36C25226Q0217_2.docx

DOCX document 30 KB Posted

Attached to
6515--Dose Calibrator (Brand Name or Equal)
Solicitation number
36C25226Q0217
Issued by

About this file

This is a Presolicitation Notice for a dose calibrator procurement by the VA Great Lakes Acquisition Center for the Edward Hines, Jr. VA Hospital in Hines, Illinois. The Department of Veterans Affairs intends to solicit one CRC-55TW dose calibrator with CRC-55TR DOS calibrator and CAPRAC-T wipe test counter, manufactured by Mirion Technologies (Capintec), Part No. B-5130-2216, or an equal alternative. The procurement is being conducted as a Small Business set-aside under FAR Parts 12, 13, and 19 with a firm fixed-price contract structure. The corresponding NAICS code is 339112 with PSC 6515.

The formal Request for Quote (RFQ) document 36C25226Q0217 will be posted on or around Wednesday, February 18, 2026, with offers due no later than 15:00 Central Time on Wednesday, February 25, 2026. The Contracting Officer is Waymon McNeal Jr. at the Department of Veterans Affairs, 36C252, 115 South 84th Street, Suite 101, Milwaukee, WI 53214-1476. For equipment qualification as an equal, vendors must meet specific salient characteristics including an 8-inch or larger color display, USB/PC communications capability, software upgrade functionality, USB printer capability, 28 programmable nuclide keys, over 80 nuclides in memory, automatic chamber quality control with linearity testing, 256-channel MCA with spectrum analysis, well counter for wipe tests with automated quality control, and automated linearity and geometry testing with F-18 measurement capability.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Presolicitation Notice

SUBJECT*
Dose Calibrator (Brand Name or Equal)

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
53214-1476
SOLICITATION NUMBER*
36C25226Q0217
RESPONSE DATE/TIME/ZONE
02-25-2026 15:00 CENTRAL TIME, CHICAGO, USA
ARCHIVE
30 DAYS AFTER THE RESPONSE DATE
SET-ASIDE
SBA
PRODUCT SERVICE CODE*
6515
NAICS CODE*
339112
PLACE OF PERFORMANCE
Edward Hines, Jr. VA Hospital

5000 South 5th Avenue

Hines IL

POSTAL CODE
60141-5000
COUNTRY
USA

CONTACT INFORMATION

CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

36C252 115 South 84th Street, Suite 101

Milwaukee WI 53214-1476

POINT OF CONTACT*

CONTRACTING OFFICER

Waymon McNeal Jr.

waymon.mcnealjr@va.gov waymon.mcnealjr@va.gov

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

DESCRIPTION

*=Required Field
Presolicitation Notice

Presolicitation Notice

The VA Great Lakes Acquisition Center intends to solicit a Brand Name or Equal Dose Calibrator to the Edward Hines, Jr. VA Hospital 5000 South 5th Avenue Hines, IL 60141-5000.

This procurement is being conducted in accordance with FAR Part 12, 13, and 19. Contract will be firm fixed price and the corresponding NAICS Code 339112. This will be a Small Business set-aside solicitation.

Please see Request for Quote (RFQ) document 36C25226Q0217 for details in the near future. This RFQ will be posted on or around Wednesday, February 18, 2026. Offers will be due no later than 15:00 Central Time, Chicago, USA on Wednesday, February 25, 2026.

Brand Name

DESCRIPTION OF SUPPLIES
Manufacturer
Part No
QTY
CRC-55TW with a CRC-55TR DOS CALIBRATOR AND CAPRAC-T WIPE TEST COUNTER
Mirion Technologies (Capintec)
B-5130-2216
1

Equal Requirements:

Salient Characteristics:

· 8 in. color display or larger

· Must have USB/PC Communications

· Must have the ability to upgrade software via USB or flash drive

· USB printer capability

· Chamber plug and play capability

· 28 programmable nuclide keys

· Built-in dose calibration, quality control and self diagnostics

· Must have over 80 nuclides in memory

· Must have automatic chamber QC including linearity

· Hard copy records of data available from optional printer

· Must have 256 channel MCA with detailed spectrum for identification analysis

· Well counter for wipe tests

· User-definable protocols for wipe testing

· Automated well QC including Chi-Square and MDA

· Manual and automatic ROI

· Must include a pre-set key for F-18 measurements

· Automated Linearity and Geometry testing

Presolicitation Notice

File details come from the government source that posted it. Updated .