36C25625Q0406.docx

DOCX document 110 KB Posted

Attached to
R428--IV Room Certification
Solicitation number
36C25625Q0406
Issued by

About this file

This is a Sources Sought notice and Performance Work Statement from the Department of Veterans Affairs seeking information about contractor capabilities to provide IV Room Certification services at the Alexandria VA Health Care System in Pineville, LA. The requirement involves environmental testing and certification of primary and secondary engineering controls used for sterile drug preparation, including testing of laminar airflow workbenches and hazardous drug storage areas in accordance with USP 797, USP 800, CETA and FDA guidelines.

The contract would be for a base year plus four option years, requiring certification services every 6 months including airflow testing, HEPA filter integrity testing, particle count testing, smoke visualization studies, and viable air/surface sampling. Responses to the Sources Sought are due January 29, 2025 at 10:00 AM CST, with questions due by January 27, 2025. All technicians must have current CETA National Board of Testing certification and be classified as Registered Cleanroom Certified Professionals for Sterile Compounding Facilities. The government anticipates a competitive firm-fixed-price contract and will use responses to determine if a small business set-aside is appropriate. Work would be performed during regular business hours Monday-Friday 8:00 AM - 4:30 PM at 2495 Shreveport Highway, Pineville, LA 71360.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SUBJECT*

IV Room Certification

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
77056-5643
SOLICITATION NUMBER*
36C25625Q0406
RESPONSE DATE/TIME/ZONE
01-29-2025 10:00AM CENTRAL TIME, CHICAGO, USA
ARCHIVE
60 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N
SET-ASIDE
SBA
PRODUCT SERVICE CODE*
R428
NAICS CODE*
541990
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

Network Contracting Office 16 5075 Westheimer Road, Suite 750

Houston TX 77056-5643

POINT OF CONTACT*

Contract Specialist Rhonda J Gibson Rhonda.Gibson2@va.gov

832) 352-2915

PLACE OF PERFORMANCE

ADDRESS
Alexandria VAHCS

2495 Shreveport HWY

Pineville LA

POSTAL CODE
71360
COUNTRY
USA

ADDITIONAL INFORMATION

AGENCY’S URL
https://www.va.gov
URL DESCRIPTION
VA Home Page
AGENCY CONTACT’S EMAIL ADDRESS
rhonda.gibson2@va.gov
EMAIL DESCRIPTION
Contract Specialist email

DESCRIPTION

1. Title: IV Room Certification

2. Purpose: The Network Contracting Office (NCO) 16 located at 5075 Westheimer Road, Galleria Financial Center, Suite 750, Houston, Texas, hereby issues the following Sources Sought to Request Information (RFI) with the intent of determining the contractor interest, capabilities, qualifications of potential businesses, nonmanufacturer rule compliance, Buy American Act compliance, Trade Agreement compliance and estimated industry pricing information for market research and estimated budgetary purpose for Television and Highspeed Internet Services.

Contractor shall provide all shipping, installation (if required), equipment, personnel, labor and testing for name brand. This Sources Sought is published for market research purposes. This Sources Sought will be used to determine if a set-aside for any small business program is appropriate. Government anticipates competitive, firm fixed price contract. Please review this announcement, respond to questions below and review all attachments in their entirety.

Government request that interested offerors complete all questions below, provide required documentation, descriptive literature and authorization letter as described below. The Government is in no way obligated to do business with or to enter into any form of contract with any person, firm or other entity that receives or responds to this announcement.

a. Contractors shall provide their point(s) of contact name, address, telephone number, and email address, and the company's business size, and SAM Unique Entity Identification Number. b. Is your company a small business, SDBs, HUBZone, or 8A concern? Please provide proof of qualifications.

b. Does your company meet all required four prong factors for the Nonmanufacturer Rule (NMR) per guidance of SBA and 13 C.F.R. § 121.406 Offer must answer all questions below.

