, or |
| [] |
| 52.232-36, Payment by Third Party |
3. INVOICES: Invoices shall be submitted in arrears:
| a. Quarterly | [] |
| b. Semi-Annually | [] |
| c. Other MONTHLY | [x] |
4. GOVERNMENT INVOICE ADDRESS: All Invoices from the contractor shall be submitted electronically in accordance with VAAR Clause 852.232-72 Electronic Submission of Payment Requests.
Department of Veterans Affairs
FMS-VA-2(101)
Financial Services Center PO Box 149971 Austin TX 78714-9971
ACKNOWLEDGMENT OF AMENDMENTS: The offeror acknowledges receipt of amendments to the Solicitation numbered and dated as follows:
| AMENDMENT NO |
| DATE |
B.2 SCHEDULE OF SERVICES
The contract agrees to complete all listed annual performance, quality assurance surveys, performance evaluations, radiation shielding design analysis, post construction shielding integrity analysis, support the Central Texas Veterans Healthcare System (CTVHCS) Radiation Safety Office personnel, regulatory compliance surveys and site health physics as described within the statement of work and the schedule listed below. The prices will be all inclusive. Travel, lodging, employee training and daily per diem will not be paid in addition to the schedule of services below – the price must take all incidental expenses into account to perform the duties described in the statement of work. The CTVHCS is the program office and point of contact for this contract. They will work directly with the vendor to schedule the services priced below on either a quarterly, semi-annual or yearly basis. Additional services may be ordered as needed. Any specific work detail not described within the schedule of services or the statement of work are not authorized unless approved by both the contractor and the contracting officer via bilateral contract modification. Payments will be made in arrears on a monthly basis and invoices shall be submitted by the vendor to the VA Financial Services Center in Austin, TX. Invoices are certified by the contracting officer’s representative that the work has been successfully performed in accordance with the statement of work and schedule of services prior to any authorized payment is made.
The sites of performance expected to be traveled to by the contractor to fulfill the tasks described in this schedule of services and the statement of work are as follows (sites of service are subject to change dependent upon addition or removal of listed equipment):
1. Olin E. Teague Veterans’ Medical Center 1901 South 1st Street Temple, TX 76504-7451
2. Austin Outpatient Clinic 7901 Metropolis Drive Austin, TX 78744
3. Brownwood Community Based Outpatient Clinic 2600 Memorial Park Drive Brownwood, TX 76801
4. Bryan College Station Community Based Outpatient Clinic 1651 Rock Prairie Road, Suite 100 College Station, TX 77845
5. Cedar Park Community Based Outpatient Clinic 600 N. Bell Boulevard Cedar Park, TX 78613
6. Doris Miller Department of Veterans Affairs Medical Center 4800 Memorial Drive Waco, TX 76711
7. LaGrange VA Outreach Clinic 890 E. Travis Street LaGrange, TX 78945
8. Palestine Community Based Outpatient Clinic 2000 South Loop 256 Suite 124 Palestine, TX 75801
9. Temple VA Clinic Annex 4501 South General Bruce Drive Suite 75 Temple, TX 76502
BASE YEAR: APRIL 1, 2018 – MARCH 31, 2019
| CLIN | DESCRIPTION OF SERVICES | QTY | UNIT | UNIT AMOUNT | |
| PRICE |
0001 Annual performance evaluation/quality 13 JOBS $________ $_______ assurance survey on radiographic x-ray machines
0002 Annual performance evaluation/quality 11 JOBS $________ $_______ assurance survey on radiographic x-ray machines located at CBOC locations in Austin, Brownwood, Bryan/College Station, Cedar Park And Palestine, TX.
| 0003 | Performance evaluations on newly installed | 10 | JOBS $________ $_______ |
| and/or repaired x-ray machines. |
| 0004 | Annual performance evaluation/quality | 12 | JOBS $________ $_______ | assurance survey on radiographic/fluoroscopic |
| x-ray units. |
| 0005 | Annual performance evaluation/quality | 5 | JOBS $________ $_______ | assurance survey on computerized tomographic (CT) |
| machines in Temple, Waco, and Austin. |
| 0006 | Annual quality assurance check and physics | 3 | JOBS $________ $_______ | inspection on Magnetic Resonance Imagers (MRI) |
| in Temple and Austin. |
0007 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation survey for Single Photon Emission Computerized Tomography (SPECT) Nuclear Medicine Cameras.
0008 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation surveys for combination SPECT/CT Nuclear Medicine Cameras.
0009 Annual performance evaluation/quality assurance 1 JOBS $________ $_______ and radiation survey for combination Positron Emission Tomography/CT (PET/CT) Nuclear Medicine Camera.
0010 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation safety survey of DEXA units in Temple and Austin.
0011 Annual performance evaluation/quality assurance 24 JOBS $________ $_______ surveys of Dental x-ray units in Temple, Waco, and Austin.
0012 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ survey on Dental panoramic x-ray units in Temple and Austin.
0013 Review of Imaging quality assurance program. 1 JOBS $________ $_______
0014 Radiation shielding design analysis for rooms that 10 JOBS $________ $_______ machine generated radiation producing imaging systems and/or nuclear medicine cameras will occupy.
0015 Post construction shielding integrity analysis in 10 JOBS $________ $_______ rooms in which machine generated radiation producing imaging systems or nuclear medicine cameras are located.
