Request For Information
1. Will the sprinkler piping and materials in this area need to be of a none-metallic materials?
If so, will it need to have its own control valve, water flow and tamper switches to monitor the area? Or will the existing metal pipe be left as is?
Sprinklers and plumbing should be left as is; no plumbing work is to be included in this.
2. Could you provide the electrical specs for the equipment? Voltage and amperage.
Electrical Requirements
M-Cabinet CXA 80 Supply Configuration: 3 phase, 3 wire power, neutral and ground. Wye.
or 3 phase, 3 wire power, ground. Delta.
Nominal Line Voltage: 400, 440, 460 or 480 VAC, 60 Hz Branch Power Requirement: 112.5 kVA minimum Circuit Breaker: 3 pole, 50 Amps (@ 480V) (Slow Blow) Residual Current Device to be installed depending on local regulations.