1. Does your company exceed the 1250 employee alternative size standard for nonmanufacturers? ___

2. Is your company primarily engaged in the retail or wholesale trade and normally sells the type of products being supplied? __________

3. Does your company must take ownership or possession of the item(s) with its personnel, equipment, or facilities in a manner consistent with industry practice? __________

4. Does your company supply the end products of a small business manufacturer or processor made in the United States? _____________

c. Is your company considered small under the NAICS Code 541990

d. Do you (offer) manufacture the product yourself or a distributor of the items being researched above?____

e. Are you an authorized distributor/reseller? Provide evidence from original equipment manufacturer (OEM).

f. Do you (offer) supply a product manufactured by another domestic small business, if it is a nonmanufacturer?

g. Do you (offer) supply the product of any sized manufacturer if SBA has granted a waiver to the nonmanufacturer rule?

h. Is your (offer) Buy American Certificate completed in SAM? Please provide evidence with response.

i. Is your (offer) Trade Agreement Certificate completed in SAM? Please provide evidence with response.

j. Information for the name brand or equal product being requested, is it manufactured outside the US-Domestic non-available?

k. Information for the name brand or equal product being requested, is it a foreign manufactured end product under the trade agreement?

l. What is your schedule of delivery after receipt of order for the name brand or equal product?

m. For the name brand or equal product required, Is your company available under any;

Government Wide Agency Contract (GWAC) General Services Administration Schedules (GSA) Indefinite Delivery Indefinite Quantity (IDIQ) Blanket Purchase Agreement (BPA) If so, please list the contract number for name brand or equal product.

n. Provide descriptive literature demonstrating ability to meet all of the requirement/salient characteristics listed if other than brand name product is being offered. Please provide list of contracts for commercial, federal, state, and local governments. Listed contracts shall provide communication on the government requirement illustrating the capability for comparison. to the name brand or equal product meeting the specific requirements per the draft Statement of Work.

o. Contractors shall provide an estimated industry pricing for name brand product for market research and budgetary purpose.

5. Opportunity: NCO 16, is seeking information from potential contractors on their ability to provide this service. THIS IS A SOURCES SOUGHT FOR REQUEST INFORMATION (RFI) ONLY. Small Business Concerns are encouraged to provide responses to this Sources Sought in order to assist the MEDVAMC in determining potential levels of competition available in the industry. Contractor shall possess the capability to provide all requirements and objectives.

6. Instructions and Response Guidelines: Sources Sought responses are due by Jan 29, 2025, at 10:00 am (CST) via email to rhonda.gibson2@va.gov.

All Questions shall be submitted by Jan 27, 2025, at 10:00 pm (CST) via email to Rhonda.gibson2@va.gov.

Telephone requests or inquires will not be accepted.

Public Information will not be provided and shall be utilized to the full extent possible.

This a New Requirement.

The subject line shall read: 36C25625Q0406-RFI-Sources Sought- IV Room Certification – Alexandria VAHCS, Pineville, LA 71360. NO SOLICITATION EXISTS AT THIS TIME. There is no page limitation on subparagraphs 5(a) - 5(o).

Please provide the information you deem relevant in order to respond to the specific inquiries of the Sources Sought. Information provided will be used solely by MEDVAMC as "market research" and will not be released outside of the MEDVAMC Purchasing and Contract Team.

This Sources Sought notice does not constitute a Request for Proposal (RFP), Invitation for Bid (IFB), or Request for Quotation (RFQ), and it is not to be construed as a commitment by the Government to enter into a contract, nor will the Government pay for the information submitted in response to this request. All information contained in this Sources Sought is preliminary as well as subject to modification and is in no way binding on the Government.

In accordance with FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the U.S. Government to form a binding contract. If a solicitation is released, it will be synopsized on Contracting Opportunities (https://beta.sam.gov/) website or GSA. It is the responsibility of the interested parties to monitor these sites for additional information pertaining to this Sources Sought, or future RFP/RFQ.

7. Contact Information:

Contract Specialist, Rhonda Gibson Email address: Rhonda.gibson2@va.gov Your responses to this notice are appreciated.

THIS NOTICE IS NOT A REQUEST FOR COMPETITIVE QUOTES/PROPOSALS; however, any firm that believes it can meet the requirements may provide answers, responses and give written notification prior to the response due date and time. Supporting evidence and responses must be furnished in enough detail to demonstrate the ability to perform the requirements.

