36C25519Q0060-0002000.docx

DOCX document 55 KB Posted

Attached to
589-19-1-4805-0004 - Surgical Light Booms for OR -WI
Solicitation number
36C25519Q0060
Issued by

About this file

36C25519Q0060 0002 36C25519Q0060 0002.docx

View the file

Other files

File Type Posted
36C25519Q0060-0003000.docx DOCX document
36C25519Q0060-0001000.docx DOCX document
36C25519Q0060-0001003.pdf PDF
36C25519Q0060-0001002.pdf PDF
36C25519Q0060-0001001.pdf PDF
36C25519Q0060-002.docx DOCX document
36C25519Q0060-001.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

5. PROJECT NUMBER (if applicable)

CODE

7. ADMINISTERED BY

2. AMENDMENT/MODIFICATION NUMBER

CODE

6. ISSUED BY

8. NAME AND ADDRESS OF CONTRACTOR

4. REQUISITION/PURCHASE REQ. NUMBER

3. EFFECTIVE DATE

9A. AMENDMENT OF SOLICITATION NUMBER

9B. DATED

PAGE OF PAGES

10A. MODIFICATION OF CONTRACT/ORDER NUMBER

10B. DATED

BPA NO.

1. CONTRACT ID CODE

FACILITY CODE

CODE

Offers must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended, by one of the following methods:

The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offers

E. IMPORTANT:

is extended,

(a) By completing Items 8 and 15, and returning __________ copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted; or (c) By separate letter or electronic communication which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY is not extended.

12. ACCOUNTING AND APPROPRIATION DATA

(REV. 11/2016)

is required to sign this document and return ___________ copies to the issuing office.

is not, A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE CONTRACT ORDER NO. IN ITEM 10A.

15C. DATE SIGNED

B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(b).

RESULT IN REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by letter or electronic communication, provided each letter or electronic communication makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.

C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:

D. OTHER

Contractor

16C. DATE SIGNED

14. DESCRIPTION OF AMENDMENT/MODIFICATION

16B. UNITED STATES OF AMERICA

Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.

15A. NAME AND TITLE OF SIGNER

16A. NAME AND TITLE OF CONTRACTING OFFICER

15B. CONTRACTOR/OFFEROR

STANDARD FORM 30

PREVIOUS EDITION NOT USABLE

Prescribed by GSA - FAR (48 CFR) 53.243 (Type or print) (Type or print) (Organized by UCF section headings, including solicitation/contract subject matter where feasible.)

(Number, street, county, State and ZIP Code) (If other than Item 6) (Specify type of modification and authority) (such as changes in paying office, appropriation date, etc.)

(If required)

(SEE ITEM 11)

(SEE ITEM 13)

(X)

CHECK

ONE

13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS,

IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.

11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS

AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT

(Signature of person authorized to sign) (Signature of Contracting Officer) 11-19-2018 589-19-1-4805-0004 n/a 36C255 Department of Veterans Affairs Network Contracting Office (NCO) 15 3450 S 4th Street Trafficway Leavenworth KS 66048 36C255 Department of Veterans Affairs Network Contracting Office (NCO) 15 3450 S 4th Street Trafficway Leavenworth KS 66048 To all Offerors/Bidders

36C25519Q0060

X X X

See CONTINUATION Page X The purpose of this amendment is to;

1) Post Questions and Answers

2) Update the Statement of Requirements and CLINS accordingly.

3) Extend the response date to Wednesday, November 28 th @ 10:00 am

*ALL CHANGES ARE HIGHLIGHTED IN RED*

Larry Zaritz

Questions and Answers for RFQ #36C25519Q0060:

1. Tables 3-5 indicate a requirement for a single universal port which is the current standard for connectivity to multiple sources, table 8 contradicts this requirement and requires individual connections to sources. Is a single universal port the requirement for this request?

3 universal laparoscopic images clownology images

1. Specific request is for HD camera requirement. The healthcare industry now provides 4K options to provide superior image clarity to improve clinical diagnoses. Would the government consider expanding the scope to include 4K throughput capabilities to align with the current technological evolution?

