36C25226Q0295_9.docx

DOCX document 23 KB Posted

Attached to
Leica DM2700 P for TL Specifically configured for Asbestos Analysis
Solicitation number
36C25226Q0295_9
Issued by

About this file

This is a Special Notice for a federal contract opportunity issued by the Department of Veterans Affairs (VA), Veterans Health Administration, Veterans Integrated Service Network 12, located in Milwaukee, Wisconsin. The solicitation seeks a Leica DM2700 P for TL microscope specifically configured for asbestos analysis. The solicitation number is 36C25226Q0295, with a response deadline of April 30, 2026, at 11:00 AM Central Time. The product service code is 6515 (Laboratory Analytical Instruments), and the NAICS code is 334516 (Analytical Laboratory Instrument Manufacturing). The contracting office is located at 115 South 84th Street, Suite 101, Milwaukee, WI 53214-1476.

The notice includes clarification regarding calibration requirements for the microscope system. The VA requires factory calibration certification with a NIST-traceable micrometer scale. However, a question-and-answer section indicates that the vendor requires further guidance on calibration methodology, specifically requesting clarification on what components will be calibrated and reported. The VA suggests testing the system in accordance with the NIOSH 7400 method using a certified phase-contrast test slide to establish calibration standards. The special notice will be archived 60 days after the response date.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Special Notice Special Notice

SUBJECT*
WLR Leica DM2700 P for TL Specifically configured for Asbestos Analysis (VA-25-00098025)

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
53214-1476
SOLICITATION NUMBER*
36C25226Q0295
RESPONSE DATE/TIME/ZONE
04-30-2026 11:00 am CENTRAL TIME, CHICAGO, USA
ARCHIVE
60 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N
PRODUCT SERVICE CODE*
6515
NAICS CODE*
334516
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

36C252 115 South 84th Street, Suite 101

Milwaukee WI 53214-1476

POINT OF CONTACT*

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

DESCRIPTION

Q&A To clarify Question Number 11 from the previous modification Q&A.

11. Will a factory calibration certificate with NIST‑traceable micrometer scale satisfy the requirement, or is third‑party calibration preferred?

Unclear what requirement is being referred to here. But Factory calibration certification with the NIST traceable micrometer scale is a requirement that is requested.

Response We can provide a NIST traceable calibration, but we need something to calibrate it to, and since there is no camera or measuring reticle on the proposal, I need guidance on what we will be calibrating and reporting the micrometer to.

I think the suggesting to test the system similar to the NIOSH 7400 method, against a certified phase-contrast test slide.

*= Required Field