***** End Word Document ******* Sources Sought Notice Sources Sought Notice

*= Required Field
Sources Sought Notice

Sources Sought Notice

PERFORMANCE WORK STATEMENT

FOR

ALEXANDRIA VA HEALTH CARE SYSTEM

IV Room Certification of Compliance with USP<797>

Environmental Monitoring Certifications for Alexandria VAHCS

1.0. Description of Service

1.1. This is a non-personnel services contract to provide environmental testing and certification of all primary engineering controls (PECs) and secondary engineering controls (SECs) used for the preparation of compounded sterile preparations (CSPs) and hazardous drugs (HDs) for Alexandria VAHCS (AVAHCS). This contract will be for a Base Year and Four-year Option Periods. The AVAHCS has two non-hazardous sterile drug compounding areas with primary engineering controls (PECs). The PECs include laminar airflow workbenches (LAFW) for all non-hazardous sterile drug compounding. The AVAHCS has a Chemotherapy Storage Closet that stores hazardous medications. The testing and certification must be performed in accordance with USP 797, USP 800, CETA and FDA guidelines.

Floor Plan

2.0. Description/Scope/Objective

2.1. The Contractor must provide all labor, equipment, tools, materials, sampling supplies, replacement parts, transportation, and travel necessary to perform all inspections, testing, certifications as required throughout the contract period, supervision and other items and services necessary to perform the work as defined in this Performance Work Statement.

2.2. Contractor must perform routine work detailed in this Performance Work Statement (PWS) on a regular schedule mutually agreed upon by the Contractor and Contracting Officer Representative (COR), or designee.

2.3. Contractor must provide unscheduled work or emergency requests within 7 days of a request from the COR, or designee. No additional funding will be provided to pay for travel for service.

2.4. The number and type of Primary Engineering Control (PEC) may change throughout the contract period. The number and types of devices are the Governments best estimate of the number and type expected to be serviced by the contract. The Government reserves the right to delete or add units as required. Table A provides the best estimate of the number and type of control expected to be tested, sampled, serviced and certified by the contractor.

2.5. If the numbers of listed items will be exceeded, no work is authorized without the prior written approval of the Contracting Officer (CO) to account for it. Work items or services not already included in this PWS, but relevant to it, may be ordered by the CO without further competition. Only the CO has the right to obligate funding or order work not expressly detailed in this PWS. The Government shall not be held responsible for payment for any work performed by the Contractor that is outside the scope of this contract or without the prior approval of the CO.

3.0. Applicable Documents

3.1. The following laws, regulations, policies, and procedures in effect on date of contract issuance and all subsequent changes or updates apply:

3.1.1. United States Pharmacopeia (USP) at http://www.usp.org/

3.1.2. Controlled Environmental Testing Association (CETA) at https://www.cetainternational.org/

3.1.3. U.S. Food and Drug Administration (FDA) at https://www.fda.gov/

4.0. Performance Requirements

4.1. The vendor shall refer to the most current versions of standards including but not limited to USP, FDA and CETA when conducting the requested monitoring and reporting. The frequencies of testing and specific procedure should be determined by the most updated version of CETA, USP 797 and 800, and per specific facility requirements and/or requests. The vendor shall be responsible for complying with the most updated versions of this guidance and will conduct all sampling per the procedures outlined within the official chapters.

4.1.1. Compounding area certification: The vendor shall provide certification of classified areas using procedures in the current Controlled Environment Testing Association (CETA) application guides. Certification must be performed at least every 6 months and as needed depending on facility requirements or requests, plan for up to two additional certifications a year. Certification must include but is not limited to the following:

4.1.1.1. Airflow testing: The air changes per hour (ACPH) from the HVAC, ACPH contributed from the PEC, and the total ACPH must be documented on the certification report.

4.1.1.2. HEPA filter integrity testing: HEPA filters must be leak tested at the factory and then leak tested again after installation and as part of recertification.

4.1.1.3. Total particle count testing: Must be performed under dynamic operating conditions using current, state-of-the-art electronic equipment. Air sampling sites must be selected in all classified areas. Measurements must be taken in each PEC at locations where there is greatest risk to the exposed CSPs, containers, and closures. Measurements of nonviable airborne particles should be taken at representative locations that reflect the quality of air in the room(s).

4.1.1.4. Smoke visualization studies: Must be performed for each PEC during dynamic operating conditions to demonstrate unidirectional airflow and sweeping action over and away from the preparation(s). Additional smoke tests will need to be performed in rooms without low air returns. A video of the smoke test must accompany the report. Any media files must be provided in USB format to comply with VHA guidance for media. No .zip files or downloadable files from vendor web sites.