Yes 4k

1. There is no specific requirement for HL7 and DICOM connectivity in the RFQ, this is typically a requirement to meet to the VA CIO’s interoperability initiatives. Would the government consider expanding the scope to include HL7 and DICOM connectivity for the solution provided?

Yes

1. As outlined in the 2017 VA Medical Device Protection Program (MDPP) and subsequent implementation of a Medical Device Isolation Architecture (MDIA), it is our understanding that a VA Directive 6550, Pre-Procurement for Medical Device/Systems, is required to be completed for any medical device that is otherwise used to capture and store patient information. Can you confirm VA Form 6550 will need to be completed in our submission?

Yes

1. Related to VA Form 6550, it is our understanding that any projects in the VA involving the acquisition of medical device equipment that captures patient information should be listed and approved in the VA ONE-VA Technical Reference Model (TRM) and comply with the architectural guidance as outlined in the 2017 VA Medical Device Protection Program (MDPP). Can you confirm any proposed solution must be currently listed in the TRM?

Trm is outdated and unreliable

1. What is the current environment/integration system in place with the other ORs in the facility? (brand/installation date)

5. Is connectivity to any existing systems, such as imaging carts that support ENT, Urology, etc., a requirement for this request?

Olympus, Yes to laberscopic scopes, colorology scopes and GE c amrs

1. Would the government consider including a requirement for booms to support the ability to integrate patient lifts in alignment with VA Safe Patient Handling Mobility Directive 1611?

No

1. Table 6 request is for a 27” bracket with a smaller monitor requested in Table 9, would the government accept a 27” monitor and a 27” bracket?

Yes

1. The Statement of Requirements Introduction references installation of “two surgical lights, two equipment booms, and two anesthesia booms for expansion of surgery to accommodate two new operating rooms.” In contrast, the drawings indicate the need for the following ceiling mounted equipment (per OR):

-1 anesthesia boom -1 equipment boom -3 LED surgical lights -2 arms to support -2 19-26” monitors. (IAW Other Questions, Max Size changed to 27”, Table #9) Your CLINS/ Price Schedule request the following:

-6 LED Lights per each Operating Room (CLINS 2, 5, 8) -2 Monitor Arms -1 Equipment Boom -1 Anesthesia Boom

1. “II Booms” references “infrastructure.” We interpret that to mean an above-ceiling support structure. Please confirm if that support structure is to be quoted, as it is not listed in the CLIN.

General Contractor is putting the supports to mount lights

1. “II Booms” references a Perfusion hookup and Table 5 lists requirements for a perfusion hookup arm. Please confirm if that is to be quoted as it is not listed in the CLIN or called out on the drawings.

It should be capable if we need to add it later. But as of right now no.

1. The Description of Work, Statement of Requirements, and Table 8 states that an integration system is a requirement. Please confirm if that is to be quoted as it is not listed in the CLIN.

We do need the integration system

1. Table 8 states a requirement for a video capture device that “must capture video and images on screen.” Please confirm if that is to be quoted as it is not listed in the CLIN.

Yes

1. The Statement of Requirements states the head shall have the ability to power on/off the light. Where should prospective offerors include the required items under the current CLINS structure?

Mount in the OR room on wall

1. Please confirm the number and type of video inputs.

1. Does the facility require room-to-room or off-site connectivity?

No

1. Table 10 states the vendor shall mount and install all products supplied using union labor. Shall the union labor portion of the cost be separate from the installation line item?

We wont be using union labor

STATEMENT OF REQUIREMENTS (SOR)

*ALL ITEMS HEREIN THAT MENTION BRAND NAME PRODUCTS MAY BE SUBSTITUTED FOR A LIKE OR SIMILAR ITEM*

1. INTRODUCTION

This solicitation is to provide the Veterans Health Administration, Wichita Facilities Management with all labor, personnel, equipment, tools, materials, parts and services necessary install three surgical lights, one equipment boom, and one anesthesia boom per room for the expansion of surgery to accommodate two new operating rooms.

2. BACKGROUND

The Wichita Robert J. Dole Veterans Health Administration Facilities Management Each room will require new surgical lighting, ceiling-mounted equipment management system (booms), and a digital OR integration system to improve information access and modernize the surgery suites in Building 34.

