This is a Sources Sought announcement issued by the Department of Veterans Affairs, Network Contracting Office -15, for ophthalmic equipment needed at the Harry S. Truman VA Medical Center's new Multi-Specialty Community-Based Outpatient Clinic (CBOC) in Rolla, Missouri.
The VA seeks three primary pieces of ophthalmic equipment: a photographic slit lamp, an automated ophthalmic perimeter with powered adjustable height table, and an optical biometer with corneal topography/pachymetry capabilities. The slit lamp must feature magnification ranges of 6x/10x/16x/25x/40x with LED light source, multiple filter options, continuously adjustable slit width (0-12mm) and length (1-12mm), and FORUM network compatibility. The automated perimeter requires seven fixation aids, 90-degree maximum temporal range, multiple threshold test strategies (including SITA Standard, SITA Fast, SITA Faster, Full Threshold, and SITA-SWAP), and progression analysis capabilities with Guided Progression Analysis functionality. The optical biometer must provide axial length measurements (14-38mm), full-length OCT imaging, swept-source biometry at 2,000 scans per second with Haigis calculation formulas, multiple imaging modes including fluorescein angiography and autofluorescence, and widefield imaging up to 267 degrees with montage capability.
All equipment must be FORUM-compatible, comply with current and future VA EHR requirements, meet federal cybersecurity standards including BitLocker encryption and TLS protocol FIPS mode validation, and include manufacturer's two-year standard warranty. Deliverables include six CLINs: the slit lamp, automated perimeter, optical biometer, software interface, white-glove installation service (not later than January 31, 2027), and training completed within 45 days of final installation with recorded electronic copies provided to the VA. Delivery is required 90 days after receipt of order to the Columbia, Missouri location. Interested vendors must respond with company information, UEI number, GSA contract status, capability statement, references with contract numbers, product brochures with specifications, authorized service provider letters from OEM, and country of manufacture data, referencing announcement number 36C25526Q0396. Prospective contractors must register in SAM.gov and complete electronic representations and certifications; Service-Disabled Veteran Owned Small Businesses and Veteran Owned Small Businesses must be verified in SBA's Small Business Search system. Point of contact is Amanda Williams at amanda.williams16@va.gov.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
This is a SOURCES SOUGHT announcement only. It is neither a solicitation announcement nor a request for proposal or quote and does not obligate the Government to award a contract. Responses to this Sources Sought must be in writing.
The purpose of this Sources Sought Announcement is for market research, to make appropriate acquisition decisions, and to gain knowledge of potentially qualified businesses capable of providing the following:
Ophthalmic Equipment *** See Statement of Requirement below for details ***
If this is within your capability, please respond with the information below:
• Company Name
• Address
• Point of Contact
• Phone, Fax, and Email
• UEI number
• GSA Contract (as applicable)
• Capability Statement that clearly and convincingly identifies your ability to fulfill this requirement.
• References, with contract numbers, demonstrating the successful completion of work like the work required in this Sources Sought, performed in the past.
• Brochures of products being offered, to include specifications.
• Authorized service provider letter from OEM.
• Country of manufacture for each piece of equipment.
Interested firms are reminded that in accordance with FAR 4.203, prospective contractors shall complete electronic annual representations and certifications to be considered for award. In addition, FAR 4.203-1(b) states that prospective contractors shall be registered in the System for Award Management (www.SAM.gov) database. All Service-Disabled Veteran Owned Small Businesses (SDVOSB) and Veteran Owned Small Businesses (VOSB) shall be verified in Small Business Search at https://search.certifications.sba.gov/.
Please include reference number 36C25526Q0396 in all correspondence regarding this Sources Sought.
Point of Contact for this Sources Sought is:
amanda.williams16@va.gov Department of Veterans Affairs Network Contracting Office -15 3450 S. 4th Street Trafficway Leavenworth, KS. 66048 https://search.certifications.sba.gov/ mailto:amanda.williams16@va.gov
STATEMENT OF REQUIREMENT
Ophthalmic Equipment
OBJECTIVE
The Harry S. Truman VA Medical Center requires a Photographic Slit Lamp, Automated Ophthalmic Perimeter, and Optical Biometer for the new Multi-Specialty Community- Based Outpatient Clinic (CBOC) in Rolla, MO.
TECHNICAL REQUIREMENTS
1. Slit Lamp Photographic
a. Magnification: 6x/10x/16x/25x/40x, with 10x eyepiece, compensation of ametropia ±8 dpt (dioptric power)
b. Interpupillary distance: Convergent tube 50 to 84 mm; Parallel tube 52 to 78 mm
c. Light source: LED
d. Slit illumination: cold-white, and warm-white options.
e. Filters: Blue, halogen, green (red-free), red, and diffusor
f. Slit width: Continuously 0 to 12 mm
g. Slit length: Continuously 1 to 12 mm
h. Slit diameter: 0.2/1/3/5/9/12 mm
i. Slit rotation: Continuously 0° to 180°; with click stop at 45°/90°/135°
j. Slit decentration (horizontal): fixable at 0°
k. Slit inclination (vertical): 0°/5°/10°/15°/20°
l. Swivel range of slit projector: >200° with scale; with click stop at
0°/±45°/±60°
m. Operation: Control panel with joystick, brightness control; Controls for
Auto View & Quick Stop
n. Adjusting range: 110 mm (side), 30 mm (height), 110 mm (length)
o. Weight: 9 to 14 kg (incl. headrest)
p. Dimensions (W x H x D): 300-400 mm x 655 mm ± 15 mm x 395 mm.
q. Power supply: 100 - 240 V AC, 50/60 Hz
r. FORUM compatible with current facility network
s. Compatible with current and future EHR and all federal cyber security requirements.