4.1.1.5. Primary engineering control (PEC) certification (laminar flow hoods, biological safety cabinets, compounding aseptic isolators, compounding aseptic containment isolators, pass through chambers etc.). The PEC must be certified to meet ISO class 5 at 0.5 μm and larger particles during dynamic operating conditions. PEC certification must include all specifications as outlined per CETA application guides and pertinent USP chapters. This must be performed at least every 6 months and as needed depending on facility requirements.

4.1.1.6. Differential pressure certification: Must be performed at least every 6 months and as needed depending on facility requirements for ante rooms and buffer rooms. Cleanrooms used for non-hazardous compounding must be positive pressure relative to the adjacent spaces. USP chapter <797> requires a minimum pressure differential of 0.020” w.c. relative to adjacent spaces. Cleanrooms used for hazardous applications should be a minimum 0.010” w.c. negative relative to adjacent spaces.

4.1.2. Viable Air Sampling: Each classified area and pass through shall have volumetric active air sampling using an impaction device during dynamic operating conditions at least every 6 months and as needed depending on facility requirements. At least 1 cubic meter or 1000 liters of air should be sampled at each designated location. The media should be examined, and the report should detail the total number of discrete colonies of microorganisms on each plate as CFU per cubic meter of air based on the sample type, sample location and sample date. The genus of any microorganism recovered must be identified with the assistance of a microbiologist if the CFU count exceeds the following: greater than 1 for ISO Class 5, greater than 10 for ISO class 7, and greater than 100 for ISO class 8. If any container demonstrates the presence of CFU that exceed the actionable levels, the preliminary results will be reported to the Chief, Pharmacy Service, the Associate Chief, Pharmacy Operations and Compounded Sterile Products Program Manager as soon as the initial result is known, and final results will be expeditiously reported after further antimicrobial testing is complete.

4.1.3. Viable Surface Monitoring: Each classified area including the interior of the PEC (including equipment contained in it), staging or work areas near the PEC, frequently touched surfaces and pass through chambers must be sampled at least every 6 months. Sampling should be performed at the end of the compounding activities, but before the area has been cleaned and disinfected. The media should be examined, and the report should detail the total number of discrete colonies of microorganisms on each device as CFU per sample based on the sample type, sample location and sample date. The genus of any microorganism recovered must be identified with the assistance of a microbiologist if the CFU count exceeds the following: greater than 3 for ISO Class 5, greater than 5 for ISO class 7, and greater than 50 for ISO class 8. If any device demonstrates the presence of CFU that exceed the actionable levels, the preliminary results will be reported to the Chief, Pharmacy Service, Associate Chief, Pharmacy Operations, and Compounded Sterile Products Program Manager as soon as the initial result is known and final results will be expeditiously reported after further antimicrobial testing is complete.

4.1.4. The contracting officer representative or designee will schedule testing with the vendor. The vendor shall provide the corresponding Pharmacy Service with all service reports and certification of the job performed. After completion of work, the vendor shall notify the contracting officer representative for signature and verification of work performed.

4.1.5. All equipment shall be maintained in proper operating condition as specified by the Manufacturer, USP, CETA and FDA standards. The vendor is to have required supplies, manuals, and schematics, which must be available on site to perform the services. All work is to be performed in accordance with the guidelines established by state and local ordinances with manufacturer’s manual and quality control manual, and with all terms, conditions, provisions, schedules and specifications provided herein.

4.2. In the event that the facility recertification or testing due to actionable results, the vendor must be able to provide retesting within 7 calendar days of the request. If vendor is not able to provide services within the required timeframe, the vendor shall propose a reasonable alternative response time.

4.3. In the event that the samples obtained during testing are damaged during transport or processing, the vendor shall annotate the damage in the report and provide retesting within 7 calendar days of the request at no additional cost to the facility. If the vendor is not able to provide services within the required timeframe, the vendor shall propose a reasonable alternative response time.