0016 Annual performance/quality assurance testing 9 JOBS $________ $_______ of Computed Radiographic (CR) readers in Temple, Waco, Austin, Brownwood, Bryan/College Station, Cedar Park, and Palestine
0017 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ testing of Mammography equipment in Temple and Austin.
0018 Support of the CTVHCS Radiation Safety Office 8 JOBS $________ $_______ by membership and/or attendance at the appropriate CTVHCS committee meetings on a quarterly or as needed basis. The CTVHCS RSO expects the diagnostic med
0019 Quarterly on site health physics/regulatory 4 JOBS $________ $_______ compliance surveys of the Nuclear Medicine, Radiation Safety Office, and of Research locations of use (when applicable).
TOTAL AMOUNT OF BASE YEAR: $____________________
OPTION YEAR ONE: APRIL 1, 2019 – MARCH 31, 2020
| CLIN | DESCRIPTION OF SERVICES | QTY | UNIT | UNIT AMOUNT | ||
| PRICE | ||||||
| 1001 | Annual performance evaluation/quality | 13 | JOBS $________ $_______ | assurance survey on radiographic x-ray |
machines
1002 Annual performance evaluation/quality 11 JOBS $________ $_______ assurance survey on radiographic x-ray machines located at CBOC locations in Austin, Brownwood, Bryan/College Station, Cedar Park And Palestine, TX.
| 1003 | Performance evaluations on newly installed | 10 | JOBS $________ $_______ |
| and/or repaired x-ray machines. |
| 1004 | Annual performance evaluation/quality | 12 | JOBS $________ $_______ | assurance survey on radiographic/fluoroscopic |
| x-ray units. |
| 1005 | Annual performance evaluation/quality | 5 | JOBS $________ $_______ | assurance survey on computerized tomographic (CT) |
| machines in Temple, Waco, and Austin. |
| 1006 | Annual quality assurance check and physics | 3 | JOBS $________ $_______ | inspection on Magnetic Resonance Imagers (MRI) |
| in Temple and Austin. |
1007 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation survey for Single Photon Emission
1008 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation surveys for combination SPECT/CT
1009 Annual performance evaluation/quality assurance 1 JOBS $________ $_______ and radiation survey for combination
Nuclear Medicine Camera.
1010 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation safety survey of DEXA units in Temple and Austin.
1011 Annual performance evaluation/quality assurance 24 JOBS $________ $_______ surveys of Dental x-ray units in Temple, Waco, and Austin.
1012 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ survey on Dental panoramic x-ray units in Temple and Austin.
1013 Review of Imaging quality assurance program. 1 JOBS $________ $_______
1014 Radiation shielding design analysis for rooms that 10 JOBS $________ $_______ machine generated radiation producing imaging systems and/or nuclear medicine cameras will occupy.
1015 Post construction shielding integrity analysis in 10 JOBS $________ $_______ rooms in which machine generated radiation producing imaging systems or nuclear medicine cameras are located.
1016 Annual performance/quality assurance testing 9 JOBS $________ $_______ of Computed Radiographic (CR) readers in Temple, Waco, Austin, Brownwood, Bryan/College Station, Cedar Park, and Palestine
1017 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ testing of Mammography equipment in Temple and Austin.
1018 Support of the CTVHCS Radiation Safety Office 8 JOBS $________ $_______ by membership and/or attendance at the appropriate CTVHCS committee meetings on a quarterly or as needed basis. The CTVHCS RSO expects the diagnostic med
1019 Quarterly on site health physics/regulatory 4 JOBS $________ $_______ compliance surveys of the Nuclear Medicine, Radiation Safety Office, and of Research locations of use (when applicable).
TOTAL AMOUNT FOR OPTION YEAR ONE $________________________________
OPTION YEAR TWO: APRIL 1, 2020 – MARCH 31, 2021
| CLIN | DESCRIPTION OF SERVICES | QTY | UNIT | UNIT AMOUNT | ||
| PRICE | ||||||
| 2001 | Annual performance evaluation/quality | 13 | JOBS $________ $_______ | assurance survey on radiographic x-ray |
machines
2002 Annual performance evaluation/quality 11 JOBS $________ $_______ assurance survey on radiographic x-ray machines located at CBOC locations in Austin, Brownwood, Bryan/College Station, Cedar Park And Palestine, TX.
| 2003 | Performance evaluations on newly installed | 10 | JOBS $________ $_______ |
| and/or repaired x-ray machines. |
| 2004 | Annual performance evaluation/quality | 12 | JOBS $________ $_______ | assurance survey on radiographic/fluoroscopic |
| x-ray units. |
| 2005 | Annual performance evaluation/quality | 5 | JOBS $________ $_______ | assurance survey on computerized tomographic (CT) |
| machines in Temple, Waco, and Austin. |
| 2006 | Annual quality assurance check and physics | 3 | JOBS $________ $_______ | inspection on Magnetic Resonance Imagers (MRI) |
| in Temple and Austin. |
2007 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation survey for Single Photon Emission
2008 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation surveys for combination SPECT/CT
2009 Annual performance evaluation/quality assurance 1 JOBS $________ $_______ and radiation survey for combination
Nuclear Medicine Camera.