2. Automated ophthalmic perimeter.
a. Unit has on screen display for rapid and accurate alignment of patient’s automatic pupil size measurement, seven different fixation aids, ergonomic chin rest, several threshold strategies, and high target intensity.
b. Must include Powered Adjustable Height Table.
c. Maximum temporal range: 90 degrees
d. Stimulus duration max: 200 milliseconds
e. Visual field-testing distance: 30-33 cm
f. Bowl Illumination: 31.5 ASB
g. Threshold Test library must include the following tests: 24-2, 30-2, 10-
2, 24-2C & 60-4, Nasal step, Macula
h. Threshold test strategies must include the following: SITA Standard, SITA Fast, SITA Faster, Full threshold and SITA-SWAP
i. Suprathreshold test library must include: C40, C76, C80c, C64, C-
Armaly, Peripheral test patterns
j. Suprathreshold test modes must include: Age corrected, Threshold related, Single intensity
k. Specialty test library must include: Social Security Disability, monocular, binocular, Esterman monocular, binocular, superior 36, 64, Custom Static patterns
l. Fixation control must include: Blind spot monitor, video eye monitor, gaze tracking x, head tracking x
m. Stimulus must include: white-on-white, red-or blue-on white x
n. General testing features must include: Stimulus sizes, foveal threshold testing, automatic pupil measurement
o. Single Field Analysis must include: STATPAC SFA, Total deviation, pattern deviation, mean deviations, glaucoma Hemifield Test (GHT)
p. Progression Analysis must include: Guided Progression Analysis, Mixed SITA, 24-2C specific GPA, Trend Analysis, VFI, Event Analysis, Glaucoma Change Probability Map
q. FORUM Data export that are compatible with the VA’s current and future EHR.
r. Remote assistance for customer service.
s. Cybersecurity: Configurable BitLocker encryption & FORUM Transport
Layer Secure (TLS) protocol Federal Information Processing Standards (FIPS) mode validated.
t. Meets all Federal cyber security requirements.
3. Optical Biometer.
a. Corneal topography/pachymetry system.
b. Provides a full-length OCT image showing anatomical details on a longitudinal cut through the entire eye.
c. Axial length measurement 14 to38mm, corneal radii 5 to11mm, anterior chamber depth 0.7 to 8mm, white to white 8 to16mm.
d. IOL calculation formulas (Haigis formula, the Haigis-L formula for post-
LVC cases, and the new Haigis-T for toric) via SWEPT Source Biometry with 2,000 scans per second. (Haigis formula, the Haigis-L formula for post-LVC cases, and the new Haigis-T for toric)
e. IOL Data interface for electronic medical record (EMR) / patient management systems (PMS) that is compliant with current and future
VA EHR
f. Data exports to USB storage media.
g. FORUM compatible
h. Ethernet port for network connection and network printer.
i. Imaging Modes:
i. True Color (with Red, Green and Blue channel split)
ii. Fluorescein Angiography
iii. Autofluorescence-Green
iv. Autofluorescence-Blue
v. Infrared reflectance
vi. External eye image (ocular surface)
vii. Stereo
j. Field of view:
i. Widefield of one image: 133 degrees
ii. Ultra-widefield of two images: 200 degrees
iii. Montage up to six images: up to 267 degrees
k. Resolution at least 7.3 micrometers
l. Minimum pupil diameter: 2.5 mm
m. Working Distance: 25 mm
n. Compensation for ametropia: - 24 D to + 20 D continuous
o. Light Sources:
i. Red LED: 585 to 640 nm
ii. Green LED: 500 to 585 nm
iii. Blue LED: 435 to 500 nm
iv. Infrared laser diode: 785 nm
p. Automatic Operations:
i. Auto-focus: Auto Montage
ii. Auto-gain: Auto-laterality
q. Acquisition speed:
i. Live IR Preview: 10 frames/second
ii. Image Capture: less than or equal to 0.2 seconds
r. Must include monitor.
4. Software Interface
a. Must be FORUM compatible.
b. Meets current and future EHR requirements.
c. Meets all federal cybersecurity requirements.
DELIVERABLES
CLIN 0001 – Lamp, Slit, Photographic
QTY: 1 EA
• In accordance with Statement of Requirement – Technical Requirements
• Manufacturer’s two-year standard warranty
• Two (2) sets of user manuals and technical manuals
CLIN 0002 – Perimeter, Automated, with Table
• In accordance with Statement of Requirement – Technical Requirements
CLIN 0003 – Optical Biometer
• In accordance with Statement of Requirement – Technical Requirements
CLIN 0004 – Software
• In accordance with Statement of Requirement – Technical Requirements
CLIN 0005 – Installation – White Glove Service
QTY: 1 JB
• NLT January 31, 2027
CLIN 0006 – Training
QTY: 1 JB
• Completed within 45 days after final installation of equipment
• Training is to be recorded, and an electronic copy provided to VA Point of
Contact for use in training additional personnel.
DELIVERY SCHEDULE
90 days after Receipt of Order
DELIVERY LOCATION
Harry S Truman Memorial Veterans Hospital Multi-Specialty CBOC – Rolla 2701 Vandiver Dr.
Columbia, MO 65202