3. SCOPE OF WORK

WORK TO BE PERFORMED:

Description of Work: The contractor shall provide surgical lights and ceiling-mounted equipment management system(booms), and a digital OR integration system. The single-phase project of two ORs (General OR Rooms) shall provide installation to include, labor, training, tools, documentation of performance, and perform all field modification needed to complete install. More details can be seen in the minimum technical requirements section and a table is available at the bottom of this document

1. Surgical Lights The term “surgical lights” is used to describe the light head, including the light bulbs, wiring and housing that provides direct illumination over the patient during surgery. Each surgical light head is affixed to an arm that is mounted to the ceiling.

1. Light head – surgical grade light

1. Camera-ready light head – light head that is capable of supporting a video camera. Light head is pre-wired for video capability and the camera can “plug and play” into the light head

1. Booms The term “booms” is used to describe the general ceiling-mounted equipment management system. The system is composed of the mounting hub and the arms. The arms are capable of multiple degrees of articulation and positioning within the operating room and terminate in a unique item, such as an equipment carrier, a surgical light, a monitor, an anesthesia hookup, or a perfusion hookup. Definition are as follows:

1. Mounting hub – Infrastructure located within the ceiling that safely supports the weight and articulation of the arms. Multiple mounting hubs may be required in each operating room and multiple arms can extend from a single hub.

1. Anesthesia Hookup Arm – provides the utilities and gases commonly used by an anesthesia machine. This arm is not a fixed column from the ceiling, but rather an arm to tether to the anesthesia machine.

1. Equipment Carrier Arm – Provides storage of the surgical equipment and utilities and gases commonly required during surgery. This usually consists of a column with multiple shelves, power outlets and medical gasses.

1. Perfusion Hookup Arm – Provides utilities and gases commonly used by a heart-lung machine. This arm is not a fixed column from the ceiling, but rather an arm to tether to the heart-lung machine, without lifting the machine off the floor.

1. Monitor Arm – An arm that supports a flat panel surgical display monitor of varying size. This does not include the monitor itself.

1. Surgical Light arm – Supports one surgical light head. Each arm terminates with one light head.

4. Minimum Technical Requirements of Equipment

1. Surgical Lights Surgical Lights are required in every operating room. The specific number of surgical lights per OR room will be specified at the time of ordering. There should be a standard surgical light line item and an in-light camera surgical light line item.

Table 1: Requirements for Surgical Lights Technical Requirements

Light technology
Lights shall utilize LED technology
Central illuminance
Lights shall be capable of achieving 160,000 lux
Head diameter
Light head shall be at least 24” in diameter
Color temperature
Lights shall have selectable color temperature

Lights shall be capable of “pure white light”

Life
Lights shall have a lifespan of at least 50,000 hours and show no diminished quality during this time

Functionality

Focal point
Light head shall be capable of adjusting the light focal point without moving the light head
Dimmer
Light head shall be capable of varying light intensity
Power switch
Light head shall have the ability to power on/off the light
Video capability
Light head should be capable of (in-light) camera functionality option

Usability

Maneuverability
Lights shall be capable of vertical travel

Lights shall be able to fully rotate 360° on all axes Lights shall have drift-free stability

Ease-of-use
Lights shall not require more force/effort than one arm (of a staff member) exerts to swing, position, or move the light

Design Quality and Safety

Sanitation
Lights shall be able to endure VA-specific cleaning procedures

Lights shall not have any contours or grooves that obstruct proper and efficient sanitation

Material durability
Lights shall be fabricated from material that will not stain, wear, rust, fray, creep, chip, bend, or leak after normal usage and scheduled maintenance
Sharp edges
Lights shall not contain any exposed sharp edges

Lights shall not contain any exposed fasteners that could compromise the effectiveness of medical gloves

1. Booms The booms should all have the same general requirements that all terminate with a unique item with additional requirements. All these requirements can be seen below.