2010 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation safety survey of DEXA units in Temple and Austin.
2011 Annual performance evaluation/quality assurance 24 JOBS $________ $_______ surveys of Dental x-ray units in Temple, Waco, and Austin.
2012 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ survey on Dental panoramic x-ray units in Temple and Austin.
2013 Review of Imaging quality assurance program. 1 JOBS $________ $_______
2014 Radiation shielding design analysis for rooms that 10 JOBS $________ $_______ machine generated radiation producing imaging systems and/or nuclear medicine cameras will occupy.
2015 Post construction shielding integrity analysis in 10 JOBS $________ $_______ rooms in which machine generated radiation producing imaging systems or nuclear medicine cameras are located.
2016 Annual performance/quality assurance testing 9 JOBS $________ $_______ of Computed Radiographic (CR) readers in Temple, Waco, Austin, Brownwood, Bryan/College Station, Cedar Park, and Palestine
2017 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ testing of Mammography equipment in Temple and Austin.
2018 Support of the CTVHCS Radiation Safety Office 8 JOBS $________ $_______ by membership and/or attendance at the appropriate CTVHCS committee meetings on a quarterly or as needed basis. The CTVHCS RSO expects the diagnostic med
2019 Quarterly on site health physics/regulatory 4 JOBS $________ $_______ compliance surveys of the Nuclear Medicine, TOTAL AMOUNT FOR OPTION YEAR TWO $______________________________
OPTION YEAR THREE: APRIL 1, 20121 – MARCH 31, 2022
| CLIN | DESCRIPTION OF SERVICES | QTY | UNIT | UNIT AMOUNT | ||
| PRICE | ||||||
| 3001 | Annual performance evaluation/quality | 13 | JOBS $________ $_______ | assurance survey on radiographic x-ray |
machines
3002 Annual performance evaluation/quality 11 JOBS $________ $_______ assurance survey on radiographic x-ray machines located at CBOC locations in Austin, Brownwood, Bryan/College Station, Cedar Park And Palestine, TX.
| 3003 | Performance evaluations on newly installed | 10 | JOBS $________ $_______ |
| and/or repaired x-ray machines. |
| 3004 | Annual performance evaluation/quality | 12 | JOBS $________ $_______ | assurance survey on radiographic/fluoroscopic |
| x-ray units. |
| 3005 | Annual performance evaluation/quality | 5 | JOBS $________ $_______ | assurance survey on computerized tomographic (CT) |
| machines in Temple, Waco, and Austin. |
| 3006 | Annual quality assurance check and physics | 3 | JOBS $________ $_______ | inspection on Magnetic Resonance Imagers (MRI) |
| in Temple and Austin. |
3007 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation survey for Single Photon Emission
3008 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation surveys for combination SPECT/CT
3009 Annual performance evaluation/quality assurance 1 JOBS $________ $_______ and radiation survey for combination
Nuclear Medicine Camera.
3010 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation safety survey of DEXA units in Temple and Austin.
3011 Annual performance evaluation/quality assurance 24 JOBS $________ $_______ surveys of Dental x-ray units in Temple, Waco, and Austin.
3012 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ survey on Dental panoramic x-ray units in Temple and Austin.
3013 Review of Imaging quality assurance program. 1 JOBS $________ $_______
3014 Radiation shielding design analysis for rooms that 10 JOBS $________ $_______ machine generated radiation producing imaging systems and/or nuclear medicine cameras will occupy.
3015 Post construction shielding integrity analysis in 10 JOBS $________ $_______ rooms in which machine generated radiation producing imaging systems or nuclear medicine cameras are located.
3016 Annual performance/quality assurance testing 9 JOBS $________ $_______ of Computed Radiographic (CR) readers in Temple, Waco, Austin, Brownwood, Bryan/College Station, Cedar Park, and Palestine
3017 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ testing of Mammography equipment in Temple and Austin.
3018 Support of the CTVHCS Radiation Safety Office 8 JOBS $________ $_______ by membership and/or attendance at the appropriate
3019 Quarterly on site health physics/regulatory 4 JOBS $________ $_______ compliance surveys of the Nuclear Medicine, TOTAL AMOUNT FOR OPTION YEAR THREE $_________________________________
OPTION YEAR FOUR: APRIL 1, 20122 – MARCH 31, 2023
| CLIN | DESCRIPTION OF SERVICES | QTY | UNIT | UNIT AMOUNT | ||
| PRICE | ||||||
| 4001 | Annual performance evaluation/quality | 13 | JOBS $________ $_______ | assurance survey on radiographic x-ray |
machines
4002 Annual performance evaluation/quality 11 JOBS $________ $_______ assurance survey on radiographic x-ray machines located at CBOC locations in Austin, Brownwood, Bryan/College Station, Cedar Park And Palestine, TX.
| 4003 | Performance evaluations on newly installed | 10 | JOBS $________ $_______ |
| and/or repaired x-ray machines. |
| 4004 | Annual performance evaluation/quality | 12 | JOBS $________ $_______ | assurance survey on radiographic/fluoroscopic |
| x-ray units. |
| 4005 | Annual performance evaluation/quality | 5 | JOBS $________ $_______ | assurance survey on computerized tomographic (CT) |
| machines in Temple, Waco, and Austin. |
| 4006 | Annual quality assurance check and physics | 3 | JOBS $________ $_______ | inspection on Magnetic Resonance Imagers (MRI) |
| in Temple and Austin. |
4007 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation survey for Single Photon Emission
4008 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation surveys for combination SPECT/CT
4009 Annual performance evaluation/quality assurance 1 JOBS $________ $_______ and radiation survey for combination
Nuclear Medicine Camera.