5.0. Additional Requirements

5.1. Scheduling of Work. Contractor must provide current documentation of required technician qualifications. Contractor must also submit a current copy of their most recent standard operating procedures (SOPs) for testing, troubleshooting, repair and certification including documentation of certification equipment calibration to National Institute of Standards and Technology (NIST)-traceable primary standards, a safety plan to include job hazard analysis, respiratory protection program, hazard communication and safety data sheets (SDS) for all chemical products that will be used during the course of this contract prior to scheduling work. Contractor must contact the COR, or designee to schedule work, prior to beginning work, and at the conclusion of all work. For services performed during normal working hours, the contractor's technician is to report to the COR, or designee to report/sign-in and obtain an identification badge, which shall be worn at all times while the technician is on station. Work must be performed during the facilities regular administrative working hours, Monday through Friday, 8:00 a.m. - 4:30 p.m. excluding- weekends and Federal holidays. Exception to this will be given when deemed necessary to accommodate evening, night, and weekend staff. Contractor must receive approval from the COR, or designee for irregular work hours testing.

5.2. Service Reports. Contractor must provide a certificate sticker for each unit that passes certification before the technician leaves the facility. Any unit that does not pass certification will be marked as such with clear signage that the unit is out of service until further notice. A complete report of services for each unit serviced and must include a listing of replacement parts, when applicable. Contractor must also contact the COR, or designeee upon completion of work to verbally report the status of the unit and any deficiencies identified. The COR, or designee, must sign the service report before the technician(s) leave the facility. Failure to follow these procedures may result in delay of payment or nonpayment. Pass/Fail notifications should be included on a per test basis where applicable. Larger formal reports as required for the pharmacy clean rooms that include viable sampling and surface sampling kits for hazardous drug detection must be submitted to the COR, or designee within fifteen (15) business days following the collection of the viable sampling and /or surface sampling kit. Results of viable samples and/or surface sampling kit must be submitted to the COR, or designee within one (1) business day of the receipt of the results from an accredited laboratory. The service reports from the contractor must contain, at a minimum, the following information:

5.2.1. Room number and/or location

5.2.2. Room type (e.g. cleanroom)

5.2.3. Date and time of the contractor's arrival on station

5.2.4. Certified technician(s) name

5.2.5. Date of testing

5.2.6. Standards used to test PEC and/or SEC

5.2.7. Diagram of room showing the location of supply/exhaust diffusers, equipment, room dimensions, etc

5.2.8. Total room area and volume

5.2.9. Room humidity and temperature

5.2.10. Air changes per hour

5.2.11. Pressure differential

5.2.12. Particle counts

5.2.13. Diagram with labeling of nonviable particle count locations

5.2.14. Viable particle sampling certificate of analysis for growth and sterility controls analysis and interpretation

5.2.15. Diagram with labeling of viable particle sampling and environmental wipe sampling test locations in room

5.2.16. Report of actions performed

5.2.17. Notes or recommendations

5.2.18. Date of next required certification

5.2.19. Service reports must also include the following as applicable to type of PEC/SEC certification or personnel assessment and testing:

5.2.19.1. List of vendor equipment used; to include manufacturer make, model, serial number, date calibrated, and date of next calibration.

5.2.19.2. SDS forms for all agents used during the certification procedures.

5.2.19.3. Certificate of analysis

5.2.19.4. Growth Control

5.2.19.5. Sterile Control

5.3. Excluded services. Contractor must not be obligated to provide services under this agreement for damage to or destruction of instrument(s) covered where such damage or destruction is a result of or caused by:

5.3.1. Fire or explosion of any origin, riot, civil commotion, aircraft, war

5.3.2. Act of God or other force majure including but not limited to lightning, windstorm, hail, flood, earthquake

5.3.3. VAMC misuse or abuse of such instrument

5.3.4. Decontamination due to spillage by VA personnel due to their own negligence

5.3.5. Installation of unauthorized field modifications.

5.4. Schedule of Insurance. Contractor must procure and maintain for the entire contract period, Workmen's Compensation and employer's general liability Insurance. Prior to contract award, the Contractor shall provide proof of insurance for the entire contract period. The parties agree that such personnel shall not be considered a VA employee for any purpose and shall be considered employees of the contractor.

5.5. Price Schedule. The contract rate is an all-inclusive rate. No separate fees will be allowed for such incidental costs of malpractice insurance coverage, supplies, equipment, etc.