Table 2: General Requirements for Arms Technical Requirements

Head Clearance
Clearance to lowest fixed point of arm shall be a minimum of seven (7) feet AFF.
Footprint
Size and arrangement of equipment shall maximize surgical workspace and accessibility to equipment
Lubrication
Arms shall not require any lubricate in addition to manufactures recommended maintenance

Arms shall not require significant deconstruction to apply lubrication of parts

Braking System
System shall use compressed air (pneumatic), electromechanical, or electric mechanisms to stop boom movement

Usability

Maneuverability
Reach and rotational capabilities permit appropriate positioning of anesthesia cart on either side of the surgical table head for effective use during surgery.

Reach and rotational capabilities permit positioning of equipment carrier around surgical table.

Arm reach and rotational capabilities allow placement of the utility arm in various locations for surgical use with rolling equipment carts.

Arm shall be capable of being positioned outside of surgical workspace in such a manner that it does not disrupt workflow when desired.

Lights shall be capable of vertical travel Arms shall have drift-free stability Adjustable rotational stops Column Height is adjustable

Ease of Use
Arms shall not require more force/effort than one staff member exerts to swing, position, or move the arm

Raceway

Arm Diameter
Large enough to support all conduit, utilities, gases and wiring for full functionality to each arm
Wiring
All wiring shall be properly grounded and in conduit

Design Quality and Safety

Sanitation
Arms shall be able to endure VA-specific cleaning procedures

Arms shall not have any contours or grooves that obstruct proper and efficient sanitation

Material durability
Arms shall be fabricated from material that will not stain, wear, rust, fray, creep, chip, bend, or leak after normal usage and scheduled maintenance
Sharp edges
Arms shall not contain any exposed sharp edges

Arms shall not contain any exposed fasteners that could compromise the effectiveness of medical gloves

Structural Interference
Arrangement of arms on the mounting hub shall include sufficient spacing to prohibit interference of arms into one another

Design renderings shall indicate clearance

Line Pressure
All pressure lines (vacuum or gases) shall be equivalent to the pressure at the wall

Table 3: Requirements Specific to the Equipment Carrier Arm Functionality

Shelving
Arm shall have at least 4 shelves

Shelves shall be movable Shelves shall be at least 36” wide Shelves shall be at least 24” deep

Weight Capacity
Shelves should be able to hold at least 75 lbs
Accessories
Arm shall have the ability for additional surgical accessories or monitors to be mounted to the arm

Usability

Maneuverability
Carrier shall be able to be positioned in multiple locations in the X-Y plane

Carrier shall be able to reach and rest adjacent to the surgical table Carrier shall be able to rotate around its Z axis

Utilities – Outlets and Ports

Min # of Med Gas Outlets
Equipment boom should be able to handle a minimum of 8 Medical Gas outlets. The various gasses include Med Air, oxygen, Vacuum, WAGD, Carbon Dioxide, Nitrouds Oxide and Nitrogen.
Electrical
Ability to add a minimum of 8 duplex power outlets
Data
Ability to add a minimum of 6 data ports
Video (for Integration System)
Should have the ability to add a minimum of 2 video inputs for the integration system
Blank Plates
Blank plates available to add future utilities

Table 4: Requirements to the Anesthesia Hookup Arm Functionality

Shelving
No shelving required for the Anesthesia Hookup Arm
Accessories
Arm shall have the ability for additional surgical accessories or monitors to be mounted to the arm
Compatibility
Arm shall be compatible with GE/Datex-Ohmeda S/5 ADU and Draeger Apollo anesthesia machines

Usability

Maneuverability
Hookup shall be able to be positioned in multiple locations in the X-Y plane

Hookup shall be able to reach such that the anesthesia unit can be positioned adjacent to either side of the surgical table Hookup shall be able to rotate around its Z axis

Utilities – Outlets and Ports

Min # of Med Gas Outlets
Equipment boom should be able to handle a minimum of 8 Medical Gas Outlets. The various gasses include Med Air, oxygen, Vacuum, WAGD, Carbon Dioxide, Nitrous Oxide and Nitrogen.
Electrical
Ability to add a minimum of 6 duplex power outlets
Data
Ability to add a minimum of 6 data ports
Video (for Integration System)
Ability to add a minimum of 1 video input for the integration system. Video input should be a universal cable for all video types.
Blank Plates
Blank plates available to add future utilities