4010 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ and radiation safety survey of DEXA units in Temple and Austin.
4011 Annual performance evaluation/quality assurance 24 JOBS $________ $_______ surveys of Dental x-ray units in Temple, Waco, and Austin.
4012 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ survey on Dental panoramic x-ray units in Temple and Austin.
4013 Review of Imaging quality assurance program. 1 JOBS $________ $_______
4014 Radiation shielding design analysis for rooms that 10 JOBS $________ $_______ machine generated radiation producing imaging systems and/or nuclear medicine cameras will occupy.
4015 Post construction shielding integrity analysis in 10 JOBS $________ $_______ rooms in which machine generated radiation producing imaging systems or nuclear medicine cameras are located.
4016 Annual performance/quality assurance testing 9 JOBS $________ $_______ of Computed Radiographic (CR) readers in Temple, Waco, Austin, Brownwood, Bryan/College Station, Cedar Park, and Palestine
4017 Annual performance evaluation/quality assurance 2 JOBS $________ $_______ testing of Mammography equipment in Temple and Austin.
4018 Support of the CTVHCS Radiation Safety Office 8 JOBS $________ $_______ by membership and/or attendance at the appropriate
4019 Quarterly on site health physics/regulatory 4 JOBS $________ $_______ compliance surveys of the Nuclear Medicine, locations of use (when applicable).
TOTAL AMOUNT FOR OPTION YEAR FOUR $_________________________________
36C25718Q0520
TOTAL AMOUNT OF BASE AND FOUR ONE YEAR OPTION PERIODS ___________________
Page 1 of
B.3 STATEMENT OF WORK
1. Scope of Work. The Contractor shall provide all personnel, transportation, equipment and software to provide health physics consultation and diagnostic medical physicist services in support of the Veterans Affairs Central Texas Veterans Health Care System (CTVHCS) and OFF SITE CLINICS on as needed basis. These services will adhere to the requirements and expectations of the Nuclear Regulatory Commission (NRC), the National Council on Radiation Protection and Measurements (NCRP), the Mammography Quality Standards Act (MQSA), The Joint Commission (TJC), VHA National Health Physics Program (NHPP), the American College of Radiology (ACR), Food and Drug Administration (FDA), and internal requirements of this facility where applicable. The Texas Department of State Health Services regulations contained in 25 TAC §289 are not in force at CTVHCS but shall be followed as closely as practical. Services shall be available Monday through Friday 8-4:30 excluding Federal holidays by performing the following tasks:
· Task 1: Annual review of CTVHCS overall Imaging Quality Assurance program for diagnostic imaging equipment. [NHPP];
· Task 2: Annual, not to exceed 14 months, quality assurance testing survey and certification of all diagnostic x-ray and imaging equipment (to include MRI and Nuclear Medicine imaging equipment). See Attachment A for list of existing equipment. This list may be modified (equipment added or removed) at any time. [TJC, NHPP];
· Task 3: Recertification of imaging equipment prior to first patient use when required due to repairs or modification that may affect the radiation output or image quality. Within 24 hours of contact, the contractor will coordinate with the affected service to have the recertification scheduled;
· Task 4: Written reports signed by the diagnostic medical physicist documenting the testing, including a summary of methods, instruments used, measurements and deficiencies identified, for all diagnostic x-ray and imaging equipment. This must be completed on new diagnostic x-ray equipment prior to first patient use. The Contractor shall provide one copy of the physicist report to the RSO and one copy to the Chief of Biomedical Services within five working days of completion of the work. The report may be in electronic format. [TJC, NHPP];
· Task 5: Deficiencies or non-conformances found during reviews or testing which represent unsafe conditions shall be reported to the RSO immediately. Verbal reports must be provided after testing to Biomedical Engineering so that CTVHCS will know that equipment can be immediately put into use. The qualified diagnostic medical physicist will perform a follow-up with Biomedical Engineering to verify compliance of any necessary corrective action performed to correct deficiencies found and report result to the RSO;
· Task 6: Dose rate surveys in unrestricted areas adjacent to the radiation areas are to be included during testing and in the final report;
· Task 7: A patient exposure chart will be provided for each unit inspected for common projections and procedures;
· Task 8: Shielding design specification calculations for new construction and renovations. [TJC, NHPP]
· Task 9: Shielding certification surveys following new construction, modifications and replacement of equipment. [TJC, NHPP]
· Task 10: Annual quality assurance testing of mammography equipment. [MQSA]
· Task 11: Support the CTVHCS Radiation Safety Office by membership and/or attendance at the appropriate CTVHCS committee meetings (i.e., Radiation Safety Committee, Computed Tomography Working Group, Fluoroscopy Safety Working Group) on a quarterly or as needed basis. The CTVHCS Radiation Safety Office expects the diagnostic medical physicist to be available to attend these meetings not less than 2 hours per quarter. [TJC]
· Task 12: Provide consultation for dose estimates calculations for patient, staff and fetal exposures and sentinel event investigations.
