36C25518Q9742-004.pdf

PDF 16 KB Posted

Attached to
Medical and Dental Air Compressors and Vacuum Maintenance Service KC VAMC
Solicitation number
36C25518Q9742
Issued by

About this file

36C25518Q9742 attachment 2 Equipment Listing and Inventory.pdf

View the file

Other files

File Type Posted
36C25518Q9742-0001000.docx DOCX document
36C25518Q9742-002.docx DOCX document
36C25518Q9742-001.docx DOCX document
36C25518Q9742-003.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Medical/Dental Vacuum and Compressor Equipment Inventory

The scheduled maintenance tasks listed below are in addition to the annual and semi-annual preventive maintenance tasks. They can be scheduled in conjunction with either the annual or semi-annual preventive maintenance.

Sub-Basement-Main Tower Control Air:

o Atlas Copco Model # GX11P-EL. Qty. 2 Units

Estimated yearly run time for each compressor: 4380 Hours

S/N: CAI636476

S/N: CAI635239

Every 4000 hours: Change oil, oil filter, air filter and condensate O-rings.

Every 8000 hours: Change oil, oil filter, air filter and condensate O-rings and separator.

Sub-Basement-Medical Vacuum:

o Powerex Medical Vacuum Model # CVPT15052RI. Qty. 1 Unit Estimated yearly run time: 5000 Hours S/N: (H) 02/13/17-6006752-556

Every 2500 hours: Replace air filters.

Every 6 months: Perform bearing service.

Boiler Plant Building 4:

o Atlas Copco Model # GA15FF

S/N: API451594

S/N: API104085

S/N: API451811

Every 4000 hours: Change oil, oil filter, air filter and condensate O-rings.

Every 8000 hours: Change oil, oil filter, air filter and condensate O-rings and separator.

Building 26 VB-720:

o Squire Cogswell Model # S1000ST3-X Estimated run time 4000 Hours A Year S/N: 149186. (Triplex 10HP) Vacuum

Every 2000 Hours: Replace Exhaust Filter and associated components.

Once a year: Replace Inlet Air Filters Every 1000 Hours: Replace Oil and Oil Filters

Dental Compressor:

o Powerex Model # SEQ200740AJ Estimated yearly run time 2500 Hours

(M) 01/09/17-6006752-274

Once a year: Replace Intake Filters.

Once a year Dryer Service: Replace Pre-Filter, After Filter, Muffler Auto Drain Pre-Filter.

V4-610 SPS Compressed Air:

Once a year: Replace CO Calibration Kit and Separator Float Assemblies Once a year Dryer Service: Replace Pre-Filter, After Filter, Muffler Auto Drain Pre-Filter.

o Powerex Model # MSQ05044PS Estimated yearly run time for each compressor (Qty. 2): 2190 hours

(H) 04/12/13-5420471-971 o Powerex Model # MSQ075554PS Estimated yearly run time for each compressor (Qty. 2): 2190 hours

(H) 01/27/17-6006752-473

Desiccant Dryers (Qty. 3):

o Speedaire: Model 4NMJ6 (Qty. 1) o Gardner Denver: Model GSRN-125A-116 (Qty. 2)

File details come from the government source that posted it.