5.6. Environment of Care. Contractor must maintain a clean and safe environment always.

5.7. Contracting with Parties Listed on the OIG List of Excluded Individuals/Entities. In accordance with HIPAA and the Balanced Budget Act (BBA) of 1977, the Office of Inspector General has established a list of parties and entities excluded from Federal health care programs. Specifically, the listed parties and entities may not receive Federal payments due to fraud and/or abuse of the Medicare and Medicaid programs. The list of excluded individuals/entities can be found on the OIG web site at www.hhs.gov/oig. With submittal of proposal, the contractor shall provide evidence than none of the contractor’s employees are on the Excluded Parties list.

5.8. Invoices/Payments. This provision shall survive the termination or ending of the contract. Payment for services rendered by the Contractor under the terms of this contract will be made upon receipt of a properly executed invoice. Invoice to be submitted within fourteen (14) workdays following completion of the Contractors rendered services. VA will verify services performed and certify invoice for payment within thirty (30) days following date of receipt of sampling reports. Any identified discrepancy(s) in billings shall be cause for extending provisions of this paragraph until the discrepancy or other identified problem with the billing invoice has been resolved. Contractor must:

5.8.1. Be electronically prepared and submitted in accordance with FAR 52.212-4(g)

5.8.2. Clearly state the terms of any discounts offered.

5.8.3. Include the current fiscal year and purchase order number.

5.8.4. Include the period of performance covered by the invoice.

5.8.5. Accept payment for services rendered under this contract as payment in full. This provision shall survive the termination or ending of the contract.

5.9. Place of Performance. Contractor must furnish services for the AVACHS located at 2495 Shreveport Hwy 71 North, Pineville, LA 71360.

5.10. Authority: This procurement is executed under FAR Part 12 – Simplified Acquisitions

5.11. Information Security Officer, Information Protection: The contractor will not have access to VA Desktop computers, nor will they have access to online resources belonging to the government while conducting services. If removal of equipment from the VA is required, any memory storage devices, such as hard drives, solid state drives and non-volatile memory units will remain in VA control and will not be removed from VA custody.

5.12. Privacy Officer: The contractor will not have access to Patient Health Information (PHI) and they will not have the capability of accessing patient information during the services provided to the VA and if removal of equipment from the VA is required, any memory storage devices, such as hard drives, solid state drives and non-volatile memory units will remain in VA control and will not be removed from VA custody. All research data available for Contractor analyses is de-identified.

5.13. Records Officer: The following standard items relate to records generated in executing the contract and should be included in a typical procurement contract:

5.13.1. Citations to pertinent laws, codes and regulations such as 44 U.S.C chapters 21, 29, 31 and 33; Freedom of Information Act (5 U.S.C. 552); Privacy Act (5 U.S.C. 552a); 36 CFR Part 1222 and Part 1228.

5.13.2. Contractor shall treat all deliverables under the contract as the property of the U.S. Government for which the Government Agency shall have unlimited rights to use, dispose of, or disclose such data contained therein as it determines to be in the public interest.

5.13.3. Contractor shall not create or maintain any records that are not specifically tied to or authorized by the contract using Government IT equipment and/or Government records.

5.13.4. Contractor shall not retain, use, sell, or disseminate copies of any deliverable that contains information covered by the Privacy Act of 1974 or that which is generally protected by the Freedom of Information Act.

5.13.5. Contractor shall not create or maintain any records containing any Government Agency records that are not specifically tied to or authorized by the contract.

5.13.6. The Government Agency owns the rights to all data/records produced as part of this contract.

5.13.7. The Government Agency owns the rights to all electronic information (electronic data, electronic information systems, electronic databases, etc.) and all supporting documentation created as part of this contract. Contractor must deliver sufficient technical documentation with all data deliverables to permit the agency to use the data.

5.13.8. Contractor agrees to comply with Federal and Agency records management policies, including those policies associated with the safeguarding of records covered by the Privacy Act of 1974. These policies include the preservation of all records created or received regardless of format [paper, electronic, etc.] or mode of transmission [e-mail, fax, etc.] or state of completion [draft, final, etc.].