· Task 13: Emergency response time on equipment tests within 48 hours.
· Task 14: Quarterly on-site health physics/regulatory compliance surveys of the Nuclear Medicine, Radiation Safety Office, and of Research locations of use (when applicable). These surveys will take place in the first two weeks of January, April, July, October and include:
· Sealed source inventories
· Sealed source leak tests (NRC, NHPP)
· Performance based audits of Nuclear Medicine practices/procedures
· Records of research acquisitions, contamination surveys, and disposals (if applicable)
· Task 15: Semi-annual audits of Nuclear Medicine Written Directive case records.
The annual (not to exceed 14 months) quality assurance testing will include all radiology and dental equipment at CTVHCS – Temple, Waco, Austin, and off site outpatient clinics – Brownwood, Bryan/College Station, Cedar Park, and Palestine. All quality assurance testing reports and mammography equipment performance evaluations reports shall be submitted to the CTVHCS Radiation Safety Office for review and approval subsequent to the completion of all testing, calibration and surveys of the equipment, facilities and procedures. The reports shall identify all items of noncompliance and include general and specific recommendations for correcting any items of noncompliance. All records (administrative and program specific) created during the period of the contract belong to VA Central Texas Health Care System(CTVHCS) and must be returned to CTVHCS at the end of the contract or destroyed in accordance to the VHA Record Control Schedule (RCS)10-1.
1.1. Quality Assurance Testing
1.1.1. All Quality Assurance Tests shall meet or exceed current and possible future changes in NRC, ACR, MQSA, NCRP, FDA and TJC requirements and recommendations. All Quality Assurance Tests shall meet or exceed the requirements in the following VHA directives and handbooks and future directives and handbooks :
· VHA Directive 1105.01, Management of Radioactive Materials
· VHA Handbook 1105.02, Nuclear Medicine and Radiation Safety Service
· VHA Handbook 1105.03, Mammography Program Procedures and Standards
· VHA Handbook 1105.04, Fluoroscopy Safety
· VHA Handbook 1105.05, MRI Safety
· VHA Directive 1129, Radiation Protection for Machine Sources of Ionizing Radiation
1.1.2. Diagnostic X-ray Equipment (Radiography and Fluoroscopy) The annual testing must include, but is not necessarily limited to, the following (as applicable):
Per VHA Handbook 1105.04, Fluoroscopy Safety:
· Measurement of radiation output parameters, including beam intensity and beam quality;
· Testing of all modes of operation used clinically, including automatic exposure rate controls of fluoroscopy systems;
· Assessment of image quality;
· Assessment of technique factors used clinically;
· Measurement of appropriate indices of patient dose output chart must be generated for each survey in accordance 25 TAC. Dose rates at typical clinical technique factors, with comparison to national standards;
Per the American College of Radiology Technical Standard for Diagnostic Medical Physics Performance Monitoring of Radiologic and Fluoroscopic Equipment:
· Integrity of unit assembly
· Collimation and radiation beam alignment
· Fluoroscopic system resolution
· Automatic exposure control system performance
· Fluoroscopic automatic brightness control performance (high-dose-rate, pulsed modes, field-of-view [FOV] variation)
· Image artifacts
· Fluoroscopic phantom image quality
· kVp accuracy and reproducibility
· Linearity of exposure versus Ma or mAs
· Exposure reproducibility
· Timer accuracy
· Beam quality assessment (half-value layer)
· Fluoroscopic entrance exposure rates
· Image receptor entrance exposure
· Equipment radiation safety functions
· Patient dose monitoring system calibration
· Video and digital monitor performance
· Digital image receptor performance If the review of clinically-used technique factors or the comparison of measured dose indices with national standards indicates that an optimum balance has not been achieved between patient dose and image quality or that the dose indices exceed national standards, the technique factors, whether posted in a chart or programmed into the fluoroscope, must be modified as necessary. [VHA Handbook 1105.04] Reports shall be generated indicating appropriate indices of patient dose, measured by a diagnostic medical physicist at clinically-used technique factors, to be posted near the controls of each fluoroscope. These indices include typical and maximal entrance skin dose or air kerma rate for each fluoroscopic mode of operation (e.g., pulse rate and magnification mode). For each image recording mode used clinically, these include the entrance skin dose or air kerma per image or the entrance skin dose or air kerma per second of imaging (e.g., cinefluorography in the cardiac catheterization laboratory) for a patient of typical thickness. [VHA Handbook 1105.04]
1.1.3. Computed Tomography
The testing must include, but is not necessarily limited to, the following (as applicable):
Per the American College of Radiology Technical Standard for Diagnostic Medical Physics Performance Monitoring of Computed Tomography (CT) Equipment:
· Alignment light accuracy
· Image localization from scanned projection radiograph (localization image)
· Table incrementation accuracy
· Radiation beam width (collimation)
· Reconstructed image thickness
· Image quality
· High-contrast (spatial) resolution
· Low-contrast sensitivity and resolution
· Image uniformity
· Noise
· Artifact evaluation
· CT number accuracy
· Acquisition workstation display
· Dosimetry
· Radiation output of CT scanner (CT dose index [CTDIvol] or equivalent)
· Patient radiation dose estimate for representative examinations
· Protocol review
· The Diagnostic Medical Physicist shall become a member of computed tomography working group (CTWG) and review the CT protocols annually. These should include head and abdomen protocols for adult and pediatric patients as applicable to the facility’s practice. In addition, facility protocols for very high dose procedures (e.g., brain perfusion) should be reviewed.