Table 5: Requirements Specific to the Perfusion Hookup Arm Functionality

Compatibility
Arm shall be compatible with Century heart lung bypass (perfusion) unit

Usability

Maneuverability
Hookup shall be able to be positioned in multiple locations in the X-Y plane

Hookup shall be able to rotate around its Z axis

Utilities – Outlets and Ports

Min # of Med Gas Outlets
Equipment boom should be able to handle a minimum of 8 Medical Gas Outlets. The various gasses include Med Air, oxygen, Vacuum, WAGD, Carbon Dioxide, Nitrous Oxide and Nitrogen.
Electrical
Ability to add a minimum of 6 duplex power outlets
Data
Ability to add a minimum of 6 data ports
Video (for Integration System)
Ability to add a minimum of 1 video input for the integration system. Video input should be a universal cable for all video types.
Blank Plates
Blank plates available to add future utilities

Table 6: Requirements Specific to the Monitor Arm Functionality

Accessories
Arm shall have the ability for additional surgical accessories or monitors to be mounted to the arm

Usability

Maneuverability
Monitor shall be able to be positioned in multiple locations in the X-Y plane

Arm shall be able to reach such that the monitor can be positioned on either side of the surgical table Arm shall be capable of vertical travel Monitor shall be able to rotate around its Z axis Arm shall include handholds to allow movement of the monitor without touching the monitor

Monitor Bracket
Bracket shall support any standard flat screen monitor up to 27” (diagonally)
Ease of Use
Monitor shall not require more force/effort than one arm (of a staff member) exerts to swing, position, or move the monitor
Integration System
Monitor should be connected to the integration system to route video sources to this monitor

Table 7: Requirements Specific to the Light Arm Functionality

Usability

Maneuverability
Light shall be able to be positioned in multiple locations in the X-Y-Z plane

Arm shall be able to reach such that the light can be positioned on either side of the surgical table Arm shall be able to reach such that the light can be positioned directly over the surgical table Arm shall be capable of vertical travel Light head shall be able to rotate around its Y axis Light head shall include a handhold(s) to allow movement of the light head without touching the housing

Ease of Use
Arm shall not require more force/effort than one staff member exerts using one arm to swing, position, or move the arm

1. Integration System The digital OR integration solution will provide integrated routing of the video, audio, and clinical information system signals in the operating room. The system will provide a graphical interface that allows clinical users to manipulate the routing of video and audio signals from a common in-room interface. This integration system should include a large monitor that will go on a wall in the OR room, monitors that can be attached to monitor arms, a computer that controls the entire integration system and an in-room camera. Additionally, the integration system should have the capability to turn on/off surgical lights, adjust the brightness of the lights, pick various sources to display on the various monitors, and connect to mobile equipment and computers to display those sources on the monitors.

Table 8 - Integration Solution Product Requirements:

Functionality Requirements

1. Must control surgical lights – turn on/off, intensity variation

1. Control entire video system – select video inputs, output video sources to the various monitors, power on/off the monitors

1. Control room camera that will be installed on one end of the room (see specs below)

1. Capture video and images displayed on screen

1. Store data on a drive on the VA network

1. Have the potential to communicate with the other ORs in the facility

Pan/Tilt/Zoom Camera Required

Location
Wall Mounted
Zoom
Ability to zoom
Wide Range Pan/Tilt
Wide Range Pan/Tilt: 340 degrees Pan / 120 degrees Tilt
Quality
4K Camera

Control Station Required

Conduit
Vendor shall provide conduit requirements for Control Station to A/V hub.
Location
Capable of being located on a shelf or desk inside each OR
Screen Size
Minimum 24” touch screen monitor
Resolution
1024 X 768 minimum required
Touch Screen
Includes touch screen
Screen Type
Flat Panel LCD/LED
Input devices
Keyboard and mouse
Voice
2 voice drops max

Audio System Required

Conduit
1” conduit run to A/V Hub max
Location
3-4 speakers per room
MP3 Players
Audio System shall include a means of connecting an MP3 player to speakers.