5.13.9. No disposition of documents will be allowed without the prior written consent of the Contracting Officer. The Agency and its contractors are responsible for preventing the alienation or unauthorized destruction of records, including all forms of mutilation. Willful and unlawful destruction, damage or alienation of Federal records is subject to the fines and penalties imposed by 18 U.S.C. 2701. Records may not be removed from the legal custody of the Agency or destroyed without regard to the provisions of the agency records schedules.

5.13.10. Contractor is required to obtain the Contracting Officer's approval prior to engaging in any contractual relationship (sub-contractor) in support of this contract requiring the disclosure of information, documentary material and/or records generated under, or relating to, this contract. The Contractor (and any sub-contractor) is required to abide by Government and Agency guidance for protecting sensitive and proprietary information.

5.14. Security: Contractor must be liable for security of all areas accessed by the technician(s). All doors, passageways, windows and accesses to equipment and service areas, must be locked and/or secured after passage. In no instance will any door, window or access be rendered unlocked, propped open, mechanism taped open or rendered to prevent securing. Contractor must be responsible to physically ensure such accesses are secured to prevent unauthorized entry.

5.15. Contractor Personnel Badges and Parking: Contractor technician must acquire an access badge upon entrance into the AVAHCS buildings. Contractor must safeguard the access badge and immediately report any lost, stolen, or destroyed badges to the facility COR, or designee. Contractor must ensure technicians properly display their access badges. Access badges must be worn at or above the waist (facing forward). Access badges must be returned to the COR, or designee at the end of each site visit. Contractor must ensure technician parks in the appropriate designated parking areas. Parking information shall be coordinated with the COR, or designee. Contractor must be aware that AVAHCS does not validate or make reimbursement for parking violations of the contractor technician under any circumstance.

5.16. Interference to Normal Function: Contractor may be required to interrupt their work at any time so as not to interfere with the normal functioning of the facility, including utility services, fire protection (detection and suppression) systems, passage of facility patients, personnel, equipment and carts and patient care activities (e.g., preparing sterile compounds). In the event of an emergency, contractor services may be stopped by the AVAHCS and rescheduled at no additional cost to the government. Contractor personnel shall inform the COR or the designee of the need to gain access to secured areas. If access is required to secure areas, prearranged scheduling will be made with the COR, or designee.

6.0. Performance Location and Hours

6.1. AVAHCS, 2495 Shreveport Hwy 71 North, Pineville, LA 71360.

6.2. Hours are from 0730 to 1600, Monday through Friday, excluding Federal holidays.

7.0. Deliverables

7.1. Report will include all components specified in 4.1.1 through 4.1.8. The structure of the report should follow specifications outlined in the CETA application guides.

7.2. Report of all testing results shall be sent via e-mail with a hard copy sent by mail to the Chief, Pharmacy Service/Contracting Officer Representative.

8.0. Contractor Personnel and Management

8.1. Contractor Personnel: The Contractor shall be responsible for employing technically qualified personnel to perform the work specified in this PWS. The Contractor shall maintain the personnel, organization and administrative control necessary to ensure that the work delivered meets the contract specifications and requirements.

8.2. Key Personnel: All work shall be completed by CETA (CNBT) and NSF Certified technicians in accordance with all applicable Federal, State, Local, and industry: standards, laws, policies, procedures, codes and any other applicable governing body for the proper performance of all equipment per manufactures specifications/calibrations. Personnel performing all certifications must have current CETA National Board of Testing (CNBT) accreditation and hold classification as a Registered Cleanroom Certified Professional for Sterile Compounding Facilities (RCCP-SCF). Proof of certification must be provided prior to the start of the contract. The contractor must provide written notification to the Contracting Officer & Contracting Officer Representative (COR) if RCCP-SCF certification lapses or is terminated.

8.3. Contractor Badges. Contractor personnel shall wear a laminated contractor identification badge at all times when performing services under this Contract, including when on Government directed travel. The badge will contain a personal picture, name of employee, and Contractor's name. Badges shall be worn on the outer garment attached to the outer shirt or jacket pocket by a button or clip or worn around the neck secured by an appropriate identification card lanyard. They must also ensure that all documents or reports produced by Contractor personnel are suitably marked as Contractor products or that Contractor participation is appropriately disclosed.