· Elements reviewed should include documentation of kVp, Ma, rotation time, detector configuration, pitch, reconstructed image thickness, and use of automatic exposure control (including ensuring documentation of reference settings used), and the indicated dose indices (CTDIvol) resulting from each examination
· Safety Evaluation
· Visual inspection
· Work load assessment
· Scatter and stray radiation measurements (if work load and other related parameters have changed since acceptance testing)
· Audible/visual signals
· Posting requirements
· Other tests as required by state or local regulations Per the TJC Revised Requirements for Diagnostic Imaging Services (Prepublication Requirements dated January 9, 2015; subject to change):
· Image uniformity
· Slice thickness accuracy
· Slice position accuracy (when prescribed from a scout image)
· Alignment light accuracy
· Table travel accuracy
· Radiation beam width
· High-contrast resolution
· Low-contrast resolution
· Geometric or distance accuracy
· CT number accuracy and uniformity
· Artifact evaluation
· Dosimetry
· The diagnostic medical physicist measures the radiation dose (in the form of volume computed tomography dose index [CTDIvol]) produced by each diagnostic CT imaging system for the following four CT protocols: adult brain, adult abdomen, pediatric brain and pediatric abdomen, as applicable. If one or more of these protocols is not used by the hospital, other commonly used CT protocols may be substituted.
· The diagnostic medical physicist verifies that the radiation dose (in the form of CTDIvol) displayed by the CT imaging system for each tested protocol is within 20 percent of the CTDIvol displayed on the CT console. The dates, results, and verifications of these measurements are documented.
· Image acquisition display monitors
· Maximum luminance
· Minimum luminance
· Luminance uniformity
· Resolution
· Spatial accuracy
1.1.4. Mammography
The annual testing must comply with the MQSA and the latest requirements posted on the ACR Website for the manufacturer of the digital mammography being inspected. It must include, but is not necessarily limited to, the following (as applicable):
· Unit assembly evaluation
· Collimation assessment
· Artifact evaluation
· kVp accuracy and reproducibility
· Bean quality assessment – HVL measurements
· Evaluation of system resolution
· Automatic Exposure Control (AEC) function performance
· Breast entrance exposure, AEC reproducibility, and average glandular dose
· Radiation output rate
· Phantom image quality evaluation
· Signal to noise ratio and contrast to noise ratio measurements
· View box luminance and room illuminance
· Review work station (RWS) tests
1.1.5. MRI Equipment
The testing must include, but is not necessarily limited to, the following (as applicable):
Per the American College of Radiology Technical Standard for Diagnostic Medical Physics Performance Monitoring of Magnetic Resonance Imaging (MRI) Equipment:
· Magnetic field homogeneity
· Slice position accuracy
· Slice thickness accuracy
· Radiofrequency (RF) calibration for all coils
· Frequency and gain/attenuator verification (prescan values)
· Image signal–to-noise ratio (SNR) for all coils
· Intensity uniformity for all volume coils
· Phase stability and image artifact assessment for all coils
· Softcopy (monitor) fidelity
· Evaluation of quality control (QC) program Per the TJC Revised Requirements for Diagnostic Imaging Services (Prepublication Requirements dated January 9, 2015; subject to change):
· Image uniformity for all radiofrequency (RF) coils used clinically
· Signal-to-noise ratio (SNR) for all coils used clinically
· Slice thickness accuracy
· Slice position accuracy
· Alignment light accuracy
· High-contrast resolution
· Low-contrast resolution (or contrast-to-noise ratio)
· Geometric or distance accuracy
· Magnetic field homogeneity
· Artifact evaluation
· Image acquisition display monitors
· Maximum luminance
· Minimum luminance
· Luminance uniformity
· Resolution
· Spatial accuracy
1.1.6. Nuclear Medicine Cameras
The annual testing must include, but is not necessarily limited to, the following (as applicable):
Per the TJC Revised Requirements for Diagnostic Imaging Services (Prepublication Requirements dated January 9, 2015; subject to change):
· Evaluation of all nuclear medicine image types by each NM scanner and include the use of phantoms to assess the following imaging metrics:
· Image uniformity/system uniformity
· High-contrast resolution/system spatial resolution
· Sensitivity
· Energy resolution
· Count-rate performance
· Artifact evaluation
· Low-contrast resolution or detectability for non-planar acquisitions (recommended not required)
1.1.7. Computed Radiography (CR) and Digital Radiography (DR) The annual testing must comply with the AAPM Report Number 93, Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems and include, but is not necessarily limited to, the following (as applicable):
· Component and imaging plate physical inspection and inventory
· Imaging plate dark noise and uniformity
· Exposure indicator calibration
· Linearity and auto-ranging response
· Laser beam function
· Limiting resolution and resolution uniformity
· Noise and low-contrast resolution
· Spatial accuracy
· Erasure thoroughness
· Aliasing/grid response
· IP throughput
· Positioning and collimation errors.