A/V Hub Required

Power
Vendor shall specify power requirements.
Cabling
Vendor shall provide all cabling for A/V hub.
Size
24” wide X 36” high X 27” deep (Maximum per Hub)
UPS
Included and mounted with system

Video Input Requirements

Video Format
Capable of routing the following video and cabling formats: 525-line interlaced monochrome composite video, 1049-line non-interlaced (progressive) monochrome composite video, NTSC composite video, NTSC S-video (Y/C), NTSC RGBs, SDTV 480i, SDTV 480p, HDTV 720p, HDTV 1080i, HDTV 1080p, RGBHV-SXGA, SDI, HD-SDI and DVI.

Video Output Requirements

Simultaneous Outputs
The system shall be capable of outputting 8 video sources of the formats listed above
Independent Outputs
The vendor shall specify the maximum number of simultaneous sources that the system can output simultaneously

Table 9 - Flat Panel Surgical Display Requirements:

The vendor shall provide medical grade LCD/LED monitors for use in the operating room. These monitors shall be capable of displaying in high resolution all video feeds from the integration solution. These monitors shall meet all necessary medical standards and be properly designed for the operating room environment. The product quoted shall meet or exceed with all specifications below.

Screen Type
Flat Panel LCD/LED
Screen Size
19” minimum – 27” maximum
Viewing Angle
+/- 85 Degrees Horizontal, +/- 85 Degrees Vertical
Boom Compatibility
Monitors shall be compatible with chosen monitor arms and mounts

Table 10 - Installation of Digital OR Integration Solution:

Following contract award vendor shall meet and coordinate with the Contracting Official Technical Representative (COTR) to review and finalize the installation schedule. The product installation services quoted shall meet or exceed with all specifications below.

Mounting
Vendor shall mount and install all products supplied. Vendor shall also provide detailed mounting information.
Configuration
Vendor shall configure each system for specified functionality at time of installation.
Misc. Items
Vendor shall provide all necessary components not specifically mentioned herein for a complete working system.
Cabling
Vendor shall furnish and install all cabling associated with the video integration solution. All cabling must be in conduit, and we prefer that 25% of the conduit is left open for potential future expansion. We also prefer fiber for all cable runs; however, this is not a requirement.
Time of Delivery
Vendor shall provide a delivery time from time of order for one room.
Time of Installation
Vendor shall provide the time required to install one integration solution.

SPECIAL WORK REQUIREMENTS:

Hours: The regular work hours for this facility is Monday through Friday, from 8:00 am to 4:30 pm (CST), excluding Federal Holidays. All maintenance shall be provided during the regular business hours of this facility. If work is required after hours prior approved will be made by department.

Response Time: None required

Preventive Maintenance Inspections: None required

Service Quality: All services provided in this contract must meet manufacturers’ performance and technical specifications, Federal Regulations, VA Regulations, and meet the requirements of The Joint Commission and Life Safety Code. The contractor shall document all maintenance and PMI actions and provide said documentation to the COR upon completion of each service action.

Noncontract Charges: The contractor shall not perform any service that will result in additional charges without prior approval from the Contracting Officer.

4. DELIVERY & LOCATION INFORMATION

Robert J. Dole Medical Facility
5300 East Kellogg

Wichita Kansas, 67218

A.1 PRICE/COST SCHEDULE

ITEM INFORMATION

CLIN#
DESCRIPTION OF SUPPLIES/SERVICES
QNTY
UNIT
UNIT PRICE
AMOUNT
0001
Handle, Devon Slip-on (E, Weighted)

MANUFACTURER PART NUMBER (MPN): 83307

4.00
EA
____________
___________
0002
F628/F628 Surgical Light (1 Per OR)

MANUFACTURER PART NUMBER (MPN): FD 2000

2.00
EA
____________
___________
0003
Preparation Camera StrykeCam 4K

MANUFACTURER PART NUMBER (MPN): CY 9600105

2.00
EA
____________
___________
0004
Multi Color Touch wall control display f

MANUFACTURER PART NUMBER (MPN): CY 1008104

2.00
EA
____________
___________
0005
Dual F628s and Single Flat Panel (1 Per OR)
2.00
EA
____________
___________
0006
Handle, Devon Slip-on (E, Weighted)