8.4. Standards of conduct:

8.4.1. Conduct of Personnel. Contractor personnel shall conduct themselves in a professional manner (i.e., timeliness, communication; spoken and written etc.). The Contracting Officer may require the Contractor to remove from the job site any employee working under this contract for reasons of suspected misconduct, a suspected security breach, or suspected to be under the influence of alcohol, drugs, or any other incapacitating agent. Contractor employees shall be subject to dismissal from the premises upon determination by the Contracting Officer that such action is necessary in the interests of the Government. The medical center director or designated representative has the authority to bar individuals from the Government facility. The removal from the job site or dismissal from the premises shall not relieve the Contractor of the requirement to provide sufficient personnel to perform the services as required by this performance work statement. The Government shall not reimburse the Contractor for travel and other expenses associated with the removal of personnel.

8.4.2. Contractor personnel are not authorized to carry or possess personal weapons to include, but not limited to, firearms and knives with a blade length in excess of three inches, while assigned under this contract.

8.4.3. Working Attire and Appearance: Contractor personnel shall present a professional appearance commensurate with standards delineated for Government civilian personnel acting in similar capacities.

9.0. Travel: N/A

10.0. Non-Personal Service Statement

10.1. Contractor employees performing services under this order will be controlled, directed, and supervised at all times by management personnel of the contractor. Contractor management will ensure that employees properly comply with the performance work standards outlined in the PWS. Contractor employees will perform their duties independent of, and without the supervision of any Government official. The tasks, duties, and responsibilities set forth in the contract may not be interpreted or implemented in any manner that results in any contractor employee creating or modifying Federal policy, obligating the appropriated funds of the United States Government, overseeing the work of Federal employees, providing direct personal services to any Federal employee, or otherwise violating the prohibitions set forth in Parts 7.5 and 37.1 of the Federal Acquisition Regulation (FAR). The Government will control access to the facility and will perform the inspection and acceptance of the completed work.

11.0. Performance Standards and Quality Measurement

11.1. Performance standards define desired services. The Government performs surveillance to determine if the Contractor exceeds, meets, or does not meet these standards. The Government shall use these standards to determine Contractor performance and shall compare Contractor performance to the Acceptable Quality Level (AQL).

Performance Standards Requirements Summary Matrix

Performance Objective
Performance Standard
Acceptable Quality Level
Surveillance Method
Incentive
Dis-incentive

Viable air and surface sampling, non-viable testing, and hood/room certification testing completed All testing is completed prior to due date 99% acceptable level Random inspection conducted by the COR Favorable Contractor Performance Evaluations Below AQL, non-acceptance of services. A Contract Discrepancy Report (CDR), will be issued for performance below the AQL. The Contractor will have five workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing the CDR, the Contractor will be re-rated. If performance does not comply with PWS, the CO will be notified, the performance record will be documented, and CO may pursue contract action.

Reports
Receive field

action report.

Receiving documents are accurate 98% of the time.

Random inspection conducted by the COR Favorable Contractor Performance Evaluations Below AQL, non-acceptance of service. A contract Discrepancy Report (CDR), will be issued. Contractor will have 5 workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing CDR, the contractor will re-rated. If performance does not comply with PWS CO may deduct from invoice 2%, 3%, 4% progressive deduction.

Certifying Personnel
Submission of

proof of current CNBT certification.

All technicians performing work maintain current

CNBT

certification as a Registered Cleanroom Certified Professional for Sterile Compounding Facilities

(RCCP-SCF)

Random inspection conducted by the COR Favorable Contractor Performance Evaluations Below AQL, non-acceptance of services. A Contract Discrepancy Report (CDR), will be issued for performance below the AQL. The Contractor will have five workdays to provide a response that includes an improvement strategy for addressing the issues. 30 days after issuing the CDR, the Contractor will be re-rated. If performance does not comply with PWS, the CO will be notified, the performance record will be documented, and CO may pursue contract action.

Table A

Room Location
Primary Engineering Control
Room
IV Area Non-Hazardous
LAFW - MFG: Nuaire

Model: NU-301-530 Serial# 181959091417 IV Room

Chemo Storage Closet
None
Hazardous Storage Area
Contingency Area
LAFW – MFG: Nuaire

Model: NU- 340-530 Serial# 196931021920 Contingency SCA

Page 1 of

Page 1 of image1.png image2.png

File details come from the government source that posted it. Updated .