1.1.8. Dental
The physics inspection shall conform to the Conference of Radiation Control Program Directors (CRCPD), Quality Control Recommendations for Diagnostic Radiography Volume 1 Dental Facilities July 2001. The annual testing must include, but is not necessarily limited to, the following (as applicable):
· Collimation
· Beam quality (HVL)
· Timer accuracy and reproducibility
· kVp accuracy and reproducibility
· mA or mAs linearity
· Exposure reproducibility
· Entrance skin exposure evaluation, with comparisons to published diagnostic reference levels and achievable doses (e.g. NCRP Report No. 172)
· Technique chart evaluation
· Image uniformity (artifact evaluation)
1.1.9. Dental CBCT
If the manufacturer has provided a phantom and procedures to perform specific tests but the tests are not included in this SOW, then the manufacturer’s machine-specific QA tests shall be performed in addition to the QA tests in this SOW. The annual performance testing must include, but is not necessarily limited to, the following (as applicable):
· Radiation output repeatability and reproducibility
· kVp accuracy, repeatability, and reproducibility
· Bean quality
· Radiation field of view (FOV)
· Image quality
· Accuracy of linear measurements
· Accuracy of patient dose metric indication
· Patient dose assessment
· Review of the technical QA program
· Display monitor evaluation
· Test pattern evaluation
· Monitor brightness
· Viewing conditions
1.1.10. PET/CT
The annual physics inspection shall conform to the ACR PET Phantom Instructions for Evaluation of PET Image., ACR Nuclear Medicine Accreditation Program PET Module. Regarding the CT component, this evaluation should include items of the CT inspection, but not be limited to the following tests (as applicable):
· Uniformity
· Spatial resolution
· SUV analysis
1.2. Structural Shielding
1.2.1. For any room in which a fluoroscopic imaging system is installed, or in which a mobile fluoroscopic imaging system is frequently used, the doses to persons in adjacent areas, including any areas above and below, must be evaluated by a qualified diagnostic medical physicist. Structural shielding must be installed as necessary to maintain doses to persons in these areas ALARA and within regulatory limits. [VHA Directive 1129 and VHA Handbook 1105.04]
1.2.2. For the structural shielding of rooms containing x-ray imaging devices, the shielding design goal must be 5 milligray (mGy) in a year to any person in a controlled area. For uncontrolled areas, the shielding design goal needs to be 1 mGy in a year to any person, and 0.02 mGy in any hour. [VHA Handbook 1105.04]
1.2.3. The qualified diagnostic medical physicist shall provide shielding design calculations for each new, replaced, or relocated x-ray imaging system. The calculations of shielding for each for and acceptance testing surveys of imaging rooms must conform to National Council on Radiation Protection and Measurements (NCRP) Report No. 147, Structural Shielding Design for Medical X-ray Imaging Facilities and for dental units, NCRP Report No. 145. The calculations and testing must be documented in a written report which includes a diagram showing the adjacent areas. The qualified diagnostic medical physicist shall perform a shielding survey to verify the structural shielding was installed per the shielding design report and complies with the design goals. A written report of the shielding survey shall be provided to the RSO within 5 working days after the shielding survey has been completed. [VHA Handbook 1105.04]
2.1. Qualifications. Prior to commencing work, all contractor employees shall meet certain criteria to perform work under this contract as a Diagnostic Medical Physicist. Qualification documents must be submitted as part of the bid to the Contracting Officer and be approved by the RSO or his/her designee. All qualifications are subject to review by the CTVHCS Radiation Safety Committee.
2.1.1. Diagnostic Medical Physicist:
2.1.1.1. The Diagnostic Medical Physicist who performs the work at CTVHCS shall meet the following education, certification, and qualification standards:
· Hold a graduate degree in physics, medical physics, biophysics, radiological physics, medical health physics, or a closely related science or engineering discipline from an accredited college or a university.
· Formal graduate level college coursework in the biological sciences with at least one course in biology or radiation biology and one course in anatomy, physiology, or a similar topic related to the practice of medical physics.
· Documented 3 years of clinical experience in CT, nuclear medicine, PET/CT, or MRI. The physicist must document the 3 years of experience for the modality being inspected.
· Hold current Board Certification by the American Board of Radiology or American Board of Medical Physics.
· Licensed by the State of Texas as a Professional Medical Physicist.
· Meet all respective NRC, ACR, MQSA, FDA and TJC requirements for a qualified diagnostic medical physicist.
· A physicist in training or under a temporary license must be accompanied at any visit by and directly supervised by a fully licensed qualified diagnostic medical physicist.
2.1.1.2. The Diagnostic Medical Physicist shall have current knowledge, understanding, and recent experience in the following:
· All Texas Department of State Health Services, Radiation Control Regulations,
· All current NRC regulations,
· All current FDA regulations (including MQSA),
· Current recommendations of the ACR and TJC on diagnostic x-ray and imaging equipment
· Current recommendations of the NCRP and AAPM on medical imaging facility design and shielding,
· Current techniques for the calibration and testing of diagnostic x-ray equipment.