MANUFACTURER PART NUMBER (MPN): 83307

4.00
EA
____________
___________
0007
Handle, Devon Slip-on FP

MANUFACTURER PART NUMBER (MPN): 77922

2.00
EA
____________
___________
0008
F628/F628/FP Surgical Light + FP Bracket,(1 Per OR)

MANUFACTURER PART NUMBER (MPN): FT 3003

2.00
EA
____________
___________
0009
Preparation Camera StrykeCam 4K

MANUFACTURER PART NUMBER (MPN): CY 9600105

2.00
EA
____________
___________
0010
Multi Color Touch wall control display f

MANUFACTURER PART NUMBER (MPN): CY 1008104

2.00
EA
____________
___________
0011
STRYKECAM 4k FOR F-GEN

MANUFACTURER PART NUMBER (MPN): P31858

2.00
EA
____________
___________
0012
Assy, camera cover for SrykeCam

MANUFACTURER PART NUMBER (MPN): P27047

4.00
EA
____________
___________
0013
HI-DEF CAMERA DECODER ASSEMBLY

MANUFACTURER PART NUMBER (MPN): P22059

2.00
EA
____________
___________
0014
CABLE ASSY DVI M TO DVI M 15 FT

MANUFACTURER PART NUMBER (MPN): 100224637

2.00
EA
____________
___________
0015
Single Flat Panel Monitor Bracket

MANUFACTURER PART NUMBER (MPN): FS 1003

2.00
EA
____________
___________
0016
Handle, Devon Slip-on FP

MANUFACTURER PART NUMBER (MPN): 77922

2.00
EA
____________
___________
0017
Equipment Boom (1 Per OR)

MANUFACTURER PART NUMBER (MPN): TP-622-C (as configu

2.00
EA
____________
___________
0018
Electric Brakes, 4/6 Series Upcharge

MANUFACTURER PART NUMBER (MPN): RS 0006157

2.00
EA
____________
___________
0019
Kit, Accessory Extension, Bottom

MANUFACTURER PART NUMBER (MPN): 5000548

2.00
EA
____________
___________
0020
Panel Ass'y., N2 Control Mounting

MANUFACTURER PART NUMBER (MPN): 5000066

2.00
EA
____________
___________
0021
Kit, Med Rail, 6 Series 3 Sides, LS

MANUFACTURER PART NUMBER (MPN): 5000579

2.00
EA
____________
___________
0022
Arm, 300 X 500MM IV Swing

MANUFACTURER PART NUMBER (MPN): 500011510

2.00
EA
____________
___________
0023
Holder, 36in Single Pole, W/20in IV Pole

MANUFACTURER PART NUMBER (MPN): 500011513

2.00
EA
____________
___________
0024
Slide, Vacuum Bottle,W/Snap Action Adpt

MANUFACTURER PART NUMBER (MPN): 500011429

4.00
EA
____________
___________
0025
Anesthesia Boom (1 Per OR)

MANUFACTURER PART NUMBER (MPN): TP-612-C (as configu

2.00
EA
____________
___________
0026
Electric Brakes, 4/6 Series Upcharge

MANUFACTURER PART NUMBER (MPN): RS 0006157

2.00
EA
____________
___________
0027
Kit, Med Rail, 6 Series 4 Sides, LS

MANUFACTURER PART NUMBER (MPN): 5000578

2.00
EA
____________
___________
0028
Slide, Vacuum Bottle,W/Snap Action Adpt

MANUFACTURER PART NUMBER (MPN): 500011429

4.00
EA
____________
___________
0029
Assy, camera cover for SrykeCam

MANUFACTURER PART NUMBER (MPN): P27047

6.00
EA
____________
___________
0030
CHROMOPHARE LIGHT HANDLE COVER (1 PACK)

MANUFACTURER PART NUMBER (MPN): 71602

1.00
EA
____________
___________
0031
Installation, Teletom TC

MANUFACTURER PART NUMBER (MPN): RS 0006139

2.00
EA
____________
___________
0032
Installation, Teletom TC

MANUFACTURER PART NUMBER (MPN): RS 0006139

2.00
EA
____________
___________
0033
Installation, Chromophare

MANUFACTURER PART NUMBER (MPN): RS 0006008

2.00
EA
____________
___________
GRAND TOTAL