CTVHCS X-Ray Machines By Service Cardiac Cath
| Manufacturer |
| Model |
Location
Phillips
Allura FD2
TEMPLE
Dental
| Manufacturer |
| Model |
Location
Gendex
GX1000
WACO
Gendex
GX770
TEMPLE
Gendex
GX770
WACO
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
| Instrumentarium |
| OP 300 |
AUSTIN
| Instrumentarium |
| OP 300-1 |
TEMPLE
Gendex
Expert DC
AUSTIN
Gendex
Expert DC
AUSTIN
Gendex
Expert DC
AUSTIN
Gendex
Expert DC
AUSTIN
Gendex
Expert DC
AUSTIN
Gendex
Expert DC
AUSTIN
Gendex
Expert DC
AUSTIN
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
GX770
TEMPLE
Gendex
DP300
WACO
Gendex
Expert DC
TEMPLE
Nuclear Medicine
| Manufacturer |
| Model |
Location
| Spectrum Dynamics |
| DSPECT L |
TEMPLE
| General Electric |
| Infinia Ha |
TEMPLE
| General Electric |
| Infinia GP |
TEMPLE
| General Electric |
| Discovery |
TEMPLE
| General Electric |
| Millineum |
TEMPLE
LUNAR
| Prodigy H8 |
| TEMPLE |
Radiology
| Manufacturer |
| Model |
Location
Phillips
DR
TEMPLE
Philips
Optimus
WACO
Phillips
DR Digital
AUSTIN
Phillips
| Optimus 50 |
| BRYAN/COLLEGE STATION |
X-Cel
MB700BC
WACO
Siemens
Sireskop 3
TEMPLE
Phillips
| Diagnost D |
| TEMPLE |
Phillips
Eleva
TEMPLE
Toshiba
Aquillion
AUSTIN
Phillips
| Optimus 50 |
| CEDAR PARK |
Phillips
| Optimus 50 |
| BROWNWOOD |
Siemens
Multi-Vert
TEMPLE
Phillips
| Optimus 50 |
| PALESTINE |
Phillips
Eleva
AUSTIN
Phillips
Eleva
AUSTIN
Phillips
Eleva
TEMPLE
Shimadzu
| DaRt Evolution |
| TEMPLE |
Shimadzu
| DaRt Evolution |
| TEMPLE |
Shimadzu
| DaRt Evolution |
| TEMPLE |
Shimadzu
| DaRt Evolution |
| TEMPLE |
Phillips
DR
TEMPLE
X-Cel
MB 900
TEMPLE
TOSHIBA
Aquilion16
TEMPLE
Toshiba
Aquilion16
WACO
Hologic
Lorad
AUSTIN
Toshiba
Aquilion16
TEMPLE
| General Electric |
| Definium 8 |
WACO
| General Electric |
| Definium 8 |
TEMPLE
TOSHIBA
Aquilion64
TEMPLE
Lunar
| Prodigy Ad |
| AUSTIN |
X-Cel
FB900
AUSTIN
X-Cel
FB900
TEMPLE
Hologic
Selenia 2D
TEMPLE
Toshiba
MRI
TEMPLE
Siemens
| Magnetom E |
| TEMPLE |
Siemens
Espree
AUSTIN
Surgical
| Manufacturer |
| Model |
Location
Phillips
Veradius
TEMPLE
| OrthoScan |
| HD1000 |
TEMPLE
| Liebel-Flarsheim |
| Hydrovision |
TEMPLE
| General Electric |
| OEC 9900 E |
| TEMPLE |
| General Electric |
| OEC 9900 E |
| TEMPLE |
| General Electric |
| OEC 9900 E |
| TEMPLE |
Urology
| Manufacturer |
| Model |
Location
| Liebel-Flarsheim |
| Unspecified |
AUSTIN
| Liebel-Flarsheim |
| Hydrovision |
TEMPLE
*****Please note that the listed machines are subject to change – additional machines may be added, machines may be taken from the list, locations of machines may also change or be added or taken away from the list above due to expansion or decrease in patient population and sites of care within the CTVHCS******
B.4 IT CONTRACT SECURITY
VA INFORMATION AND INFORMATION SYSTEM SECURITY/PRIVACY
1. GENERAL
Contractors, contractor personnel, subcontractors, and subcontractor personnel shall be subject to the same Federal laws, regulations, standards, and VA Directives and Handbooks as VA and VA personnel regarding information and information system security.
2. ACCESS TO VA INFORMATION AND VA INFORMATION SYSTEMS
a. A contractor/subcontractor shall request logical (technical) or physical access to VA information and VA information systems for their employees, subcontractors, and affiliates only to the extent necessary to perform the services specified in the contract, agreement, or task order.
b. All contractors, subcontractors, and third-party servicers and associates working with VA information are subject to the same investigative requirements as those of VA appointees or employees who have access to the same types of information. The level and process of background security investigations for contractors must be in accordance with VA Directive and Handbook 0710, Personnel Suitability and Security Program. The Office for Operations, Security, and Preparedness is responsible for these policies and procedures.
c. Contract personnel who require access to national security programs must have a valid security clearance. National Industrial Security Program (NISP) was established by Executive Order 12829 to ensure that cleared U.S. defense industry contract personnel safeguard the classified information in their possession while performing work on contracts, programs, bids, or research and development efforts. The Department of Veterans Affairs does not have a Memorandum of Agreement with Defense Security Service (DSS). Verification of a Security Clearance must be processed through the Special Security Officer located in the Planning and National Security Service within the Office of Operations, Security, and Preparedness.
d. Custom software development and outsourced operations must be located in the U.S. to the maximum extent practical.
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it.